K243623 · Surgical Theater, Inc. · SBF · Dec 24, 2024 · Neurology
Device Facts
Record ID
K243623
Device Name
SpineAR SNAP (SyncAR Spine)
Applicant
Surgical Theater, Inc.
Product Code
SBF · Neurology
Decision Date
Dec 24, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
SpineAR SNAP is intended for use for pre-operative surgical planning on-screen and in a virtual environment, and intra-operative surgical planning and visualization on-screen and in an augmented environment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling. SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 and HoloLens2 AR headsets. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
Device Story
Software-based system for pre-operative planning and intra-operative visualization; runs on high-performance COTS desktop PC. Inputs: 2D medical images; navigation data (tool position/orientation) from 3rd party systems (e.g., Medtronic StealthStation S8). Transforms 2D images into interactive 3D scenes; provides guidance tools (dotted line, orientation line, ILS crosshairs) with color-coded alignment cues. Outputs: 3D visualization on HD touch screen or streamed to AR headsets (HoloLens2, Magic Leap 1). Used by surgeons in OR; AR display projected above patient. Enhances spatial awareness during pedicle screw placement; does not replace 2D stereotaxic navigation. Benefits: improved visualization of surgical plan relative to anatomy.
Clinical Evidence
Bench testing only. Accuracy of pedicle screw placement in thoracic region assessed using spine model and post-surgical CT scans. Compared HoloLens2 performance against predicate Magic Leap 1. Results: Mean positional error 0.76mm (vs 1.04mm), Max positional error 1.47mm (vs 1.90mm), Mean angular error 1.73° (vs 1.92°), Max angular error 2.96° (vs 2.80°). All metrics met pre-defined accuracy requirements.
Technological Characteristics
Software-based; runs on COTS high-performance desktop PC. Interfaces with 3rd party navigation systems (e.g., Medtronic StealthStation S8). Visualizes 3D scenes on HD touch screen or AR headsets (HoloLens2, Magic Leap 1). Provides guidance via visual cues (dotted/orientation lines, crosshairs). No custom hardware.
Indications for Use
Indicated for spinal stereotaxic surgery in patients requiring spinal implant procedures (e.g., Pedicle Screw Placement) in lumbar and thoracic regions. Used with Magic Leap 1 or HoloLens2 AR headsets.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K233725 — Spine Navigation System · Medivis, Inc. · Jul 26, 2024
K220146 — VisAR · Novarad Corporation · May 27, 2022
K211254 — ARAI Surgical Navigation System · Surgalign Spine Technologies · Jan 14, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
December 24, 2024
Surgical Theater, Inc. Kevin Murrock Sr. Director of Quality & Regulatory 23645 Mercantile Road Suite M Beachwood, Ohio 44122
Re: K243623
Trade/Device Name: SpineAR SNAP (SyncAR Spine) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: SBF, LLZ Dated: November 21, 2024 Received: November 25, 2024
Dear Kevin Murrock:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}------------------------------------------------
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243623
Device Name SpineAR SNAP
#### Indications for Use (Describe)
SpineAR SNAP is intended for use for pre-operative surgical planning on-screen and in a virtual environment, and intraoperative surgical planning and visualization on-screen and in an augment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling.
SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 and HoloLens2 AR headsets.
The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in orange and red gradient, followed by the words "SURGICAL THEATER" in red and orange. The words are stacked on top of each other, with "SURGICAL" on top and "THEATER" on the bottom.
## 510(k) Premarket Notification
## 510(k) SUMMARY
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 8807.92.
## Date Prepared: 23-Dec-2024
510(k) Type: Special 510(k) for Device Modification
#### Manufacturer/Submitter:
Surgical Theater, Inc. 23645 Mercantile Road, Suite M Beachwood, Ohio 44122 Phone: (330) 472-6520
Establishment Registration Number: 3010197287
#### Contact Person:
Kevin M. Murrock Sr. Director of Quality and Regulatory 23645 Mercantile Road, Suite M Beachwood, Ohio 44122 Phone: (330) 472-6520 Email: kmurrock@surgicaltheater.com
## Name of Device
- Trade Name: SpineAR SNAP ●
- Other Device Trade Names:
- Common Name: ●
- Classification Name:
- Regulation Number:
- Product Code:
- Regulatory Classification: ●
- Device Panel: ●
SyncAR Spine Augmented Reality System Orthopedic Stereotaxic Instrument 21 CFR 882.4560 SBF, LLZ II Orthopedic
## Predicate Device
Device Name: SpineAR SNAP Manufacturer: Surgical Theater, Inc. 510(k) Number: K213034
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Surgical Theater, which includes a stylized "S" shape in gradient colors from red to yellow. Below the logo, the words "SURGICAL THEATER" are written in a combination of red and yellow. Underneath the logo and text, there is a double underline, followed by the words "510(k) Premarket Notification" in bold font.
SpineAR SNAP
# Device Description:
The SpineAR SNAP does not require any custom hardware and is a software-based device that runs on a high-performance desktop PC assembled using "commercial off-the-shelf" components that meet minimum performance requirements.
The SpineAR SNAP software transforms 2D medical images into a dynamic interactive 3D scene with multiple point of views for viewing on a high-definition (HD) touch screen monitor. The surgeon prepares a pre-operative plan for stereotaxic spine surgery by inserting guidance objects such as directional markers and virtual screws into the 3D scene. Surgical planning tools and functions are available on-screen and when using a virtual reality (VR) headset. The use of a VR headset for preoperative surgical planning further increases the surgeon's immersion level in the 3D scene by providing a 3D stereoscopic display of the same 3D scene displayed on the touch screen monitor.
By interfacing to a 3rd party navigation system such as a Medtronic StealthStation S8, the SpineAR SNAP extracts the navigation data (i.e. tool position and orientation) and presents the navigation data into the advanced interactive, high quality 3D image, with multiple point of views on a high-definition (HD) touch screen monitor. Once connected, the surgeon can then execute the plan through the intraoperative use of the SpineAR SNAP's enhanced visualization and guidance tools.
The SpineAR SNAP supports three (3) guidance options from which the surgeon selects the level of guidance that will be shown in the 3D scene. The guidance options are dotted line (indicates deviation distance), orientation line (indicates both distance and angular deviation), and ILS (indicates both distance and angular deviation using crosshairs). Visual color-coded cues indicate alignment of the tracker tip to the guidance object (e.g. green = aligned).
The 3D scene with guidance tools can also be streamed into an AR wireless headset (Magic Leap 1 or HoloLens2) worn by the surgeon during surgery. The 3D scene and guidance shown within the AR headset is projected above the patient and does not obstruct the surgeon's view of the surgical space.
## Intended Use / Indications for Use:
SpineAR SNAP is intended for use for pre-operative surgical planning on-screen and in a virtual environment, and intra-operative surgical planning and visualization on-screen and in an augmented environment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling.
SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 and HoloLens2 AR headsets.
The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
{6}------------------------------------------------
## Summary of Technological Characteristics
The modified SpineAR SNAP has identical technological characteristics, principle of operation, and intended use of the predicate device. No design changes are included in the device modification for this 510(k) submission as only non-clinical performance testing was necessary to support the expanded indications for use of the SpineAR SNAP with the HoloLens2 AR headset from spinal implant procedures in lumbar region only to also include spinal implant procedures in thoracic region.
The following table is a comparison of the technological characteristics of the subject and predicate devices included in the scope of this 510(k).
| Feature | Predicate Device:<br>SpineAR SNAP | Subject Device:<br>SpineAR SNAP |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Surgical Theater, Inc. | Surgical Theater, Inc. |
| Indications for Use | SpineAR SNAP is intended for use<br>for pre-operative surgical planning<br>on-screen and in a virtual<br>environment, and intra-operative<br>surgical planning and visualization<br>on-screen and in an augmented<br>environment using the HoloLens2<br>and Magic Leap 1 AR headset<br>displays with validated navigation<br>systems as identified in the device<br>labeling.<br><br>SpineAR SNAP is indicated for<br>spinal stereotaxic surgery, and<br>where reference to a rigid<br>anatomical structure, such as the<br>spine, can be identified relative to<br>images of the anatomy. SpineAR<br>is intended for use in spinal<br>implant procedures, such as<br>Pedicle Screw Placement, in the<br>lumbar and thoracic regions with<br>the Magic Leap 1 AR headset, and<br>in the lumbar region with the<br>HoloLens2 AR headset.<br><br>The virtual display should not be<br>relied upon solely for absolute<br>positional information and should<br>always be used in conjunction with | SpineAR SNAP is intended for use<br>for pre-operative surgical planning<br>on-screen and in a virtual<br>environment, and intra-operative<br>surgical planning and visualization<br>on-screen and in an augmented<br>environment using the HoloLens2<br>and Magic Leap 1 AR headset<br>displays with validated navigation<br>systems as identified in the device<br>labeling.<br><br>SpineAR SNAP is indicated for<br>spinal stereotaxic surgery, and<br>where reference to a rigid<br>anatomical structure, such as the<br>spine, can be identified relative to<br>images of the anatomy. SpineAR is<br>intended for use in spinal implant<br>procedures, such as Pedicle Screw<br>Placement, in the lumbar and<br>thoracic regions with the Magic<br>Leap 1 and HoloLens2 AR headsets.<br><br>The virtual display should not be<br>relied upon solely for absolute<br>positional information and should<br>always be used in conjunction with<br>the displayed 2D stereotaxic<br>information. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in red and orange gradient, followed by the words "SURGICAL THEATER" in red and yellow gradient. There is a trademark symbol after the word "THEATER", and two black lines underneath the words.
## 510(k) Premarket Notification
| Feature | Predicate Device:<br>SpineAR SNAP | Subject Device:<br>SpineAR SNAP |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | the displayed 2D stereotaxic information. | |
| Magic Leap 1 Intra-operative Indications for Use | Spinal implant procedures, such as<br>Pedicle Screw Placement, in both<br>the lumbar and thoracic regions. | Spinal implant procedures, such as<br>Pedicle Screw Placement, in <b>both</b> the<br>lumbar and thoracic regions. |
| HoloLens2 Intra-operative Indications for Use | Spinal implant procedures, such as<br>Pedicle Screw Placement, in the<br>lumbar region <b>only</b> . | Spinal implant procedures, such as<br>Pedicle Screw Placement, in <b>both</b> the<br>lumbar and thoracic regions. |
## Performance Data
Performance testing was conducted to demonstrate the SpineAR with HoloLens2 AR headset is substantially equivalent to the use of the Magic Leap 1 AR headset cleared in the predicate SpineAR SNAP when performing spinal implant procedures in the thoracic region. The final placement of each screw in a spine model was assessed via a post-surgical CT scan and compared to the pre-surgical plan to calculate the absolute displacement of the screw and absolute angular error of the screw. The HoloLens2 navigation accuracy test results in Table 1 demonstrate performance in 3D positional and trajectory accuracy that meets the accuracy of the predicate SpineAR SNAP using the Magic Leap 1 AR headset.
| | Accuracy<br>Requirement | Magic Leap 1 AR<br>Headset Results | HoloLens2 AR<br>Headset Results |
|------------------------------------|-------------------------|------------------------------------|---------------------------------|
| Mean Positional/Displacement Error | $\leq$ 2.0 mm | 1.04 mm | 0.76 mm |
| Max Positional/Displacement Error | $\leq$ 3.0 mm | 1.90 mm | 1.47 mm |
| Mean Trajectory/Angular Error | $\leq$ 2° | 1.92° | 1.73° |
| Max Trajectory/Angular Error | $\leq$ 3° | 2.80° | 2.96° |
Table 1 - Performance Test Results: Accuracy of Pedicle Screw Placement in Thoracic Region
Performance data demonstrated the SpineAR SNAP functions as intended without raising new safety or effectiveness issues, and does not impact accuracy or clinical workflow. The test results support a finding that the use of the HoloLens2 AR headset in the subject device is substantially equivalent to the use of the Magic Leap 1 AR headset in the predicate SpineAR SNAP when performing spinal implant procedures in the thoracic region.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in orange and red gradient, followed by the words "SURGICAL" in red and "THEATER" in yellow. There are two horizontal lines under the text.
510(k) Premarket Notification
## Substantial Equivalence Conclusion
The SpineAR SNAP is substantially equivalent to its predicate. Both the subject and predicate devices have identical intended use, technological characteristics, and principles of operation. The indications for use are similar with the only difference being the expanded use of the HoloLens2 AR headset for spinal implant procedures in the thoracic region. Previously the use of the HoloLens2 was only indicated for spinal implant procedures in the lumbar region, while the Magic Leap 1 AR headset was indicated for use in both the lumbar and thoracic regions. The modified indications for use did not require any design change, only the acquisition of additional performance data supporting the use of the HoloLens2 with SpineAR for spinal implant procedures in the thoracic region.
Any differences between the subject and predicate device do not raise any new concerns regarding safety and effectiveness, and performance testing demonstrates that the subject device is at least as safe and effective as the predicate device, and functions as intended.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.