DEEPWAVE PERCUTANEOUS NEUROMODULATION PAIN THERAPY SYSTEM

K061166 · Biowave Corporation · NHI · Aug 15, 2006 · Neurology

Device Facts

Record IDK061166
Device NameDEEPWAVE PERCUTANEOUS NEUROMODULATION PAIN THERAPY SYSTEM
ApplicantBiowave Corporation
Product CodeNHI · Neurology
Decision DateAug 15, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Deepwave Percutaneous Neuromodulation Pain Therapy System is comprised of a percutaneous electrode array and the Deepwave Neuromodulation Pain Therapy Device. The Deepwave Percutaneous Neuromodulation Pain Therapy System is indicated for: - Symptomatic relief of chronic, intractable pain, post surgical and post-traumatic acute pain; - Symptomatic relief of post-traumatic pain; - Symptomatic relief of post-operative pain.

Device Story

Battery-powered electrical muscle and nerve stimulator; provides PENS therapy. System includes control unit, standard feed pad electrode, and sterile, single-use percutaneous electrode array (PEA). PEA features 1014 medical-grade 316L stainless steel microneedles (736 microns length) on 1.5-inch diameter array. Clinician prescribes therapy; patient controls signal amplitude via buttons on device face; LCD displays amplitude. Device sends biphasic electrical signal from feed pad through body to PEA placed over pain source. Used for pain management; intended to provide symptomatic relief.

Clinical Evidence

No clinical data provided. Substantial equivalence based on technical comparison of design, output specifications, and intended use.

Technological Characteristics

Battery-powered (12V NiMH) electrical stimulator. Materials: 316L stainless steel microneedles. Output: Biphasic waveform, 100-200 Hz beat frequency. Form factor: 7.4" x 5.6" x 2.25", 3 lbs. Connectivity: Standalone. Software: Driven by internal program for pain reduction. Sterilization: PEA is sterile, single-use.

Indications for Use

Indicated for symptomatic relief of chronic, intractable pain, post-surgical acute pain, post-traumatic acute pain, and post-operative pain.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary for the Biowave Deepwave Percutaneous Neuromodulation Pain Therapy System #### SPONSOR 1. Biowave Corporation 16 Knight Street Norwalk. CT 06851 Contact Person: Brad Siff (203) 247-9020 Telephone: Date Prepared: July 19, 2006 ### DEVICE NAME 2. | Proprietary Name: | Deepwave Percutaneous Neuromodulation Pain Therapy<br>System | |-----------------------|--------------------------------------------------------------| | Common/Usual Name: | Electrical Muscle and Nerve Stimulator | | Classification Names: | Percutaneous Electrical Nerve Stimulation Devices | ### 3. PREDICATE DEVICES - Vertis NeuroScience Inc., Percutaneous Neuromodulation Control Unit and . Accessories (K022241) - Biowave Corporation's Deepwave Neuromodulation Pain Therapy Device . (K052289) ### 4. INTENDED USE The Deepwave Percutaneous Neuromodulation Pain Therapy System is comprised of a percutaneous electrode array and the Deepwave Neuromodulation Pain Therapy Device. The Deepwave Percutaneous Neuromodulation Pain Therapy System is indicated for: - Symptomatic relief of chronic, intractable pain, post surgical and post-. traumatic acute pain; {1}------------------------------------------------ - . Symptomatic relief of post-traumatic pain; - . Symptomatic relief of post-operative pain. ### ട്. DEVICE DESCRIPTION Biowave Corporation's Deepwave Percutaneous Neuromodulation Pain Therapy System is essentially identical to Biowave's Deepwave Neuromodulation Pain Therapy Device cleared under K052289 except that the proposed device includes a percutaneous electrode array instead of standard surface electrodes. The percutaneous electrode array (PEA) is a sterile, single-use microneedle patch which facilitates the delivery of electrical signals through the skin into tissue. The size of the PEA patch is 2.5 inches in diameter. In the center of the patch is a 1.5 inch diameter array of 1014 microneedles made from medical grade 316L stainless steel which are 736 microns (0.736 millimeters) in length. When placed onto the skin the PEA feels like sandpaper or Velcro to the touch. The Feed Pad is a standard electrode that is 2" x 4" or 5" x 8" in size. The Deepwave Percutaneous Neuromodulation device sends a signal from the Feed Pad through the body to the PEA. The Deepwave Percutaneous Neuromodulation Pain Therapy Device is a batterypowered device intended to provide clinicians with the ability to prescribe PENS therapy. The device measures approximately 7.4 inches wide, 5.6 inches long, and 2.25 inches deep, weighs about 3 pounds, and operates on a 12 volt rechargeable NiMH battery that is enclosed within the unit. The unit will not operate while it is plugged into the wall to recharge the battery. The patient controls the amplitude of the signal with two buttons (a Plus (+) Button and a Minus (-) Button) on the face of the device. An LCD displays the amplitude of the signal in numerical format. Two wires emanate from the unit. One wire is attached to a large disposable electrode pad placed opposite the source of pain ("Feed Pad"). The second wire is attached to a smaller microneedle patch called a Percutaneous Electrode Array ("PEA") placed directly over the source of the pain (the treatment site). ## BASIS FOR SUBSTANTIAL EQUIVALENCE 6. The Deepwave Percutaneous Neuromodulation Pain Therapy System, the Vertis Percutaneous Neuromodulation Therapy device and the Deepwave Neuromodulation Pain Therapy device are identical in indications in that they are all indicated for the symptomatic relief of chronic, intractable pain, post surgical and post-traumatic acute {2}------------------------------------------------ pain and the symptomatic relief of post-operative pain. Biowave's Deepwave Percutaneous Neuromodulation Pain Therapy System and the Vertis device are PENS devices that use percutaneous electrodes whereas Biowave's Deepwave Neuromodulation Pain Therapy device is a TENS that uses a standard TENS electrode. The use of Biowave's percutaneous electrode array (PEA) does not affect safety or effectiveness since the Biowave PEA penetrates less depth of the skin than the Vertis PEA. The Basic Unit Characteristics and Output Specifications are similar for the Neuromodulation Pain Therapy and the predicate device according to the requirements of the FDA Guidance Document for Powered Muscle Stimulator 510(k) s, Attachment II, Guidance for Reporting Technological Characteristics. Biowave Corporation's Deepwave Percutaneous Neuromodulation Pain Therapy System and the Vertis PENS device are similar in design and function. Both devices offer percutaneous electrodes, a biphasic waveform and a beat frequency in the range of 100-200 Hz. Both the proposed and predicate devices are software driven PENS units that provide the user with a treatment program for pain reduction. The conclusion of this technical comparison is that Biowave Corporation's Deepwave Percutaneous Neuromodulation Pain Therapy System is substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles a caduceus, with three figures representing people. The overall design is simple and professional, reflecting the department's role in public health and human services. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## AUG 1 5 2006 Biowave Corporation % Medical Device Consultants, Inc. Ms. Mary McNamara-Culliname, RAC Staff Consultant 49 Plain Street North Attleboro, Massachusetts 02760 Re: K061166 Trade/Device Name: Deepwave Percutaneous Neuromodulation Pain Therapy System Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NHI, GZJ Dated: July 19, 2006 Received: July 20, 2006 Dear Ms. McNamara-Culliname: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Ms. Mary McNamara-Culliname, RAC forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Barbara Buchup Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 3 -- Ms. Mary McNamara-Culliname, RAC ## Indications for Use 510(k) Number (if known): Device Name: Indications For Use: Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED ) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): Deepwave Percutaneous Neuromodulation Pain Therapy System Device Name: Indications for Use: The Deepwave Percutaneous Neuromodulation Pain Therapy System is comprised of a percutaneous electrode array and the Deepwave Neuromodulation Pain Therapy Device. The Deepwave Percutaneous Neuromodulation Pain Therapy System is indicated for: - Symptomatic relief of chronic, intractable pain, post surgical and post-traumatic . acute pain; - . Symptomatic relief of post-traumatic pain; - . Symptomatic relief of post-operative pain. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Barbara Paueun Benson **(Division Sign-Off)** **510(k) Number** K061106
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...