← Product Code [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX) · K153548

# VS3 STEREOSCOPIC HIGH DEFINITION VISION SYSTEM (K153548)

_Visionsense, Ltd. · HRX · Feb 22, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K153548

## Device Facts

- **Applicant:** Visionsense, Ltd.
- **Product Code:** [HRX](/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX.md)
- **Decision Date:** Feb 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

The system is intended for viewing internal surgical sites during general surgical procedures, visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and large joint arthroscopic procedures.

## Device Story

VS3-ETV is a rigid metal stereoscopic endoscope; transmits high-definition visual data from surgical sites to external display. Operates via proximal sensor module on camera block; captures left/right images through optical transmission; fiber optic threads provide illumination. Used in OR by surgeons for visualization during general, neurological, spinal, and arthroscopic procedures. Provides stereoscopic view to assist surgical navigation and decision-making. Modification from predicate includes addition of 3 working channels for instrument passage/irrigation. System architecture, CCU, and display components remain identical to predicate VS3.

## Clinical Evidence

Bench testing only. Validation included reuse, cleaning, disinfection, and sterilization testing to ensure the addition of working channels did not negatively impact device processing or performance compared to the predicate.

## Technological Characteristics

Rigid metal shaft endoscope; stereoscopic optical transmission; fiber optic illumination. Dimensions: 4-5.5mm diameter, 175-300mm length. Field of view: 70°-95°. Working distance: 8mm-70mm. Connectivity: CCU, display monitors, and illumination sources. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- VS3 Stereoscopic High Definition Vision System, Model VS3 ([K141002](/device/K141002.md))

## Reference Devices

- Aesculap Minop System ([K983365](/device/K983365.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles incorporated into the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 22, 2016

Visionsense Ltd % Raymond Kelly Licensale Inc 57 Lazy Brook Rd Monroe, Connecticut 06468

Re: K153548

Trade/Device Name: VS3 Stereoscopic High Definition Vision System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, GWG Dated: December 25, 2015 Received: January 8, 2016

Dear Mr. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K153548

Device Name

VS3-ETV

Indications for Use (Describe)

The system is intended for viewing internal surgical sites during general surgical procedures, visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, and foraminotomy, and shoulder and large joint arthroscopic procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Visionsense VS3-ETV

## Submitter's Name, Address, Telephone Number,

Visionsense Ltd. 20 Hamagshimim Street, Petach Tikva 49348 Israel Phone: 972-3-9244339 Facsimile: +972-3-9244-335

#### Contact Person and Date Prepared

Raymond Kelly 57 Lazy Brook Rd Monroe, CT 06468 USA Phone: (203) 880-4091

Date Prepared: October 25, 2015

## Name of Device

VS3-ETV

Common or Usual Name / Classification Name Neurological Endoscope / Arthroscope

| Product Code / | Regulation Number   |
|----------------|---------------------|
| GWG / HRX      | 882.1480 / 888.1100 |

Review Panel Neurology / Orthopedic

Device Class: Class II

### Predicate Device

VS3 Stereoscopic High Definition Vision System, Model VS3 (K141002)

### Intended Use / Indications for Use

The system is intended for viewing internal surgical sites during general surgical procedures, visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and large joint arthroscopic procedures.

### Principles of Operation / Conditions of Use

The VS3 ETV endoscope is the same endoscope as the predicate VS3 with the addition of 3 working channels. The endoscope is designed and manufactured the same way. The endoscope is a rigid metal shaft that uses a proximal sensor module which is mounted onto a camera block on the proximal side of the endoscope (the handle). The endoscope uses stereoscopic images which are transmitted from the visual field at the distal end of the scope through an optical transmission (for left and right channels). The endoscope camera sensor module picks up the left and right images of the scene inside the body and delivers them to the external display. The endoscope incorporates fiber optic threads that transfer the light from the illumination unit to the surgical scene. The VS3 ETV endoscope contains 3 working channels. The endoscope utilizes the same principle of operation, software, system architecture, camera and visual viewing technologies, illumination sources, display monitors, carts, CCU and couplers as the predicate VS3.

{4}------------------------------------------------

## Technological Characteristics / Performance Data

| Standards<br>No. | Standards<br>Organization | Standards Title                                                                                                                                                                                              |
|------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 60601-1-2        | IEC                       | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>collateral standard: electromagnetic compatibility -<br>requirements and tests (Edition 3). |
| 60601-1          | IEC                       | Medical electrical equipment - Part 1: General requirements<br>for basic safety and essential performance.                                                                                                   |
| 60601-1-4        | IEC                       | Medical electrical equipment - Part 1: General requirements<br>for safety, Collateral standard: Programmable electrical<br>medical systems.                                                                  |

The subject device conforms to the following recognized standards:

Performance testing consisting of reuse validation, cleaning and disinfection validation, autoclave and sterilization were performed to demonstrate the additional working channels do not present changes in reuse handling and processing compared to the predicate device. No new biocompatibility assessments or electrical safety or EMC testing was required as the only modification to the VS3 system was the addition of 3 working channels.

| Feature/Parameter                  | Proposed VS3-ETV                                                                   | Predicate VS3 (K141002)                                                            |
|------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Able to function with no<br>camera | No                                                                                 | No                                                                                 |
| Working distance range             | 8mm - 70mm                                                                         | 8mm - 70mm                                                                         |
| Field Of View                      | 70° - 95°                                                                          | 70° - 95°                                                                          |
| Direction of view                  | 0º - 70º                                                                           | 0° - 70°                                                                           |
| Horizontal resolution              | >199 Ipf                                                                           | >199 Ipf                                                                           |
| Vertical resolution                | >199 lpf                                                                           | >199 lpf                                                                           |
| Identification                     | Support functionality to permit<br>the system to identify scope                    | Support functionality to permit<br>the system to identify scope                    |
| Irrigation Sheath                  | Support an optional irrigation<br>sheath with a standard Storz-<br>style connector | Support an optional irrigation<br>sheath with a standard Storz-<br>style connector |
| Endoscope diameter                 | 4 - 5.5 mm                                                                         | 4 - 5.5 mm                                                                         |
| Endoscope length                   | 175 - 300 mm (±5 mm)                                                               | 175 - 300 mm (±5 mm)                                                               |
| Depth of field                     | 7 - 30mm and 15 - 60mm                                                             | 7 - 30mm and 15 - 60mm                                                             |
| Working Channels                   | yes                                                                                | no                                                                                 |

#### Substantial Equivalence

### Working/Irrigation Channels Reference Device

K983365 (Aesculap Minop System) cleared December 1998 is used as reference device to show precedence of the irrigation and working channels.

#### Conclusions

The VS3-ETV is substantially equivalent to the predicate device. Performance testing demonstrates that the modified device performs substantially equivalent to the predicate device, and any differences in technological characteristics do not raise different questions of safety or efficacy compared to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K153548](https://fda.innolitics.com/submissions/OR/subpart-b%E2%80%94diagnostic-devices/HRX/K153548)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
