← Product Code [OBP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/OBP) · K251210

# Ampa One System (AMPA-001) (K251210)

_Neuromodulatory Devices & Applications · OBP · Jun 27, 2025 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/OBP/K251210

## Device Facts

- **Applicant:** Neuromodulatory Devices & Applications
- **Product Code:** [OBP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/OBP.md)
- **Decision Date:** Jun 27, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.5805
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

## Device Story

The Ampa One System is a non-invasive repetitive transcranial magnetic stimulation (rTMS) device used in clinical settings to treat Major Depressive Disorder. It consists of an Ampa Pulse Generator, L and M stimulation coils, a neuronavigation cap, a support arm, and a control pad with a mobile app interface. The system operates by discharging a capacitor through the stimulation coil to generate a time-varying magnetic field, which induces electrical currents in the cerebral cortex to excite neurons. The operator uses the control pad to manage treatment protocols, including stimulation intensity (based on motor threshold), frequency, and pulse patterns. The device targets the left dorsolateral prefrontal cortex (DLPFC) or bilateral prefrontal cortex (DMPFC) depending on the coil used. By delivering targeted electromagnetic pulses, the device aims to modulate neural activity, potentially benefiting patients by providing an alternative treatment for medication-resistant depression.

## Clinical Evidence

No clinical data was provided. Substantial equivalence was demonstrated through non-clinical bench testing, including safety (function and compatibility) and effectiveness (magnetic field strength and spatial characteristics) testing in accordance with the FDA Special Controls Guidance for Transcranial Magnetic Stimulator devices.

## Technological Characteristics

Electromagnetic rTMS system. Components: Ampa Pulse Generator, figure-of-eight air-core coils (L and M), neuronavigation cap, control pad, and support arm. Energy source: Capacitor-based pulse generator. Connectivity: Mobile app interface on control pad. Electrical safety: IEC 60601-1, IEC 60601-1-2. Waveform: 210 μs biphasic sinusoid. Treatment parameters: 50 Hz (L-coil) or 18 Hz (M-coil) repetition rates.

## Regulatory Identification

A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.

## Predicate Devices

- Ampa One System ([K243319](/device/K243319.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 26, 2025

Neuromodulatory Devices &amp; Applications
% Matthieu Kirkland
Head of Regulatory Affairs
Avio Medtech Consulting
2300 Myrtle Ave, Suite 200
St. Paul, Minnesota 55114

Re: K251210
Trade/Device Name: Ampa One System (AMPA-001)
Regulation Number: 21 CFR 882.5805
Regulation Name: Repetitive transcranial magnetic stimulation system
Regulatory Class: Class II
Product Code: OBP
Dated: May 29, 2025
Received: May 29, 2025

Dear Matthieu Kirkland:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251210 - Matthieu Kirkland
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K251210 - Matthieu Kirkland
Page 3

Sincerely,

**PAMELA D. SCOTT -S**

Digitally signed by PAMELA D. SCOTT -S
Date: 2025.06.26 23:32:03 -04'00'

Pamela D. Scott
Assistant Director
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K251210

Device Name

Ampa One System (AMPA-001)

Indications for Use (Describe)

The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

|  510(k) #: | 510(k) Summary | Prepared on: 2025-06-05  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | Neuromodulatory Devices & Applications |   |
|  Applicant Address | 501 W. Broadway Suite 1025 San Diego CA 92101 United States |   |
|  Applicant Contact Telephone | 619-846-5566 |   |
|  Applicant Contact | Dr. Don Vaughn |   |
|  Applicant Contact Email | don@ampahealth.com |   |
|  Correspondent Name | Avio Medtech Consulting |   |
|  Correspondent Address | 2300 Myrtle Ave Suite 200 St. Paul MN 55114 United States |   |
|  Correspondent Contact Telephone | 310-748-2539 |   |
|  Correspondent Contact | Mr. Matthieu Kirkland |   |
|  Correspondent Contact Email | Matthieu@aviomedtech.com |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | Ampa One System (AMPA-001) |   |
|  Common Name | Repetitive transcranial magnetic stimulation system |   |
|  Classification Name | Transcranial Magnetic Stimulator |   |
|  Regulation Number | 882.5805 |   |
|  Product Code(s) | OBP |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K243319 | Ampa One System | OBP  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  Please note the subject device is nearly identical to the device cleared in K243319. The only change to the System is that the Bridge Box for connection to the MagVenture MagPro R30 pulse generator and the MagVenture MagPro R30 pulse generator itself have been replaced with an Ampa Pulse Generator. The Ampa One System with the Ampa Pulse Generator is substantially equivalent to its previously cleared design. The Ampa One System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic fields to induce electrical currents in brain tissues to target specific regions of the cerebral cortex. The system contains two different coils for targeting different regions of the brain. The L Coil is designed to target the left dorsolateral prefrontal cortex (DLPFC). The M Coil is designed to target the bilateral prefrontal cortex (DMPFC). Both coils are used within the Ampa One System to treat Major Depressive Disorder (MDD) in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.  |   |   |

{5}

Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons.

The Ampa One System is an integrated system composed of the following hardware and software components with associated functions to enable its intended use:
- Ampa L and M Coils - MT determination and Depression treatment
- Ampa Pulse Generator - Provide Electrical power to the coils
- Ampa Control Pad - The software interface for operators to control TMS treatment
- Axon Cable - Provides connections between the Ampa Pulse Generator and Ampa Control Pad
- Ampa Magic Arm - Support system to securely hold the coils in place during treatment
- Neuronavigation Cap - Fabric cap worn by patient with integrated markings for determination of cerebral cortex treatment targets

The system comes preloaded with a mobile app on the Ampa Control Pad containing the software interface for operation of the system.

|  **Intended Use/Indications for Use** | **21 CFR 807.92(a)(5)**  |
| --- | --- |
|  The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.  |   |
|  **Indications for Use Comparison** | **21 CFR 807.92(a)(5)**  |
|  The Ampa One System and the predicate device have the same indications for use.  |   |
|  **Technological Comparison** | **21 CFR 807.92(a)(6)**  |
|  The subject device has nearly identical technological characteristics to the predicate device.  |   |
|  Both devices have identical treatment protocol parameters (e.g., repetition rate, pulses per train, number of trains, number of pulses, inter train interval, treatment time) and consist of the same general TMS system components (e.g., coils, pulse generator or pulse generator connector, navigation cap, tablet with software, coil holder arm). The only difference is that rather than using a bridging connector (i.e., Bridge Box) to connect to an external pulse generator (i.e., MagVenture MagPro R30), the subject device includes an Ampa Pulse Generator and does not require a bridging connector.  |   |
|  The Ampa Pulse Generator produces pulsed electrical current, which generates a time-varying magnetic field through the coil. It was intentionally designed & developed to closely match the well-known MagVenture R30, utilizing a nearly identical electrical topology to achieve functionally equivalent output.  |   |
|  All coils are unmodified within the subject device and identical to those in the predicate device clearance. Additional performance testing has been conducted using the well-established methods in the rTMS Special Controls guidance to evaluate the change and ensure both devices provide the same reasonable expectation of safety and effectiveness (e.g., methods, protocols, and acceptance criteria used to support clearance of K243319).  |   |
|  Please note that there is an attached Substantial Equivalence Table after the conclusions section with a complete comparison and analysis of device parameters.  |   |
|  **Non-Clinical and/or Clinical Tests Summary & Conclusions** | **21 CFR 807.92(b)**  |
|  Nonclinical testing evaluated the Ampa One System's performance to be equivalent in terms of safety and effectiveness compared to the predicate device. The suite of testing to demonstrate substantial equivalence can be divided into two aspects: safety (function and compatibility) and effectiveness (magnetic characteristics).  |   |
|  Safety testing established that the Ampa One System is able to function in its configuration with the Ampa One pulse generator to  |   |

{6}

perform its intended use. Safety is defined as functioning with risks assessed of TMS therapy mitigated through design. Effectiveness testing established that the Ampa One System produced substantially equivalent magnetic characteristics compared to the predicate devices. Specifically, testing demonstrated that the amplitude of the magnetic field strength as a function of power input, and the spatial characteristics of the magnetic field produced were substantially equivalent between the subject and predicate device.

Development and testing of the Ampa One System was performed in conformance with the FDA Special Controls Guidance on Transcranial Magnetic Stimulator devices.

Clinical Data was not applicable for this submission.

Testing in accordance with the FDA Special Controls Guidance and voluntary consensus standards demonstrate that the Ampa One System achieves substantially equivalent safety and effectiveness compared to the legally marketed predicate devices.

{7}

|   | Modified Device: Ampa One System | Predicate Device: Ampa One System (K243319) | Comments  |
| --- | --- | --- | --- |
|  Product Code, Device Risk Classification | OBP
Class II | OBP
Class II | Identical  |
|  Indications for Use | The Ampa One System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode. | The Ampa One System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode. | Identical  |
|  Target Population | Adult subjects with Major Depressive Disorder | Adult subjects with Major Depressive Disorder | Identical  |
|  Energy Used / Delivered | Electromagnetic Energy is delivered | Electromagnetic Energy is delivered | Identical  |
|  Area of Brain to Stimulated | Ampa L-Coil: Head – theta burst stimulation to the left dorsolateral prefrontal cortex (DLPFC)

Ampa M-Coil: Head – stimulation to the dorsolateral prefrontal cortex (DLPFC) | Ampa L-Coil: Head – theta burst stimulation to the left dorsolateral prefrontal cortex (DLPFC)

Ampa M-Coil: Head – stimulation to the dorsolateral prefrontal cortex (DLPFC) | Substantially Equivalent  |
|  Components | Ampa L and M Coils | Ampa L and M Coils | Identical  |
|   | Ampa Pulse Generator | Bridge Box | Substantially Equivalent. The Bridge Box, which allowed connection to the MagVenture MagPro R30 Pulse Generator, has been replaced with an Ampa Pulse Generator.  |
|   | Neuronavigation Cap | Neuronavigation Cap | Identical  |
|   | Ampa Control Pad | Ampa Tablet | Identical – Name Change  |
|   | Ampa Magic Arm | Ampa Magic Arm | Identical  |
|  Applicator: Configuration Core Material | Ampa L-Coil:
Figure-of-eight coil
Air core

Ampa M-Coil:
Figure-of-eight coil
Air Core | Ampa L-Coil:
Figure-of-eight coil
Air core

Ampa M-Coil:
Figure-of-eight coil
Air Core | Identical  |

{8}

|   | Modified Device: Ampa One System | Predicate Device: Ampa One System (K243319) | Comments  |
| --- | --- | --- | --- |
|  Performance of Treatment Protocol  |   |   |   |
|  Physical unit of amplitude setting (e.g., coil current, peak magnetic field) at coil and its relation to the SMT unit | L Coil
0 – 1.7 SMT

M Coil
0 – 1.4 SMT | L Coil
0 – 2.0 SMT

M Coil
0 – 1.7 SMT | Substantially Equivalent  |
|  Patient Contacting Materials | Treatment Cap | Treatment Cap | Identical  |
|  Coil Positioning System | Cap positioned in relation to nasion | Cap positioned in relation to nasion | Identical  |
|  Electrical safety | Complies with IEC 60601-1 and IEC 60601-1-2 | Complies with IEC 60601-1 and IEC 60601-1-2 | Identical  |
|  Treatment stimulation parameters | L Coil
Waveforms: 210 μs biphasic sinusoid
Magnetic field strength: 120% of patients’ hand-twitch motor threshold (MT)
Repetition rate: 50 Hz
Train duration: 2 sec*
Inter-train interval: 8 secs
Burst pulses: 3
Bursts: 200
Inter-pulse interval: 20 msec
Number of trains: 20
Numbers of pulses/session: 600
Total duration: 3 min 9 sec

M Coil
Waveform: 210 μs biphasic sinusoid
Magnetic Field Strength: 120% of patients’ hand-twitch motor threshold (MT)
Repetition rate: 18 Hz
Train duration: 2 sec*
Inter-train interval: 20 sec
Number of trains: 55
Magnetic Pulses per Session: 1980
Treatment Session Duration: ~ 20 min | L Coil
Waveforms: 290 μs biphasic sinusoid
Magnetic field strength: 120% of patients’ hand-twitch motor threshold (MT)
Repetition rate: 50 Hz
Train duration: 2 sec*
Inter-train interval: 8 secs
Burst pulses: 3
Bursts: 200
Inter-pulse interval: 20 msec
Number of trains: 20
Numbers of pulses/session: 600
Total duration: 3 min 9 sec

M Coil
Waveform: 290 μs biphasic sinusoid
Magnetic Field Strength: 120% of patients’ hand-twitch motor threshold (MT)
Repetition rate: 18 Hz
Train duration: 2 sec*
Inter-train interval: 20 sec
Number of trains: 55
Magnetic Pulses per Session: 1980
Treatment Session Duration: ~ 20 min | Substantially Equivalent  |

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/OBP/K251210](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/OBP/K251210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
