← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K253217

# MiiS Horus Arthroscope  (EJA 100); MiiS Horus Endoscope Display System -Tablet (EDS 500); MiiS Horus Endoscope Display System -Video Box (EVS 700) (K253217)

_Medimaging Integrated Solution, Inc (Miis) · HRX · Dec 23, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K253217

## Device Facts

- **Applicant:** Medimaging Integrated Solution, Inc (Miis)
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Dec 23, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

MiiS Horus Arthroscope: The MiiS Horus Arthroscope is a sterile and single-use device designed for operation with its compatible display system. It is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization of an interior cavity of the body through a surgical opening. The device is intended for use in a hospital or medical office environment and is designed for use in adults. MiiS Horus Endoscope Display System -Tablet: The MiiS Horus Endoscope Display System -Tablet is intend for use with compatible endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation. MiiS Horus Endoscope Display System -Video Box: The MiiS Horus Endoscope Display System -Video Box is intend for use with compatible endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation.

## Device Story

System comprises sterile, single-use arthroscope (EJA 100) and reusable display units (Tablet EDS 500 or Video Box EVS 700). Arthroscope captures endoscopic video signals; transmits via cable to display system. Tablet displays live images/snapshots; Video Box acts as standalone transfer system for projection to external computers. Used in hospital/medical office by clinicians for diagnostic/operative visualization. Output allows real-time observation of interior cavities to guide surgical/diagnostic decision-making. Benefits include minimally invasive visualization of body cavities.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), sterilization validation (ISO 11135), biocompatibility (ISO 10993), optical/color performance, and endoscopic performance (ISO 8600).

## Technological Characteristics

Sterile, single-use arthroscope with distal LED light source. FOV 120°, DOF 3-50 mm. Resolution 720x720. Reusable tablet/video box display units. Connectivity via cable. Sterilization via EO. Biocompatibility per ISO 10993-1 for limited duration contact. Complies with IEC 60601-1, IEC 60601-2-18, and IEC 62471.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Arthrex NanoScope System ([K201134](/device/K201134.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 23, 2025

Medimaging Integrated Solution, Inc (MiiS)
Sandy Chou
Regulatory Affairs
3f, # 24-2, Industry E. Rd. Iv, Hsinchu Science Park
Hsinchu, 30077
Taiwan

Re: K253217
Trade/Device Name: MiiS Horus Arthroscope (EJA 100); MiiS Horus Endoscope Display System - Tablet (EDS 500); MiiS Horus Endoscope Display System -Video Box (EVS 700)
Regulation Number: 21 CFR 888.1100
Regulation Name: Arthroscope
Regulatory Class: Class II
Product Code: HRX
Dated: September 25, 2025
Received: September 29, 2025

Dear Sandy Chou:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253217 - Sandy Chou
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K253217 - Sandy Chou
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JESSE MUIR-S
Digitally signed by JESSE
MUIR-S
Date: 2025.12.23 13:11:21
-05'00"

Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253217 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  MiiS Horus Arthroscope (EJA 100);
MiiS Horus Endoscope Display System -Tablet (EDS 500);
MiiS Horus Endoscope Display System -Video Box (EVS 700)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  MiiS Horus Arthroscope:
The MiiS Horus Arthroscope is a sterile and single-use device designed for operation with its compatible
display system.
It is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide
visualization of an interior cavity of the body through a surgical opening.
The device is intended for use in a hospital or medical office environment and is designed for use in adults.

MiiS Horus Endoscope Display System -Tablet:
The MiiS Horus Endoscope Display System -Tablet is intend for use with compatible endoscopes to provide
visualization for endoscopic diagnosis, treatment, and video observation.

MiiS Horus Endoscope Display System -Video Box:
The MiiS Horus Endoscope Display System -Video Box is intend for use with compatible endoscopes to
provide visualization for endoscopic diagnosis, treatment, and video observation.  |   |   |
|  Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |

{4}

Page 1

K253217

510(k) Summary

Prepared: September 25, 2025

Submitter/Owner’s: Medimaging Integrated Solution Inc. (MiiS)

Name/Address: 3F., No. 24-2, Industry E. Rd. IV, Hsinchu Science Park, Hsinchu, 30077, Taiwan, R.O.C.

Contact Person: Sandy Chou

Telephone: +886-3-5798860 Ext: 1701

Fax: +886-3-5798011

Address: ra@miis.com.tw

{5}

Page 2

# Device Identification:

Trade/Device Name: Arthroscope: MiiS Horus Arthroscope (EJA 100)
Tablet: MiiS Horus Endoscope Display System -Tablet (EDS 500)
Video Box: MiiS Horus Endoscope Display System -Video Box (EVS 700)

Common Name: Arthroscope
Regulation Name: Arthroscope
Regulation Number: 21 CFR 888.1100
Regulatory Class: Class II
Product Code: HRX

# Predicate Device:

K201134
Trade/Device Name: Arthrex NanoScope System
Regulation Number: 21 CFR 876.1500 and 888.1100
Regulatory Class: Class II
Product Code: GCJ, HRX

# Description of Device:

MiiS Horus Arthroscope is a sterile and single-use device designed to be operated with its compatible display system. The device provides visualization for endoscopic procedures of an interior cavity of the body through a surgical opening.

The Arthroscope needs to connect with its compatible display system MiiS Horus Endoscope Display System -Tablet or MiiS Horus Endoscope Display System -Video Box and the signal captured was then transmitted to the cable-connected display system. Both the Tablet and Video box are reusable devices.

MiiS Horus Endoscope Display System –Tablet is a tablet which showing live image or snapshot from Arthroscope.

MiiS Horus Endoscope Display System -Video Box is a standalone imaging transfer system; it can be connected to computer to project live imaging. If needed, the pictures or videos can be storage to Tablet or computer.

{6}

Page 3

# Indications for Use

## MiiS Horus Arthroscope:

The MiiS Horus Arthroscope is a sterile and single-use device designed for operation with its compatible display system.

It is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization of an interior cavity of the body through a surgical opening.

The device is intended for use in a hospital or medical office environment and is designed for use in adults.

## MiiS Horus Endoscope Display System -Tablet:

The MiiS Horus Endoscope Display System -Tablet is intended for use with compatible endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation.

## MiiS Horus Endoscope Display System -Video Box:

The MiiS Horus Endoscope Display System -Video Box is intended for use with compatible endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation.

# Substantial Equivalence Summary

The MiiS Horus Arthroscope and compatible display system is substantially equivalent in intended use, technical specifications, performance and principles of operation to the predicate device; Arthrex Nanoscope System (K201134).

The indications for use for the MiiS Horus Arthroscope and compatible display system represent a subset of the cleared indications for the predicate Arthrex Nanoscope System. As a subset of the indications for the cleared device, this difference does not raise any new risks or issues related to safety or efficacy.

Below is a summary table that provides a top-level overview of the substantial equivalent comparison between proposed and predicate devices.

|  Item | Subject Device
MiiS Horus Arthroscope
(EJA 100) | Predicate Device
Arthrex NanoScope System
(K201134)  |
| --- | --- | --- |
|  Submitter | Medimaging Integrated Solution Inc. | Arthrex Inc.  |
|  Model name | Arthroscope:
MiiS Horus Arthroscope

Tablet:
MiiS Horus Endoscope Display System
-Tablet | Arthrex NanoScope System  |

{7}

|  Item | Subject Device
MiiS Horus Arthroscope
(EJA 100) | Predicted Device
Arthrex NanoScope System
(K201134)  |
| --- | --- | --- |
|   | Video Box:
MiiS Horus Endoscope Display System
-Video Box |   |
|  K Number | --- | K201134  |
|  Product code & Classification | HRX
Class II | HRX
Class II  |
|  Principle System Components
(Sterile, Disposable, Single Use) | MiiS Horus Arthroscope | Nanoscope  |
|  Principle System Components
(Reusable) | MiiS Horus Endoscope Display System
-Tablet
MiiS Horus Endoscope Display System
-Video Box | Nanoscope Control Unit  |
|  Intended use | MiiS Horus Arthroscope:
The MiiS Horus Arthroscope (EJA 100) is a sterile and single-use device designed for operation with its compatible display system.
It is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization of an interior cavity of the body through a surgical opening.
The device is intended for use in a hospital or medical office environment and is designed for use in adults.

MiiS Horus Endoscope Display System
-Tablet:
The MiiS Horus Endoscope Display System -Tablet (EDS 500) is intend for use with compatible endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation.

MiiS Horus Endoscope Display System
-Video Box:
The MiiS Horus Endoscope Display System -Video Box (EVS 700) is intend for use with compatible | The Arthrex NanoScope System is intended to be used as an endoscopic video camera in a variety of endoscopic diagnostic and surgical procedures, including but not limited to:
orthopedic, spine, laparoscopic, urologic, sinuscopic, and plastic surgical procedures. The device is also intended to be used as an accessory for microscopic surgery.  |

Page 4

{8}

|  Item | Subject Device
MiiS Horus Arthroscope
(EJA 100) | Predicted Device
Arthrex NanoScope System
(K201134)  |
| --- | --- | --- |
|   | endoscopes to provide visualization for endoscopic diagnosis, treatment, and video observation. |   |
|  Technical Features  |   |   |
|  Lens Specification | FOV: 120°
DOV: 0° (EJA 100-0D)/ 30° (EJA 100-30D)
DOF: 3-50 mm | FOV: 120°
DOV: 0°
DOF: 3 mm × 100 mm  |
|  Resolution | 720x720 | 400x400  |
|  Illuminance | >100 Lux at 50 mm | 49 Lux at 50 mm  |
|  Mechanical Specification | Working length: 125 mm
Scope outer diameter: 2.2 mm (EJA 100-0D)/3.4 mm (EJA 100-30D) | Working length: 125 mm
Scope outer diameter: 1.9 mm  |
|  LED Light source | At distal tip | Fiber at distal tip with LED in the handle  |
|  Bench test | ISO 8600-1
ISO 8600-3
ISO 8600-4 | ISO 8600-1
ISO 8600-3
ISO 8600-4  |
|  Sterilization | EO Sterilization | EO Sterilization  |
|  Electrical Safety  |   |   |
|  EMC | IEC 60601-1-2 | IEC 60601-1-2  |
|  Safety | IEC 60601-1
IEC 60601-2-18 | IEC 60601-1
IEC 60601-2-18  |
|  Materials  |   |   |
|  Biocompatibility of Materials
Meets ISO 10993-1 requirements | Meets ISO 10993-1 requirements for tissue/bone/dentin contact of limited duration (<24 hrs) | Meets ISO 10993-1 requirements for tissue/bone/dentin contact of limited duration (<24 hrs)  |

## Summary of the technological characteristics in comparison to the predicate devices

## Similarities

"MiiS Horus Arthroscope" and its compatible display system (MiiS Horus Endoscope Display System - Tablet (EDS 500) or MiiS Horus Endoscope Display System - Video Box (EVS 700)) and the predicate device, Arthrex Inc. Arthrex NanoScope System (K201134), have the same intended use in arthroscopic diagnostics and procedures. Both devices share the same key mechanical specifications such as working length and field of view (FOV). For model EJA 100-

{9}

0D, the direction of view (DOV) is also the same. In addition, the subject and predicate device have successfully passed the same testing, including EMC, safety, bench, package integrity, transportation, aging, and biocompatibility tests. Both devices are sterilized using the same EO method and comply with EO/ECH residue limits.

## Differences

The differences pertain to the scope’s outer diameter and the 30° direction of view (DOV) of model EJA 100-30D. The outer diameter is slightly larger but remains within the range of commonly used arthroscope sizes. The larger diameter enhances optical performance and structural strength. The 30° DOV, which is commonly available in arthroscopes, has been verified to meet specifications and performance requirements. Regarding depth of field (DOF), the subject device has a range of 3–50 mm, compared to 3 mm × 100 mm for the predicate device. Bench testing shows MiiS Horus Arthroscope provides comparable image resolution and illuminance. These differences do not introduce any new risks or concerns regarding safety or effectiveness. Therefore, the MiiS Horus Arthroscope is considered substantially equivalent to its predicate devices.

## Nonclinical Tests

The following tests have been performed in support of the substantial equivalence determination:

- Electrical safety per IEC/EN 60601-1
- Electromagnetic Compatibility per IEC/EN 60601-1-2
- Sterilization Validation per ISO 11135
- Package Integrity per ISO 11607-1
- Transportation test per ASTM D4169
- Aging test per ASTM F1980
- Biocompatibility test per ISO 10993-1
- Cytotoxicity test per ISO 10993-5
- Sensitization test per ISO 10993-10
- Irritation test per ISO 10993-23
- Acute systemic toxicity and Pyrogen test per ISO 10993-11
- Optical and Color performance testing
- Software validation
- Cybersecurity test
- Endoscopic Performance test per ISO 8600
- EO/ECH residue test per ISO 10993-7

Page 6

{10}

- Light hazard (photobiological safety) test per IEC 62471

## Clinical Tests

No clinical studies were performed.

## Conclusion

The substantial equivalence comparison, together with bench performance testing, supports the conclusion that the MiiS Horus Arthroscope is substantially equivalent to the predicate device, the Arthrex NanoScope System (K201134).

Page 7

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K253217](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K253217)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
