MEDTRONIC PS MEDICAL INNERVISION VENTRICULAR CATHETER, 15CM MODEL 99102
K983331 · Medtronic PS Medical · HCA · Dec 16, 1998 · Neurology
Device Facts
Record ID
K983331
Device Name
MEDTRONIC PS MEDICAL INNERVISION VENTRICULAR CATHETER, 15CM MODEL 99102
Applicant
Medtronic PS Medical
Product Code
HCA · Neurology
Decision Date
Dec 16, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innervision Ventricular Catheter with BioGlide is designed to be used with the Neuropen endoscope or other compatible Medtronic PS Medical endoscope when the physician desires direct vision from the tip of a ventricular catheter during its placement. Direct vision facilitates placement of the catheter tip at a specific intraventricular location. The catheter is designed to be used as the proximal component of a CSF-Flow Control Shunt for use in shunting cerebrospinal fluid from the ventricles of the brain into the right atrium of the heart or to the peritoneal cavity.
Device Story
Translucent silicone elastomer ventricular catheter with barium-impregnated stripe and BioGlide surface modification; features slit tip. Designed for use with Neuropen or compatible endoscopes to provide direct visualization during placement within brain ventricles. Functions as proximal component of CSF-Flow Control Shunt system. Used by neurosurgeons in clinical settings to facilitate precise catheter tip positioning. Direct visualization aids in navigating intraventricular anatomy, potentially improving placement accuracy and reducing complications associated with blind catheter insertion. Device serves as conduit for CSF drainage to peritoneal cavity or right atrium.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and material characteristics.
Technological Characteristics
Translucent silicone elastomer; barium-impregnated stripe; BioGlide surface modification; slit tip design. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients requiring cerebrospinal fluid (CSF) shunting from the brain ventricles to the right atrium or peritoneal cavity, where direct visualization of the catheter tip placement is desired by the physician using a compatible endoscope.
Regulatory Classification
Identification
A ventricular catheter is a device used to gain access to the cavities of the brain for injection of material into, or removal of material from, the brain.
Medtronic PS Medical Ventricular Catheter with BioGlide (K951258)
Related Devices
K981046 — MEDTRONIC PS MEDICAL EXTERNAL DRAINAGE AND MONITORING (EDM) VENTRICULAR CATHETER, 20CM AND 35 CM MODEL # 96115 AND 96118 · Medtronic PS Medical · Jun 16, 1998
K110751 — CODMAN BACTISEAL ENDOSCOPIC VENTICULAR CATHETER · Codman & Shurtleff, Inc. · May 9, 2011
K200950 — Ventricular Catheter · Aesculap, Inc. · May 8, 2020
K063836 — RIVULET VENTRICULAR AND SNAP SHUNT AND CONVERTIBLE VENTRICULAR CATHETERS, MODELS 41701 AND 41704-41715 · Medtronic Neurosurgery · Aug 16, 2007
K073139 — SMALL LUMEN PERITONEAL CATHETER · Medtronic Neurosurgery · Jan 30, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
DEC I 6 998
983331
## Safety and Effectiveness Summary X.
This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| Establishment Registration Number: | 2021898 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Address of Manufacturer: | Medtronic PS Medical Corporation<br>125 Cremona Drive<br>Goleta CA, 93117<br>(805) 968-1546 ext. 1770<br>Fax: (805) 968-5038 |
| Contact Person: | Jeffrey Henderson: |
| Date: | August 17, 1998 |
| Trade or Proprietary Name: | Medtronic PS Medical Innervision Ventricular<br>Catheter w/BioGlide |
| Common usual or Classification Name: | Ventricular Catheter (882.4100) |
Medtronic PS Medical Innervision Ventricular Predicate Device Identification: Catheter submitted under the Premarket Notification for the Catheter Placement Kit (K940096), Medtronic PS Medical Ventricular Catheter with BioGlide submitted under the Premarket Notification for the BioGlide Shunt Kit (K951258)
The Medtronic PS Medical Innervision Description: Ventricular Catheter with BioGlide is manufactured of translucent silicone elastomer w/barium impregnated stripe. The catheter incorporates a BioGlide surface modification on the interior and exterior of the catheter. The catheter incorporates a slit tip.
Intended Use: The Innervision Ventricular Catheter with BioGlide is designed to be used with the Neuropen endoscope or other compatible Medtronic PS Medical endoscope when the physician desires direct vision from the tip of a ventricular catheter during its placement. Direct vision facilitates placement of the catheter tip at a specific intraventricular location. The catheter is designed to be used as the proximal component of a CSF-Flow Control Shunt for use in shunting cerebrospinal fluid from the ventricles of the brain into the right atrium of the heart or to the peritoneal cavity.
Intended Use predicate device: The Innervision Ventricular Catheter is designed to be used with the Innervision endoscope when the physician desires direct vision from the tip of a ventricular catheter during its placement. Direct vision facilitates placement of the catheter tip at a specific intraventricular location. The catheter is designed to be used as the proximal component of a CSF-Flow Control Shunt for use in shunting cerebrospinal fluid from the ventricles of the right atrium of the heart or to the peritoneal cavity.
{1}------------------------------------------------
Medtronic PS Medical submits that the Technological comparison: materials of fabrication, intended uses, performance characteristics and design specifications of the Innervision Ventricular Catheter with BioGlide are substantially equivalent to those of the predicate devices. Based upon the summary above, Medtronic PS Medical determines substantial equivalence, safety, and efficacy of the Innervision Ventricular Catheter with BioGlide based upon the predicate and currently marketed devices.
1. 1. 1. 1. 1.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 16 1000
Mr. Jeffrey Henderson Vice-President. Quality Medtronic PS Medical 125 Cremona Drive Goleta, California 93117-5500
Re: K983331
Trade Name: Medtronic PS Medical Innervision Ventricular Catheter With Bioglide Regulatory Class: II Product Code: HCA Dated: September 18, 1998 Received: September 22, 1998
Dear Mr. Henderson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Jeffrey Henderson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Device Name: Innervision Ventricular Catheter with BioGlide 510(k) Number (if known): 《♀♀3333
Indications for Use:
"The Innervision Ventricular Catheter with BioGlide is designed to be used with the Neuropen endoscope or other compatible Medtronic PS Medical endoscope when the physician desires direct vision from the tip of a ventricular catheter during its placement. Direct vision facilitates placement of the catheter tip at a specific intraventricular location. The catheter is designed to be used as the proximal component of a CSF-Flow Control Shunt for use in shunting cerebrospinal fluid from the ventricles of the brain into the right atrium of the heart or to the peritoneal cavity."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Over the Counter Use: ` or Prescription Use: (Per 21 CFR 801.109)
ōōely
(Division Sign-Off) Division of General Restorative Devices 298333 510(k) Number
(optional format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.