Browse hierarchy Ophthalmic (OP) Subpart E — Surgical Devices 21 CFR 886.4300 Product Code KYB K063155 — MONARCH III IOL DELIVERY SYSTEM
MONARCH III IOL DELIVERY SYSTEM
K063155 · Alcon Manufacturing, Ltd. · KYB · Mar 27, 2007 · Ophthalmic
Device Facts
Record ID K063155
Device Name MONARCH III IOL DELIVERY SYSTEM
Applicant Alcon Manufacturing, Ltd.
Product Code KYB · Ophthalmic
Decision Date Mar 27, 2007
Decision SESE
Submission Type Traditional
Regulation 21 CFR 886.4300
Device Class Class 1
Attributes Therapeutic
Indications for Use
The MONARCH® III IOL Delivery System is used for implantation of ALCON® qualified ACRYSOF® intraocular lenses into the eye following cataract removal.
Device Story
Monarch III IOL Delivery System is a surgical instrument for intraocular lens (IOL) implantation. System comprises a disposable D-cartridge and reusable H4 handpiece. Device functions by folding and injecting Alcon AcrySof IOLs into the eye during cataract surgery. Operation involves manual manipulation by a surgeon to guide the lens through the cartridge nozzle into the eye. Modification from Monarch II system includes updated nozzle tip sizing while maintaining original inner lumen design and lens guiding mechanisms. Device is used in clinical settings by ophthalmologists. Benefits include controlled, consistent delivery of IOLs during cataract procedures.
Clinical Evidence
Bench testing only. The device conforms to the FDA Intraocular Lens Guidance Document (Draft 1999) Section C, Mechanical Testing, specifically folding and injection testing.
Technological Characteristics
System consists of a disposable cartridge (D) and reusable handpiece (H4). Utilizes progressive folding mechanism for IOL delivery. Mechanical device; no energy source or software. Designed for use with Alcon AcrySof IOLs.
Indications for Use
Indicated for implantation of Alcon AcrySof intraocular lenses into the eye following cataract removal.
Regulatory Classification
Identification An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Monarch® II IOL Delivery System (K003768 )
Monarch® II IOL Delivery System (K001157 )
Related Devices
K112977 — MONARCH III IOL DELIVERY SYSTEM, MONARCH III C CARTRIDGE, MONARCH III HANDPIECE (H4) · Alcon Research, Ltd. · Mar 27, 2012
K073010 — ROYALE AE 9036 LSP INJECTOR · Asico, LLC · Nov 7, 2007
K981571 — MONARCH IOL DELIVERY SYSTEM · Alcon Laboratories · Jul 9, 1998
K101774 — MONARCH III IOL DELIVERY SYSTEM; MONARCH III P CATRIDGE; MONARCH III HANDPIECE (H4) · Alcon Manufacturing, Ltd. · Oct 13, 2011
K182965 — BL-Cart IOL Delivery Cartridge · Ast Products, Inc. · Mar 29, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
## 5. 510(K) SUMMARY
# MAR 2 7 2007
### 510(k) Summary
This summary document is being prepared in accordance with section 21 CFR 807.92(c).
The submitter of the 510(k) is:
Catherine G. Goble Senior Manager, Regulatory Affairs Alcon Research, Ltd. 6201 South Freeway Fort Worth, TX 76134 Phone: (817) 551-6816 Fax: (817) 551-4630
Date Summary Prepared: October 9, 2006
Device Subject to this 510(k):
| Trade Name: | Monarch® III IOL Delivery System<br>(D Cartridge and H4 Handpiece) |
|----------------------|--------------------------------------------------------------------|
| Common Name: | Intraocular Lens Guide |
| Classification Name: | 21 CFR 886.4300 |
### 1. Predicate Devices:
The legally marketed device(s) to which we are claiming substantial equivalence are:
| 510(k) Number | Device |
|---------------|----------------------------------------------------|
| K003768 | Monarch® II IOL Delivery System (A & C Cartridges) |
| K001157 | Monarch® II IOL Delivery System (B Cartridge) |
### 2. Device Description:
The Monarch© III IOL Delivery System is a modification to the previously cleared Alcon Monarch(r) II IOL Delivery System (K003768) consisting of a disposable cartridge and reusable handpiece. The Monarch II cartridges are designated as A, B, and C, and feature various nozzle tip sizes that correspond to the various AcrySof® IOL models. As with the cleared cartridges, the
{1}------------------------------------------------
Monarch© III IOL Delivery System presented in this submission utilizes a one-piece disposable cartridge (designated as D) with progressive folding in combination with a reusable handpiece (H4). This new cartridge possesses the same basic inner lumen design and lens guiding mechanisms as the existing Monarch*) II A, B, and C cartridges, but differs in nozzle tip sizing.
#### 3. Indications for Use:
The MONARCH® III IOL Delivery System is used for implantation of ALCON® qualified ACRYSOF® intraocular lenses into the eye following cataract removal.
### 4. Brief Summary of Nonclinical Test and Results:
Alcon has provided a Statement of Conformance (Attachment J) that the Monarch® III IOL Delivery System Cartridge D and Handpiece H4 intended to be introduced, will conform in all respects with the requirements set forth in the FDA Intraocular Lens Guidance Document (Draft released 10/14/1999) Section C. Mechanical Testing, 5. Folding and Injection Testing prior to marketing the product.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alcon Research, Ltd. Ms. Catherine G. Goble Senior Manager, Regulatory Affairs 6201 South Freeway Fort Worth, TX 76134
Re: K063155
Trade Name: · Monarch® III IOL Delivery System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: KYB Dated: February 28, 2007 Received: March 1, 2007
MAR 2 7 2007
Dear Ms. Goble:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Ms. Catherine G. Goble
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
M.B. Eichler, M.D.
Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 3 -- Ms. Catherine G. Goble
# Indications for Use
510(k) Number (if known): K063155
Device Name: Monarch III IOL Delivery System
Indications For Use:
The MONARCH® III IOL Delivery System is used for implantation of
ALCON® qualified ACRYSOF® intraocular lenses into the eye
following cataract removal.
Prescription Use YES AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )
Concurrence of CDRH, Office of Device Evaluation (ODE)
Keria Alexander
(Division Sign-Off)
**Division of Ophthalmic Ear
Nose and Throat Devises**
510(k) Number K6/3155
Page 1 of 1