21 CFR 882.5700 — Thermal System For Insomnia

Neurology (NE) · Part 882 Subpart F—Neurological Therapeutic Devices · § 882.5700

Identification

A thermal system for insomnia is a prescription device for use in patients with insomnia that is used to apply a specified temperature to the skin surface.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
PLUThermal System For Insomnia21

Special Controls

PLU — Thermal System For Insomnia

In combination with the general controls of the FD&C Act, the Thermal System for Insomnia is subject to the following special controls: 1. The patient-contacting components of the device must be demonstrated to be biocompatible. 2. Performance testing must demonstrate electromagnetic compatibility and electrical safety. 3. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: a. Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions. Mechanical testing to demonstrate the device can withstand forces under b. anticipated use conditions. c. Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions. 4. Software verification, validation, and hazard analysis must be performed. 5. Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.

De Novo Order DEN140032

PLU — Thermal System For Insomnia

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Performance testing must demonstrate electromagnetic compatibility and electrical safety. (3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: (i) Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions. (ii) Mechanical testing to demonstrate the device can withstand forces under anticipated use conditions. (iii) Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions. (4) Software verification, validation, and hazard analysis must be performed. (5) Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.

eCFR

PLU — Thermal System For Insomnia

(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Performance testing must demonstrate electromagnetic compatibility and electrical safety. (3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: (i) Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions. (ii) Mechanical testing to demonstrate the device can withstand forces under anticipated use conditions. (iii) Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions. (4) Software verification, validation, and hazard analysis must be performed. (5) Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.

Ecfr Llm

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...