OMNICUT RESECTION BLADE

K130912 · Depuy Mitek, A Johnson & Johnson Company · HRX · Jun 14, 2013 · Orthopedic

Device Facts

Record IDK130912
Device NameOMNICUT RESECTION BLADE
ApplicantDepuy Mitek, A Johnson & Johnson Company
Product CodeHRX · Orthopedic
Decision DateJun 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The FMS Blades and Burrs are an accessory to the FMS Fluid Management Systems. FMS Blades and Burrs are intended to provide controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery.

Device Story

OMNICUT Resection Blade; accessory to FMS Fluid Management Systems; used in arthroscopic surgery. Device consists of stainless steel outer tube and rotating inner tube with plastic hubs; distal tips feature sharp edges for cutting. Operated by surgeon during arthroscopic procedures. Device facilitates mechanical removal of bone and soft tissue. Performance verified via bench testing comparing soft tissue/bone cutting and particulate generation against predicate devices. Lubricated design; modified cutting window configuration.

Clinical Evidence

Bench testing only. Studies included soft tissue cutting comparison, bone cutting comparison, and blade shed (particulate) testing against predicate K953695. Results demonstrated equivalent performance.

Technological Characteristics

Stainless steel outer tube; rotating inner tube; plastic hubs. Mechanical cutting/shaving/abrading principle. Lubricated. Dimensions/form factor consistent with arthroscopic shaver blades. No electronic components or software.

Indications for Use

Indicated for patients undergoing arthroscopic surgery requiring controlled cutting, burring, shaving, and abrading of bone and tissue.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130912 Page 1 of 2 ・ ## 510(k) SUMMARY . ## JUN 1 4 2013 ・ . | Submitter's<br>Name and<br>Address | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Tatyana Korsunsky<br>Regulatory Affairs Specialist<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA<br>Telephone: 508-828-3122<br>Facsimile: 508-977-6911<br>e-mail: tkorsuns@its.jnj.com | | Name of<br>Medical Device | Proprietary Name: OMNICUT™ Resection Blade<br>Common Name: Blade | | Substantial<br>Equivalence<br>Facility | The OMNICUT™ blades are substantially equivalent to:<br>K953695: Dyonics Disposable Arthroscopic Blades (Full Radius Bonecutter blade) K954465: FMS DUO® pump and shaver system (FMS blades) K041824: FMS NeXtra™ arthroscopic pump and shaver system (FMS blades) | | Device<br>Classification | Classification Name: Arthroscope<br>Classification Number: 21 CFR 888.1100 Class II<br>Product Code: HRX | | Device<br>Description | OMNICUT Blades are used with FMS™ Handpieces and Fluid Management Systems<br>during arthroscopic surgery. OMNICUT Blades consist of stainless steel outer tube<br>and a rotating inner tube connected to an inner and outer plastic hubs. Distal tips of<br>the tubes contain sharp edges to facilitate the cutting. | | Indications for<br>Use | The FMS Blades and Burrs are an accessory to the FMS Fluid Management Systems.<br>FMS Blades and Burrs are intended to provide controlled cutting, burring, shaving<br>and abrading of bone and tissue for use in arthroscopic surgery. | | Nonclinical<br>Testing | Design verification activities, such as soft tissue and bone tissue removal, as well as<br>blade particulate generation testing, were performed on OMNICUT blades and their<br>predicate devices. | | Safety and<br>Performance | While there are no required tests for arthroscopy blades, the following<br>preclinical studies were submitted to demonstrate substantial equivalence:<br>Blade soft tissue cutting comparison with K953695 predicate. Blade bone cutting comparison with K953695 predicate. Blade shed testing comparison with K953695 predicate. The OMNICUT blades passed all tests and showed equivalent performance. | {1}------------------------------------------------ Results of performance and safety testing have demonstrated that the proposed devices are suitable for their intended use. Based on the indications for use, technological characteristics, and comparison to predicate devices, the proposed OMNICUT Blades have been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act. Comparison to The proposed OMNICUT Blades have the same intended use and are similar in the predicate design and materials to the predicate Smith & Nephew's Full Radius Bonecutter devices blades (K953695) and Mitek's FMS blades (K954465/ K041824). However, the proposed OMNICUT blades have a modified cutting window configuration and are provided lubricated. The proposed OMNICUT device can cut both bone and soft tissue, similar to a predicate Smith & Nephew's Full Radius Bonecutter blade. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 14, 2013 DePuy Mitek % Ms. Tatyana Korsunsky Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts 02767 Re: K130912 Trade/Device Name: OMNICUT Resection Blade Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: April 11, 2013 Received: April 12, 2013 Dear Ms. Korsunsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Tatyana Korsunsky comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. FOR # Peter Dinkumm -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation. Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### INDICATIONS FOR USE 510(k) Number (if known): K130912 Device Name: OMNICUT® Resection Blade Indications for Use: The FMS Blades and Burrs are an accessory to the FMS Fluid Management Systems. FMS Blades and Burrs are intended to provide controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery. | Prescription Use | AND/OR | |------------------|--------| | ↓ | | (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 | DSD—DIVISION SIGN-OFF | Joshua C. | |------------------------------|-----------| | Division of Surgical Devices | Nipper-S | | 510(k) Number: | K130912 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...