← Product Code [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET) · K974763

# DEROYAL INDUSTRIES DISPOSABLE SURGICAL TROCAR/CANNULA (K974763)

_Deroyal Industries, Inc. · HET · Jul 10, 1998 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K974763

## Device Facts

- **Applicant:** Deroyal Industries, Inc.
- **Product Code:** [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET.md)
- **Decision Date:** Jul 10, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The DeRoyal Industries Disposable Surgical Trocar/Cannula is a single patient use disposable device intended for use during operative and/or diagnostic surgery to puncture/penetrate the abdominal wall and to serve as a port of entry for laparoscopic instruments (i.e. laparoscope, forceps, scissors, aspiration/irrigation cannula, etc.).

## Device Story

Disposable surgical trocar/cannula system; used for abdominal wall penetration during laparoscopic procedures. Components include stainless steel, aluminum, and biocompatible polymers; available in 5, 8, 10/11, and 12 mm diameters. Features include pyramidal tip with tri-segmented safety shield or blunt tip for open procedures; integrated reducer diaphragm for smaller instruments; luer fitting for insufflation; manual flapper valve for desufflation. Operated by surgeons in clinical/OR settings. Device provides access port for instruments like laparoscopes, forceps, and scissors; facilitates abdominal insufflation/desufflation. Benefits include standardized port access for minimally invasive surgery.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and design comparison to legally marketed predicate devices.

## Technological Characteristics

Materials: surgical grade stainless steel, aluminum, biocompatible polymers, fiberglass, silicone rubber. Design: 5-12mm diameter cannulas, 75-100mm length. Features: pyramidal tip with tri-segmented safety shield or blunt tip; luer fitting for insufflation; manual flapper valve for desufflation. Radiotranslucent shaft. Sterile, single-patient use.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- Auto Suture Surgiport Disposable Trocar & Sleeve ([K862611](/device/K862611.md), [K874879](/device/K874879.md), [K900487](/device/K900487.md), [K903419](/device/K903419.md))
- Ethicon Endopath Disposable Surgical Trocar and Sleeve ([K912398](/device/K912398.md), [K924035](/device/K924035.md), [K932282](/device/K932282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K974763

\$\rho_{193}\$

## 510(k) SUMMARY: DEROYAL INDUSTRIES, INC. DISPOSABLE SURGICAL TROCAR/CANNULA

- (1) DeRoval Industries 200 DeBusk Lane Powell. TN 37849
Camille Matlock Contact Person: (423) 938-7828 Telephone: Date Summary Prepared: December 18, 1997

- Trade or Proprietary Name: none established (2)
Common Name (s):

Disposable Surgical Trocar/Cannula

Disposable Surgical Blunt tip Trocar/Cannula

Classified Name:

Endoscope and accessories, & 876.1500, Class II

- Auto Suture® Surgiport® Disposable Trocar (3) Predicates: & Sleeve (510(k) nos. K862611. K874879, K900487, K903419)
Ethicon Endopath Disposable Surqical Trocar and Sleeve (510(k) nos. K912398, K924035, K932282)

- (4) Description of Device:
The DeRoyal Disposable Surgical Trocar/Cannula is a disposable single patient use device fabricated from surgical grade stainless steels, surgical grade aluminum, and biocompatible medical grade polymers.

{1}------------------------------------------------

K974763

p2017

The device comes in 5, 8, 10/11, and 12 mm diameter cannulas. Cannula length ranges from 75 mm - 100 mm. The cannula diameter can be used with equivalent diameter instruments or scopes, but is also equipped with a detachable integrated reducer diaphragm on all large sizes to bring the compatible diameter down to 5 mm. The trocar/cannula is supplied with either a pyramidal tip cutting trocar with safety shield or with a blunt tip trocar/cannula without safety shield for use with open laparoscopic procedures. The cutting trocar/cannula has a tri-segmented safety shield which is comprised of three shield segments that retract and advance independently.

The cannula body has a luer fitting to accommodate insufflation apparatus and an external lever for manual opening of the internal flapper valve for rapid desufflation. The cannula shaft is radiotranslucent.

- (5) Intended Use:
The DeRoval Disposable Surgical Trocar/Cannula is a single patient use disposable device intended for use during operative and/or diagnostic surgery to puncture/penetrate the abdominal wall and to serve as a port of entry for laparoscopic instruments (i.e., laparoscope, forceps, scissors, aspiration/irrigation cannula, etc.).

| Characteristics                               | Predicate Devices                                                                                                                                  | DeRoyal Industries                                                       |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Intended Use                                  | Application in thoracic, gynecologic<br>laparoscopy, and other abdominal<br>procedures to establish a path of entry<br>for endoscopic instruments. | Same                                                                     |
| Pyramidal Tip                                 | Yes                                                                                                                                                | Same                                                                     |
| Safety Shield                                 | Yes                                                                                                                                                | Same                                                                     |
| Reducer to Accommodate<br>Smaller Instruments | Yes                                                                                                                                                | Same                                                                     |
| Sterility                                     | Sterile                                                                                                                                            | Same                                                                     |
| Materials                                     | Polymers, silicone rubber, stainless steel,<br>and aluminum.                                                                                       | Polymers, silicone rubber, fiberglass,<br>stainless steel, and aluminum. |

- (6) Technological Characteristics:

{2}------------------------------------------------

03073

The DeRoyal Disposable Surgical Trocar/Cannula has similar/same technological characteristics as the predicate devices in that they are comprised of similar design, materials, and are intended to be used as a port of entry for laparoscopic instruments.

## Conclusion: (7)

The proposed device has the same intended use and the same basic technology as the legally marketed predicate devices identified in the premarket notification submission. The proposed device contains, in some combination, similar/same features, materials, and design as the predicate devices and does not pose any new questions concerning safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 0 1998

Ms. Camille Matlock Regulatory Affairs DEROYAL Industries, Inc. 200 DeBusk Lane Powell. TN 37849

Re: K974763

Disposable Surgical Trocar/Cannula Dated: April 16, 1998 Received: April 17, 1998 Regulatory Class: II 21 CFR 884.1720/Procode: 85 HET

Dear Ms. Matlock:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours.

Lillian Yin, Ph.D.

Director. Division of Reproductive. Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

K974763 510(k) Number (if known): Disposable Surgical Trocar/Cannula Device Name:

Indications for Use:

The DeRoyal Industries Disposable Surgical Trocar/Cannula is a single patient use disposable device intended for use during operative and/or diagnostic surgery to puncture/penetrate the abdominal wall and to serve as a port of entry for laparoscopic instruments (i.e. laparoscope, forceps, scissors, aspiration/irrigation cannula, etc.).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Robert R. Satting
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number K974763

Prescription Use (Per 21 CFR § 801.109)

OR

Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K974763](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K974763)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
