AMB101 SYSTEM, AMBLYZ, LCG SYSTEM, XPAND

K132042 · X6d, Ltd. · HOY · Apr 28, 2014 · Ophthalmic

Device Facts

Record IDK132042
Device NameAMB101 SYSTEM, AMBLYZ, LCG SYSTEM, XPAND
ApplicantX6d, Ltd.
Product CodeHOY · Ophthalmic
Decision DateApr 28, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4750
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Amblyz™ System is an electronic shutter module intended to provide occlusions of the eye in an intermittent fashion.

Device Story

Amblyz™ is an electronic shutter eyewear system for intermittent occlusion of the sound eye to treat amblyopia. Device consists of spectacle frame with liquid crystal (LC) lenses, microcontroller, battery, and charging port. Optician performs one-time setup using a programming button to select the eye for occlusion. During operation, LC lens toggles between transparent (OFF) and opaque (ON) states via electronic control, providing fixed 30-second occlusion/30-second transparent cycles. Device is worn by patient; operation is autonomous after initial setup. LED indicator provides battery status. System benefits patient by automating occlusion therapy, ensuring consistent treatment delivery without user tampering. Healthcare providers use the device to manage amblyopia treatment regimens.

Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1, IEC 60601-1-2), software verification/validation, lens impact resistance, drop testing, climatic testing, cosmetic/sweat/cleaning resistance, thermal shock, and biocompatibility (ISO 10993-5). All tests met acceptance criteria.

Technological Characteristics

Liquid crystal shutter module; spectacle frame made of biocompatible plastics (PC Makrolon 2458, Texin RXT50D, Elastosil). Powered by internal battery; rechargeable via micro-USB. Fixed 30s/30s occlusion cycle. Dimensions: 129x129x40mm; weight 36.4g. Software-controlled via onboard firmware.

Indications for Use

Indicated for patients requiring intermittent occlusion of the eye for therapeutic purposes. No specific age or gender restrictions provided.

Regulatory Classification

Identification

An ophthalmic eye shield is a device that consists of a plastic or aluminum eye covering intended to protect the eye or retain dressing materials in place.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## APR 2 8 2014 #### 510(K) SUMMARY #### 510(K) Number K132042 | Applicant's Name: | X6D Ltd.<br>195, Arch. Makariou III<br>Neocleous House<br>3030 Limassol, Cyprus<br>Tel: +357 25 110000<br>Fax: +357 25 110010 | |-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Neta Landman<br>Duet Medical Consulting Ltd.<br>21 Hanafa Street<br>Zur- Moshe 42810, ISRAEL<br>P.O.Box 536<br>T +972 54 443 6180<br>F +972 153 52 833 7207<br>Email: neta.l@duet-medical.com | | Date Prepared: | 05 November 2013 | | Trade Name: | Amblyz™ | | Classification: | Classification Name: Shield, eye, ophthalmic<br>Medical Specialty: Ophthalmic<br>Product Code: HOY<br>Device Class: I<br>Regulation Number: 886.4750<br>Panel: Ophthalmic | #### Predicate Devices: - LCG (Liquid Crystal Glasses) System; cleared under K050856; Product Code . HOY; Regulation Number 886.4750 #### Intended Use / Indication for Use: The Amblyz™ System is an electronic shutter module intended to provide occlusions of the eye in an intermittent fashion. #### Technological Characteristics: The Amblyz™ system uses operation principles equivalent to those of the legally marketed LCG System. The device utilizes a Liquid Crystal and electronics technology to create an electronic shutter module that enables the exercise of the amblyopic eye combining with rest periods of the sound eye. There is no control mechanism change {1}------------------------------------------------ and no change in energy type. All changes that do exist are presented in the comparison table presented below under 'substantial equivalence'. #### Device Description: The Amblyz™ device is an optical eyewear using a Liquid Crystal lens and electronics technology that acts as an intermittent shutter. The shutter incorporated in the optical lens performs an occlusion of the sound eye in an intermittent mode by changing the lens from a clear to an opaque state via electronic control. The system incorporates the following components (as detailed in chapter 12 of this submission): #### 1. Spectacle Frame The Spectacle frame designed to hold both the liquid crystal lens and the refractive lens. The frame is made of biocompatible durable plastics. #### 2. Liquid Crystal Shutter (LCS) The LCS lens, which functions as an electronic shutter is bundled in the glasses frame. Controlled by the microcontroller on the PCB, the designated LCS performs repetitive occlusion and transparent sessions during which the lens is fully transparent ("OFF" state) when no voltage is applied and turns opaque ("ON" state) when voltage is applied. #### 3. Earpieces The ear pieces are the part of the spectacle frame that extends from the hinges and to the back end of the device #### 4. Nose Piece Nosepiece increases the distance between the nose and the spectacle frame, The nosepiece is detachable from the spectacle frame and the user can choose if to use it or not. - 5. Clip to Hold Refractive Lens A plastic clip that holds the optic lenses (if needed). #### 6. Micro-USB Connector During normal operation the connector is used only to charge the glasses. X6D technicians also use this connector to upload firmware to the electronic control for tests and/or updates. #### 7. LED Indicator The Amblyz™ is equipped with an orange LED. The LED should be OFF when in normal operation and ON when properly connected to the USB charger and charging. Once the battery is fully charged, the LED indicator will turn OFF, even if Amblyz™ glasses are still connected to the USB charger. A flashing LED light indicates that the battery is low and charging is needed. #### 8. ON/OFF Switch A mechanical rotary switch used only to power ON the glasses, an action that is performed only once, by the optician who initially sets up the glasses. #### 9. Programming Button The programming button is a simple push button, located next to the ON switch. Push button is used for setting the eye to be occluded. This operation is performed only once, by the optician who initially sets up the glasses. ### 10 Amblyz™ onboard Software {2}------------------------------------------------ An onboard software (referred to as "firmware") controls the Amblyz [ glasses An onboard sonware (referred to as "limitate") Sontere the Amblyz™ glasses unit, is to allow the initial setting, and then after continuously deliver the therapy sequence without any option for user tampering. Other "house holding" functions of the software are to alert on low battery state, support the "watch-dog" component that resets the glasses in case of error. ## 11 Amblyz™ Technician Software Amblyz™ PC Maintenance (Technician) Software is intended to be used only by qualified X6D personnel for the purpose of reprogramming / upgrading the onboard Amblyz™ software. #### Substantial Equivalence: The table below presents the comparison between the Amblyz™ System and the legally marketed device - the LCG System. | | Amblyz™ System | LCG System (Predicate) | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications | The Amblyz™ System is an electronic<br>shutter module intended to provide<br>occlusions of the eye in an intermittent<br>fashion.<br>(As the LCG) | The LCG (Liquid Crystal Glasses)<br>System is an electronic shutter<br>module intended to provide<br>occlusions of the eye in an<br>intermittent fashion. | | Technological<br>Characteristics | Computer derived, electrical and Liquid<br>Crystal technology. | Computer derived, electrical and<br>Liquid Crystal technology. | | | A Liquid Crystal Module (LCM) lens<br>covers a spectacle unit lens, which is<br>intermittently occluded when voltage is<br>applied.<br>(As the LCG) | A Liquid Crystal Module (LCM)<br>lens covers a spectacle unit lens,<br>which is intermittently occluded<br>when voltage is applied. | | Materials | The Amblyz™ System's components<br>that come with direct contact with the<br>patient for a continuous duration of use<br>are the Spectacle Frame (comprised of<br>front frame, nose piece, ear handles<br>and clip).<br>• The front frame is made of durable<br>and biocompatible plastic PC<br>Makrolon 2458.<br>• The nose piece is made of Texin<br>RXT50D<br>• The Amblyz™ System's handle<br>comprises Elastosil Colour Paste Pt<br>Yellow Ral 1021 and Elastosil R<br>401/70<br>All the above materials conform to | The LCG System's parts that come<br>in direct contact with the patient<br>and intended for a continuous<br>duration of use are the Spectacle<br>Frame and its Strap.<br>• The Spectacle Frame is made<br>of a durable and biocompatible<br>plastic Grilamid TR 90.<br>• The Strap is made of<br>biocompatible Thermoplastic<br>polyurethanes (TPU)<br>Carbothane®. | | | Amblyz™ System | LCG System (Predicate) | | | applicable biocompatibility and flammability requirements. | | | Mode of Action | • The patient is wearing the AMBLYZ™ (spectacle) Unit.<br>• Intermittent occlusions are performed.<br>• ON/OFF switch is used by technician only – in course of user operation the glasses are always ON | • The patient is wearing the LCG (spectacle) Unit.<br>• Intermittent occlusions are performed.<br>• ON/OFF switch is activated by a touch sensor on the glasses that is activated upon wearing the glasses | | Set-Up | One time only "press button" set-up - done by healthcare professional | Set-up via dedicated PC software. | | Charging | Mini-USB connection to computer or any other USB port or wall charger (not included in system) | Mini-USB connection to dedicated charging box | | Compliance monitoring | None | Glasses mounted sensor monitors wear time | | Occlusion Pattern | 1) Occlusion is performed in an intermittent fashion, enabling rest periods<br>2) Total occlusion time is approximately 50% of the time<br>3) Occlusion pattern is fixed: 30 seconds and rest period is 30 seconds. | 1) Occlusion is performed in an intermittent fashion, enabling rest period<br>2) Total occlusion time is approximately 50% of the time:<br>3) Occlusion pattern is quasi-random, with occlusion periods varying between 12 up to 35 seconds. | | Indicators | (On glasses only) A flashing LED light indicates when battery is low.<br>Constant LED light indicates charging.<br>Shutter Lenses rapid flashing indicate correct connection to charger. | 1. None on glasses<br>2. The reporting box indicates actual and prescribed wear time and present smiley/sadly icons that indicate compliance. | | Design | 1. Frame dimensions: L129 x W129 x H40 mm<br>2. Weight: 36,4 g<br>3. The electronics and battery are housed within the frame<br>4. Optic lenses on clip, can be detached for cleaning and/or replacement<br>5. Viewing angle (pantoscopic angle) provides a 6 degrees | 1. Frame dimensions: L120xW120xH34mm<br>2. Weight: ~35gr<br>3. The electronics and battery are housed in the spectacles strap.<br>4. Optic lenses are coupled (glued) to device.<br>5. Viewing angel zero (pantoscopic angel not | | | Amblyz™ System | LCG System (Predicate) | | | pantoscopic angle | applied) | | Performance | 1. Operation time without charge: over 48 hours<br>2. Charging time: less than 2 hours<br>3. Lens transparency: 40%<br>4. Contrast ratio: Over 400 | 1. Operation time without charge: 20 hours<br>2. Charging time: 8 hours<br>3. Lens transparency: 40%<br>4. Contrast ratio: 1:300 | {3}------------------------------------------------ . . : ォ {4}------------------------------------------------ ## Safety and Performance Testing A series of safety and performance testing were performed: - Electrical safety and electromagnetic compatibility testing according teg. IEC . 60601-1 (and amendments), and IEC 60601-1-2 standards. The Amblyz™ has found to comply with the requirements of these standards. Certificate of Compliance to these standards is attached to this submission. - Software verification and validation testing . - A set of in vitro (bench) performance testing: . | Test Description | Results | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Lens Impact Resistance/Drop Ball Test | The lenses cracked but did not shatter into pieces<br>and therefore comply with standard requirements | | Destructive Test | The LC lens glass was cracked and the liquid<br>crystal was spread across the glass. Nevertheless,<br>the laminated films completely held the LCM glass<br>pieces. No change in the overall package and no<br>glass residues were found. Test was performed on<br>four lenses, all resulting in same findings | | Charge/Discharge Test | Tests show, that glasses have more than enough<br>operating time with the used battery. | | Charging Time Test | All ten (10) test cases comply with the acceptance<br>criteria | | Climatic Testing | The samples sustained the specified<br>environmental testing, no damage was noted. | | Unpacked Drop Test | All samples have passed all visual and<br>operating/performance checks after the drop | | Cosmetic Resistance | No visual change on tested samples after the test<br>was completed. | | Sweat Resistance | No visual change observed on tested surfaces in<br>all tested samples | {5}------------------------------------------------ | Test Description | Results | |---------------------------------|---------------------------------------------------------------------------------------| | Cleaning Resistance | No visual change observed on tested surfaces in<br>all tested samples | | Thermal Shock | All samples have passed all visual and<br>operating/performance checks after the test | | Electrical Characteristics Test | All tests passed successfully and meet acceptance<br>criteria. | Packaging Validation including the following tests: . | Test Description | Results | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Fall Test - Master Packaging | Functional and cosmetic inspections have passed<br>successfully. | | Fall Test - Retail Packaging | All samples passed and meet acceptance criteria | | Transportation Vibration Test | After the completion of Transportation Vibration<br>Test, units inside the carton box were subjected to<br>functionality check. All test units passed the<br>functionality check with no functionality and<br>cosmetic problem. | | Bump Test | After the completion of Bump Test, units inside the<br>carton box were subjected to functionality check.<br>All test units passed the functionality check with no<br>functionality and cosmetic problem. | ## Biocompatibility Testing According to the Blue Book Memo, G95-1, Use of International Standard ISO-10993, and ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, the components of the Amblyz™ glasses are categorized as surface device with Intact skin permanent (>30days) contact. The parts that come in direct contact with the patient and intended for a continuous duration of use are the Spectacle Nosepiece, Front frame (Case left & right) prescription frame, and the Amblyz™ Insert left & right part of the handle. Cytotoxicity testing was conducted on all the above, based on the requirements of ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. Also, hazard analysis including risk level and solutions performed in compliance with ISO 14971:2012 for the entire system and for the software. In conclusion, a series of in vitro, software, electrical safety, and biocompatibility test results demonstrate that the product has acceptable safety and performance characteristics. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized bird or a human figure with outstretched arms. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 28, 2014 X6D Ltd. % Ms. Neta Landman Duct Medical Consulting Ltd. 21 Hanafa Street Zur- Moshe 42810, Israel P.O. Box 536 Re: K132042 Trade/Device Name: Amblyz™M Regulation Number: 21 CFR 886.4750 Regulation Name: Ophthalmic Eye Shield Regulatory Class: Class I Product Code: HOY Dated: March 21, 2014 Reccived: March 31, 2014 Dear Ms. Landman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Page 2 - Ms. Neta Landman Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Deborah L. Falls -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |-----------------------------------------|--| | Food and Drug Administration | | | Indications for Use | | Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K132042 Device Name Amblyz Indications for Use (Describe) The Ambly2TM System is an electronic shutter module intended to provide occlusions of the eye in an internittent fashion. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | (Division Sign-Off) | | |---------------------------------------------------------|---------| | Division of Ophthalmic and Ear, Nose and Throat Devices | | | 510(k) Number | K132042 | Andrew Yang -S 2014.04.25 13:21:30 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden the for this Gallection of micronization of maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (1/14) Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...