K020482 · Micro Specialties, Inc. · HNO · Dec 16, 2002 · Ophthalmic
Device Facts
Record ID
K020482
Device Name
KERATOME BLADE 200200 & 600600
Applicant
Micro Specialties, Inc.
Product Code
HNO · Ophthalmic
Decision Date
Dec 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4370
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The MICROspecialties, keratome blades are designed to produce a corneal flap when used with the following Moria microkeratomes: MICROspecialties 200200 Blade for use on Moria LSK Microkeratome . MICROspecialties 600600 Blade for use on Moria CB Microkeratome
Device Story
Disposable stainless steel keratome blades; designed for use with Moria LSK and CB microkeratome systems. Blades function as mechanical cutting components to create corneal flaps during ophthalmic surgery. Operated by surgeons in clinical settings. Device provides physical interface for tissue resection; performance validated through compatibility testing with Moria systems and nonclinical evaluations of handling and ease of use. Gamma radiation sterilized; packaged in protective plastic cases.
Clinical Evidence
No clinical data. Nonclinical evaluation performed by independent physicians and clinics assessing package handling, ease of use, and performance on Moria microkeratome systems. Design validation confirmed appearance and compatibility.
Technological Characteristics
Medical grade stainless steel; single cutting edge. Dimensions/form factor specific to Moria LSK and CB microkeratome heads. Sterilization via gamma radiation. Mechanical cutting principle. No software or electronic components.
Indications for Use
Indicated for use in adult patients undergoing surgical procedures to create a corneal flap for vision correction.
Regulatory Classification
Identification
A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.
Related Devices
K993122 — MSP M/K BLADE · Medical Sterile Products, Inc. · Mar 16, 2000
K030891 — SURGISTAR MICROKERATOME BLADE, MODEL NO. 2400-LSK · Surgistar, Inc. · Jun 17, 2003
K980508 — KERATOME BLADE, CHIROM MODEL ALK 500500; KERATOME BLADE, S.C.M.D. MODEL TUROKERATOME 400400 · Micro Specialties, Inc. · May 1, 1998
K022205 — DISPOSABLE N-PE MICROKERATOME BLADES · Oasis Medical, Inc. · Oct 3, 2002
K030401 — DISPOSABLE M2-PE MICROKERATOME BLADES · Oasis Medical, Inc. · Jun 19, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
MICROspecialties,Inc. 264 Quarry Road, Milford, CT 06460
Tel: 203-874-1832
Fax: 203-877-3762
DEC 1 6 2002
# K020482 510(k) Summary (ref. 807.87h)
This summary of the 510(k) safety and effectiveness information is being submitted for two styles of microkeratome blades for use on Moria microkeratome systems. The MICROspecialties 200200 Blade is designed for use on the Moria, Model LSK microkeratome and the MICROspecialties 600600 Blade for use on the Moria, Model CB microkeratome.
The blades are manufactured using the same materials and packaging. The method of sterilization is commonly used in these devices and their packaging.
### Content and Format of a 510(k) Summary (807.92a)
- 1. Submitter: Gaston Levesque, President MICROspecialties, Inc. 264 QUARRY Road Milford. CT 06460 Phone: 203/874/1832 Fax: 203-877-3762
- 2. Keratome Blade
- 3. 200200 Keartome Blade 600600 Keratome Blade
- 4. The disposable 200200 microkeratome blade is designed to be used with the Moria LSK microkeratome system. The disposable 600600 microkeratome blade is designed to be used with the Moria CB microkeratome system. Both of these blades are designed to produce a corneal flap. The blades are packaged in plastic cases to protect the cutting edge and the placed in a Tyvek/poly pouch. They are sold as ten blades per box or in larger quantities of thirty (three ten-packs). Labeling identifies each quantity per box. The blades are then gamma radiation sterilized.
The blades are made from medical grade stainless steel with a single cutting edge. They are assembled with their compatible head and driver systems.
- 5. The 200200 and 600600 keratome blades are designed to produce a corneal flap. This is part to correct correct vision of a surgical procedure in the the the adult population.
{1}------------------------------------------------
MICROspecialties, Inc. 264 Quarry Road, Milford, CT 06460
Tel: 203-874-1832
## 510(k) Summary - cont. (ref. 807.87h)
- 5. The MICROspecialties 200200 and 600600 keratome blades are designed for use with Moria Microkeratome systems. The 200200 blade is for use with the Moria, Model LSK Microkeratome and the 600600 for use with the Moria, Model CB Microkeratome.
A risk analysis was performed to provide a systematic approach to the evaluation of hazards associated with the keratome blades.
Independent physicians and clinics were used to evaluate the performance of the 200200 and 600600 microkeratome blades as part of a nonclinical evaluation The evaluation included package handling, ease of use, performance on the respective Moria microkeratomes.
Engineering, manufacturing and inspection specifications were finalized based on this evaluation. A design validation was conducted to validate the appearance, compatibility with the Moria microkeratomes and performance within these systems. The blades go through 100% inspection to insure that they meet the product specification and follow design controls of MICROspecialties.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles three overlapping, curved lines.
DEC 16 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MICROspecialties, Inc. c/o Ms. Leigh Ayres, Director Quality Assurance and Regulatory Affairs 264 Quarry Road Milford, CT 06460
K020482 Re:
Trade/Device Name: MICROspecialties 200200 Blade for the Moria LSK Microkeratome MICROspecialties 600600 Blade for the Moria CB Microkeratome
Regulation Number: 21 CFR 886.4370 Regulation Name: Keratome Regulatory Class: Class I Product Code: HNO Dated: November 20, 2002 Received: November 21, 2002
Dear Ms. Ayres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Leigh Ayres, Director
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### Indications for Us MICROspecialties. Inc. Disposable Keratome Blades
Page 1 of 1
K020482 510(k) Number (if known):
Device Name: Disposable Keratome Blades for use on Moria Microkeratomes
The MICROspecialties, keratome blades are designed to produce a corneal flap when used with the following Moria microkeratomes:
- MICROspecialties 200200 Blade for use on Moria LSK Microkeratome .
- MICROspecialties 600600 Blade for use on Moria CB Microkeratome
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ar Callaway
(Bivision Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises
K020482 510(k) Number_
(Optional Format 3-10-98)
Prescription Use
(Per 21 CFR 801.109)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.