← Product Code [GXQ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ) · K161278

# Cerafix Dura Substitute (K161278)

_Acera Surgical, Inc. · GXQ · Aug 8, 2016 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K161278

## Device Facts

- **Applicant:** Acera Surgical, Inc.
- **Product Code:** [GXQ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ.md)
- **Decision Date:** Aug 8, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5910
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Cerafix Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 4.4 in2 (28.3 cm2) or less in area. For example, 4.0 in x 1.1 in (10.1 cm x 2.8 cm) would be an acceptable defect size.

## Device Story

Cerafix Dura Substitute is a resorbable, porous polymer matrix implant used for dural repair. Surgeon cuts implant to size; places over dural defect; secures with tensionless sutures (2-3 mm from edge); ensures at least 1 cm overlap of remaining dura. Used in surgical settings. Provides scaffold for neoduralization; resorbs over time; facilitates tissue healing. Benefits patient by restoring dural integrity and preventing CSF leakage.

## Clinical Evidence

Bench testing only; no human clinical data. Equivalence supported by side-by-side canine bilateral duraplasty model (n=6 animals total, 2 defects per animal). Endpoints included clinical health, neurological evaluation, CSF analysis, and histopathology per ISO 10993-6 at 4 weeks. Results showed no CSF leaks, similar neoduralization, and equivalent tissue reaction between test and predicate devices.

## Technological Characteristics

Porous PGLA/PDO polymer matrix; resorbable. Supplied sterile (SAL 10^-6) in double sterile nested pouch. No specific orientation required. Dimensions vary by size (up to 5"x7").

## Regulatory Identification

A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).

## Predicate Devices

- Cerafix Dura Substitute ([K153613](/device/K153613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of the human profile in triplicate.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2016

Acera Surgical, Inc. % Linda Braddon, Ph.D. President and CEO Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188

Re: K161278

Trade/Device Name: Cerafix Dura Substitute Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: May 5, 2016 Received: May 10, 2016

Dear Dr. Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J. Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K161278

Device Name Cerafix Dura Substitute

#### Indications for Use (Describe)

The Cerafix Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 4.4 in2 (28.3 cm2) or less in area. For example, 4.0 in x 1.1 in (10.1 cm x 2.8 cm) would be an acceptable defect size.

| Type of Use (Select one or both, as applicable)                                                          |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                                                            |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary of Safety and Effectiveness

Image /page/3/Picture/1 description: The image shows the logo for Acera Surgical. The logo consists of a green triangle on the left, with a blue curved shape at the bottom. To the right of the triangle is the word "ACERA" in green, with the word "SURGICAL" below it in a smaller font.

In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the Acera Surgical Cerafix® Dura Substitute is provided below.

| Date Summary Prepared    | August 1, 2016                                                                                                                                                                            |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by             | Acera Surgical, Inc.<br>10880 Baur Blvd<br>St. Louis, MO 63132<br>Phone 844-879-2237                                                                                                      |
| 510(k) Contact           | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>855-MED-DEV1 (office)<br>LGB@SecureBME.com |
| Trade Name               | Cerafix® Dura Substitute                                                                                                                                                                  |
| Common Name              | Dura substitute                                                                                                                                                                           |
| Code -Classification     | GXQ 21 CFR 882.5910 : Class II                                                                                                                                                            |
| Primary Predicate Device | K153613 Cerafix® Dura Substitute                                                                                                                                                          |

#### Device Description

Cerafix® Dura Substitute is a resorbable implant for repair of dural defects and is to be used with tensionless sutures. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.

#### Indications for Use

The Cerafix® Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 4.4 in (28.3cm²) or less in area. For example, 4.0 in x 1.1 in (10.1 cm x 2.8 cm) would be an acceptable defect size.

{4}------------------------------------------------

### Technological Characteristics

The component material used in the subject device is the exact same as the component material used in the predicate device. The only difference is finished device size.

Based on test results included in this submission, a maximum allowable defect size has been prescribed for the subject device. The subject device has the same technological characteristics as the predicate device in terms of principles of operation, materials of construction, material performance, and biocompatibility. Additionally, side-by-side animal studies show the subject device is equivalent for the indicated use of a dura substitute for the repair of dura mater. The subject device has the same technological characteristics as the predicate device as follows:

| Characteristic              | Cerafix® Dura Substitute<br>(subject device)                                                                                                                                                                                                                                       | Cerafix® Dura Substitute<br>(predicate device)                                                                                                                                                                                                                                     | Comparison                                                                                    |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| 510(k)                      | K161278                                                                                                                                                                                                                                                                            | K153613                                                                                                                                                                                                                                                                            | N/A                                                                                           |
| Principles of<br>Operation  | Device can be cut by<br>surgeon and placed on<br>dural defect with<br>tensionless suture<br>application. Suture line<br>should be 2-3 mm from<br>edge of implant. Implant<br>should be large enough to<br>overlap edge of the<br>remaining dura by at least<br>one (1) centimeter. | Device can be cut by<br>surgeon and placed on<br>dural defect with<br>tensionless suture<br>application. Suture line<br>should be 2-3 mm from<br>edge of implant. Implant<br>should be large enough<br>to overlap edge of the<br>remaining dura by at<br>least one (1) centimeter. | Equivalent to<br>predicate device                                                             |
| Material of<br>Construction | Porous polymer matrix                                                                                                                                                                                                                                                              | Porous polymer matrix                                                                                                                                                                                                                                                              | Equivalent to<br>predicate device                                                             |
| Indications for<br>Use      | Indicated as a dura<br>substitute for the repair<br>of dura mater. This device<br>is indicated for defects of<br>4.4 in2 (28.3cm²) or less<br>in area. For example, 4.0<br>in x 1.1 in (10.1 cm x 2.8<br>cm) would be an<br>acceptable defect size.                                | Indicated as a dura<br>substitute for the repair<br>of dura mater. This<br>device is indicated for<br>defects of 1.9 in2<br>(12.5cm²) or less in area.<br>For example, 1.2 in x 1.6<br>in (3 cm x 4 cm) would<br>be an acceptable defect<br>size.                                  | Equivalent to<br>predicate device                                                             |
| Size                        | 1"x1"<br>1"x3"<br>2"x2"<br>3"x3"<br>4"x5"<br>5"x7"                                                                                                                                                                                                                                 | 1"x1"<br>1"x3"<br>2"x2"<br>3"x3"                                                                                                                                                                                                                                                   | Subject device is cut<br>into larger finished<br>device sizes compared<br>to predicate device |
| Material<br>Composition     | Porous PGLA / PDO matrix                                                                                                                                                                                                                                                           | Porous PGLA / PDO<br>matrix                                                                                                                                                                                                                                                        | Equivalent to<br>predicate device                                                             |

{5}------------------------------------------------

| Surgical<br>Application<br>Restrictions | Device does not have<br>requirement for specific<br>orientation                       | Device does not have<br>requirement for specific<br>orientation                       | Equivalent to<br>predicate device |
|-----------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------|
| Sterility                               | Sterile, SAL 10-6                                                                     | Sterile, SAL 10-6                                                                     | Equivalent                        |
| Packaging                               | Double sterile pack.<br>Nested pouch<br>configuration within a<br>chipboard envelope. | Double sterile pack.<br>Nested pouch<br>configuration within a<br>chipboard envelope. | Equivalent to<br>predicate device |
| Pyrogenicity                            | Non-pyrogenic                                                                         | Non-pyrogenic                                                                         | Equivalent                        |
| Resorbable                              | Yes                                                                                   | Yes                                                                                   | Equivalent to<br>predicate device |
| Biocompatibility                        | Biocompatible                                                                         | Biocompatible                                                                         | Equivalent                        |

The following technological differences exist between the subject and predicate devices:

- . Subject device and predicate device are manufactured from the exact same component material; however, the subject device will be cut into larger finished device sizes.
Despite the size differences, the subject device is equivalent in function, indication for use, device classification product code, environment of use, and principles of operation to the predicate device. The sizes of the subject device are equivalent to the reference device.

{6}------------------------------------------------

#### Non-Clinical Testing — Mechanical

No new mechanical testing was conducted for this submission. All mechanical testing for the Cerafix® Dura Substitute device was submitted in K153613. The subject device is cut from the same component material as the predicate device, and therefore has equivalent mechanical properties per unit area.

#### Non-Clinical Testing - Biocompatibility

No new biocompatibility testing was conducted for this submission. All biocompatibility testing for the Cerafix® Dura Substitute device was submitted in K153613. The subject device is cut from the same component material as the predicate device, and therefore has equivalent biocompatibility.

#### Non-Clinical Testing – Side-by-Side Animal Study Comparison

The subject device (test) was compared to the predicate device (control) using a canine bilateral duraplasty model with terminal assessments at 4-weeks post-operatively. Each time point evaluated 3 test and 3 control animals, each with 2 defects. Animals were observed daily and weighed weekly for signs of general clinical health. Particular attention was paid to the implant sites to assess the presence of hydrocephalus, cerebral spinal fluid (CSF) leakage, hemorrhage, infection, and any other adverse side effects. At each end-point, CSF was obtained from each animal and submitted for analysis. Tissue samples from each defect site were processed by histopathology techniques and analyzed for dural integrity / neoduralization and local tissue reactions according to ISO 10993-6.

All animals in both test and control groups appeared healthy with normal neurological evaluations at both time points. No CSF leaks were observed in either group throughout the duration of both studies. Test and control articles were shown to be similar in the mechanism of neoduralization and absorption, independent of the size of the induced dural defect at both time points. Test and control articles demonstrated signs of resorption with infiltration of fibrovascular connective tissue and successful neoduralization.

In total, both the test and control articles showed equivalent clinical performance at each time point and successfully repaired the induced dural defects independent of defect size. Therefore, the results of these studies indicate that the subject device demonstrates equivalent safety and efficacy when compared to the predicate device.

#### Conclusions

The subject and predicate device underwent non-clinical evaluation that confirmed device equivalency in the indication for use, device classification, product code, biocompatibility, safety, efficacy, environment of use, and the principles of operation. Therefore, the subject device demonstrates equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K161278](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K161278)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
