V-VITFREEZE is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. V-VITWARM is intended for use in the thawing of vitrified of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Device Story
V-VITFREEZE and V-VITWARM are sets of reproductive media used in IVF laboratories for cryopreservation and thawing of oocytes and embryos. V-VITFREEZE (Equilibration and Vitrification solutions) uses cryoprotectants (ethylene glycol, DMSO, trehalose) to replace water in cells prior to liquid nitrogen storage. V-VITWARM (Thawing, Diluent, and Washing solutions) uses a stepwise process to remove cryoprotectants post-thaw. Used by embryologists in clinical settings; output is the successful preservation or recovery of viable embryos/oocytes for subsequent transfer. Benefits include enabling embryo storage for future fertility treatment.
Clinical Evidence
Bench testing only. Includes aseptic filtration validation (ISO 13408), radiation sterilization validation (ISO 11137), shelf-life testing (ASTM F1980), transportation testing (ASTM D4169), and Mouse Embryo Assay (MEA) showing ≥ 80% development to expanded blastocyst at 96 hours.
Technological Characteristics
Media components: water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, hydroxypropyl cellulose, polymeric substance, gentamicin, phenol red. Cryoprotectants: ethylene glycol, DMSO, trehalose. Packaging: polypropylene vials with high-density polyethylene caps. Sterilization: aseptic filtration; vials sterilized via radiation. Shelf-life: 12 months at 2-8°C.
Indications for Use
Indicated for vitrification and thawing of oocytes (MII), pronuclear (PN) zygotes, day 3 cleavage stage embryos, and blastocyst stage embryos in IVF procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
K190152 — Vit Kit- Freeze NX and Vit Kit- Warm NX · Fujifilm Irvine Scientific, Inc. · Jun 21, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2024
VITROMED GmbH c/o Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92606
Re: K240176 Trade/Device Name: V-VITFREEZE and V-VITWARM Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Received: August 8, 2024
Dear Greg Holland:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael T. Bailey -S
For
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240176
Device Name V-VITFREEZE and V-VITWARM
Indications for Use (Describe)
V-VITFREEZE is intended for use in the vitrification of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
V- VITWARM is intended for use in the thawing of vitrified of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">□</span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K240176
#### SUBMITTER I.
| Applicant: | VITROMED GmbH |
|---------------------|-----------------------------------------------|
| Address: | Raiffeisenstr. 15a, 40764 Langenfeld, Germany |
| Phone: | +49 2173-20041-30 |
| Fax: | +49 2173-20041-58 |
| Contact Person: | Deepesh Changat |
| Email: | qm@vitromed.com |
| Consultant Contact: | Greg Holland, Regulatory Specialists, Inc. |
| Email: | greg@regulatoryspecialists.com |
Date Prepared: September 10, 2024
#### II. DEVICE
| Trade Name: | V-VITFREEZE; V-VITWARM |
|--------------------|-----------------------------------|
| Common Name: | Assisted Reproduction Media |
| Regulation Name: | Reproductive Media and Supplement |
| Regulation Number: | 884.6180 |
| Product Code: | MQL (Media, Reproductive) |
| Regulatory Class: | II |
#### III. PREDICATE DEVICE
Vitrification Kit and Thawing Kit (K171748) from Kitazato Corporation. The predicate device has not been subject to a design-related recall.
### DEVICE DESCRIPTION IV.
V-VITFREEZE AND V-VITWARM include a set of five media intended for use in the vitrification and thawing of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos as part of InVitro Fertilization (IVF) procedures.
V-VITFREEZE (2mL vials) includes two media components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol, trehalose (only in VS), and dimethyl sulfoxide. During the vitrification process, embryos are first exposed to ES and then to VS.
{5}------------------------------------------------
Using this methodology, the permeating cryoprotectants can replace water in the oocyte or PN through blastocyst stage embryos prior to vitrification and storage in liquid nitrogen.
V-VITWARM (4mL vials) is composed of three media used stepwise for thawing and removing cryoprotectants from vitrified oocytes and PN through blastocyst stage embryos. It is composed of Thawing Solution (TS), Diluent Solution (DS) and Washing Solution (WS).
The base medium for the vitrification and thawing media includes water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, hydroxypropyl cellulose, polymeric substance, gentamicin, and phenol red. Cryoprotectants in the media include ethylene glycol, (7.5% in ES and 15% in VS), DMSO (7.5% in ES and 15% in VS), and trehalose in VS, TS, and DS.
The five solutions in the V-VITFREZE and V-VITWARM kit are aseptically filtered (storage vials sterilized by radiation) and provided in polypropylene vials with high density polyethylene screw caps. They have a shelf-life of 12 months when stored at 2-8℃. Media in the vials can be used for up to seven days after vial opening.
#### V. INDICATIONS for USE
V-VITFREEZE is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
V-VITWARM is intended for use in the thawing of vitrified of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
### VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A comparison of the intended use and technological features of the subject and predicate devices are described in the table below:
| Comparison Item | K240176<br>Subject Device | K171748<br>Predicate Device | Comparison |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | V-VITFREEZE is intended<br>for use in the vitrification of<br>oocytes (MII), pronuclear<br>(PN) zygotes through day 3 | The Vitrification Kit is<br>indicated for use in the<br>preparation, vitrification,<br>and storage of oocytes<br>(MII), pronuclear (PN) | There are differences in<br>the wording of the<br>indications for use<br>statements for the subject<br>and predicate device; |
| Comparison Item | K240176<br>Subject Device | K171748<br>Predicate Device | Comparison |
| | cleavage stage embryos and<br>blastocyst stage embryos.<br><br>V-VITWARM is intended for<br>use in the thawing of vitrified<br>of oocytes (MII), pronuclear<br>(PN) zygotes through day 3<br>cleavage stage embryos and<br>blastocyst stage embryos. | zygotes through day 3<br>cleavage stage embryos,<br>and blastocyst stage<br>embryos.<br><br>The Thawing Kit is<br>indicated for use in the<br>preparation and thawing of<br>vitrified oocytes (MII),<br>pronuclear (PN) zygotes<br>through day 3 cleavage<br>stage embryos, and<br>blastocyst stage embryos. | however, the intended<br>uses of the subject and<br>predicate device media<br>are the same (i.e.,<br>vitrification and warming<br>of oocytes and embryos). |
| Conditions for Use | Prescription Use Only | Prescription Use Only | Same |
| Kit components | V-VITFREEZE consists of<br>Equilibration Solution (ES)<br>and Vitrification Solution<br>(VS). V-VITWARM includes<br>TS (Thawing Solution), DS<br>(Dilution Solution) and WS<br>(Washing Solution). | The Vitrification Kit<br>includes three media<br>components, Basic<br>Solution (BS),<br>Equilibration Solution (ES)<br>and Vitrification Solution<br>(VS). The Vitrification Kit<br>comes prepackaged with 4<br>Cryotop devices (Cryotop<br>CL, Cryotop SC, or<br>Cryotop US), and 2 Repro<br>Plates.<br><br>The Thawing Kit is<br>composed of TS<br>(Thawing Solution), DS<br>(Dilution Solution) and<br>WS (Wash Solution). The<br>Thawing Kit comes pre-<br>packaged with one Repro | The predicate device<br>includes an additional<br>basic solution component,<br>equipment to conduct the<br>procedure (plates), and a<br>closed cryopreservation<br>storage device that are<br>different than the subject<br>device. These differences<br>do not raise different<br>questions of S&E. |
| Comparison Item | K240176<br>Subject Device | K171748<br>Predicate Device | Comparison |
| | | Plate, and two 35 mm<br>dishes. | |
| Freeze Kit Formulation | Base medium (see Device<br>Description section of<br>summary) with<br>cryoprotectants Ethylene<br>Glycol, DMSO, and<br>Trehalose | Medium 199, Ethylene<br>Glycol, DMSO, Sucrose,<br>Dextran Serum<br>Supplement, Gentamicin | Different: The<br>formulations of the<br>subject and predicate<br>devices are not the same.<br>Differences in device<br>formulations do not raise<br>different questions of<br>safety and effectiveness<br>(S&E). |
| Thaw Kit Formulation | Base medium (see Device<br>Description section of<br>summary) with<br>cryoprotectant Trehalose | Medium 199; Sucrose,<br>Dextran Serum<br>Supplement, Gentamicin | Different: The<br>formulations of the<br>subject and predicate<br>devices are not the same.<br>Differences in device<br>formulations do not raise<br>different questions of<br>S&E. |
| pH | WS; DS; TS: 7.20 - 7.45<br>VS: 7.50 - 7.70<br>ES: 7.30 - 7.60 | 7.2-7.6 | Different: The subject<br>device has a higher pH<br>range than the predicate<br>device. These differences<br>in pH specifications do<br>not raise different<br>questions of S&E. |
| Osmolality (mOsm/kg) | ES - 2300-3100 (1:10<br>dilution)<br>VS: 4600-6150 (1:20<br>dilution)<br>TS: 1250-1600 (1:4 dilution)<br>DS - 730-930 (1:2 dilution)<br>WS - 240-300 (No dilution) | Not available publicly | Different: The subject<br>device and predicate<br>devices have differences<br>in osmolality<br>specifications. These<br>differences in osmolality<br>specifications do not raise |
| Comparison Item | K240176<br>Subject Device | K171748<br>Predicate Device | Comparison |
| Bacterial Endotoxin | < 0.25 EU/mL | ≤ 0.25 EU/mL | Same |
| Mouse Embryo Assay | 1-Cell MEA: ≥ 80% embryos<br>developed to expanded<br>blastocyst at 96 hours. | 1-Cell MEA: ≥ 80%<br>expanded blastocyst after<br>96 hours | Same |
| Sterilization Method | Aseptic Filtration<br>Vials are sterilized via<br>radiation | Aseptic Filtration<br>Vials are sterilized via<br>radiation | Same |
| Shelf-Life | 12 months | 12 months | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# K240176 Page 4 of 6
{8}------------------------------------------------
As shown in the table above, there are differences in the indications for use statements and technological characteristics of the subject and predicate devices. However, as stated in the differences in indications for use do not represent a new intended use and the differences in technological characteristics do not raise different questions of safety and effectiveness.
#### VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
The following studies have been conducted in support of the substantial equivalence to the predicate device.
- Aseptic filtration and aseptic filling validation, per ISO 13408-1:2008 & A1:2013 and ISO 13408-● 2:2018.
- Vial radiation sterilization, per ISO 11137-1:2006 (including: Amendment 1 (2013) and Amendment ● 2 (2018)) Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2:2013 -Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose [Including Amendment 1 (2022)].
- Shelf-life testing was conducted to support a 12-month shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after accelerated aging in accordance with ASTM F1980-21:
- Appearance: All of the solutions should be pink to rose color without precipitates A
- > pH per USP <791>: WS, DS, TS: 7.20 7.45; VS: 7.50 7.70; ES: 7.30 7.60
{9}------------------------------------------------
- ≫ Osmolality per USP <785>: ES 2300-3100 mOsm/kg (1:10 dilution), VS: 4600-6150 mOsm/kg (1:20 dilution), TS: 1250-1600 mOsm/kg (1:4 dilution), DS - 730-930 mOsm/kg (1:2 dilution), WS - 240-300 mOsm/kg (No dilution)
- Sterility per USP <71>: No microbial growth A
- > Bacterial endotoxin per USP <85>: < 0.25 EU/mL
- A MEA per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: 1-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours.
- Transportation testing per ASTM D4169-22 and cap/seal leak testing using a method equivalent to ● USP <1207.2> on transportation-conditioned devices.
#### VIII. CONCLUSION
The results of the performance testing described above demonstrate that V-VITFREEZE and V-VITWARM kit is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.