← Product Code [MQK](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK) · K180740

# VitriGuard (K180740)

_Origio A/S · MQK · May 4, 2018 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K180740

## Device Facts

- **Applicant:** Origio A/S
- **Product Code:** [MQK](/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK.md)
- **Decision Date:** May 4, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.6160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

## Device Story

VitriGuard is a single-use, disposable cryopreservation storage device used in assisted reproduction laboratories. It consists of a two-piece polystyrene assembly comprising a hexagonal-shaped stick and a cap. During vitrification, embryos are loaded onto a trough on the stick tip; the cap is then applied to create a seal. The device is plunged into liquid nitrogen for cryopreservation. Visual markings on the stick and tip assist with orientation. The device is used by embryologists to store embryos, facilitating subsequent thawing and clinical use. It provides a controlled environment for embryo containment during the vitrification process.

## Clinical Evidence

No clinical data. Bench testing included Mouse Embryo Assay (MEA) on unaged and one-year accelerated aged devices (ASTM F1980-07), demonstrating ≥80% blastocyst development at 96 hours. Additional testing included sterilization validation, bioburden testing, shipping/distribution (ISTA 3A 2008), and package integrity (ASTM F1929-15, ASTM F88/F88M-15, ASTM F1886/F1866M-09).

## Technological Characteristics

Polystyrene construction; two-piece stick and cap assembly with taper seal; hexagonal stick geometry; black marker bands for orientation. Sterilization via radiation (SAL 10^-6). Endotoxin limit ≤2.0 EU/device. Cooling rate: -2,271°C/min; Warming rate: 36,377°C/min. Shelf life: one year.

## Regulatory Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- ORIGIO a/s VitriGuard ([K162833](/device/K162833.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 4, 2018

ORIGIO a/s % Roaida Johnson Director, RA New Product Development CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611

Re: K180740

Trade/Device Name: VitriGuard™ Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: April 24, 2018 Received: April 25, 2018

Dear Roaida Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Form Approved: OMB No. 0910-0120 |
|----------------------------------|
| Expiration Date: 06/30/2020      |
| See PRA Statement below.         |

**DEPARTMENT OF HEALTH AND HUMAN SERVICES**  

Food and Drug Administration  

**Indications for Use**

| 510(k) Number (if known)       | K180740                                                                                                                                                                   |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | VitriGuard™                                                                                                                                                               |
| Indications for Use (Describe) | VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. |

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRASstaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

PSC Publishing Services (301) 443-6740

000009 EF

»

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "origio", which is written in a stylized purple font. To the right of the logo, it says "a CooperSurgical company" in a smaller, also purple, font. The logo is simple and modern.

# 510(k) Summary K180740

#### Submitter Information

| Company Name:    | ORIGIO a/s                             |
|------------------|----------------------------------------|
| Company Address: | Knardrupvej 2<br>2760 Måløv<br>Denmark |
|                  | Telephone: +45 46 79 02 00             |
|                  | Fax: +45 46 79 03 02                   |

## Submission Correspondent Information

| Company Name:    | CooperSurgical Inc.                      |
|------------------|------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611 |
| Telephone:       | 203-601-5200 Ext. 3325                   |
| Fax:             | 203-601-9870                             |

Contact Person: Roaida Johnson

Date Prepared: May 3, 2018

## Device Information

| Trade Name:        | VitriGuard™                          |
|--------------------|--------------------------------------|
| Common Name:       | Cryopreservation Storage Device      |
| Regulation Number: | 21 CFR 884.6160                      |
| Regulation Name:   | Assisted Reproduction Labware        |
| Product Code:      | MQK (Labware, Assisted Reproduction) |
| Regulatory Class:  | II                                   |

#### Predicate Device Information

ORIGIO a/s VitriGuard (K162833).

The predicate device has not been subject to a design-related recall.

#### Device Description

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. It is single-use, disposable, and provided sterile. The device consists of a two-piece polystyrene assembly that includes a hexagonalshaped stick and cap. As part of the vitrification procedure, the embryos to be stored are loaded on the tip, and capped prior to plunging the device in liquid nitrogen and for subsequent storage. The tip has a trough 

{4}------------------------------------------------

area for loading, maintaining, and securing the embryos. The stick and cap include a taper design that creates a seal when assembled. Markings at the end of the stick and the tip of the device provide visual aid for proper device orientation.

Product specifications are listed in Table 1 below:

## Table 1: Product Specifications

| Parameter                | Specification          |
|--------------------------|------------------------|
| Cooling rate             | -2,271°C/min           |
| Warming rate             | 36,377°C/min           |
| Sterilization            | Radiation, SAL 10-6    |
| Endotoxin                | ≤2.0 EU/device         |
| Mouse Embryo Assay (MEA) | ≥80% blastocyst at 96h |

## Indications for Use

VitriGuard is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.

## Substantial Equivalence Discussion

Table 2 provides a comparison of the subject and predicate device.

| Property                             | Subject VitriGuard                                                                                                                                                                                                                  | Predicate VitriGuard                            |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| 510(k) Number                        | K180740                                                                                                                                                                                                                             | K162833                                         |
| Indications for Use                  | VitriGuard is a cryopreservation<br>storage device that is intended for use<br>in vitrification procedures to contain<br>and maintain human 4-8 cell and<br>blastocyst stage embryos                                                | Same                                            |
| Fundamental<br>Scientific Technology | The device is composed of a two-piece<br>assembly that includes a stick and cap.<br>As part of the vitrification procedure,<br>the embryos to be stored are loaded on<br>the tip of the stick and capped for<br>subsequent storage. | Same                                            |
| Material                             | Polystyrene (clear, blue, green, yellow,<br>orange, pink, lime green, and purple)                                                                                                                                                   | Polystyrene (clear, blue, green,<br>and yellow) |
|                                      | Black marker bands                                                                                                                                                                                                                  | Black marker bands                              |
| Sterilization Method                 | Radiation                                                                                                                                                                                                                           | Same                                            |
| Number of Uses                       | Single Use, Disposable                                                                                                                                                                                                              | Same                                            |
| Packaging                            | Box of 20 (individually pouched) or<br>box of 50 (5 per pouch)                                                                                                                                                                      | Box of 20 (individually<br>pouched)             |
| Shelf Life                           | One Year                                                                                                                                                                                                                            | Three Years                                     |

Table 2: Comparison of the Subject VitriGuard to the Predicate VitriGuard

The subject and predicate devices have intended use and are technologically comparable. Differences between the predicate and subject device includes the subject device having additional color 

{5}------------------------------------------------

options of the stick component, a package configuration including five VitriGuard devices in a pouch, and a reduced shelf-life as compared to the predicate device. These differences do not raise different questions of safety and effectiveness as compared to the predicate device.

# Non-Clinical Performance Testing

As part of demonstrating substantial equivalence to the predicate, the following tests were performed:

- . MEA was performed on VitriGuard devices that were unaged and after one year of accelerated aging per ASTM F1980-07(2011). One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group. The subject device met the predetermined acceptance criteria (see Table 1).
- Risk assessment to apply sterilization validation and bioburden testing on the predicate device to the subject device
- Shipping and Distribution Testing per ISTA 3A 2008
- . Package Integrity Testing Following Accelerated Aging
	- ASTM F1980-07(2011) o
	- O ASTM F1929-15
	- ASTM F88/F88M-15 O
	- o ASTM F1886/F1866M-09(2013)

# Conclusion

The results of the performance testing provided demonstrated that the subject devices are as safe and effective as the predicate devices and supports substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K180740](https://fda.innolitics.com/submissions/OB/subpart-g%E2%80%94assisted-reproduction-devices/MQK/K180740)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
