← Product Code [GXP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP) · K241027

# Permatage Settable Bone Putty (K241027)

_Orthocon, Inc. · GXP · Sep 26, 2024 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K241027

## Device Facts

- **Applicant:** Orthocon, Inc.
- **Product Code:** [GXP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP.md)
- **Decision Date:** Sep 26, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5300
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Orthocon Permatage Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage Settable Bone Putty should be used only in skeletally mature individuals.

## Device Story

Permatage Settable Bone Putty is a sterile, nonabsorbable, biocompatible material for cranial defect repair. Device consists of two putty-like components containing granular calcium phosphate, paraffin oil, vitamin E acetate, triglyceride, and polyether-based polymers. Components are manually mixed immediately prior to surgery to form a cohesive, settable material that adheres to bone surfaces. Applied manually by surgeons in a clinical setting; cures within 10 minutes without appreciable exothermic reaction. Provides a stable, radiopaque filler for burr holes and craniotomy cuts. Benefits include effective cranial defect reconstruction in skeletally mature patients.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence supported by bench testing (visual inspection, stiffness, leak testing, temperature sensitivity, water uptake/swelling/dissolution) and in vivo animal testing evaluating histopathologic response in a critical-sized cranial defect model.

## Technological Characteristics

Two-part putty-like material; components: granular calcium phosphate, paraffin oil, vitamin E acetate, triglyceride, polyether-based polymers. Radiopaque. Nonabsorbable. Manual mixing (45 seconds). Settable within 10 minutes; 2-minute working time. No exothermic reaction. Sterilized via gamma irradiation. Single-patient use.

## Regulatory Identification

Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.

## Predicate Devices

- Stryker HydroSet Injectable Cement ([K060763](/device/K060763.md))

## Reference Devices

- Orthocon, Inc. HBP7 Settable Hemostatic Bone Putty ([K202363](/device/K202363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 26, 2024

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, which includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Orthocon, Inc. Howard Schrayer Regulatory Consultant 700 Fairfield Avenue, Suite 1 Stamford, Connecticut 06902

Re: K241027

Trade/Device Name: Permatage Settable Bone Putty Regulation Number: 21 CFR 882.5300 Regulation Name: Methyl Methacrylate For Cranioplasty Regulatory Class: Class II Product Code: GXP Dated: August 27, 2024 Received: August 27, 2024

Dear Howard Schrayer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Adam D. Adam D. Pierce -S Pierce -S Date: 2024.09.26 13:53:13 -04'00'

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K241027

Device Name

#### Permatage Settable Bone Putty

Indications for Use (Describe)

Orthocon Permatage Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage Settable Bone Putty should be used only in skeletally mature individuals.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) SUMMARY- K241027 (Per 21 CFR 807.92)

#### General Company Information

| Name:                      | Orthocon, Inc.                                      |
|----------------------------|-----------------------------------------------------|
| Contact:                   | Howard Schrayer<br>Regulatory Affairs Consultant    |
| Address:                   | 700 Fairfield Avenue, Suite 1<br>Stamford, CT 06902 |
| Telephone:                 | (855) 475 - 9175                                    |
| Date                       | September 26, 2024                                  |
| General Device Information |                                                     |

| Product Name:   | Permatage Settable Bone Putty        |
|-----------------|--------------------------------------|
| Common Name:    | Methyl Methacrylate for Cranioplasty |
| Classification: | Class II                             |
| Product codes:  | GXP                                  |
| Regulation:     | 21 CFR 882.5300                      |

# Predicate Devices: Primary Predicate:

| Stryker | HydroSet Injectable Cement |
|---------|----------------------------|
|         | [510(k) Number K060763]    |

## Reference Devices:

Orthocon, Inc. HBP7 Settable Hemostatic Bone Putty [510(k) Number K202363]

## Device Description

Permatage Settable Bone Putty is a sterile, biocompatible, nonabsorbable material of putty-like consistency for use in repair of cranial defects. The single use Permatage device contains two separate components of putty-like consistency comprised of granular calcium phosphate, paraffin oil, vitamin E acetate, a triglyceride, and a mixture of nonabsorbable, polyether-based polymers. When mixed together, the components of the Permatage device form a nonabsorbable cohesive, putty-like material that adheres to the bone surface and remains in place following application. The resulting hardening material is primarily calcium phosphate and nonabsorbable polymer materials. Permatage components must be mixed immediately prior to use.

{5}------------------------------------------------

#### Indications for Use

Orthocon Permatage Settable Bone Putty is a self-setting cement indicated for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. Permatage Settable Bone Putty should be used only in skeletally mature individuals.

The following table shows comparisons of characteristics of Permatage Settable Bone Putty and the predicate device.

## SUBSTANTIAL EQUIVALENCE INFORMATION

Orthocon, Inc. Permatage Settable Bone Putty 510(k) — K241027

Stryker Injectable Cement HydroSet

## 510(k) - K060763

#### Comparisons of Technological Characteristics

| Permatage Settable Bone Putty is intended<br>for use in the repair of neurosurgical burr<br>holes, contiguous craniotomy cuts and<br>other cranial defects.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Stryker® Injectable Cement is intended for<br>use in the repair of neurosurgical burr<br>holes, contiguous craniotomy cuts and<br>other cranial defects.                                                                                                                                                                                                                                                                               |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code GXP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Product Code GXP                                                                                                                                                                                                                                                                                                                                                                                                                       |
| At the time of application, device is in the<br>form of a spreadable material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | At the time of application, device is in the<br>form of a spreadable material                                                                                                                                                                                                                                                                                                                                                          |
| Device is designed to be manually applied<br>to the cranial defect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Device is designed to be applied to the<br>cranial defect with a manually operated<br>dispenser                                                                                                                                                                                                                                                                                                                                        |
| Permatage Settable Bone Putty is<br>formulated as a two-part putty/putty device<br>that forms a "settable" (hardening) material<br>when mixed at the time of surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Stryker Injectable cement is formulated as<br>a two-part powder/liquid device that forms<br>a "settable" (hardening) material when<br>mixed at the time of surgery                                                                                                                                                                                                                                                                     |
| Permatage Settable Bone Putty is a sterile,<br>biocompatible, nonabsorbable material of<br>putty-like consistency for use in repair of<br>cranial defects. The single use Permatage<br>device contains two separate components of<br>putty-like consistency comprised of granular<br>calcium phosphate, paraffin oil, vitamin E<br>acetate, a triglyceride, and a mixture of<br>nonabsorbable, polyether-based polymers.<br>When mixed together, the components of the<br>Permatage device form a nonabsorbable<br>cohesive putty-like material that adheres to the<br>bone surface and remains in place following<br>application. The resulting hardening material is<br>primarily comprised of calcium phosphate and<br>nonabsorbable polymer materials. | Sterile mixture of two separate<br>components, a powder comprised of<br>dicalcium phosphate dihydrate,<br>tetracalcium phosphate and tri-sodium<br>citrate; and a liquid comprised of sodium<br>phosphate, polyvinylpyrrolidone and<br>water. Stryker Injectable Cement is to be<br>manually mixed immediately prior to use.<br>Resulting settable material from the two<br>components is primarily comprised of<br>calcium phosphate. |

{6}------------------------------------------------

| Radiopaque - Contains hydroxyapatite and β-<br>tricalcium phosphate                                                                 | Radiopaque - Contains hydroxyapatite and β-<br>tricalcium phosphate                                                                                  |
|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implanted device is nonabsorbable.                                                                                                  | Implanted device is resorbable in greater<br>than 2 years primarily due to presence of<br>calcium phosphate.                                         |
| Single-patient-use device is provided sterile<br>by gamma irradiation                                                               | Single-patient-use device is provided<br>sterile by gamma irradiation and ethylene<br>oxide                                                          |
| The bone putty is available in individual<br>sizes of 5cc.                                                                          | The device is available in individual;<br>and/or multi-pack patient use sizes of 3, 5,<br>10 and 15cc.                                               |
| The putty is provided in two foil packages<br>containing putty-like components that are<br>mixed together immediately prior to use. | Each kit contains one liquid-filled glass<br>syringe and one plastic bowl of powder<br>packaged within a double pre-formed tray<br>with a Tyvek lid. |
| Mixing for homogeneity takes 45 seconds                                                                                             | Mixing for homogeneity takes 45 sec.                                                                                                                 |
| Material is settable within 10 minutes of<br>application                                                                            | Material is settable within 10 minutes of<br>application                                                                                             |
| Material provides a working time of 2<br>minutes.                                                                                   | Material provides a working time of 5<br>minutes.                                                                                                    |
| Device cures with no appreciable<br>exothermic reaction.                                                                            | Device cures with no appreciable<br>exothermic reaction                                                                                              |

## Testing Completed

# Biocompatibility Testing

Testing was conducted to evaluate biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiation sterilized device and in accordance with the GLP requirements: cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, local tissue toxicity, hemolysis, pyrogenicity and neurotoxicity.

# Performance Data

Performance testing included a series of laboratory evaluations and in vivo testing. These evaluations are summarized below.

{7}------------------------------------------------

# Bench Testing

| Test                                         | Description                                          | Conclusions                                          |
|----------------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Visual Inspection                            | Evaluated paste color<br>using a reference scale     | Paste color<br>met specification                     |
| Paste Stiffness                              | Evaluated paste stiffness<br>using a reference scale | Paste stiffness met<br>specification                 |
| Package Leak Test                            | Bubble emission leak test                            | All test articles passed                             |
| Temperature Sensitivity                      | Acceptable maximum temperature<br>increase observed  | Acceptable maximum<br>temperature increase observed  |
| Water Uptake,<br>Swelling<br>and Dissolution | Measured volume and mass<br>changes over time        | Acceptable water uptake,<br>swelling and dissolution |

# In-Vivo Testing

In-vivo animal testing was used to demonstrate substantial equivalence of Permatage Settable Bone Putty in the repair of a critical sized cranial bone defect in an animal model compared to the comparator device. Substantial equivalence was assessed based on histopathologic evaluation.

## Clinical Testing

No clinical studies have been conducted in support of this 510(k).

## Conclusions

This submission supports the position that Orthocon Permatage Settable Bone Putty is substantially equivalent to the predicate device.

The information provided establishes that similar legally marketed devices, including the primary predicate, have been used for the same clinical application as Orthocon Permatage Settable Bone Putty and that Substantial Equivalence to the predicate device has been established. Each of the tests conducted passed the requirements as stated in the protocols and in recognized standards. The materials from which the Orthocon device is fabricated have an established history of use, and the devices have been tested in accordance with applicable FDA quidelines. Any differences between Permatage and the predicate do not raise new concerns or risks.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K241027](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K241027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
