CRANIAL LOOP, CRANIAL LOOP L AND CRANIAL LOOP XL AND CRANIAL BONE FIXATION SYSTEM
Applicant
Neos Surgery S.L
Product Code
GXR · Neurology
Decision Date
Sep 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation. In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
Device Story
Biocompatible, postoperative cranial bone fixation system; fixes bone flap to skull. Device consists of upper and lower platforms with ratcheting mechanism; no specific surgical instruments required for implantation. Cranial LOOP and LOOP L used in calvarial gaps; Cranial LOOP XL used for 14 mm burr holes. Surgeon implants device during craniotomy; device provides mechanical fixation of bone flap. Benefits include rapid, simple application; low profile; adaptation to varying cranium curvatures and thicknesses. Mechanical locking mechanism secures bone flap; prevents lateral movement. Sterile, single-use device.
Clinical Evidence
No clinical data required. Evidence provided via bench testing: functional verification of locking mechanism, cadaveric testing for ease of use and anatomical adaptation, and biomechanical testing (push-in, pull-out, and dynamic load testing for 0.5 million cycles). Accelerated aging testing (ASTM F 1980) confirmed shelf-life integrity.
Technological Characteristics
Biocompatible cranial bone fixation system. Consists of upper and lower platforms with ratcheting locking mechanism. Dimensions vary by model (LOOP, LOOP L, LOOP XL). Sterile, single-use. No software or energy source. Packaging validated per ISO 11607.
Indications for Use
Indicated for post-craniotomy bone flap fixation in patients requiring cranial bone fixation. Cranial LOOP and Cranial LOOP L are for use within the calvarial gap; Cranial LOOP XL is for covering a standard 14 mm cranial burr hole.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
Predicate Devices
Cranial LOOP Cranial Bone Fixation System (K101235)
Related Devices
K101235 — CRANIAL LOOP CRANIAL BONE FIXATION SYSTEM · Neos Surgery S.L · Oct 25, 2010
K240137 — Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System · Neos Surgery S.L · Feb 16, 2024
K221606 — Absorbable Cranial Flap Fixation System · Chendu Medart Medical Scientific Co., Ltd. · Nov 21, 2023
K070901 — MODIFIED PLATE DESIGN FOR BIOPLATE ZIP CRANIOTOMY FIXATION SYSTEM · Bioplate, Inc. · Jun 15, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
K132044
## 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System 510(k) premarket notification.
| Applicant: | Neos Surgery S.L. C/Ceramistes. 2 08290-Cerdanyola del Vallès Barcelona, Spain |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3008974316 |
| US Contact: | Cherita James M Squared Associates, Inc. 901 King Street, Suite 101 Alexandria, VA 22314 Ph: 703-562-9800 ext 257 Fax: 703-562-9797 E-mail: CJames@msquaredassociates.com |
| Date submitted: | August 27, 2013 |
| Proprietary Name: | Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System |
| Common Name: | BurrHole cover |
| Classification Status: | II |
| Regulation: | 21 CFR 882.5250 Burrhole cover |
| Panel: | Neurology |
| Product Code: | GXR |
| Predicate Device: | Cranial LOOP Cranial Bone Fixation System K101235 |
| Device Description: | The Cranial LOOP. Cranial LOOP L and Cranial LOOP XL Cranial Bone |
SEP 3 0 2013Fixation System is a biocompatible, postoperative cranial bone fixation system that fixes the bone flap to the skull, without any specific surgical instrument for its handling or implantation. It is provided sterile, for single use. This submission includes a line extension for the Cranial LOOP L
Cranial LOOP L and Cranial LOOP XL
{1}------------------------------------------------
| N° | Test | Methodology |
|----|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A | Functional verification | Test 1: Functionality of the implantable parts (P01 (lower platform with ties) + P02 (upper platform)): Determine the breaking strength of P01+P02 union (ratchets). A load was applied to the platforms, in an opposite direction, in order to assess the strength of the locking mechanism. Test 2: Functionality of the handle: Determine the breaking strength of P01 (lower platform with ties) + P03 (handle) union (ratchets). A load was applied to the lower platform and the handle, in an opposite direction, in order to assess the strength of the locking mechanism. |
| B | Cadaveric test | Three different craniotomies were tested: Frontal-Parietal-Sphenoid wing craniotomy (left side) Frontal lateral (right side) Frontal-parietal (right side). One burr hole on the temporal bone. The craniotomies were obtained using Linvatec cranitome with saws (router) from 1.4 to 2.2 mm cutting diameter and 14 mm diameter perforator.FC050200 (Cranial LOOP XL ) are applied in each of the craniotomies according the IFUs. The following aspects were evaluated, Sufficient space for device placement No danger for the surgeon during device application. Correct positioning of the device Rapid and simple application of the device with minimum instruments No danger for duramater. Finally the duramater and the brain will be examined after the removal of the cranial fixation, detecting any possible incidence. Low epicranial and subcranial platform profiles. Adaptation to different cranium curvatures. Adaptation to different cranium thicknesses. Qualitative evaluation of the lateral movement. Qualitative evaluation of cranial fixation strength. Evaluation of the ease with which the system is removed. |
| C | Biomechanical tests | Test 1: Strength to bone flap compression (Push-in). A push in load was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use. Test 2: Pull-out strength test. A pull out load was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use. Test 3: Strength to dynamic load over the bone flap. A push in load to produce a displacement of 0.5 mm on the bone flap, was applied to a model which simulates the cranium and a bone flap with three Cranial LOOPs representing clinical use for 1/2M cycles. |
| D | Accelerated ageing test | The following parameters were verified after an aging: 'Blister seal strength' (5 samples). 'Leak integrity' (5 samples). 'Sterility' (5 samples). 'Tensile strength' (5 samples) The accelerated aging was conducted according to the ASTM F 1980 standard. Q10 Factor = 2.0 Real Time aging. RT = 5 years. AAT temperature = 60 °C, AAT = 131,02 days to simulate 5 years of real aging time |
・
{2}------------------------------------------------
and Cranial LOOP XL which have new device dimensions and design features that reduce lateral movement after implantation..
Cranial LOOP and Cranial LOOP L are applied in the craniotomy gap. It can fix cranial thicknesses ranging from 1.5 mm to 24 mm and gaps ranging between 1.7 mm and those made using a craniotome standard cranial router. Cranial LOOP XL is applied in a burr hole made using a standard drill 14 mm. They can fix cranial thicknesses ranging from 4 mm to 24 mm.
Intended Use: The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
#### Summary of Technological Characteristics:
The claim of substantial equivalence of the Cranial Loop to the predicate devices is based on the comparison of the intended use, product technical characteristics, performance characteristics and product handling.
Summary of Nonclinical Testing: Mechanical and various performance testing confirms the Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System performs as intended and is substantially equivalent to the predicate device. A Declaration of Conformity with design controls is provided.
{3}------------------------------------------------
| Nº | Test | Methodology |
|----|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| E | Profile and range of<br>bone thickness<br>covered. | Dimensional test are performed in order to validate:<br>Epicranial and subcranial profile. The epicranial profile protruding from the plane of the bone was<br>measured to ensure it does not exceed the established limits. And<br>the subcranial profile protruding from the plane of the bone was<br>measured to ensure it does not exceed the established limits Range of bone thickness The upper platform (P02) is lowered as much as possible towards the<br>lower platform (P01). Measure the distance between both platforms in<br>order to check the minimum bone thickness required.<br>The distance between platforms on ready-to-use device is measured in<br>order to check the maximum bone thickness |
| F | Packaging integrity | Packaging validation according ISO 11607. |
Summary of Clinical Testing: Clinical data was not required in support of this submission.
### Substantial Equivalence Discussion:
The line extension for the Cranial LOOP L and Cranial LOOP XL has the following similarities to the previously cleared Cranial LOOP Cranial Bone Fixation System- K101235:
- . has the same indicated use.
- uses the same operating principle, .
- incorporates the same basic device design, .
- incorporates the same materials .
- incorporates comparable manufacturing and packaging processes .
- incorporates the same sterilization process. .
Based on these similarities, and mechanical and performance tests outcomes, the Cranial LOOP L and Cranial LOOP XL are substantially equivalent to the predicate Cranial LOOP Cranial Bone Fixation System.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### September 30, 2013
Neos Surgery S.L. c/o Cherita James, Regulatory Consultant M Squared Associates, Inc. 901 King Street, Suite 101 Alexandria, VA 22314
Re: K132044
Trade/Device Name: Cranial LOOP, Cranial LOOP L, Cranial LOOP XL Cranial Fixation System Regulation Number: 21 CFR 882.5250 Regulation Name: Burr hole cover Regulatory Class: Class II Product Code: GXR Dated: June 21, 2013 Received: July 2, 2013
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. James
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
### 510(k) Number: K132044
Cranial LOOP, Cranial LOOP L, and Cranial LOOP XL Cranial Bone Device Name: Fixation System
Indications For Use:
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP, Cranial LOOP L and Cranial LOOP XL, are long-term implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP L (FC050100) are for use within the calvarial gap while the Cranial LOOP XL (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang -S
Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K132044
Page 1 of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.