21 CFR 890.5800 — Virtual Reality Behavioral Therapy Device For Pain Relief

Neurology (NE) · Part 890 Subpart F—Physical Medicine Therapeutic Devices · § 890.5800

Identification

A virtual reality behavioral therapy device for pain relief is a device intended to provide behavioral therapy for patients with pain. Therapy is administered via a virtual reality display that utilizes a software program containing the behavioral therapy content.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
QRAVirtual Reality Behavioral Therapy Device For Pain Relief26SaMD

Special Controls

QRA — Virtual Reality Behavioral Therapy Device For Pain Relief

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under the labeled conditions for use must validate the model of behavioral therapy as implemented by the device and evaluate all adverse events. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Software verification, validation, and hazard analysis must be performed. (4) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed. (5) Labeling must include the following: (i) A warning regarding the risk of nausea and motion sickness; (ii) A warning regarding the risk of discomfort from the device; and (iii) A summary of the clinical testing with the device.

eCFR

QRA — Virtual Reality Behavioral Therapy Device For Pain Relief

(1) Clinical performance testing under the labeled conditions for use must validate the model of behavioral therapy as implemented by the device and evaluate all adverse events. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Software verification, validation, and hazard analysis must be performed. (4) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed. (5) Labeling must include the following: (i) A warning regarding the risk of nausea and motion sickness; (ii) A warning regarding the risk of discomfort from the device; and (iii) A summary of the clinical testing with the device.

Ecfr Llm

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...