The Lenstec IOL Injection System is intended for use in implantation of the Lenstec Softec Posterior Chamber Intraocular Lens into the capsular bag following extracapsular extraction.
Device Story
System facilitates implantation of foldable Softec intraocular lenses (IOLs) into capsular bag during cataract surgery. Comprises reusable titanium syringe-based injector and single-use disposable cartridges (LC604220 or LC604240) with silicone cushion. Operated by ophthalmologist in clinical setting. Injector delivers folded IOL through small incision (2.2mm or 2.8mm) into eye. Reusable components are autoclavable; disposable components provided sterile. System ensures controlled delivery of IOL, minimizing incision size and potential trauma to ocular tissues.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on all contact materials. System performance verified using Softec intraocular lenses. No clinical data required.
Technological Characteristics
System components: reusable titanium injector (autoclavable), single-use lubricated polypropylene cartridge, and medical-grade polypropylene/silicone cushion tip. Manual mechanical actuation. Dimensions vary by cartridge for 2.2mm or 2.8mm incision sizes. No software or electronic components.
Indications for Use
Indicated for implantation of Lenstec Softec Posterior Chamber Intraocular Lens into the capsular bag following extracapsular extraction.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Medicel MultiJect Injector for IOLs and MicroGlide Cartridge (K040837)
STAAR Surgical Foam Tip™ Injector System (K980696)
Lenstec Intraocular Lens (IOL) Injection System (K050638)
Related Devices
K231466 — RxSight Insertion Device · Rxsight, Inc. · Jun 12, 2023
K050638 — LENSTEC INTRAOCULAR LENS (IOL) INJECTION SYSTEM FOR TETRAFLEX INTRAOCULAR LENSES · Lenstec, Inc. · May 31, 2005
K961242 — AMO PHACOFLEX II INTERSECTION SYSTEM · Allergan, Inc. · Jun 17, 1996
K172228 — pioli IOL Delivery System · Ast Products, Inc. · Dec 15, 2017
K122848 — LENSTEC LC INJECTION SYSTEM · Lenstec, Inc. · Oct 18, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Ko60533
Image /page/0/Picture/1 description: The image shows the logo for "LENS-TEC". The logo is in a bold, blocky font with a black outline and a white fill. The letters are arranged in two rows, with "LENS" on the top row and "TEC" on the bottom row. To the right of the word "TEC" is a circle with a white outline and a black fill. The letters are slightly distorted, giving the logo a unique and modern look.
APR 1 4 2006
2870 Scherer Drive · Suite 300 · St. Petersburg, FL 33716 · USA Telephone: (727) 571-2272 · Fax: (727) 571-1792 · E-mail: lenstec@lenstec.com
### 510(K) Summary
## Date Summary Prepared: February 12, 2006
## Lenstec, Inc. 510(k) Premarket Notification Submission Lenstec Injector System for Softec Intraocular Lenses 510(K) Premarket Notification Summary
Trade/Device Name: Lenstec Injector System for Softec Intraocular Lenses Regulation Number: 21 CFR 886.1850 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS
#### Labeling:
Federal (United States) Law restricts this device to sale by or on the order of a physician.
- 1. Applicant Information:
- a. Name: Lenstec, Inc.
- b. Address: 2870 Scherer Drive, Suite 300 St. Petersburg, FL 33716 Telephone Number: 727 571-2272 Fax: 727 571-1792
- c. Contact Person: Luis A. Ruiz, Vice President, Quality Assurance
- d. Address: 2870 Scherer Drive, Suite 300 St. Petersburg, FL 33716 Telephone Number: 727 571-2272 Fax: 727 571-1792 Email: Iruiz@lenstec.com
2. Name of Device:
- a. Trade Name: Lenstec Injector System for Softec Intraocular Lenses
- b. Common Name: Intraocular lens guide.
- c. Classification Name: Folders and injectors, intraocular Iens (IOL) (MSS, 886.4300)
# 000018
{1}------------------------------------------------
- 3. Substantially Equivalent legally-marketed devices:
- Medicel MultiJect Injector for IOLs and MicroGlide Cartridge K040837. a. June 17, 2004.
- b. STAAR Surgical Foam Tip™ Injector System K980696, Sept 11, 1998.
- c. Lenstec Intraocular Lens (IOL) Injection System K050638, May 31, 2005.
- 4. Device Description:
The system consists of the following components:
| Cartridge | Injector | Nominal<br>Incision Size (mm) | Used with Softec<br>Intraocular Lenses<br>Power Range (D) |
|------------|----------|-------------------------------|-----------------------------------------------------------|
| LC604220 | I 9000 | 2.8 | 26.5D to 42.0D |
| LC604240 | I 9007 | 2.2 | -5.0D to 26.0D |
| W/LS604500 | | | |
Two types of injectors are provided: both injectors are syringe based and are reusable and autoclavable. The cartridges/silicone cushion are single-use and provided sterile.
- 5. Use:
The Lenstec Injector System is to be used solely to insert the foldable Softec intraocular lenses manufactured by Lenstec and is used in conjunction vitte the Medicel cartridges, MicroGlide LC604220 and ViscoGlide LC604240 (w/ilicone Cushion LS604500.
- 6. Indications for use:
The Lenstec IOL Injection system is intended for use in implantation of the Lenstec Softec Posterior Chamber Intraocular Lens into the capsular bag following extracapsular extraction.
### 7. Technological characteristics:
The primary system has two major components: a reusable injector and a disposable cartridge LC604220. A secondary system also has three major components: a reusable injector and a disposable cartridge LC604240 and includes the use of a Silicone Cushion (LS604500) supplied with the LC604240 cartridge.
- The injector is manufactured of titanium and can be autoclaved. a.
- The cartridge is manufactured of lubricated polypropylene and is Single Use. b.
- The silicone tip is manufactured of medical grade polypropylene body and a c. silicone tip and also is Single Use.
- 8. Performance data:
- a. Non-clinical tests
All contact materials have been tested for biocompatibility. The system was tested with Softec intraocular lenses.
{2}------------------------------------------------
- 9. Clinical tests: Not required
- 10. Conclusions
The Lenstec Injector System with Medicel cartridges are substantially equivalent in safety and efficacy to the legally marketed predicate device.
Luis A. Ruiz
Vice President, Quality Assurance
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Lenstec, Inc. Mr. Luis Ruiz Vice President, Quality Assurance 2870 Scherer Dr. Suite 300 St. Petersburg, FL 33716
APR 1 4 2006
Re: K060533
Trade/Device Name: Lenstec Intraocular Lens (IOL) Injection System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guides Regulatory Class: Class I Product Code: MSS Dated: February 12, 2006 Reccived: February 28, 2006
Dear Mr. Ruiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Luis Ruiz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eychus - M.D.
Malvina B. Eydelman, M.D Division Director Division of Ophthalmic and Far, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Lenstec Intraocular Lens Injection System
Indications For Use: The Lenstec IOL Injection System is intended for use in implantation of the Lenstec Softec Posterior Chamber Intraocular Lens into the capsular bag following extracapsular extraction.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Jivision Sign-Off
rivision of Ophthalmic Ear Jose and Throat Devises
510(k) Number ________________________________________________________________________________________________________________________________________________________________
Page 1 of _ 1 ________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.