K071335 · Porex Surgical, Inc. · HPZ · Aug 8, 2007 · Ophthalmic
Device Facts
Record ID
K071335
Device Name
MEDPOR ATTRACTOR IMPLANT
Applicant
Porex Surgical, Inc.
Product Code
HPZ · Ophthalmic
Decision Date
Aug 8, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MEDPOR ATTRACTOR Implant is intended for patients who require replacement of volume of an enucleated or eviscerated orbit and who wish to gain improved prosthetic eye motility by coupling the MEDPOR ATTRACTOR Implant to the prosthetic eye.
Device Story
Device consists of porous polyethylene orbital implant containing titanium nitride-coated stainless steel screw/insert; functions as magnetic attractor. Ocularist embeds gold-plated magnets into posterior of custom prosthetic eye. Surgeon implants device into enucleated/eviscerated orbit; covered by tissue. Magnetic coupling between implant and prosthetic eye improves motility. Used in clinical setting by surgeons; prosthetic eye fabrication performed by ocularists. Benefits patient through enhanced movement of prosthetic eye compared to uncoupled implants.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Porous polyethylene implant; titanium nitride-coated stainless steel screw/insert; gold-plated magnets. Mechanical coupling via magnetic force. Dimensions/form factor designed for orbital implantation. Non-electronic; no software.
Indications for Use
Indicated for patients requiring orbital volume replacement following enucleation or evisceration who desire improved prosthetic eye motility.
Regulatory Classification
Identification
An eye sphere implant is a device intended to be implanted in the eyeball to occupy space following the removal of the contents of the eyeball with the sclera left intact.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ocular peg which is supplied sterile only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
K040851 — MEDPOR ATTRACTOR MAGNETIC COUPLING SYSTEM · Porex Surgical, Inc. · Jul 16, 2004
K960859 — MEDPOR OCULAR PEG SYSTEM AND ACCESSORIES · Porex Technologies Corp. of Georgia · May 29, 1996
K974203 — PERRY-KOLBERG (PK) TITANIUM MOTILITY/SUPPORT SYSTEM · Integrated Orbital Implants, Inc. · Feb 4, 1998
K980822 — TITANIUM PEG SYSTEM · Fci Ophthalmics, Inc. · Jul 9, 1998
K971583 — MEDPOR OCULAR SCREW AND ACCEOSSORIES · Porex Technologies Corp. · Jun 30, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
## 510 (k) SUMMARY
Manufacturer and Submitter
Porex Surgical, Inc. 15 Dart Road Newnan, GA 30265
Tel: (678) 479-1610 Fax: (678) 479-4495
Contact: Jerri Davis E-mail: jerri.davis@porex.com
Date: May 8, 2007
Trade Name: Device Name: MEDPOR® ATTRACTOR™ Implant
Class: II
:
Product Code: MQU
510 (k) Number_KD 7 (335
Substantially equivalent to:
A. MEDPOR ATTRACTOR Magnetic Coupling System B. MEDPOR Surgical Implant Material; Preformed Cranial & Facial Implants
Device description:
MEDPOR Orbital Implant made of porous polyethylene with a titanium nitride coated stainless steel screw or insert. The screw or insert acts as an attractor to a magnet that is embedded into the posterior of a prosthetic eye. The magnets are gold plated and designed small enough to remain entirely within the prosthesis material of the prosthetic eye, and powerful enough to provide a coupling force between the implant and the prosthesis.
Indications for Use:
The MEDPOR ATTRACTOR Implant is intended for patients who require replacement of volume of an enucleated or eviscerated orbit and who wish to gain improved prosthetic eye motility by coupling the MEDPOR ATTRACTOR Implant to the prosthetic eye.
{1}------------------------------------------------
## Technological Characteristics:
A prosthetic eye, which is not part of this device, is created by an ocularist. The ocularist would fit the prosthetic eye in the usual manner. It is advantageous to mechanically couple the prosthetic eye to the implant in some way to improve motility over an uncoupled prosthetic eye. The device that is the subject of this premarket notification uses magnetic force to accomplish the coupling between the MEDPOR ATTRACTOR Implant and the prosthetic eye. During the custom fabrication of the prosthetic eye, by an ocularist, one or more small magnets are placed into the posterior of the prosthetic eye so that the magnets are covered with a thin layer of acrylic. The MEDPOR ATTRACTOR Implant is covered with tissue when the surgeon implants it in the patient's enucleated or eviscerated orbit. When the prosthetic eye is placed on the implant, the magnet force between the magnets in the prosthetic eye and the MEDPOR ATTRACTOR Implant provide a coupling force to improve the movement of the prosthetic eye.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 8 2007
Porex Surgical, Inc. c/o Mr. Jerri Davis 15 Dart Road Newman, GA 30265
Re: K071335
Trade/Device Name: MEDPOR® ATTRACTOR™ Implant Regulation Number: 21 CFR 886.3320 Regulation Name: Implant with Titanium Coated Stainless Steel Screw or Insert Regulatory Class: II Product Code: HPZ Dated: July 30, 2007 Received: July 31, 2007
Dear Mr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M.B. Eglehus, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Applicant: Porex Surgical Inc. 15 Dart Road Newnan, GA 30265
Tel: (678) 479-1610 Fax: (678) 423-1437
510(k) Number: KO71335
Trade Name: Device Name: MEDPOR® ATTRACTOR™ Implant
Indications for Use:
The MEDPOR ATTRACTOR Implant is intended for patients who require replacement of volume in an enucleated or eviscerated orbit and who wish to gain improved prosthetic eye motility by coupling the MEDPOR ATTRACTOR Implant to the prosthetic eye.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE)
H. Viman
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
**510(k) Number** K071335
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.