This catheter is for use on patients requiring intrapartum, intrauterine pressure monitoring.
Device Story
Disposable, sterile, single-patient-use intrauterine pressure monitoring system; used during labor and delivery. Device features pressure sensor at catheter tip, soft elastomeric tip to reduce perforation risk, and lumen with female luer connector for amniotic fluid access. Operated by clinicians in clinical settings. Provides real-time intrauterine pressure signal to external monitoring equipment via external zeroing. Soft tip design enhances patient safety; disposable nature prevents cross-contamination. Clinical utility includes monitoring pressure, accessing amniotic space, and maintaining signal integrity for labor management.
Clinical Evidence
Clinical studies and physician evaluations conducted. Results demonstrated the device is as effective or more effective than predicate devices regarding intrauterine pressure signal integrity, ease of use, amniotic space access, device retention, and zeroing/drift signal stability. Bench testing on fresh placenta confirmed lower perforation risk compared to predicates. Mechanical integrity and biocompatibility verified via independent laboratory testing.
Indicated for patients requiring intrapartum, intrauterine pressure monitoring.
Regulatory Classification
Identification
An intrauterine pressure monitor is a device designed to detect and measure intrauterine and amniotic fluid pressure with a catheter placed transcervically into the uterine cavity. The device is used to monitor intensity, duration, and frequency of uterine contractions during labor. This generic type of device may include the following accessories: signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
MX 4042 Series Intrauterine Pressure Catheter (Medex)
K982565 — EASY TRANS TIP · Carl Schuh · Sep 22, 1998
K961290 — IUP - TIP TRANS, IUP - PROXI TRANS · Scientific Device Manufacturer, LLC · Oct 15, 1996
Submission Summary (Full Text)
{0}
K954955
K954955
# 510(k) Summary
4PR 17 1996
Submitter: Clinical Innovations, Inc.
Name: Wm. Dean Wallace
Address: 6477 So. Cottonwood St., Murray, UT 84107
Telephone: (801) 268-8200
Trade Name: STORC Intrauterine Pressure Catheter
Common Name: Intrauterine Pressure Catheter
Classification Name: Catheter, Intrauterine and Introducer
The legally marketed devices to which equivalence is claimed include the MX 4042 Series Intrauterine Pressure Catheter from Medex, the Meditrace Disposable Intrauterine Catheter from Graphic Controls, the Intrauterine Catheter from Telos, and the IUC - Intrauterine Catheter (INTRAN) from Utah Medical Products.
Description of the device: STORC™ is a Disposable Intrauterine Pressure Monitoring System. This product is a sterile, single patient use catheter.
Intended use: This catheter is for use on patients requiring intrapartum, intrauterine pressure monitoring.
The STORC Intrauterine Pressure Catheter is substantially equivalent to the predicate devices because:
It has the same intended uses, namely,
- intrauterine pressure measurement and monitoring
- port for amniotic fluid access,
It has the same basic technological characteristics as predicate devices, namely,
- pressure sensor located at catheter tip,
- soft tip for lower risk of perforation,
- markings for catheter insertion,
- low-cost disposable in order to avoid cross contamination,
- port and lumen with female luer connector for access to amniotic space,
- external zeroing of pressure.
It uses the same or similar materials, all of which have been shown to be biocompatible and to function well in the intended application,
{1}
The soft elastomeric tip is a further advantage that adds safety.
The safety and effectiveness are similar or better than existing devices as demonstrated in the laboratory and clinical testing.
## Safety
Laboratory testing has shown that in the following areas the STORC Intrauterine Pressure Catheter is safe:
- Perforation: bench testing on fresh placenta showed that the STORC is less likely to perforate than predicate devices.
- Mechanical integrity: laboratory testing and basic design assure that no parts will come loose and be left in the patient.
- Biocompatibility and infection: independent lab testing show that the materials used in the STORC are safe for use in this application.
In addition, clinical studies including physician evaluation verified the safe nature of the STORC Intrauterine Pressure Catheter.
## Effectiveness
The clinical trials showed that the STORC Intrauterine Pressure Catheter was as effective or more effective than the predicate devices in the following areas:
- integrity of intrauterine pressure signal
- ease of use
- amniotic space access
- device retention
- zeroing and drift signal stability
The laboratory testing verified the performance standards in terms of sensor accuracy, mechanical integrity, and overall performance.
Wm. Dean Wallace, M.D., Ph.D.
10-23-95
Date
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.