Galilei G6 Lens Professional

K182659 · Sis Ag, Surgical Instrument Systems · MXK · Jul 25, 2019 · Ophthalmic

Device Facts

Record IDK182659
Device NameGalilei G6 Lens Professional
ApplicantSis Ag, Surgical Instrument Systems
Product CodeMXK · Ophthalmic
Decision DateJul 25, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2

Indications for Use

The Galilei G6 Lens Professional is designed to take images of the anterior segment of the eye, which includes cornea, iris, pupil, anterior chamber, and crystalline lens. To evaluate: • Corneal shape • Pachymetry (corneal thickness) • Position of the cornea relative to iris and lens • Anterior and posterior opacity • Anterior chamber angle • Anterior chamber depth • Volume of the anterior chamber • White-to-white distance • Pupil size • Condition and position of implants (e.g. IOL, phakic IOLs, intracorneal rings) • Location of cataracts (nuclear, sub capsular and or cortical), using cross slit imaging with densitometry • Condition of the lens (opaque crystalline lens) • Lens shape • Crystalline lens thickness The Galilei G6 Lens Professional is designed to additionally evaluate: • Axial length The Galilei G6 Lens Professional) also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.

Device Story

Galilei G6 Lens Professional is a tabletop ophthalmic diagnostic device for anterior segment imaging and biometry. It captures images using dual Scheimpflug cameras (45° angle) and a Placido disk (20 concentric rings). Inputs include optical reflections from the eye; the device reconstructs a 3D model of the anterior chamber. An integrated 'EBR Accessory' performs partial coherence interferometry (880 nm SLED) to measure axial length. Operated by clinicians in a clinic setting, the device provides automated measurements and IOL power calculations to assist in surgical planning. Output is displayed on a monitor to aid physicians in assessing corneal shape, pachymetry, cataract location, and lens condition. The device benefits patients by providing precise biometric data for intraocular lens selection and anterior segment evaluation.

Clinical Evidence

Prospective, observational, one-center study of 105 eyes (105 subjects, mean age 40.7 years) including normal eyes, cataracts, high myopia/hyperopia, and post-refractive surgery. Primary endpoints: inter-device/inter-operator repeatability and reproducibility, and agreement with Pentacam AXL. Results showed repeatability/reproducibility CVs generally <1% (comparable to Pentacam AXL). Agreement was substantially equivalent for all parameters except WtW, which is managed via device-specific offsets. No adverse effects observed.

Technological Characteristics

Tabletop device; Scheimpflug imaging (470 nm LED); Placido topography (750 nm LED); Partial coherence interferometry biometry (880 nm SLED, Laser Class 1/3B). Connectivity: PC-based. Software: Major level of concern. Standards: IEC 60601-1 (safety), IEC 60825-1 (laser safety), IEC 60601-2 (EMC), IEC 62304 (software lifecycle).

Indications for Use

Indicated for imaging and evaluation of the anterior segment of the eye (cornea, iris, pupil, anterior chamber, crystalline lens) and axial length measurement in patients requiring ocular assessment, including those with cataracts, high myopia, high hyperopia, or post-keratorefractive surgery conditions.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 25, 2019 SIS AG, Surgical Instrument Systems % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, CO 80127 ## Re: K182659 Trade/Device Name: Galilei G6 Lens Professional Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: MXK, HJO Dated: June 24, 2019 Received: June 25, 2019 ## Dear Kevin Walls: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Bradley Cunningham Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182659 Device Name Galilei G6 Lens Professional Indications for Use (Describe) The Galilei G6 Lens Professional) is designed to take images of the anterior segment of the eye, which includes cornea, iris, pupil, anterior chamber, and crystalline lens. To evaluate: - · Corneal shape - · Pachymetry (corneal thickness) - Position of the cornea relative to iris and lens - · Anterior and posterior opacity - · Anterior chamber angle - · Anterior chamber depth - · Volume of the anterior chamber - · White-to-white distance - · Pupil size - · Condition and position of implants (e.g. IOL, phakic IOLs, intracorneal rings) - · Location of cataracts (nuclear, sub capsular and or cortical), using cross slit imaging with densitometry - · Condition of the lens (opaque crystalline lens) - · Lens shape - · Crystalline lens thickness The Galilei G6 Lens Professional is designed to additionally evaluate: • Axial length The Galilei G6 Lens Professional) also performs calculations to assist physicians in determining the power of the intraocular lens for implantation. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Ziemer Ophthalmolgy on the left and a red abstract design on the right. Below the logo is a horizontal line, and below that is the text "Page 1/13". The logo is in grayscale, while the abstract design is in red. ## 510(k) Summary GALILEI Lens Professional #### l. SUBMITTER SIS AG, Surgical Instrument Systems Allmendstrasse 11 CH-2502 Port Switzerland Frank Ziemer, CEO Phone: +41 32 332 70 70 +41 32 332 70 71 Fax: Email: frank.ziemer@ziemergroup.com Contact Person: Mr. Kevin Walls Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127 Phone: 720-962-5412 Fax: 720-962-5413 Email: kevin@reginsight.com Date Prepared: July 24, 2019 #### DEVICE ll. Name of Device: Galilei G6 Lens Professional Common or Usual Name: G6. Galilei Classification Name: AC-powered slit lamp biomicroscope Requlatory Class: Class II Product Code: MXK, HJO Regulation Number: 886.1850 ## III. PREDICATE DEVICE The following two predicate devices were chosen: - Galilei G4 Dual Scheimpflug Analyzer (K051940) for topography/ tomography 1. measurements only and - 2. Pentacam® AXL (K152311) for topography/tomography as well as biometry measurements The Galilei G6 Lens Professional is (in hardware and software) identical to the Galilei G4 Dual Scheimpflug Analyzer but includes an additional biometry module (hardware and software) referred to as an "EBR Accessory" for evaluating axial length. The Galilei G4 Dual Scheimpflug Analyzer is already cleared by the FDA (K051940) and was audited by the FDA in 2017 without deviations. The Pentacam® AXL was chosen as a predicate device because it is the only FDA approved device on the market that has substantially similar measurement principles to the Galilei G6. Both the Galilei G6 Lens Professional and the Pentacam® AXL devices use a combination of Scheimpflyg topography/tomography and axial length measurements based on partial coherence interferometry. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Ziemer Ophthalmology, which includes the word "ziemer" in a sans-serif font, with the word "OPHTHALMOLOGY" in a smaller font underneath. To the right of the word "ziemer" is a stylized red swoosh. Below the logo is the text "Page 2/13", indicating that this is page 2 of a 13-page document. ## IV. DEVICE DESCRIPTION The Galilei G6 Lens Professional is (in hardware and software) identical to the Galilei G4 Dual Scheimpflug Analyzer but features an additional biometry module (hardware and software) referred to as "EBR Accessory". The Galilei G6 Lens Professional consists of the following functional units: - Measurement Head Container for the cameras, light sources, monitor drivers and electron-● ics. The Placido disk and Dual-Scheimpflug imaging is integrated into the Measurement Head, which performs a 180-degree rotation during data acquisition. - Main Monitor Display and navigation through the software, selection of functions. ● - PC box Container of the power supply and the computer. Periphery (Main monitor, mouse ● and keyboard) is connected directly to the computer. - . Elevation Table - Height-adjustable instrument table with locking wheels. - EBR Accessory Container of the EBR main printed circuit board as well as the optical and ● mechanical components such as a scanner and a partial coherence interferometer used for biometry measurements. Mounted inside the PC box. The GALILE! G6 device takes images of the anterior segment of the eye, which includes the cornea, iris, pupil, limbus, anterior chamber and crystalline lens. Topography and anterior segment tomography are calculated from those images. A pair of slit light images are recorded simultaneously with two cameras placed at opposite sides at an angle of 45°. Due to the Scheimpflug principle, an angled orientation of the camera's sensor allows a sharp focus over the entire image in spite of the 45° recording angle. The images are then analyzed and anterior cornea, anterior lens and iris surfaces are detected. This information is then used to reconstruct a three-dimensional model of the anterior chamber. Twenty (20) concentric rings in the Placido are reflected on the anterior surface of the cornea and recorded by a top-view camera in the center of the measurement head. The sizes and shapes of the recorded rings are used to calculate the curvature of the anterior surface of the eye. Both the Placido and Scheimpflug information are then merged to a single model of the eye. The EBR Accessory enables the Galilei G6 to take an optical A-scan by means of partial coherence interferometry. A beam of partial coherence infrared light is directed along the optical axis into the eye. Whenever it passes a transition between layers with different refractive indices (e.g., corneal surfaces, crystalline lens surfaces, retinal surfaces), a portion of the light is reflected back towards the reflected light is compared to a reference beam passing through a light path of adjustable optical length. The length of the reference arm is varied by a scanner. When the optical lengths of sample arm and reference arm match to within the coherence length of the partial coherence light source, an interference peak is detected and the corresponding layer within the eye is deduced. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Ziemer Ophthalmology. The logo consists of the word "ziemer" in a lowercase sans-serif font, with the word "OPHTHALMOLOGY" in a smaller font below it. To the right of the word "ziemer" is a stylized red "S" shape. Below the logo is the text "Page 3/13". #### INDICATIONS FOR USE V. The Galilei G6 Lens Professional is designed to take images of the anterior segment of the eye, which includes cornea, iris, pupil, anterior chamber, and crystalline lens. To evaluate: - Corneal shape ● - Pachymetry (corneal thickness) ● - Position of the cornea relative to iris and lens ● - Anterior and posterior opacity - Anterior chamber angle ● - Anterior chamber depth - Volume of the anterior chamber - White-to-white distance ● - Pupil size ● - Condition and position of implants (e.g. IOL, phakic IOLs, intracorneal rings) ● - Location of cataracts (nuclear, sub capsular and or cortical), using cross slit imaging with densitometry - Condition of the lens (opaque crystalline lens) - Lens shape - Crystalline lens thickness The Galilei G6 Lens Professional is designed to additionally evaluate: - Axial length ● The Galilei G6 Lens Professional) also performs calculations to assist physicians in determining the power of the intraocular lens for implantation. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES The extent of equivalence to the two predicate devices (see section III) is documented in the following tables {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Ziemer Ophthalmology, with the word "ziemer" in black, followed by "OPHTHALMOLOGY" in smaller letters below it. To the right of the word is a red abstract symbol. Below the logo is the text "Page 4/13" in black. #### Table 1: Comparison Table of Galilei G6 Lens Professional and Galilei G4 Dual Scheimpflug Analyzer | Characteristics | Galilei G6 Lens Professional | Galilei G4 Dual Scheimpflug Analyzer | Difference | |-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) number | K182659 (pending) | K051940 | n/a | | Intended use | The Galilei G6 Lens Professional is designed to take im-<br>ages of the anterior segment of the eye, which includes<br>cornea, iris, pupil, anterior chamber, and crystalline lens.<br>To evaluate:<br>• Corneal shape<br>• Pachymetry (corneal thickness)<br>• Position of the cornea relative to iris and lens<br>• Anterior and posterior opacity<br>• Anterior chamber angle<br>• Anterior chamber depth<br>• Volume of the anterior chamber<br>• White-to-white distance<br>• Pupil size<br>• Condition and position of implants (e.g. IOL, phakic<br>IOLs, intracorneal rings)<br>• Location of cataracts (nuclear, sub capsular and or<br>cortical), using cross slit imaging with densitometry<br>• Condition of the lens (opaque crystalline lens)<br>• Lens shape<br>• Crystalline lens thickness<br><br>The GALILEI G6 Lens Professional is designed to<br>additionally evaluate:<br>• Axial length | The Galilei G4 Dual-Scheimpflug Analyzer is a device<br>intended to take images of the anterior segment of the<br>eye, which includes cornea, iris, pupil, anterior chamber,<br>and lens, to evaluate and analyse:<br>• Corneal shape<br>• Lens shape<br>• Pachymetry (thickness of the cornea)<br>• Pupil size<br>• Lens thickness<br>• Condition of the lens<br>○ Location of cataracts (nuclear, subcapsular<br>and or cortical), using Scheimpflug slit<br>imaging with densitometry<br>○ State of the lens (opaque crystalline lens)<br>• Condition and position of implants (e.g. IOLs, phakik<br>IOLs, intracorneal rings)<br>• Anterior chamber (size, volume and angle)<br>• Scheimpflug Image<br>• Position of the cornea relative to iris and lens | No difference for meas-<br>urements of the anterior<br>chamber.<br><br>The Galilei G6 Lens<br>Professional additionally<br>evaluates axial length.<br>Safety and effective-<br>ness are addressed in a<br>clinical study with re-<br>spect to the second<br>predicate device. | | Measurement<br>principle<br>anterior<br>segment &<br>Topography | Scheimpflug photography for cross-sectional anterior im-<br>aging of the anterior chamber.<br><br>Scheimpflug photography and Placido ring imaging for<br>anterior corneal surface topography | Scheimpflug photography for cross-sectional anterior<br>imaging of the anterior chamber.<br><br>Scheimpflug photography and Placido ring imaging for<br>anterior corneal surface topography | No difference | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for "ziemer ophthalmology" on the left side of the image. To the right of the logo is a red abstract symbol. Below the logo and symbol is the text "Page 5/13". #### Galilei G4 Dual Scheimpflug Analyzer Difference Characteristics Galilei G6 Lens Professional Keratometry 4.5 - 13.5 mm (25 - 75 D) 4.5 - 13.5 mm (25 - 75 D) No difference range Imaging Digital camera Digital camera No difference Technology LED LED Source: Source: Scheimpflug No difference Wavelength: 470 nm Wavelength: 470 nm Source: LED Source: LED Placido No difference Wavelength: 750 nm Wavelength: 750 nm No difference Device concept Tabletop Tabletop Operator guided by monitor and joystick Operator guided by monitor and joystick Alignment No difference Operating No difference Distance 58 mm 58 mm No difference LED LED Source: Source: Fixation Target Wavelength: 617 nm Wavelength: 617 nm Power Re-No difference 110/220 VAC, 50/60 Hz 110/220 VAC, 50/60 Hz quirement {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Ziemer Ophthalmology, which includes the word "ziemer" in a bold, sans-serif font, with the word "OPHTHALMOLOGY" in a smaller font underneath. To the right of the wordmark is a stylized red graphic. Below the logo is the text "Page 6/13", indicating that this is page 6 of a 13-page document. Table 2: Comparison Table of Galilei G6 Lens Professional and Pentacam® AXL | Characteristics | Galilei G6 Lens Professional | Pentacam® AXL | Difference | Comment | |-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) number | K182659 (pending) | K152311 | n/a | n/a | | Intended use | The GALILEI G6 Lens Professional is designed to<br>take images of the anterior segment of the eye,<br>which includes cornea, iris, pupil, anterior cham-<br>ber, and crystalline lens. To evaluate:<br>Corneal shape Pachymetry (corneal thickness) Position of the cornea relative to iris and lens Anterior and posterior opacity Anterior chamber angle Anterior chamber depth Volume of the anterior chamber White-to-white distance Pupil size Condition and position of implants (e.g. IOL,<br>phakic IOLs, intracorneal rings) Location of cataracts (nuclear, sub capsular<br>and or cortical), using cross slit imaging with<br>densitometry Condition of the lens (opaque crystalline lens) Lens shape Crystalline lens thickness | The Pentacam® AXL is designed to take photos<br>of the anterior segment of the eye, which includes<br>cornea, pupil, anterior chamber and lens of the<br>eye. To evaluate:<br>corneal shape, analyze condition of the lens (opaque<br>crystalline lens), analyze the anterior chamber angle, analyze anterior chamber depth, analyze the volume of the anterior chamber, analyze anterior or posterior cortical opacity, analyze the location of cataracts (nuclear, sub<br>capsular and or cortical), using cross slit im-<br>aging with densitometry corneal thickness, axial length, white-to-white distance. | Minor<br>difference | The Pentacam® AXL<br>does not provide lens<br>thickness, lens shape<br>and pupil size.<br><br>Lens thickness, lens<br>shape and pupil size<br>are not used for in-<br>traocular lens calcu-<br>lation. The difference<br>doesn't alter the two<br>intended uses that<br>are new compared to<br>the G4 (evaluation of<br>axial length, deter-<br>mining the power of<br>the intraocular lens<br>for implantation) and<br>is therefore not ex-<br>pected to affect safe-<br>ty and effectiveness. | | | The GALILEI G6 Lens Professional is designed to<br>additionally evaluate:<br>Axial length<br>The Galilei G6 Lens Professional also performs<br>calculations to assist physicians in determining the<br>power of the intraocular lens for implantation. | The Pentacam® AXL also performs calculations<br>to assist physicians in determining the power of<br>the intraocular lens for implantation. | | | | Characteristics | Galilei G6 Lens Professional | Pentacam® AXL | Difference | Comment | | | Anterior Chamber | | | | | Measurement<br>principle an-<br>terior seg-<br>ment & To-<br>pography | Scheimpflug photography for cross-sectional<br>anterior imaging of the anterior chamber.<br>Scheimpflug photography and Placido ring im-<br>aging for anterior corneal surface topography | Scheimpflug photography for cross-sectional<br>anterior imaging of the anterior chamber<br>Scheimpflug photography for anterior corneal<br>surface topography | Minor<br>difference | The Pentacam® AXL<br>does not feature<br>Placido ring imaging for<br>conducting topography.<br>Anterior corneal surface<br>topography may be as-<br>sessed with either<br>technology. | | | | | | Safety and effective-<br>ness are addressed in<br>a clinical study with<br>respect to the second<br>predicate device. | | Keratometry<br>range | 4.5 - 13.5 mm (25 - 75 D) | 3 - 38 mm (9 - 99 D) | Minor<br>difference | Keratometry range of<br>both devices is suffi-<br>ciently large to cover<br>any corneal shape in<br>any human subject.<br>The difference is<br>therefore not ex-<br>pected to affect safety<br>and effectiveness. | | Imaging<br>Technology | Digital camera | Digital camera | No<br>difference | n/a | | Scheimpflug | Source: LED<br>Wavelength: 470 nm | Source: LED<br>Wavelength: 475 nm | No<br>difference | n/a | | Placido | Source: LED<br>Wavelength: 750 nm | n/a | Minor<br>difference | The Pentacam® AXL<br>does not feature<br>Placido ring imaging for<br>conducting topography.<br>Safety with respect to<br>eye safety is guaranteed<br>by compliance with the<br>respective standards, | | Characteristics | Galilei G6 Lens Professional | Pentacam® AXL | Difference | Comment | | Biometry | | | | | | Measurement<br>principle axial<br>biometry | Partial coherence interferometry | Partial coherence interferometry | No<br>difference | n/a | | Biometry | Source: SLED<br>Wavelength: 880 nm<br>Laser class: 1<br>Laser class embedded: 3B | Source: SLED<br>Wavelength: 880 nm<br>Laser class: 1<br>Laser class embedded: 3B | No<br>difference | n/a | | Axial length<br>range | 14 - 40 mm | 14 - 40 mm | No<br>difference | n/a | | Device | | | | | | Device concept | Tabletop | Tabletop | No<br>difference | n/a | | Alignment | Operator guided by monitor and joystick | Operator guided by monitor and joystick | No<br>difference | n/a | | Operating<br>Distance | 58 mm | 80 mm | Minor<br>difference | Exact focusing distance<br>depends on the respec-<br>tive imaging optics.<br>Both 58mm and 80mm<br>represent safe distanc-<br>es in terms of not acci-<br>dentally touching the<br>eye.<br>The difference is<br>therefore not ex-<br>pected to affect safety<br>and effectiveness. | | Fixation Target | Source: LED<br>Wavelength: 617 nm | Source: SLED<br>Wavelength: 880 nm | Minor<br>difference | Both wavelengths are<br>suitable for fixation as<br>they are visible to the<br>human eye under direct<br>fixation.<br>The difference is<br>therefore not ex-<br>pected to affect safety<br>and effectiveness. | | Characteristics | Galilei G6 Lens Professional | Pentacam® AXL | Difference | Comment | | Power Requirement | 110/220 VAC, 50/60 Hz | 110/220 VAC, 50/60 Hz | No difference | n/a | {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for Ziemer Ophthalmology. The logo consists of the word "ziemer" in a gray, sans-serif font, with the word "OPHTHALMOLOGY" in a smaller, red font underneath. To the right of the word "ziemer" is a stylized, red letter "S" that appears to be in motion. Page 7/13 {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for Ziemer Ophthalmology. The word "ziemer" is in a dark gray sans-serif font on the left side of the image. Below the word "ziemer" is the word "OPHTHALMOLOGY" in a smaller, red sans-serif font. To the right of the word "ziemer" is a red abstract symbol that resembles the letter "S". #### Page 8/13 {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for Ziemer Ophthalmology. The word "ziemer" is in lowercase, sans-serif font, and is in gray. Below the word "ziemer" is the word "OPHTHALMOLOGY" in a smaller, sans-serif font, and is also in gray. To the right of the word "ziemer" is a red, stylized "S" shape. {12}------------------------------------------------ Image /page/12/Picture/1 description: The image shows the logo for Ziemer Ophthalmology on the left side of the image. The logo is in black and red. The text "Page 10/13" is at the bottom of the image. #### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. #### Non-clinical testing #### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Galilei G6 Lens Professional. The device complies with the IEC 60601-1 and IEC 60825-1 standards for safety and 60IEC 60601-2 standard for EMC. #### Bench performance testing Bench performance tests demonstrating accuracy and precision of the measured parameters of the eye were performed with our device. #### Software Verification and Validation testing As recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", Software verification and validation was performed to confirm that the software performs as intended and the corresponding documentation was provided. The software for this device is rated as "major" level of concern, as inaccurately measured or displayed ocular parameters may mislead the physician in the clinical assessment. The device complies with the IEC 62304 for Medical Device Software life-cycle processes. #### Clinical Studies #### Comparison Galilei G6 Lens Professional with Pentacam® AXL #### Study description: A clinical study was conducted to support the substantial equivalence of the Galilei G6 Lens Professional to the predicate device, the Pentacam® AXL. Since the Galilei G4 Dual Scheimpflug Analyzer and Galilei G6 Lens Professional are identical except that the Galilei G6 Lens Professional has an additional EBR Module for measuring axial length, only the Galilei G6 Lens Professional and the Pentacam® AXL were included in the study. The primary objectives of this clinical study were to: - evaluate the inter-device repeatability, inter-operator repeatability and reproducibility of the 1. GALILEI G6 Lens Professional in repeated measurements of anterior segment geometry and axial intraocular distances and, - 2. demonstrate substantial equivalence through examining agreement of the GALILEI G6 Lens Professional measurements to those of the Pentacam® AXL. #### Studv design: The clinical study was a prospective, observational, one-center study. Of the total 117 enrolled and measured subjects, 12 subjects were discontinued from the study because they required repeat measurements that would have exceeded the recommended maximum number of daily measurements for patient safety and per protocol. Reasons for discontinuation included dense cataract, inability to hold gaze, see the fixation target or keep the eye open during the measurement, and excessive head or eve movements, all of which prevented extensive data collection with PAXL and Galliei G6. The mean age of the subjects involved in this study was 40.7±16.3 years. A total of 105 eyes of 105 subjects were assessed, 49 being right eyes and 56 being left eyes. Only one eye of each subject {13}------------------------------------------------ Image /page/13/Picture/1 description: The image shows the logo for Ziemer Ophthalmology. The logo consists of the word "ziemer" in a gray sans-serif font, with the word "OPHTHALMOLOGY" in a smaller font below it. To the right of the word "ziemer" is a red abstract shape. Below the logo is the text "Page 11/13". - 1) normal eyes (phakic eyes without cataracts or corneal disease), - 2) eyes with varying degrees of cataract, - eves with high myopia. 3) - 4) eyes with high hyperopia, - 5) eves with post-keratorefractive surgery condition. Additional 5 analyzed eyes consisted of 2 eyes with severe keratoconus and 3 eyes with prior crosslinking treatment. Of the 105 assessed subjects, 63 were males and 42 were females. Ethnicity was composed of 100 Non-Hispanics and 5 Hispanics. Race consisted of 98 Whites, 5 Asians, and 2 American/Alaskan Natives. The 20 patients with cataract who completed all measurements included not only eyes with moderate cataract, but also eyes with severe cataract with grades as high as 4 in the Lens Opacification Classification System (LOCS) III, in order to ensure that cataracts of various severity grades, particularly of severe cataract, can be accurately and precisely assessed with the Galilei G6 In the LOCS III, cataract severity ranges from grade 1 (very early stage) to grade 5 (severe). #### Results: - In terms of agreement, the Galilei G6 demonstrates substantially equivalent performance to the predicate device, except for WtW. - . WtW is known to be different between state-of-the-art devices, and device specific offsets are commonly used in clinical practice to account for inter-device differences. Inter-device differences between the state-of-the-art devices and Galilei G6 are substantially equivalent to those between state-of-the-art devices and the PAXL. - Repeatability and reproducibility attained with the Galilei G6 are substantially equivalent to ● those attained with the PAXL. - . Minor technological differences between the Galilei G6 and the Pentacam AXL raise no new issues of safety or effectiveness. - . No adverse effects and complications were observed or are to be expected. #### Summary: Performance data demonstrates that the Galilei G6 Lens Professional is as safe and effective as the predicate device, the Pentacam® AXL. Minor technological differences between the Galilei G6 and the predicate device raise no new issues of safety or effectiveness. No adverse effects or complications were observed or are to be expected. #### In-house Precision Testing: Results An internal clinical study was conducted - to evaluate precision in terms of repeatability (intra-device, intra-operator) and reproducibility (inter-1) device, inter-operator) of the GALILEI G6 Lens Professional for repeated measurements and - 2) to compare the precision of the Galilei G6 Lens Professional to the reported values of the Pentacam® AXL for normal eyes. Anterior segment geometry and axial intraocular distances under assessment included: - 1. AL - 2. ССТ - 3. R flat - 4. R steep - 5. Rm - 6. CC - 7. A flat - 8. AC…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...