The intrauterine pressure monitoring catheter (IUP) is for use on patients requiring intrapartum intrauterine monitoring, amniofusion, and sampling of amniotic fluid.
Device Story
Intrauterine pressure transducer catheters (IUP-TipTrans and IUP-ProxiTrans) monitor pressure during labor; input consists of intrauterine pressure waves. TipTrans model features distal tip transducer; ProxiTrans model uses fluid column to transmit pressure to proximal transducer. Both utilize thick film/ceramic strain gage technology. Devices include fluid infusion/sampling lumen and electronic zeroing switch for calibration without removal. Catheters are soft, flexible polyurethane with blunt tips to prevent perforation; supplied with split-sheath polypropylene introducer and thigh-attachment device. Used in clinical labor/delivery settings by physicians or nurses. Output is electrical signal transmitted via cable to fetal monitor; enables continuous pressure monitoring to guide clinical management of labor and patient safety.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: polyurethane (catheter), polypropylene (introducer). Sensing: thick film/ceramic strain gage. Energy: electrical (transducer). Form factor: catheter with distal tip or proximal transducer, fluid lumen, and connector. Connectivity: wired to fetal monitor. Sterilization: not specified.
Indications for Use
Indicated for patients requiring intrapartum intrauterine pressure monitoring, amniofusion, and amniotic fluid sampling.
Regulatory Classification
Identification
An intrauterine pressure monitor is a device designed to detect and measure intrauterine and amniotic fluid pressure with a catheter placed transcervically into the uterine cavity. The device is used to monitor intensity, duration, and frequency of uterine contractions during labor. This generic type of device may include the following accessories: signal analysis and display equipment, patient and equipment supports, and component parts.
K974730 — SOFTRANS PLUS INTRAUTERINE PRESSURE CATHETER SYSTEM (IUP 5000, SOFTRANS PLUS IUPC) · Graphic Controls Corp. · Jul 14, 1998
Submission Summary (Full Text)
{0}
scientific device manufacturer, inc.
SCD
OCT 15 1996
999 Andersen Drive, Suite 110
San Rafael, California 94901
Phone 415-454-9370 Fax 415-454-9380
# Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: 1290
## Applicant Information:
Date Prepared: April 1, 1996
Name: Scientific Device Manufacturer, LLC.
Address: 999 Andersen Drive, Suite 120
San Rafael, California 94901
Contact Person: Richard C. Ball
Phone Number: 415-454-9370
Fax Number: 415-454-9380
## Device Information:
Trade Name: SDM IUP-TipTrans and IUP-ProxiTrans
Common Name: Intrauterine Pressure Transducer Catheter
Classification Name: Intrauterine pressure Transducer
## Equivalent Device:
IntranPlus (Utah Medical)
Fetal Monitoring Intrauterine Pressure Kit, #14099C (Hewlett Packard)
## Classification:
In preparation of this PreMarket Notification, it was determined that devices of this generic type have been previously classified as Class II devices.
## Product Description:
This 510(k) Notification is being submitted prior to marketing introduction of Scientific Device Manufacturer's Intrauterine Tip Transducer Catheter and its insertion device. Two models will be offered, one with a pressure transducer located at the proximal end using a fluid column to transmit pressure waves, the other has the pressure transducer located in the distal tip. In both cases the transducer is a thick film/ceramic strain gage connected to a proximal connector. The connector interfaces with a reusable cable which is designed for proper connection to the monitor being used. Both devices will also have a separate fluid infusion and sampling lumen terminating in the side of the catheter near the distal end and integral to the connector at the proximal end.
{1}
sdm
scientific device manufacturer, inc.
999 Andersen Drive, Suite 110
San Rafael, California 94901
Phone 415-454-9370 Fax 415-454-9380
Incorporated in the tip transducer catheter is a signal interruption switch for electronic zeroing of the system following insertion and prior to commencement of use. This switch is also used for periodic rezeroing of the system during use without the need for removal and replacement of the catheter.
Both catheter tips have rounded, blunt ends while the catheter material is soft, flexible polyurethane. This design minimizes the possibility of uterine perforation while at the same time providing maximum comfort for the patient.
The catheter introducer, made of a somewhat stiffer polypropylene, is designed for easy insertion of the flexible catheter. It is of a split sheath design for simple removal following insertion. Also, a hook and loop attachment device is supplied for fixing the connector end of the catheter to the thigh following placement in order to prevent inadvertent dislodgement during use.
## Intended Use:
The intrauterine pressure monitoring catheter (IUP) is for use on patients requiring intrapartum intrauterine monitoring, amniofusion, and sampling of amniotic fluid.
## Comparison To Predicate Device:
Scientific Device Manufacturer's Intravascular Pressure Transducer Catheters are substantially equivalent to currently marketed IUP devices with regard to intended use, materials, dimensions and mechanical properties. Predicate devices include those made by Utah Medical, Corometrics, and Hewlett Packard. No significant changes or modifications were made from those predicate devices. SDM, LLC, therefore posits that its devices are equivalent in safety and effectiveness to those devices.
Signed: Richard C. Ball
Name: Richard C. Ball
Position: VP Regulatory Affairs and Quality Assurance
Date: April 1, 1996
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.