21 CFR 882.5892 — Non-Invasive Vagus Nerve Stimulator - Headache

Neurology (NE) · Part 882 Subpart F—Neurological Therapeutic Devices · § 882.5892

Identification

An external vagal nerve stimulator for headache is a prescription device used to apply an electrical current to a patient's vagus nerve through electrodes placed on the skin for the treatment of headache.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
PKRNon-Invasive Vagus Nerve Stimulator - Headache29
QAKNon-Invasive Vagus Nerve Stimulator For Migraine Headache20

Special Controls

PKR — Non-Invasive Vagus Nerve Stimulator - Headache

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm 2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed. (3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed. (4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC). (5) The labeling must include: (i) Instructions for proper use of the device, including placement of the device on the patient; and (ii) Instructions on care and cleaning of the device.

eCFR

PKR — Non-Invasive Vagus Nerve Stimulator - Headache

(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm 2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed. (3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed. (4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC). (5) The labeling must include: (i) Instructions for proper use of the device, including placement of the device on the patient; and (ii) Instructions on care and cleaning of the device.

Ecfr Llm

QAK — Non-Invasive Vagus Nerve Stimulator For Migraine Headache

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm 2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed. (3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed. (4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC). (5) The labeling must include: (i) Instructions for proper use of the device, including placement of the device on the patient; and (ii) Instructions on care and cleaning of the device.

eCFR

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...