PLANCON MICROLAMELLAR KERATOME (SINGLE PIECE KERATOME HEAD WITH ACCESSORIES)
Applicant
Plancon Instruments
Product Code
HNO · Ophthalmic
Decision Date
Mar 31, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4370
Device Class
Class 1
Attributes
Therapeutic
Intended Use
To perform microlamellar keratoplasty for the correction of myopia or hyperopia
Device Story
The Plancon Instruments Microlamellar Keratome is a surgical instrument used by ophthalmologists to perform microlamellar keratoplasty. The system consists of a gas-powered turbine motor, a fixed single-piece keratome head, a keratome blade, pneumatic fixation rings, and applanation lenses. The device is powered by a vacuum pump unit that regulates gas pressure. During surgery, the pneumatic fixation ring secures the eye, and the keratome head, equipped with a blade, creates a lamellar corneal incision. The fixed-head design limits the cut depth to a single predetermined value, enhancing safety compared to adjustable-head models. The device is used in a clinical or surgical setting to facilitate refractive surgery, potentially improving patient vision by correcting myopia or hyperopia.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System includes gas-powered turbine motor (0-20,000 RPM), stainless steel/titanium/aluminum construction, pneumatic fixation rings, and applanation lenses. Operates via gas pressure regulation. Fixed single-piece keratome head design. No gear system.
Indications for Use
Indicated for patients requiring microlamellar keratoplasty for the correction of myopia or hyperopia.
Regulatory Classification
Identification
A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.
K040297 — ONEUSE- PLUS MICROKERATOME · Moria SA · Mar 31, 2004
K022560 — M2 SINGLE USE MICROKERTOME · Moria SA · Sep 27, 2002
Submission Summary (Full Text)
{0}
MAR 31 1997
K970377
# SAFETY AND EFFECTIVENESS SUMMARY K970377 p. 1/2
Plancon Instruments, 15 Rue Georges Besse, 92160 Antony, France
Contact person: Mr. Alain DUPRAT, General Manager
Date: December 13, 1996
Tel.: (33-1) 46 74 46 74
| | Predicated device ①
510 K Document number
K 960395 | Predicated device ②
510 document number :
Not known | Predicated device ③
510 K document number :
K 935342 | Submitted device |
| --- | --- | --- | --- | --- |
| Designation / Feature | PLANCON INSTRUMENTS
Microlamellar Keratome | EYE TECHNOLOGY | SCMD | PLANCON INSTRUMENTS
Microlamellar Keratome
(single piece head) |
| Intended use | To perform microlamellar keratoplasty for the correction of myopia or hyperopia | To perform micro lamellar keratoplasty for the correction of myopia or hyperopia | To perform micro lamellar keratoplasty for the correction of myopia or hyperopia | To perform microlamellar keratoplasty for the correction of myopia or hyperopia |
| System composition | Power unit
Turbine motor
Adjustable keratome head
Keratome blade
Pneumatic fixation rings
Applanation lenses
Plate checking gauge | Power unit
Turbine motor
Adjustable keratome head
Keratome blade
Pneumatic fixation rings
Applanation lenses
Plate checking gauge | Power unit
Turbine motor
Adjustable keratome head
Keratome blade
Pneumatic fixation rings
Applanation lenses
Plate checking gauge | Power unit
Turbine motor
Fixed keratome head
Keratome blade
Pneumatic fixation rings
Applanation lenses |
| Power unit | Vacuum pump
Battery
Vacuum release valve
Knob to regulate gas pressure
Gas & vacuum pressure gauges
Vacuum & turbine quick connectors
Battery charger unit | Vacuum pump
Battery
Vacuum release valve
Knob to regulate gas pressure
Gas & vacuum pressure gauges
Vacuum & turbine quick connectors
Battery charger unit | Vacuum pump
Battery
Vacuum release valve
Knob to regulate gas pressure
Gas & vacuum pressure gauges
Vacuum & turbine quick connectors
Battery charger unit | Vacuum pump
Battery
Vacuum release valve
Knob to regulate gas pressure
Gas & vacuum pressure gauges
Vacuum & turbine quick connectors
Battery charger unit |
{1}
K970377
SAFETY AND EFFECTIVENESS SUMMARY K970377 p. 2/2
| | Predicated device ①
510 K Document number
K 960395 | Predicated device ②
510 document number :
Not known | Predicated device ③
510 K document number :
K 935342 | Submitted device |
| --- | --- | --- | --- | --- |
| Designation / Feature | PLANCON INSTRUMENTS
Microlamellar Keratome | EYE TECHNOLOGY | SCMD | PLANCON INSTRUMENTS
Microlamellar Keratome
(single piece head) |
| Turbine motor | Gas powered
Adjustable speed 0 to 20 000 RPM (recommended 14 000 RPM)
Material : stainless steel, titanium and aluminium.
No gear system | Gas powered
Adjustable speed
Material : stainless steel
No gear system | Gas powered
Adjustable speed from 10 to 24 000 RPM
Pre-set speed 14 000 RPM
Material : stainless steel and brass
No gear system | Gas powered
Adjustable speed 0 to 20 000 RPM (recommended 14 000 RPM)
Material : stainless steel, titanium and aluminium.
No gear system |
| Keratome | Material : stainless steel
Adjustable
Safety : maximum cut depth, and checking gauge | Material : stainless steel
Adjustable
Permanent adjustable head with adjustment knob
Safety : maximum cut depth and checking gauge | Material : stainless steel
Adjustable
Safety : maximum cut depth and checking gauge | Material : stainless steel
Fixed single piece head
Safety : only one cut depth allowed by head principle |
| Pneumatic fixation rings | Material : stainless steel
Double dovetails | Material : stainless steel
Double dovetails | Material : stainless steel
Double dovetails | Material : stainless steel
Double dovetails |
| Applanation lenses | Material : clear plastic + stainless handle
Engraved inner reticle diameter | Material : clear plastic + stainless handle
Engraved inner reticle diameter | Material : clear plastic + stainless handle
Engraved inner reticle diameter | Material : clear plastic + stainless handle
Engraved inner reticle diameter |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.