BAERVELT PARS PLANA GLAUCOMA IMPLANT
Device Facts
| Record ID | K955455 |
|---|---|
| Device Name | BAERVELT PARS PLANA GLAUCOMA IMPLANT |
| Applicant | Pharmacia Iovision, Inc. |
| Product Code | KYF · Ophthalmic |
| Decision Date | Feb 18, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.3920 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Baerveldt Pars Plana Glaucoma Implant is intended for use in patients with medically uncontrollable glaucoma with poor surgical prognosis.
Device Story
Non-valved artificial drainage device; shunts aqueous humor via silicone tube to episcleral plate; implanted in equatorial regions of globe. Device features silicone plate with 13 mm convex radius; barium-impregnated for radiopacity (CT, MRI, plain film). Includes silicone tube with elbow attachment for posterior chamber insertion; flange with fixation holes for scleral attachment. Used by ophthalmic surgeons in surgical setting to manage intraocular pressure in refractory glaucoma patients; provides alternative drainage pathway to reduce pressure; prevents vision loss.
Clinical Evidence
No clinical data provided; substantial equivalence based on design comparison and surgical performance similarity to predicate.
Technological Characteristics
Medical grade silicone construction; barium-impregnated for radiopacity; 13 mm convex radius plate; non-valved drainage mechanism; includes silicone tube with elbow connector.
Indications for Use
Indicated for patients with medically uncontrollable glaucoma and poor surgical prognosis.
Regulatory Classification
Identification
An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and (3) “Aqueous Shunts—510(k) Submissions.”
Predicate Devices
- Baerveldt Glaucoma Implant
Related Devices
- K182518 — AHMED ClearPath Glaucoma Drainage Device · New World Medical, Inc. · Jan 18, 2019
- K152996 — Molteno3 Glaucoma Implant · Innovative Ophthalmic Products, Inc. (Iop) · Dec 17, 2015
- K062252 — MOLTENO3 · Iop, Inc. · Sep 27, 2006
- K991072 — AHMED GLAUCOMA VALVE IMPLANT · New World Medical, Inc. · Jul 13, 1999
- K171451 — Ahmed Glaucoma Valve Model FP8 · New World Medical, Inc. · Aug 8, 2017