← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K252458

# Kyphoplasty Balloon Dilatation Catheters (K252458)

_Shanghai Lange Medtech Co., Ltd. · HRX · Nov 3, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K252458

## Device Facts

- **Applicant:** Shanghai Lange Medtech Co., Ltd.
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Nov 3, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Kyphoplasty Balloon Dilatation Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements).

## Device Story

Inflatable balloon catheter used in percutaneous kyphoplasty (PKP) to treat spinal fractures; consists of inner stylet, balloon catheter, and protection tube. Physician inserts catheter into vertebral body; inflates balloon to restore vertebral height and create cavity in cancellous bone; deflates balloon; cavity facilitates controlled filling of PMMA bone cement. Procedure reduces risk of cement leakage and complications. Radiopaque markers at tip assist in fluoroscopic positioning. Device is single-use, sterile, and operated by physicians in clinical settings.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench performance testing (burst pressure, burst volume, fatigue, deflation time, tensile force) and biocompatibility testing per ISO 10993.

## Technological Characteristics

Materials: PC, stainless steel, TPU, PE, Pebax, Pt/Ir. Components: Inner stylet, balloon catheter, PP protection tube. Sterilization: EO (ISO 11135). Biocompatibility: ISO 10993-1 (limited contact). Connectivity: None. Form factor: Straight or curved tip, various lengths.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Kyphoplasty Balloon Catheter ([K223709](/device/K223709.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 3, 2025

Shanghai Lange MedTech Co., Ltd
% Chen Kevin
Official Correspondent
Shanghai Mil-Medshare Medical Technology Co., Ltd
Room 711, East Building, BHC Central Center, No. 2218 Hunan
Road, Pudong New Area
Shanghai,
China

Re: K252458
Trade/Device Name: Kyphoplasty Balloon Dilatation Catheters
Regulation Number: 21 CFR 888.1100
Regulation Name: Arthroscope
Regulatory Class: Class II
Product Code: HRX, HXG
Dated: August 1, 2025
Received: August 5, 2025

Dear Chen Kevin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252458 - Chen Kevin
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252458 - Chen Kevin
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JESSE
MUIR -S

Digitally signed by
JESSE MUIR -S
Date: 2025.11.03
11:47:03 -05'00'

Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. |  | ?  |
|  Please provide the device trade name(s). |  | ?  |
|  Kyphoplasty Balloon Dilatation Catheters  |   |   |
|  Please provide your Indications for Use below. |  | ?  |
|  Kyphoplasty Balloon Dilatation Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements).  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

1 / 7
K252458

510(k) summary
K252458

I Submitter
Device submitter: Shanghai Lange MedTech Co., Ltd.
1st floor, Building 22, 1500 Duhui Road, Minhang District, 201108
Shanghai, P.R. China Shanghai 201108 China.

Contact person: ZHOU Xin
Management representative
Phone: +86 18802178896
Fax: 021-67657131
Email: RA@langemedtech.com

Date: Oct 31, 2025

II Device
Trade Name of Device: Kyphoplasty Balloon Dilatation Catheters
Common Name: Inflatable BonSe Tamp
Regulation Number: 21 CFR 888.1100
Regulation Name: Arthroscope
Regulatory Class: II
Product code: HRX, HXG
Review Panel: Orthopedic

III Correspondent
Shanghai Mil-Medshare Medical Technology Co., Ltd.
Room 711, East Building, BHC Central Center, No. 2218 Hunan Road, Pudong New
Area, Shanghai
Contact person: Kevin Chen
Email: Kevin.chen@mil-medshare.com

IV Device description
The proposed device, Kyphoplasty Balloon Dilatation Catheters are inflatable balloon catheters used in percutaneous kyphoplasty (PKP). It consists of Inner Stylet, Balloon
Catheter, and polypropylene (PP) Protection tube. The radiopaque markers located at
the balloon catheter tip to reflect the balloon position during positioning.

The proposed device, Kyphoplasty Balloon Dilatation Catheters are inflatable balloon catheters used in percutaneous kyphoplasty (PKP). It consists of Inner Stylet, Balloon
Catheter, and polypropylene (PP) Protection tube. The device is made of PC, stainless
steel, TPU, PE, Pebax and Pt/Ir.

The inner stylet is made from a stylet attached to a 6% Luer cap. The stylet enhances the
stiffness of the balloon catheter to facilitate a smooth insertion of balloon catheter
through the established working pathway during the percutaneous procedure. The PP
Protection tube protects the balloon from damage during the production and packaging.

{5}

The balloon catheter is composed of six sections: Inflation Port, Stylet Port, Inner Tube, Outer Tube, Radiopaque Markers, Inflatable Component. The Stylet Port is used to provide a support for the Stylet. The Inflation Port provide an adapter for a balloon inflation Injector. The Inflatable Component plays a critical role in the procedure. The Inflatable Component is to restore the height of a compression-fractured vertebral body and also leave a cavity in the fractured area after the Inflatable Component deflation. The cavity is therefore to facilitate safer filling of the bone cement. The cavity in the fractured area left by the inflated balloon reduces the filling pressure of bone cement. This technology greatly reduces the risk of uncontrollable bone cement leakage and spread, and thus lowers complications and minimizes the risk of the procedure.

The radiopaque markers located at the balloon catheter tip to reflect the balloon position during positioning.

Kyphoplasty Balloon Dilatation Catheters are supplied sterilized, single-use.

## Model and specifications

There are generally two types of Kyphoplasty Balloon Dilatation Catheters, one is straight type which has a straight balloon catheter tip, and the other one is curved type which has a curved shape of tip of the balloon catheter. These catheters are available in different effective length. The detailed specifications are listed in Table 1.

Table 1 Specifications of Kyphoplasty Balloon Dilatation Catheters

|  Type | Specifications | Effective length of Inflatable Component L (mm) | Diameter Of Outer Tube D1 (mm) | Bending angle A (°)  |
| --- | --- | --- | --- | --- |
|  Straight type | S01-1610 | 10±2 | 2.0±0.5 | /  |
|   |  S01-1615 | 15±2 | 2.4±0.5  |   |
|  Curved type | S01-1620 | 20±2 |   | 90±20  |

## V Indications for use

Kyphoplasty Balloon Dilatation Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements).

## VI Substantial Equivalence

The proposed Kyphoplasty Balloon Dilatation Catheters is substantially equivalent to the predicate device:

K252458

{6}

|  Trade name: | Kyphoplasty Balloon Catheter  |   |
| --- | --- | --- |
|  Common name: | Inflatable Bone Tamp  |   |
|  Classification name: | Product Code: | Classification:  |
|  Arthroscope | HRX | Class II, 21CFR 888.1100  |
|  Tamp | HXG | Class I, 21CFR 888.4540  |
|  Premarket Notification: | K223709  |   |
|  Manufacturer: | Jiangsu Changmei Medtech Co., Ltd.  |   |

Comparison of technological characteristics with the predicate devices

K252458

{7}

|  Device feature | Subject Device (Kyphoplasty Balloon Dilatation Catheters) | Predicate Device K223709 (Kyphoplasty Balloon Catheter) | Comment  |
| --- | --- | --- | --- |
|  Product Code | HRX
HXG | HRX
HXG | Equivalent  |
|  Regulation No. | 21 CFR 888.1100
21 CFR 888.4540 | 21 CFR 888.1100
21 CFR 888.4540 | Equivalent  |
|  Class | Class II | Class II | Equivalent  |
|  Indication for use | Kyphoplasty Balloon Dilatation Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements). | Kyphoplasty Balloon Catheter is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements). | Equivalent  |
|  Balloon Size | 10.0mm,15.0mm
20.0mm | 10.0mm,15.0mm
20.0mm | Equivalent  |
|  Balloon Brust Pressure | 400 psi | 400 psi | Equivalent  |
|  Balloon Burst Volume | ≥6ml | ≥6ml | Equivalent  |
|  Balloon Inflation Behavior | Balloon Diameter:18±2 mm
Balloon working length: 28±3mm | Balloon Diameter:18.3±3.0 mm
Balloon working length: 28.0±3.0 | Different 1 see explanation below  |
|  Tensile Force Balloon | ≥15 N | ≥15 N | Equivalent  |

K252458

{8}

|  Balloon Repeated Inflation | The balloon must not rupture or leak within 20 inflation/deflation cycles | The balloon must not rupture or leak within 20 inflation/deflation cycles | Equivalent  |
| --- | --- | --- | --- |
|  Balloon Deflation Time | ≤3 s | ≤3 s | Equivalent  |
|  Shapes | cylindrical | Peanut, cylindrical | Different 2 see explanation below  |
|  Inflation medium | Contrast medium | Contrast medium | Equivalent  |
|  Single/double use of catheter | Single | Single | Equivalent  |
|  Sterility | Yes | Yes | Equivalent  |
|  Biocompatibility | Confirm to the requirements of ISO 10993 series standards | Confirm to the requirements of ISO 10993 series standards | Equivalent  |

## Discussion:

### Different 1-Balloon Inflation Behavior

The difference in the size does not raise additional questions for safety and effectiveness of the device. The performance test of the subject devices has been performed on the final finished device. The test results show pass the requirements.

### Different 2- Shapes

The shape of predicate device has a Peanut type, and the proposed device haven't. The shape of subject device is within the scope of predicate device. Therefore, the differences on balloon shape do not raise new questions about safety and effectiveness.

K252458

{9}

6 / 7
K252458

# Conclusion:

The subject and predicate device have the same intended use. The differences in technological characteristics between the subject and predicate devices (i.e. Balloon Inflation Behavior and shape) do not raise different questions of safety and effectiveness. The proposed device is substantially equivalent to the predicate device.

# VII Summary of non-clinical testing

## Performance testing

The following tests were performed in support of the substantial equivalence determination.

|  Test performed | Acceptance criteria  |
| --- | --- |
|  Balloon burst pressure (Constrained) | ≥400 psi  |
|  Balloon burst volume (Constrained) | ≥6 mL  |
|  Balloon Behavior (Unconstrained Balloon Compliance) Inflation | Inflated with 6 ml, the balloon diameter should be 18±2.0 mm, and the balloon working length should be 28.0±3.0 mm.  |
|  Tensile force | ≥15 N  |
|  Fatigue | There shall be no rupture or leakage after 20 times of inflation/deflation  |
|  Balloon deflation time limits | ≤3 seconds  |

## Biocompatibility testing

In accordance with ISO 10993-1, the Kyphoplasty Balloon Dilatation Catheters is classified as: Externally Communicating Device, Intact bone/tissue, Limited Contact (&lt; 24 hours).

The following tests were performed on the final product with a surface area of 41.08cm² as recommended:

Cytotoxicity ISO 10993-5:2009

Sensitization ISO 10993-10:2021

Irritation ISO 10993-23:2021

Acute Systemic toxicity ISO 10993-11:2017

Pyrogenicity USP&lt;151&gt;

## Sterilization and shelf life testing

- The product is sterilized by EO to maintain sterility assurance level (SAL) of 10^6. The sterilization method has been validated to ISO 11135, ISO 11138-1, ISO 11737-1 and ISO 11737-2, which has thereby determined the routine control and monitoring parameters.

According to ISO 10993-7, the devices are belongs to limited exposure to human body tissue/bone, so the limit of residual EO should be less than 4mg; the limit of residual ECH should be less than 9mg. The EO/ECH residue test report shows the EO residue is less than 4mg and ECH is less than 9mg.

- Transportation test per ASTM D4169
- Packaging validation per ISO 11607-1/-2

{10}

&gt; The 3 years shelf life of the device is determined based on stability study which includes real time aging test and simulated transportation test.
&gt; Bacterial Endotoxin Testing per USP-NF:2023 &lt;85&gt;

## VIII Clinical testing

N/A-No clinical test were conducted for this submission.

## IX Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate.

K252458

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K252458](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K252458)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
