Durepair is indicated as a dura substitute for the repair of the dura mater.
Device Story
Durepair Dura Regeneration Matrix is a collagen implant derived from processed fetal bovine materials; used for repair of dura mater defects. Device supplied sterile in double-peel packaging for single-use; surgeon cuts matrix to desired shape/size for application. Suturability allows securing to tissue. Device functions as a scaffold for dural regeneration. Bench testing confirms physical attributes (tensile strength, stiffness, suture retention, hydration rate) and biocompatibility (non-pyrogenic, non-cytotoxic, non-sensitizing). No clinical data; substantial equivalence established via bench and biocompatibility testing comparing manufacturing process changes (removal of organic solvents) to predicate.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility studies. Bench tests included tensile strength/stiffness, suture retention, pore size, hydration rate, and histology. Biocompatibility testing included cytotoxicity (ISO 10993-5), sensitization/irritation (ISO 10993-10), systemic toxicity (ISO 10993-11), hemolysis (ISO 10993-4), intramuscular implantation (ISO 10993-6), and genotoxicity (ISO 10993-3). All tests met pre-established acceptance criteria.
Technological Characteristics
Collagen matrix derived from processed fetal bovine materials. Dimensions vary (e.g., 1"x1" to 5"x8") with ±5% tolerance. Suturable. Hydration rate ≤ 3 minutes. Sterilized. Manufacturing process uses no organic solvents. Testing per ISO 10993 standards for biocompatibility.
Indications for Use
Indicated for the repair of the dura mater in patients requiring a dura substitute.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
K141608 — COLLAGEN DURA MEMBRANE · Collagen Matrix, Inc. · Jan 16, 2015
K063117 — MODIFICATION TO DUREPAIR DURA REGENERATION MATRIX · Medtronic Neurosurgery · Nov 3, 2006
K251191 — Collagen Dura Regeneration Membrane - Repair · Collagen Matrix, Inc. · Nov 18, 2025
K150825 — Collagen Dural Regeneration Matrix · Collagen Matrix, Inc. · Nov 20, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a sense of community and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Medtronic Neurosurgery Manas Lele Senior Regulatory Affairs Specialist 125 Cremona Drive Goleta, California 93117
Re: K161370
Trade/Device Name: Durepair Dura Regeneration Matrix Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: September 27, 2016 Received: September 28, 2016
Dear Mr. Lele:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Michael J. Hoffmann -A
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161370
#### Device Name Durepair Dura Regeneration Matrix
Indications for Use (Describe) Durepair is indicated as a dura substitute for the repair of the dura mater.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Owner: | Medtronic Neurosurgery<br>125 Cremona Drive<br>Goleta, CA 93117-5503 USA |
|----------------------------|----------------------------------------------------------------------------------------------------------------------|
| Contact Name: | Manas Lele<br>Senior Regulatory Affairs Specialist<br>Telephone: (805) 571-8956<br>Email: Manas.M.Lele@Medtronic.com |
| Date Summary Prepared: | October 31, 2016 |
| Trade or Proprietary Name: | Durepair Dura Regeneration Matrix |
| Common Name: | Dura Substitute |
| Classification Name: | Dura Substitute<br>(21 CFR §882. 882.5910, Product Code GXQ) |
| Predicate Device: | Durepair Dura Regeneration Matrix (K063117<br>K041000 and K052211) |
### Device Description:
Durepair® Dura Regeneration Matrix is a collagen implant for the repair of defects in the dura mater. Durepair is supplied sterile, in a double-peel package, and is intended for single (one-time) use-only. Durepair is available in a variety of sizes intended to be cut by the surgeon to the desired shape.
## Indications for Use:
Durepair is indicated as a dura substitute for the repair of the dura mater.
### Summary of Technological Characteristics Compared to the Predicate Device:
The proposed Durepair Dura Regeneration Matrix incorporates the same basic technological characteristics as the predicate device. The manufacturing process change relative to the predicate device did not affect the device design, indications of use, material or the fundamental technology of the device.
{4}------------------------------------------------
| | | Table 1 - Summary of Technological Characteristics Comparison between Predicate | | |
|--|----------------------------------------------|---------------------------------------------------------------------------------|--|--|
| | Durepair Device and Proposed Durepair Device | | | |
| Technological<br>Characteristic | Predicate Durepair Device | Proposed Durepair Device |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Fundamental<br>Technology | Collagen implant for the repair of<br>the defects in the dura mater | Same. The Fundamental Technology has<br>not changed. |
| Sizes | Durepair is available in different<br>sizes<br>1" x 1"<br>1" x 3"<br>2" x 2"<br>3" x 3"<br>4" x 5"<br>5" x 8" | Same. The length or width sizes or<br>tolerances have not changed |
| Suturability | Use of suture allowed to secure<br>device | Same. This characteristic has not<br>changed. |
| Material | Collagen matrix harvested from<br>processed fetal bovine materials | Same. The proposed device is made<br>from the collagen matrix |
| Cut product to size | Durepair can be cut ot size based<br>on user need | Same. This characteristic has not<br>changed. |
| Manufacturing<br>Process (Use of<br>Solvents) | Use of organic solvents | No use of organic solvents. Replacement<br>with equivalent processing step. |
# Table 2 – Summary of Bench Top Testing:
The following bench testing was submitted in support of substantial equivalence.
| Test | Test Method Summary for Proposed Durepair device | Results for Proposed Durepair device |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sizes | The specified length or width had a tolerance of $\pm$ 5%. Measured with digital calipers and were inspected 100%. | All samples met the acceptance criteria, demonstrating that there are no concerns regarding the physical attributes of the proposed device in comparison to the predicate device. |
| Tensile<br>Strength | Tensile Strength must be an average 5 MPa minimum. All skins were sampled at the two thinnest corners of each skin representing worst case. | All samples met the acceptance criteria, demonstrating that there are no concerns regarding the strength and stiffness attributes of the proposed device in comparison to the predicate device. |
| Tensile<br>Stiffness | Tensile Stiffness must have an average 225 MPa maximum. All skins were sampled at the two thinnest corners of each skin representing worst case. | All samples met the acceptance criteria, demonstrating that there are no concerns regarding the strength and stiffness attributes of the proposed device in comparison to the predicate device. |
| Test | Test Method Summary for<br>Proposed Durepair device | Results for Proposed Durepair device |
| Suture<br>Retention<br>Strength | At a pull rate of 20mm/min, 3 mm<br>suture bite using polypropylene 4-0<br>suture, suture strength must be a<br>Minimum of 5 N. Two suture retention<br>samples were taken from the thinnest<br>(weakest) areas of each skin. | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the ability to maintain a suture<br>on the proposed device in comparison to<br>the predicate device. |
| Wet Shrink<br>Temperature | Wet shrink Temperature - Hydrated<br>specimens must be 58° - 67° C (an in<br>process specification). A Differential<br>Scanning Colorimeter was used.<br>Indicator that the scaffold has not<br>been damaged during processing. | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the ability to maintain a suture<br>on the proposed device in comparison to<br>the predicate device. |
| Pore Size | No visible through pores should be<br>seen and this test helps determine<br>the material porosity | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the physical attributes of<br>proposed device in comparison to the<br>predicate device. |
| Hydration Rate | Time to hydrate must be less than or<br>equal to 3 minutes. Saline solution is<br>used to hydrate at room temperature | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the hydration rate time of<br>proposed device in comparison to the<br>predicate device. |
| Histology (Wet<br>EBM) | No cells or cellular/nuclear debris<br>evident.<br>To demonstrate the EBM is free of<br>cells and cellular/nuclear debris | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the contents of the proposed<br>device in comparison to the predicate<br>device. |
| Safety | Must be non-pyrogenic (Less than or<br>equal to 2.15 EU/device)<br>No bacterial endotoxins verified per<br>each production lot. | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the contents of the proposed<br>device in comparison to the predicate<br>device. |
| Bioburden | No bioburden observed in the final<br>rinse water, 0 CFUs for each lot.<br>Final rinse water samples collected<br>for each lot. | All samples met the acceptance criteria,<br>demonstrating that there are no concerns<br>regarding the contents of the proposed<br>device in comparison to the predicate<br>device. |
| Test | Test Method Summary | Result |
| Calcification | No calcification. Samples implanted into<br>weanling rats for 4 weeks, explants were<br>examined grossly and microscopically.<br>Samples were tested for potential for calcification. | Pass.<br>No calcification was<br>present in the<br>samples. |
| Cytotoxicity | Per ISO 10993-5<br>Test item considered non-cytotoxic if none of<br>cultures exposed to test item show greater than<br>mild reactivity (grade 2)<br>Lots tested for biological reactivity by exposing<br>mouse fibroblasts to a MEM elution of product. | Pass.<br>None of the cultures<br>showed greater than<br>grade 2 of reactivity. |
| Skin Sensitization Study<br>(Saline & Cottonseed Oil<br>Extraction) | Per ISO 10993-10<br>No significant dermal contact sensitization.<br>Extract of EBM was tested for allergenic potential<br>via the guinea pig maximization test | Pass.<br>All test animals<br>scored a 0 and had<br>no significant dermal<br>contact sensitization. |
| Irritation Study,<br>Intracutaneous Injection<br>(Saline & Cottonseed Oil<br>Extraction) | Per ISO 10993-10<br>Mean reaction scores for the test articles must be<br>< 1.0<br>Extract of EBM was evaluated for its potential to<br>produce irritation after Intracutaneous injection in<br>rabbits | Pass.<br>Mean reaction<br>scores for the test<br>articles were <1.0. |
| Acute Systemic Toxicity,<br>Systemic Injection<br>(Saline & Cottonseed Oil<br>Extraction) | Per ISO 10993-11<br>No evidence of mortality or acute systemic<br>toxicity.<br>Extract of EBM was evaluated for its potential to<br>cause acute toxicity after intravenous and intra<br>peritoneal injection in mice. | Pass.<br>No test or control<br>animals showed<br>signs of toxicity. |
| Hemolysis, Rabbit Blood | Per ISO 10993-4<br>Corrected % hemolytic index for direct and<br>indirect contact must be < 5%.<br>The samples were evaluated for hemolytic activity<br>on rabbit blood via both direct and indirect<br>contact. | Pass.<br>Corrected %<br>hemolytic index for<br>direct and indirect<br>contact was < 5%. |
| Intramuscular<br>Implantation Study (4<br>weeks & 12 weeks) | Per ISO 10993-6<br>Must be non-reactive.<br>Samples evaluated for local tissue responses and<br>the potential to induce local toxic effects after<br>implantation in a rat intramuscular model | Pass.<br>No<br>reaction<br>was<br>seen in the samples. |
| Genotoxicity, Ames<br>Reverse Mutation Assay | Per ISO 10993-3<br>No 2X, or greater, increase in the mean number of<br>revertants compared to the negative control.<br>Extracts of EBM were evaluated for potential<br>mutagenicity in certain bacteria via a change in their<br>dependence for histidine or tryptophan. | Pass.<br>None of the test<br>article extracts<br>induced a significant<br>increase in number<br>of<br>revertants<br>as |
| Test | Test Method Summary | Result |
| | | compared to the negative control. |
{5}------------------------------------------------
In all cases, the results of bench testing met applicable pre-established acceptance criteria and raised no concerns regarding safety and effectiveness relative to the predicate device. Therefore, the bench testing summarized above supports the substantial equivalence of Durepair® Dura Regeneration Matrix and the predicate device.
{6}------------------------------------------------
Biocompatibility Testing:
# Table 3 – Summary of Biocompatibility Testing:
{7}------------------------------------------------
In all cases, Durepair® passed biocompatibility testing, demonstrating that the Durepair chemical change process does not raise any biocompatibility concerns relative to the predicate device. Therefore, the biocompatibility testing summarized above supports the substantial equivalence of Durepair and the predicate device.
## Conclusion:
Based on the indications of use, design and technology similarities, performance testing including bench and biocompatibility testing performed on the proposed device, it can be concluded that the proposed Durepair® Dura Regeneration Matrix is substantially equivalent to the currently marketed Durepair® cleared devices under K063117, K052211 and K041000.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.