CLEARPATH DS-120(R) LENS FLUORESCENCE BIOMICROSCOPE

K112880 · Freedom Meditech, Inc. · MYC · Jan 31, 2013 · Ophthalmic

Device Facts

Record IDK112880
Device NameCLEARPATH DS-120(R) LENS FLUORESCENCE BIOMICROSCOPE
ApplicantFreedom Meditech, Inc.
Product CodeMYC · Ophthalmic
Decision DateJan 31, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1570
Device ClassClass 2

Indications for Use

The ClearPath DS-120® Lens Fluorescence Biomicroscope is indicated for use to detect autofluorescence of the crystalline lens.

Device Story

ClearPath DS-120 is a scanning biomicroscope used to measure crystalline lens autofluorescence. Input: eye illuminated by blue LED; near-infrared LEDs used for pupil positioning. Operation: blue light beam scans in one dimension (depth) through the lens; elastic (Rayleigh) and inelastic (fluorescent) scattering interactions occur. Collection optics isolate blue and green light signals via filters; photomultiplier detects signals; A/D converter digitizes data. Software controls scanning, performs automatic pupil tracking, rejects invalid scans (blinks, anomalies), and computes the ratio of lens autofluorescence to scattered light. Output: fluorescence ratio displayed on screen; printable report. Used in clinical settings; operated by trained personnel. Healthcare providers use the ratio as a measurement tool; clinical utility/prognosis not established by this device.

Clinical Evidence

Prospective single-site precision study (n=27 subjects, ages 21-68). Evaluated repeatability and reproducibility of lens fluorescence/scattering ratio measurements. Three operators and three devices used. Total 405 successful scans. Results: Mean ratio 0.17899; Repeatability SD 0.01034 (5.779% CV); Reproducibility SD 0.01153 (6.443% CV). No adverse events observed. Study concluded device is safe and provides consistent measurements within the 0.07-0.33 range. No diagnostic or prognostic utility established.

Technological Characteristics

Optoelectronic biomicroscope; blue LED light source; photomultiplier detector; motorized chin rest; laptop PC interface. Connectivity: digital data output. Software: automated pupil tracking, scan validation, and ratio computation. Materials: standard optical/electronic components. Sterilization: not applicable (non-contact optics).

Indications for Use

Indicated for detection of crystalline lens autofluorescence in patients.

Regulatory Classification

Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 3 1 2013 K112880 ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. SUMMARY PREPARED: January 25, 2013 1.1 SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED APPLICANT: Freedom Meditech, Inc. 10455 Pacific Center Court San Diego, CA 92121, USA Tel: (858)-638-1433 > Contact Person: Craig H. Misrach Chairman & CEO #### 1.2 NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME ClearPath DS-120® Lens Fluorescence Biomicroscope TRADE NAME: ClearPath DS-120® Lens Fluorescence Biomicroscope COMMON NAME: CLASSIFICATION NAME: Ophthalmoscope (Class II) DEVICE CLASSIFICATION: 21 CFR 886.1570 PRODUCT CODE: MYC PREDICATE DEVICE: Optos PIC P200MAAF Ophthalmoscope, K102492 #### SUBSTANTIALLY EQUIVALENT TO 1.3 | 510(K) NUMBER | TRADE OR PROPRIETARY OR<br>MODEL NAME | MANUFACTURER | |---------------|---------------------------------------|--------------| | K102492 | PIC P200MAAF<br>OPHTHALMOSCOPE | OPTOS | {1}------------------------------------------------ #### 1.4 DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION ## 1.4.1 MAJOR COMPONENTS The major functional components of the ClearPath DS-120® Lens Fluorescence Biomicroscope are an optics unit and a laptop personal computer. The only components that contact the patient are a manually adjustable headrest (in/out) and a motorized adjustable chin rest (up/down). The adjustable optics window does not contact the patient. ### 1.4.2 PRINCIPLES OF OPERATION The patient is positioned with the forehead centered on the headrest. The eye is illuminated by near-infrared lights and observed by an IR sensitive video camera. An image of the eye is displayed on the computer screen to assist the operator in the alignment of the patient's eve. The headrest is adjusted manually so that the eye is in the camera image that is visible on the computer screen. The patient is instructed to look at a red blinking fixation light. The operator can adjust the chin rest and optics window positions to enable the patient to sit comfortably with the fixation target properly positioned. The software includes an automatic tracking program for positioning the pupil. The blue light source and the collection optics are confocally aligned within a volume of measurement that is scanned through the lens. In the eye, this blue light is scattered by elastic (Rayleigh scattering) and inelastic (fluorescent) interactions with lens proteins. Filters in the collection optics reject light from the positioning lights and isolate blue and green light signals, which are detected by a highly sensitive photomultiplier and sent to an A/D converter and then to the computer. Under software control, the volume of measurement is scanned through the lens and back. Computer software records both scattered and fluorescent light and computes the average of the ratio of lens autofluorescence to scattered light in the central portion of the lens. Software rejects scans that are outside of physiological limits or have anomalies due to eye blinks, or to excessive differences between the forward and reverse scans. For valid scans, the fluorescence ratio is reported on the screen and can be printed for the patient and/or for the patient's file. #### 1.4.3 MAINTENANCE AND CALIBRATION There are no user serviceable parts or adjustments in the optics unit. The system self-tests when it is turned on. There is no maintenance required from the clinician other than cleaning the head and chin rest and optics window, as described in the Instruction for Use. {2}------------------------------------------------ KI12880 #### 1.5 INDICATION FOR USE The ClearPath DS-120® Lens Fluorescence Biomicroscope is indicated for use to detect autofluorescence of the crystalline lens. #### 1.6 PREDICATE DEVICE COMPARISON TABLE | | ClearPath DS-120® Lens<br>Fluorescence Biomicroscope | P200MAAF Ophthalmoscope<br>(K102492) | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. Intended Use | To examine the eye via detection<br>of autofluorescence from ocular<br>structures | To examine the eye via detection<br>of autofluorescence from ocular<br>structures | | 2. Indications<br>for Use | The ClearPath DS-120® Lens<br>Fluorescence Biomicroscope is<br>indicated for use to detect<br>autofluorescence of the crystalline<br>lens. | The P200MAAF scanning laser<br>ophthalmoscope is intended to be<br>used as a wide field and retinal<br>fluorescence and autofluorescence<br>imaging ophthalmoscope to aid in<br>the diagnosis and monitoring of<br>diseases or disorders that manifest<br>in the retina. | | | ClearPath DS-120® Lens<br>Fluorescence Biomicroscope | P200MAAF Ophthalmoscope<br>(K102492) | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 3. Method of<br>operation | The instrument consists of an<br>optoelectronic unit and a<br>computerized system of data<br>acquisition and processing. | The instrument consists of an<br>optoelectronic unit and a<br>computerized system of data<br>acquisition and processing. | | | | The instrument is a scanning<br>biomicroscope that uses a low<br>power blue LED beam to scan in<br>one dimension (depth) over the<br>lens. | The P200MAAF is a conventional<br>scanning laser ophthalmoscope<br>that uses low power blue, red and<br>green laser beams to scan in two<br>dimensions over the retina. | | | | The returned light is detected and<br>used to generate a digital image<br>with a computer or electronic<br>imaging device. | The reflected (or returned) light is<br>detected and used to generate a<br>digital image with a computer or<br>electronic imaging device. | | | ClearPath DS-120® Lens<br>Fluorescence Biomicroscope | P200MAAF Ophthalmoscope<br>(K102492) | | | | The device uses a blue LED as a<br>light source that is scanned by a<br>deflection system in one axis<br>(depth) to generate a one-<br>dimensional image. | The device uses blue, green, and<br>red lasers as light sources that are<br>scanned by a deflection system in<br>two axes across the retina to<br>generate an image. | | | | The scanning of the beams on one<br>axis is done by a stepper motor<br>driven displacement of a volume<br>of measurement at the intersection<br>of focused illumination and<br>collection optics. | The scanning of the beams on the<br>two axes is done using a rotating<br>polygon for the fast vertical scan<br>and a motor driven mirror for the<br>slower horizontal scan. | | | | The generation of the image is<br>performed using light detectors,<br>the output of which is digitized,<br>and the data collected in a<br>computer for reconstruction,<br>display, and storage. | The generation of the image is<br>performed using light detectors,<br>the output of which is digitized,<br>and the data collected in a<br>computer for reconstruction,<br>display, and storage. | | | | An alignment system helps ensure<br>that the patient's eye is correctly<br>positioned. | An alignment pattern helps ensure<br>that the patient's eye is correctly<br>positioned. | | | | The returned light then travels<br>back through the device to a<br>rotating filter wheel and a discrete<br>detector that converts the light to<br>an electrical signal (effectively a<br>blue and green channel). | The returned light then travels<br>back through the device to an<br>array of discrete detectors that<br>converts the light to an electrical<br>signal (red, green and blue<br>channels). | | | | This electrical signal is digitized<br>and used to build up an electronic<br>picture in a computer and<br>displayed on a computer display. | This electrical signal is digitized<br>and used to build up an electronic<br>picture in a computer and<br>displayed either on a cathode ray<br>tube or a liquid crystal display. | | | | The signal strength varies as the<br>beam is scanned through the lens,<br>allowing an image to be created<br>and recorded, revealing the<br>variation in its constituent material<br>and structures. | The signal strength varies as the<br>laser beam is scanned across the<br>eye, allowing an image to be<br>created and recorded, revealing the<br>variation in its constituent material<br>and structures. | | | | 4. | Exposure parameters | ClearPath DS-120® Lens Fluorescence Biomicroscope | P200MAAF Ophthalmoscope (K102492) | | | | Illumination is provided by a blue LED and filter, three 880 nm LEDs, and a red fixation target. | Illumination is provided by red and blue lasers. | | | | Compliance to electrical safety (including EMC), light emitting products, programmable devices and biocompatibility standards are met. | Compliance to electrical safety (including EMC), light emitting products, programmable devices and biocompatibility standards are met. | | 5. | Data Collection and Display | Image analysis software is integrated with the data collection software. | The images can be exported from the device as "tiff" image files for analysis on a personal computer. | | | | The images produced are displayed as separate blue channel and green channel images. The green channel image shows the natural fluorescence (autofluorescence) of the lens of the eye. | The images produced can be displayed as separate green channel and a red channel images. In one imaging mode, the eye is illuminated using the green laser, and a red channel image shows the natural fluorescence (autofluorescence) of the eye. In another mode, only the blue laser illuminates the retina and the green channel fluorescence is recorded. | | | | Software detects the lens two images of the lens (forward and reverse scans), and integrates the blue and green signal intensities. The ratio of green and blue integrated measurements is displayed on the computer monitor. | Device software is designed to display images. The tiff files could be read by general purpose image analysis software (e.g., NIH Image), which would enable blue and green image intensities to be measured in user-selected image areas. | | 6. | Flammability of Materials | The low power LEDs do not present ignition hazard. Materials are standard optical and electronic components. | Not available | | | | ClearPath DS-120® Lens<br>Fluorescence Biomicroscope | P200MAAF Ophthalmoscope<br>(K102492) | | 7. | Maximum<br>temperature<br>of parts of the<br>device held by<br>the operator<br>or accessible<br>to the patient | All accessible parts operate near<br>ambient temperature. | Not available | | 8. | Brightness<br>controls | Brightness is controlled by the<br>instrument. | Not available | {3}------------------------------------------------ # よ//2880 {4}------------------------------------------------ ## · KII 2880 {5}------------------------------------------------ Both instruments have similar configurations and principles of operation. Light with selected wavelength ranges is focused on eye structures and the returned light is measured at selected wavelengths. The Optos P200MAAF uses three lasers for illumination. The ClearPath DS-120 Lens Fluorescence Biomicroscope uses an eye-safe blue LED for illumination that is focused on the lens. No new safety issues are raised by this difference as both devices are within optical radiation safety limits. Both instruments scan the illumination beam and record an image. The Optos scans in two directions on the retina surface. The ClearPath DS-120 Lens Fluorescence Biomicroscope scans in one direction through the depth of the lens to generate a depth profile (a one-dimensional image). ## Optical Equivalency and Radiation Safety Scanning laser ophthalmoscopes using Class 1 lasers are exempt from this requirement. The low power LED light source in this instrument does not raise new safety questions. Optical safety is further discussed below. #### 1.7 BRIEF SUMMARY OF NONCLINICAL AND CLINICAL TESTS AND RESULTS #### 1.7.1 BENCH Bench testing has been performed for radiation safety and for linearity of the photodetector response. ## OPTICAL RADIATION SAFETY The patient is exposed to three optical sources. Three of the optical sources, the blue LED probe, the red blinking LED fixation light, and the three near-infrared illumination LEDs for pupil positioning, are actively used while the patient interacts with the instruments. The ClearPath DS-120® Lens Fluorescence Biomicroscope has been independently reviewed for optical radiation hazard. The instrument operates at all wavelengths and {6}------------------------------------------------ KI.2880 emission levels that would not produce any ocular injury - even within foreseeable misuse conditions. ## DETECTOR LINEARITY To demonstrate linear photometer response in the clinically relevant dynamic range of the signal, mixtures of coumarin (a fluorescent material) and microspheres (scattering) were prepared and dilutions of this mixture were measured by placing a cuvette with the mixtures in the optical path. The detector response vs. dilution factor was linear in both the fluorescence and scatter channels. We also compared the detector response observed in this experiment with the range of detector responses observed in our clinical study to confirm that these linear detector responses correspond to the dynamic range observed during clinical use. #### 1.7.2 Clinical Precision Study ## SUMMARY OF PROTOCOL A clinical study was performed to evaluate the repeatability and reproducibility of the ratio of the lens fluorescence and scattering response of human crystalline measured by the ClearPath DS-120. This was a prospective single-site study with a recruitment goal of twenty one (21) participants minimum to be enrolled based on the study eligibility criteria. Subjects who signed an informed consent form and met all the inclusion and exclusion underwent a ClearPath DS-120 examination. Three operators (A, B, and C) and three units of ClearPath DS-120 (1, 2, and 3) were included in the study. In order to evaluate the inter-operator variability, each of three operators used Unit 1 to take three successful lens fluorescence measurements on each test subject. The three Operator/Device configurations for the inter-operator variability are A/1, B/1, and C/1, Therefore nine successful measurements were collected on each test subject for the interoperator evaluation. In order to evaluate the inter-device variability, Operator A used each of the three ClearPath DS-120 units to take three successful lens fluorescence measurements on each test subject. The three Operator/Device configurations for the inter-device variability are A/1, A/2, and A/3. Therefore nine successful measurements were collected on each test subject for the inter-device evaluation. The measurement order of the five configurations was randomized. Subjects were repositioned between all scans. {7}------------------------------------------------ ## RESULTS Demographics of the enrolled study population are shown in the table below. The gender of study subjects was reasonably distributed and the age of the study subjects ranged from 21 years to 68 years, with a median age of 46 years. The range of Lens Fluorescence measurements obtained from these 27 precision study subjects was 0.07 to 0.33. | | n/N | % | |------------------------|-------|-------| | Gender | | | | Male | 14/27 | 51.9% | | Female | 13/27 | 48.1% | | Age (year) | | | | N | | 27 | | Mean | | 43.9 | | SD | | 14.9 | | Median | | 46.0 | | Min | | 21 | | Max | | 68 | | 20 - 29 | 6/27 | 22.2% | | 30 - 39 | 5/27 | 18.5% | | 40 - 49 | 6/27 | 22.2% | | 50 - 59 | 5/27 | 18.5% | | 60 - 69 | 5/27 | 18.5% | | Ethnicity | | | | Asian/Pacific Islander | 8/27 | 29.6% | | Caucasian | 16/27 | 59.3% | | Hispanic | 1/27 | 3.7% | | Multi-Race | 2/27 | 7.4% | ## DEMOGRAPHIC INFORMATION {8}------------------------------------------------ This table summarizes the precision data from each device and from all three devices combined. | Statistics | All Devices (3) | Devices 1 | Devices 2 | Devices 3 | |----------------------------|-----------------|-----------|-----------|-----------| | Number of Subjects | 27 | 27 | 27 | 27 | | Number of Successful Scans | 405 | 243 | 81 | 81 | | Mean | 0.17899 | 0.18029 | 0.17383 | 0.18025 | | Standard Deviation | 0.04718 | 0.04669 | 0.04865 | 0.04741 | | Repeatability SD' | 0.01034 | 0.01015 | 0.01170 | 0.01005 | | Repeatability %CV1 | 5.779 | 5.630 | 6.731 | 5.576 | | Reproducibility SD2 | 0.01153 | - | - | - | | Reproducibility %CV2 | 6.443 | - | - | - | ## PRECISION SUMMARY 1 Reveatability SD = Estimate of the standard deviation among measurements taken on the same eye using the same operator and device in the same testing session with repositioning. Repeatability %CV = Repeatability SD =Mean. x100. 2 Reproducibility SD = Estimate of the standard deviation among measurements taken on the same eve using different operators and devices, including repeatability. Reproducibility %CV = Reproducibility SD +Mean. ×100. No adverse events were observed during the study. ## STUDY CONCLUSION This clinical precision study showed the device to be safe and characterized well the variability of the measurements provided by the Clear Path DS-120 device for the range of 0.07-0.33. The variability of the autofluorescence measurements provided by the ClearPath DS-120 device shows the device is fit for its purpose as a measurement tool. The purpose of this study was not to establish the diagnostic or prognostic utility of the device; therefore, these results do not imply any correlation with a specific clinical diagnosis or prognosis. #### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE: 1.8 As described in this 510(k) Summary, all testing deemed necessary was conducted on the ClearPath DS-120® Lens Fluorescence Biomicroscope to ensure that the device is safe and effective for its intended use when used in accordance with its Instructions for Use. Minor differences between the Indications for Use statement and that of the predicate device do not alter the intended diagnostic effect or affect safety or effectiveness for the intended use. The minor differences in technical characteristics between the ClearPath DS-120® Lens Fluorescence Biomicroscope and the predicate Optos P200MAAF ophthalmoscope do not affect safety or effectiveness for the intended use. The physics of the measurement and basic system functions are the same as the Optos P200MAAF. The descriptive rationale for substantial equivalence is supported by bench and clinical testing as described above. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure embracing a globe, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the perimeter of the circle. January 31, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Freedom Meditech, Inc. % Mr. Craig H. Misrach, MBA, CPA Chairman & CEO 10455 Pacific Center Court San Diego, CA 92121 Re: K112880 Trade/Device Name: ClearPath DS-120® Lens Fluorescence Biomicroscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: January 25, 2013 Received: January 28, 2013 Dear Mr. Misrach: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {10}------------------------------------------------ Page 2 - Mr. Craig H. Misrach, MBA, CPA CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Y. Alexander -S for Malvina B. Eydelman, M.D. Director Division of Opthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ K112880 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ ClearPath DS-120® Lens Fluorescence Biomicroscope Device Name: Indications for Use: The ClearPath DS-120® Lens Fluorescence Biomicroscope is indicated for use to detect autofluorescence of the crystalline lens. Prescription Use AND/OR Over-The-Counter Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) CLC (Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...