The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.
Device Story
The OPTO-IFL1000 is a CO2 insufflation device used in clinical settings to create and maintain pneumoperitoneum during laparoscopic surgery. It injects CO2 gas into the abdominal cavity to distend the wall and create an operating space. The device is operated by healthcare professionals; it monitors intra-abdominal pressure and gas flow, providing visual and audible alarms for overpressure or overheating conditions. It supports continuous leakage compensation and includes an optional heating function via compatible insufflation tubes (OPTO-T1000H). The device output assists surgeons in maintaining a stable surgical field of view. Benefits include controlled abdominal distension for improved visualization and access during minimally invasive procedures.
Clinical Evidence
Bench testing only. Performance testing verified gas supply, pressure accuracy, leakage compensation, overpressure/overheating alarms, flow accuracy, and heating function. Compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2, and IEC 60601-1-8 was demonstrated. Software verification and validation were performed for a 'Major' level of concern.
Technological Characteristics
CO2 insufflator; pressure range 5-25 mmHg; power 110-240 V~; dimensions 370mm x 410mm x 166.5mm; weight 11kg. Features include visual/audible alarms for overpressure (>4 mmHg above nominal) and overheating (>41°C). Reprocessing via steam sterilization. Connectivity includes compatible insufflation tubes (OPTO-T1000/T1000H).
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic laparoscopic procedures requiring pneumoperitoneum via abdominal cavity gas distension.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
K090652 — ALPHA DUO LAP INSUFFLATOR · Hippokratec GmbH · Sep 25, 2009
K070783 — SOPRO MODEL 640 LAPAROSCOPIC INSUFFLATOR · Sopro · Dec 13, 2007
K153513 — Insufflator 50L FM134 · W.O.M. World of Medicine GmbH · Mar 4, 2016
K120151 — NEBULAE I · Northgate Technologies, Inc. · Aug 17, 2012
K962683 — SNOWDEN-PENCER COMPUTERIZED HIGH FLOW INSUFFLATOR · Deknatel, Inc. · Jan 10, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2023
Guangdong OptoMedic Technologies, Inc. Weijuan Guo Regulatory Affairs Engineer Suite 503, Building A. Golden Valley Intellicreation Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan, 528200 China
Re: K231342
> Trade/Device Name: Insufflator (OPTO-IFL1000) Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: August 28, 2023 Received: August 28, 2023
Dear Weijuan Guo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K231342
Device Name Insufflator (OPTO-IFL1000)
Indications for Use (Describe)
The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circular design on the left, followed by the word "OptoMedic" in black text. The circular design is made up of two curved lines that start and end at the same point, creating a circular shape. The word "OptoMedic" is written in a bold, sans-serif font.
# 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
Date Prepared: September 14, 2023
## I. General Information
| 510(k) Submitter/Owner: | Guangdong OptoMedic Technologies, Inc. |
|---------------------------|-----------------------------------------------------------------|
| | Suite 503, Building A, Golden Valley Intellicreation Community, |
| | No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan, |
| | Guangdong, 528200, P.R. China |
| | Establishment Registration Number: Not yet registered |
| Primary Contact Person: | Weijuan Guo |
| | Regulatory Affairs Engineer |
| | Tel: +86 (757) 8670 2920 |
| | Email: guoweijuan@optomedic.com |
| Secondary Contact Person: | Alice Lau |
| | Regulatory Affairs Engineer |
| | Tel: +86 (757) 8670 2920 |
| | Email: liuli@optomedic.com |
## II. Device Identification
| Device Trade Name: | Insufflator (OPTO-IFL1000) |
|-----------------------|----------------------------|
| Common or Usual Name: | Insufflator |
| Model: | OPTO-IFL1000 |
| Regulation Name: | Laparoscopic Insufflator |
| Regulation Number: | 21 CFR 884.1730 |
| Regulatory Class: | Class II |
| Product Code: | HIF |
## III. Predicate Device
510(k) Number: K030837 40 L High Flow Insufflator F108 Product Name:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular design on the left, followed by the word "OptoMedic" in bold, black letters. The circular design appears to be made of three curved lines that overlap each other.
#### IV. Device Description
Insufflator (OPTO-IFL1000) is a CO2 insufflation device for creating and maintaining a pneumoperitoneum during laparoscopic examinations and operations. It is capable of establishing the surgical field of view and operating space. CO2 gas can be injected into abdominal cavity by the device, and the gas separates the abdominal wall from the internal organs of the abdominal cavity, forming a space for the operation and visual field. The device is to be used with the following insufflation tubes:
- 1. OPTO-T1000H (with heating function)
- 2. OPTO-T1000 (without heating function)
#### V. Indications for Use
The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.
#### VI. Comparison of Technological Characteristics with The Predicate Device
| Description | Subject Device | Predicate Device (K030837) |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 C.F.R. § 884.1730 | 21 C.F.R. § 884.1730 |
| Product Code | HIF | HIF |
| Device class | Class II | Class II |
| Indication for use | The Insufflator (OPTO-IFL1000) is intended to<br>generate and maintain<br>pneumoperitoneum by<br>filling the abdominal<br>cavity with gas to distend<br>it during diagnostic or<br>therapeutic laparoscopic<br>procedures. | The 40 L High Flow Insufflator<br>F108 is a device intended to<br>facilitate the use of the<br>laparoscope by filling the<br>peritoneal cavity with gas to<br>distend it. |
| Prescription/<br>Over-the-counter use | Prescription | Prescription |
| Distension<br>Medium | CO2 | CO2 |
| Pressure range: | 5-25 mmHg | 1-30 mmHg |
| Overpressure alarm | When the nominal<br>pressure is exceeded by<br>more than 4 mmHg,<br>visual and audible | When the nominal pressure is<br>exceeded by more than 4<br>mmHg and with a pressure >29<br>mmHg, visual and audible |
#### Table 1 General Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circle on the left, followed by the word "OptoMedic" in black, bold letters. The circle is a gradient of green and yellow, with the green on the bottom and the yellow on the top.
| Overheating alarm | At >41°C, visual and<br>acoustic alarm | At >42°C, a warning signal is<br>audible |
|-----------------------|----------------------------------------|------------------------------------------|
| Reprocessing Method | Steam sterilization | Steam sterilization |
| Dimensions<br>(W*H*D) | 370mm*410mm<br>*166.5mm | 267 x 138 x 410 (mm) |
| Weight | 11Kg (N.W.) | approx. 7 kg |
| Power supply | 110-240 V~ | 100-240 V~ |
The differences technological characteristics do not raise different questions of safety and effectiveness.
#### VII. Performance data
Non-clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ANSI/AAMI ES 60601-1: 2005+A2 (R2012) +A1 Medical Electrical Equipment- Part 1: . General requirements for basic safety and essential performance.
- . IEC 60601-1-2 Edition 4.1 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
- IEC 60601-1-8 Edition 2.2 Medical electrical equipment Part 1-8: General requirements for . basic safety and essential performance - Collateral Standard: General requirements tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- AAMI TIR 30:2016 A compendium of processes, materials, test methods, and acceptance . criteria for cleaning reusable medical devices
- AAMI TIR 12:2020 Designing, testing, and labeling reusable medical devices for reprocessing . in health care facilities: A guide for medical device manufacturers
The software verification and validation testing were conducted and the test results demonstrated the software function met the requirements. The software for this device was considered a "Major" level of concern.
Performance testing were also conducted and demonstrate that the proposed system performs according to specifications and functions as intended. And the test result shows that the preset acceptance criteria are met.
- 1. Gas Supply Indication
- 2. Accuracy of the Pressure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circle on the left side and the word "OptoMedic" in black on the right side. The circle is made up of two curved lines that form a complete circle, with the top line being a lighter shade of green than the bottom line.
- 3. Accuracy of the Pressure-Under Leak Condition (Continuous leakage compensation testing)
- 4. Overpressure Alarm
- న్. Overpressure Reduction
- 6. Under-pressure Replenishment (Transient leakage compensation testing)
- 7. Accuracy of the Flow
- 8. Heating Function
- 9. Overheating Alarm
- 10. Accuracy of Gas Consumption Display
#### VIII. Conclusions
The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device to support a substantial equivalence determination.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.