The Fentec Laser System for Capsulotomy is indicated for anterior capsulotomy during cataract surgery.
Device Story
Femtec Laser System for Capsulotomy is a precision ophthalmic surgical laser used in cataract surgery. Device uses ultrashort, high-frequency laser pulses to perform anterior capsulotomy via micro-photodisruption of tissue. System includes a sterile, disposable Patient Interface (contact lens and suction clip) to fix the eye and provide a reference surface for depth control. Surgeon operates the system in an OR setting, using pre-programmed patterns to create incisions of predetermined diameter and height. Output is a continuous capsular incision created by scanning thousands of individual pulses. Clinical benefits include improved circularity and centration of the capsulorhexis compared to manual techniques, facilitating intraocular lens placement.
Clinical Evidence
Prospective clinical trial evaluated anterior capsulotomy performance. Results: 100% success rate in capsulotomy completion; no radial tears observed; intraocular lenses successfully centered in all eyes. Compared to manual capsulorhexis, the Femtec system demonstrated significantly higher circularity and improved pupil-centered outcomes. Bench testing in porcine eyes and optical phantoms confirmed accuracy and reproducibility of incision size, shape, and depth.
Technological Characteristics
Ophthalmic femtosecond laser; ultrashort, high-frequency pulses; photodisruption mechanism. Includes sterile, disposable Patient Interface (contact lens/suction clip) for eye fixation and depth control. System utilizes pre-programmed scanning patterns for incision geometry. Class II device.
Indications for Use
Indicated for patients undergoing anterior capsulotomy during cataract surgery.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
K090633 — LENSAR LASER SYSTEM · Lensar, Inc. · May 13, 2010
K141379 — VICTUS FEMTOSECOND LASER PLATFORM · Technolas Perfect Vision GmbH · Jan 16, 2015
K252682 — LenSx Laser System (8065000944) · Alcon Laboratories, Inc. · Sep 24, 2025
K122829 — LENSAR LASER SYSTEM - FS 3D FOR CORNEAL INCISIONS · Lensar, Inc. · Dec 3, 2012
K094052 — LENSX 550 LASER SYSTEM. MODEL 550 · Lensx Lasers, Inc. · Apr 23, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Technolas Perfect Vision GmbH<br>Messerschmittstr. 1 +3<br>80992 Munich, Germany<br>Tel: 49 174 3494062 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Betty M. Johnson<br>Sr. Manager, Regulatory Affairs<br>Technolas Perfect Vision GmbH<br>PO Box 17190<br>Anaheim, CA 92817<br>Ph: 714-273-8254<br>Fax: 714-855-2794<br>Email: b.johnson@technolaspv.com |
| TRADE NAME: | Femtec Laser System for Capsulotomy |
| COMMON NAME: | Ophthalmic Laser |
| CLASSIFICATION<br>NAME: | Ophthalmic Femtosecond Laser |
| DEVICE<br>CLASSIFICATION: | Class II |
| PRODUCT CODE | OOE (Ophthalmic Femtosecond Laser) |
| PREDICATE<br>DEVICES: | LenSx 550 Laser System, LensAR Laser System for Anterior<br>Capsulotomy |
#### PREDICATE DEVICES:
| 510(K) NUMBER | PRODUCT TRADE NAME | MANUFACTURER |
|---------------|-------------------------------------------------|-------------------------------|
| K082947 | LenSx 550 Laser System | LenSx Lasers / Alcon |
| K090633 | LensAR Laser System for<br>Anterior Capsulotomy | LensAR, Inc. |
| K060372 | IntraLase FS Laser | Abbott Medical Optics |
| K040204 | Zyoptix XP Microkeratome | Technolas Perfect Vision GmbH |
| K022560 | Moria 2 Microkeratome | Moria |
{1}------------------------------------------------
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
## Description of the Device Subject to Premarket Notification:
The Femtec Laser System for Capsulotomy is a precision ophthalmic surgical laser indicated for use in patients undergoing anterior capsulotomy during cataract surgery. Capsular dissection is achieved through precise individual micro-photodisruption of tissue, measuring a few microns in diameter, created by tightly focusing ultrashort laser pulses into the targeted capsular tissue. Surgical effects are produced by scanning thousands of individual pulses, producing a continuous incision. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location. Pre-programmed patterns produce capsular resections of predetermined diameter and height.
Laser pulses are delivered through a sterile (disposable) Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam.
## Indication for Use:
The Femtec Laser System for Capsulotomy is indicated for anterior capsulotomy during cataract surgery. The proposed indication for use is identical to the previously cleared LenSx 550 Laser (K082947) and LensAR Laser System for Anterior Capsulotomy (K090633).
## Technical Characteristics Comparison:
The Femtec Laser System for Capsulotomy mode of operation and technology are very similar to the previously cleared LenSx 550 Laser (K082947) and LensAR Laser System for Anterior Capsulotomy (K090633). The Femtec Laser System for Capsulotomy utilizes ultrashort, high frequency laser pulses to create photodisruption in the anterior capsule, as do the predicate devices. The Femtec Laser system utilizes a curved patient interface as does the LenSx 550 laser predicate, while the LenAR Laser System utilizes a fluid-fluid interface. However, both patient interface configurations have the same function, which to provide a reference surface through which the laser pulses are delivered to the anterior capsule and are therefore substantially equivalent.
## Brief Summary of Preclinical and Clinical Performance Test Results:
Technolas Perfect Vision developed the Femtec Laser system for use in anterior capsulotomy. Pre-clinical and clinical testing has included bench testing and a prospective clinical trial.
#### Performance Data:
Testing and analyses included accuracy and reproducibility of capsulotomy incisions in porcine eyes, as well as in plastic and agar gel optical phantoms. The data demonstrated that the Femtec Laser System produces anterior capsulotomies that are accurate and predictable in size, shape, and centration over a range of depths. The capsular edge created by the Femtec Laser system is also at least as smooth as that created during anterior capsulotomy.
{2}------------------------------------------------
#### Clinical Trial:
A clinical trial of the Femter Laser System was conducted to evaluate the performance of the laser system in the creation of anterior capsulotomy during cataract surgery. Anterior capsulotomy was successfully performed in all eyes using the Femtec Laser System; i.e., the capsulotomy was complete in all eyes, with no radial tears observed intraoperatively or postoperatively, and with an intraocular lens placed in each capsular bag. Postoperatively, the course of follow-up in the study population was unremarkable. The intraocular lens was centered in all study eyes, and no posterior capsule tears were observed. All capsulotomies were judged to be well-centered by the surgeon using visual inspection in the operating microscope. While shown to be equally safe as manual capsulorhexis, the Femtec Laser System anterior capsulotomy procedure produced significantly more circular capsulorhexis than the manual technique, and significantly more pupil-centered capsulorhexis compared to manual capsulotomy.
#### Basis for Determination of Substantial Equivalence:
The technological and performance characteristics of the Femtec Laser System for Capsulotomy are substantially equivalent to the technological and performance characteristics of the LenSx 550 Laser System and the LensAR Femtosecond Laser, cleared under K082947 and K090633, respectively, for anterior capsulotomy during cataract surgery.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Technolas Perfect Vision GMBH c/o Ms. Betty M. Johnson Sr. Manager, Regulatory Affairs, Americas PO box 17190 Anaheim, CA 92817
NOV 2 2 2 2011
Re: K110427
Trade/Device Name: Femtec Laser System for Capsulotomy Regulation Number: 21 CFR 886.4390 Regulation Name: Opthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: November 7, 2011 Received: November 8, 2011
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Ms. Betty M. Johnson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kevin Alexander
for
Mälvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _ & 1 / 0 4 L 7
- **
Device Name(s): Femtec Laser System for Capsulotomy
Indications for Use:
The Fentec Laser System for Capsulotomy is indicated for anterior capsulotomy during cataract surgery.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page of
K11042.7
SIMIA Numb TECHNOLAS PERFECT VISION GMBH FEMTEC LASER SYSTEM FOR CAPSULOTOMY
PAGE 18 PREMARKET NOTIFICATION
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.