ExcelsiusXR™

K241525 · Globus Medical, Inc. · SBF · Jul 8, 2025 · Neurology

Device Facts

Record IDK241525
Device NameExcelsiusXR™
ApplicantGlobus Medical, Inc.
Product CodeSBF · Neurology
Decision DateJul 8, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

ExcelsiusXR™, when used in conjunction with ExcelsiusHub™ and/or ExcelsiusGPS®, is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. This system is indicated for the placement of spinal and orthopedic (Sacroiliac, Ulna, and Tibia) bone screws, and interbody fusion devices. The ExcelsiusXR™ Headset displays 2D stereotactic images and 3D virtual anatomy images, and displays the virtual instrument location in relation to the virtual anatomy to assist in percutaneous visualization and trajectory planning. This headset should not be relied upon solely for absolute positional information and should always be used in conjunction with the primary stereotactic display.

Device Story

Head-mounted navigation device (headset) used with ExcelsiusHub/ExcelsiusGPS; inputs include preoperative CT, intraoperative CT, and fluoroscopy images. System tracks compatible surgical instruments and uses hand-tracking cameras for display manipulation. Headset projects 2D stereotactic and 3D virtual anatomy images, plus virtual instrument location, onto surgeon's retina via transparent near-eye display. Used in operating room by surgeons to visualize anatomy and instrument trajectory in real-time. Complements primary stereotactic display; assists in percutaneous visualization and trajectory planning. Benefits include improved precision in locating anatomical structures and positioning implants/instruments during orthopedic and spinal procedures.

Clinical Evidence

No clinical data. Evidence consists of non-clinical system, software, and instrument verification and validation; surgical simulations on phantom models; qualitative and quantitative validation; and optical bench testing for head-mounted display image quality. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1:2020 and IEC 60601-1-2:2020.

Technological Characteristics

Head-mounted display (headset) with optical tracking and hand-tracking cameras. Integrates with ExcelsiusHub/GPS for navigation. Software-based visualization of 2D/3D radiological data. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 62304 (software lifecycle), and IEC 62366-1 (usability engineering).

Indications for Use

Indicated for surgeons performing open or percutaneous spinal and orthopedic (Sacroiliac, Ulna, and Tibia) bone screw placement and interbody fusion device implantation, requiring identification of fiducial markers and rigid anatomy via CT or fluoroscopy.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 8, 2025 Globus Medical, Inc. Jennifer Antonacci Senior Group Manager, Regulatory Affairs Valley Forge Business Center 2560 General Armistead Ave Audubon, Pennsylvania 19403 Re: K241525 Trade/Device Name: ExcelsiusXR™ Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument. Regulatory Class: Class II Product Code: SBF, Dated: March 21, 2025 Received: March 24, 2025 Dear Jennifer Antonacci: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K241525 - Jennifer Antonacci Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K241525 - Jennifer Antonacci Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K241525 | | | Device Name ExcelsiusXR™ | | | Indications for Use (Describe) ExcelsiusXR™, when used in conjunction with ExcelsiusHub™ and/or ExcelsiusGPS®, is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. This system is indicated for the placement of spinal and orthopedic (Sacroiliac, Ulna, and Tibia) bone screws, and interbody fusion devices. The ExcelsiusXR™ Headset displays 2D stereotactic images and 3D virtual anatomy images, and displays the virtual instrument location in relation to the virtual anatomy to assist in percutaneous visualization and trajectory planning. This headset should not be relied upon solely for absolute positional information and should always be used in conjunction with the primary stereotactic display. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K241525- page 1 of 4 # 510(k) Summary: ExcelsiusXR™ Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon, PA 19403 610-930-1800 Primary Contact: Jennifer Antonacci, Ph.D. Director, Regulatory Affairs Date Prepared: July 7, 2025 Device Name: ExcelsiusXR™ Common Name: Computer-assisted surgical device Classification: Per 21 CFR as follows: §882.4560 Stereotaxic instrument Product Code: SBF (Primary), Regulatory Class: II Primary Predicate: Augmedics xvision Spine System (K220905) Additional Predicates: ExcelsiusHub™ (K211616) ExcelsiusGPS® (K171651, K191100) # Purpose: The purpose of this submission is to request clearance for ExcelsiusXR™ and additional instruments. # Device Description: ExcelsiusXR™ is a head-mounted navigation device, or headset, that is used in conjunction with ExcelsiusHub, and ExcelsiusGPS if robotic guidance is desired, as an aid for precisely locating anatomical structures in open or percutaneous procedures, and for precisely positioning compatible surgical instruments or implants (screws and interbody devices) during surgery. ExcelsiusXR™ includes hardware and software that enables real-time surgical visualization using radiological patient images (preoperative CT, intraoperative CT, and fluoroscopy), provides tracking and planning capabilities for a series of compatible instruments, and contains hand tracking cameras for manipulation of the head-mounted display by the user. The Headset displays 2D stereotactic images and provides a 3D visual, or virtual image, of the patient anatomy in the lower region. The 2D data and 3D model, along with tracking information, are projected to the surgeon's {5} retina from the transparent near-eye-display Headset, allowing the surgeon to look at the patient and the navigation data at the same time. ## Indications for Use: ExcelsiusXR™, when used in conjunction with ExcelsiusHub™ and/or ExcelsiusGPS®, is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. This system is indicated for the placement of spinal and orthopedic (Sacroiliac, Ulna, and Tibia) bone screws, and interbody fusion devices. The ExcelsiusXR™ Headset displays 2D stereotactic images and 3D virtual anatomy images, and displays the virtual instrument location in relation to the virtual anatomy to assist in percutaneous visualization and trajectory planning. This headset should not be relied upon solely for absolute positional information and should always be used in conjunction with the primary stereotactic display. ## Technological Characteristics: ExcelsiusXR™ has similar technological characteristics to the predicate devices including function, workflow, user interface, software features, and design. ExcelsiusXR™ is comparable to the predicates in terms of intended use, fundamental scientific technology, technological characteristics and principle of operation. Comparison of Principles of Operation and Technological Characteristics | Device | Subject ExcelsiusXR™ | Predicate Augmedics xvision (K220905) | Predicates ExcelsiusHub (K211616) & ExcelsiusGPS (K171651, K191100) | | --- | --- | --- | --- | | Intended Use Environment | Operating Room | Operating room | Operating room | | Principle of Operation | • Intraoperative/ Preoperative images • Patient registration • Surgical planning (optional) • Real-time tracking of navigated instruments • Manual, free hand or robotic guidance of instruments • Hand tracking | • Patient Preparation • System Set-Up • Interoperative Scan • Scan Import • Patient Registration • Navigation | • Intraoperative/ Preoperative images • Patient registration • Surgical planning (optional) • Real-time tracking of navigated instruments • Manual, free hand or robotic guidance of instruments | | Integrated Planning Software | Excelsius Planning and Navigation Application software | XVS Software | Excelsius Planning and Navigation Application software | | Save/load Planning | Yes | No | Yes | | Merge Images Functionality | Yes | No | Yes | K241525- page 2 of 4 {6} | Device | Subject ExcelsiusXR™ | Predicate Augmedics xvision (K220905) | Predicates ExcelsiusHub (K211616) & ExcelsiusGPS (K171651, K191100) | | --- | --- | --- | --- | | Image-guided | Yes | Yes | Yes | | Patient Registration Method | Intra-Op CT: Registration Fixture Pre-Op CT: Fluoroscopic to Pre-Op CT Merge Fluoroscopy: Registration Fixture Excelsius3D™ CT Excelsius3D™ fluoroscopy | Intra-Op CT: Registration Fixture | Intra-Op CT: Registration Fixture Pre-Op CT: Fluoroscopic to Pre-Op CT Merge Fluoroscopy: Registration Fixture Excelsius3D™ CT Excelsius3D™ fluoroscopy | | Accuracy Verification on Anatomical Landmarks | Yes | N/A | Yes | | Real Time Display of the Instrument Position | Yes | Yes | Yes | | Patient Fixation | Reference is fixed to patient's bony structure such as a long bone, iliac crest, spinous process, vertebra, etc. for tracking system. | Reference is fixed to patient's spinous process for tracking system. | Reference is fixed to patient's bony structure such as a long bone, iliac crest, spinous process, vertebra, etc. for tracking system. | ## Performance Testing: Verification and validation testing were conducted on ExcelsiusXR™ to confirm that the device meets performance requirements under the indications for use and to ensure safety and efficacy of the system: - Non-clinical system, software, and instrument verification and validation - demonstrated compliance with user needs and corresponding design inputs - Surgical simulations conducted on phantom models - Qualitative and quantitative validation to confirm intended use and accuracy - Optical bench testing to evaluate the image quality characteristics of the head mounted display - Electrical Safety and Electromagnetic Compatibility - Testing was performed to assure compliance with recognized safety standards for electrical safety and electromagnetic compatibility - IEC 60601-1:2020 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2020 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - IEC 60601-1-6:2020 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability K241525- page 3 of 4 {7} - IEC 62304:2015 Medical device software - Software lifecycle processes - IEC 62366-1:2020 Medical devices - Part 1: Application of usability engineering to medical devices ## Software Verification and Validation Testing: Software verification and validation testing was performed in accordance with the FDA Guidance for the Content of Premarket Submissions for Device Software Functions, June 14, 2023. The software for this device was determined to require "Enhanced Documentation". ## Basis of Substantial Equivalence: ExcelsiusXR™ has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices. K241525- page 4 of 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...