← Product Code [MVN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MVN) · K150630

# Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear, Eye Secret 55 UV Aspheric ( methafilcon A) Soft (hydrophilic) Contact lens for Daily Wear (K150630)

_Yung Sheng Optical Co., Ltd. · MVN · Jun 15, 2015 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MVN/K150630

## Device Facts

- **Applicant:** Yung Sheng Optical Co., Ltd.
- **Product Code:** [MVN](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MVN.md)
- **Decision Date:** Jun 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for the correction of refractive ametropia (myopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye. The eye care professionals may prescribe the lens for single use daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.

## Device Story

Soft hydrophilic contact lens; polymacon material; cast-molded; aspheric design. Corrects myopia; provides UV protection. Prescribed by eye care professionals for daily wear or frequent replacement. Supplied in sterile isotonic phosphate buffered saline with sodium hyaluronate and trehalose. Patient wears lens directly on eye; requires chemical or hydrogen peroxide disinfection for planned replacement. Benefits include refractive correction and UV radiation blocking.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing including acute systemic injection, ocular irritation, cytotoxicity (ISO 10993-5), extractables, light transmittance, refractive index, water content, and mechanical properties.

## Technological Characteristics

Material: Polymacon (HEMA/EGDMA copolymer). UV-absorbing monomer included. Tint: C.I. Reactive Blue No.4. Water content: 38%. Refractive index: 1.440. Manufacturing: Cast molding. Packaging: Sterile isotonic phosphate buffered saline with sodium hyaluronate and trehalose. Non-electronic, non-software device.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Eye Secret 38 UV Aspheric Color (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear ([K143052](/device/K143052.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 15, 2015

Yung Sheng Optical Co., Ltd. Wen-Han Chen, Regulatory Affairs Supervisor 3F-1, No.6, Jhongke Road, Daya District, Taichung City, 42881 Taiwan

Re: K150630

> Trade/Device Name: Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: April 10, 2015 Received: April 16, 2015

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -A

for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose. and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150630

### Device Name

Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear

### Indications for Use (Describe)

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for the correction of refractive ametropia (myopia) in aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye.

The eye care professionals may prescribe the lens for single use daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be dising a chemical or hydrogen peroxide disinfecting system.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

- 1. Type of Submission: Traditional 510(k)

| 2. | Submitter:                         | Yung Sheng Optical Co., Ltd.                                       |  |
|----|------------------------------------|--------------------------------------------------------------------|--|
|    | Address:                           | 3F-1, No.6, Jhongke Road, Daya District, Taichung                  |  |
|    |                                    | City 42881 Taiwan                                                  |  |
|    | Manufacturing<br>facility:         | No.8, Keya 2nd Road, Daya District, Taichung City<br>42881, Taiwan |  |
|    | Phone:                             | (04) 25658384 #156 、157                                            |  |
|    | Fax:                               | (04) 25658387                                                      |  |
|    | Contact:                           | Wen-Han Chen / Tsung-Jen Yeh                                       |  |
|    | Date prepared:                     | April 10, 2015                                                     |  |
|    | Establishment Registration Number: | 3004021238                                                         |  |

#### 3. Identification of the Device

| Proprietary/Trade name: | Eye Secret 38 UV Aspheric (polymacon) Soft<br>(hydrophilic) Contact Lens for Daily Wear |                                     |
|-------------------------|-----------------------------------------------------------------------------------------|-------------------------------------|
| Common Name:            | Contact Lens                                                                            |                                     |
| Classification Name:    | Lenses, Soft Contact, Daily Wear                                                        |                                     |
| Device Classification:  | II                                                                                      |                                     |
| Regulation Number:      | 886.5925                                                                                |                                     |
| Panel:                  | Ophthalmic                                                                              |                                     |
| Product Code:           | LPL for Lenses, Soft Contact, Daily Wear                                                | MVN for Lens, Contact, (Disposable) |

#### 4. Identification of the Predicate Device

| Predicate Device                        | Eye Secret 38 UV Aspheric Color (polymacon)     |
|-----------------------------------------|-------------------------------------------------|
| Name:                                   | Soft (hydrophilic) Contact Lens for Daily Wear; |
| Manufacturer:                           | Yung Sheng Optical Co., Ltd.                    |
| 510(k) Number or Clearance Information: | K143052                                         |

{4}------------------------------------------------

## 5. Intended Use and Indications for Use of the subject device

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for the correction of refractive ametropia (myopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye.

The eye care professionals may prescribe the lens for single use daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.

## 6. Device Description

The Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is manufactured by using cast molding method. The lens material, polymacon, is a random copolymer of 2-hydroxyethyl methacrylate (HEMA) crosslinked with ethylene glycol dimethacrylate (EGDMA). A UV absorbing monomer is used to block UV radiation. The average transmittance characteristics are less than 5% in the UVB range of 280 to 315 nm and less than 50% in the UVA range of 316 to 380 nm. The lenses are tinted blue for visibility purposes with the color additives, C.I. Reactive Blue No.4.

The lenses are available as aspheric lenses. Each finished lens is supplied in a plastic blister container with a sterile isotonic phosphate buffered saline solution containing sodium hyaluronate and trehalose wetting agents.

## 7. Summary of Clinical Study

Polymacon lenses have been used widely. Their safety and effectiveness have been well documented and cleared by FDA "Eye Secret 38 UV Aspheric Color (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear; Eye Secret 38 UV Aspheric Color (Polymacon) 1-Day Soft(Hydrophilic) Contact Lens (K143052)" submitted by Yung Sheng Optical Co., Ltd. are examples.

{5}------------------------------------------------

Clinical study for the subject device is not required for the premarket notification as the USAN name and process are the same as the above mentioned predicate devices.

## 8. Non-Clinical Testing

A series of preclinical testing were performed to demonstrate the safety and effectiveness of the Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear. The results of all testing demonstrated that the safety and effectiveness of the Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear are equivalent to the Eye Secret 38 UV Aspheric Color (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear (K143052). The following tests were conducted as recommended by the FDA Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, February 27, 1997:

- · Toxicity
	- 1. Acute Systemic Injection Study: The lens material meets the requirements of the systemic injection test and is considered non-toxic.
	- 2. White Rabbit Ocular Irritation Test: Ocular irritation test was performed and produced no ocular irritation.
	- 3. Cytotoxicity Test: The test article meets the requirements of ISO 10993-5.
- Extractables (Leachability) .
- . Finished Lens Parameters
- Light Transmittance .
- . Refractive Index
- . Water Content
- Shelf-life .
- . Mechanical Properties Comparative Testing
- Specific Gravity Comparative Testing
- Physical Compatibility Test with Contact Lens Care Solution and Packaging . Solution

The results of the non-clinical testing demonstrated that the Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is substantially equivalent to the predicate device.

{6}------------------------------------------------

### 9. Substantial Equivalence Determination

The information presented in this submission establishes the substantial equivalence of the subject device "Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear" to predicate devices. The following table summarized the substantial equivalence comparison information.

|                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K143052                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                     | Subject Device                                                                             | K143052                                                                                                                                                                                                                                                                   |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name             | Eye Secret 38 UV Aspheric<br>(polymacon) Soft (hydrophilic)<br>Contact Lens for Daily Wear                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Eye Secret 38 UV Aspheric Color<br>(polymacon) Soft (hydrophilic) Contact<br>Lens for Daily Wear                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Device Name         | Eye Secret 38 UV Aspheric<br>(polymacon) Soft (hydrophilic)<br>Contact Lens for Daily Wear | Eye Secret 38 UV Aspheric Color<br>(polymacon) Soft (hydrophilic) Contact<br>Lens for Daily Wear                                                                                                                                                                          |
| Intended Use            | The Eye Secret 38 UV is indicated<br>for the correction of refractive<br>ametropia (myopia) in aphakic and<br>not-aphakic persons with<br>non-diseased eyes. The lenses may<br>be worn by person who exhibit<br>astigmatism of 2.00 diopters or less<br>that does not interfere with visual<br>acuity. The Eye Secret 38 UV helps<br>protect against transmission of<br>harmful UV radiation to the cornea<br>and into the eye. The eye care<br>professionals may prescribe the<br>lens for single use daily disposable<br>or daily wear in a Frequent<br>Replacement Program. As<br>prescribed for planned<br>replacement, the lens should be<br>disinfected using a chemical or<br>hydrogen peroxide disinfecting<br>systems. | The Eye Secret 38 UV Color is indicated<br>for the correction of ametropia<br>(myopia) in aphakic and not-aphakic<br>persons with non-diseased eyes. The<br>lenses may be worn by persons who<br>exhibit astigmatism of 2.00 diopters or<br>less that does not interfere with visual<br>acuity. Eye Secret 38 UV Color helps<br>protect against transmission of<br>harmful UV radiation to the cornea<br>and into the eye. For Eye Secret 38 UV<br>Color, the eye care professionals may<br>prescribe the lens for single use daily<br>disposable or daily wear in a Frequent<br>Replacement Program. As prescribed<br>for planned replacement, the lens<br>should be disinfected using a chemical<br>or hydrogen peroxide disinfecting<br>systems. | UV-A                | < 50 %                                                                                     | < 50 %                                                                                                                                                                                                                                                                    |
| Material                | Polymacon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Polymacon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | UV-B                | < 5 %                                                                                      | < 5 %                                                                                                                                                                                                                                                                     |
| Manufacturing<br>Method | Cast Molded                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Cast Molded                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Refractive<br>Index | 1.440 (± 0.005 nd)                                                                         | 1.440 (± 0.005 nd)                                                                                                                                                                                                                                                        |
| Water Content           | 38 %                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 38 %                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Base Curve          | 7.85 ~ 10.00 (± 0.2 mm)                                                                    | 8.60 (± 0.20 mm)                                                                                                                                                                                                                                                          |
| Powers                  | -0.50 D ~ -20.00 D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | -0.00 ~ -20.00 D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Diameter            | 12.00 ~ 15.00 (± 0.2 mm)                                                                   | 14.00 (± 0.20 mm)                                                                                                                                                                                                                                                         |
| Light<br>Transmittance  | 95 % (± 5 %)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 95 % (± 5 %)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Tint                | C.I. Reactive Blue #4 for handling<br>purpose                                              | C.I. Reactive Blue #4 for handling<br>purpose and printed with an<br>intermittent coating a combination of<br>the following approved pigments:<br>Copper Phthalocyanine, Reactive Black<br>5, Iron Oxide, Titanium Dioxide,<br>Carbazole Violet and D&C Yellow No.<br>10. |
| Packaging<br>solution   | Phosphate buffered saline solution<br>containing sodium hyaluronate and<br>trehalose                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Phosphate buffered saline solution<br>containing sodium hyaluronate and<br>trehalose                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                     |                                                                                            |                                                                                                                                                                                                                                                                           |

{7}------------------------------------------------

Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear

## 10. Conclusion

After analyzing the bench tests and safety testing data, it can be concluded that the Eye Secret 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MVN/K150630](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/MVN/K150630)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
