SYNTHES AIR PEN DRIVE (APD) SYSTEM

K093361 · Synthes (Usa) · HBB · Apr 14, 2010 · Neurology

Device Facts

Record IDK093361
Device NameSYNTHES AIR PEN DRIVE (APD) SYSTEM
ApplicantSynthes (Usa)
Product CodeHBB · Neurology
Decision DateApr 14, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4370
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Synthes Air Pen Drive (APD) is indicated for screw insertion, pin and wire placement, cutting of bone and metal, drilling, reaming, decorticating, shaping and smoothing of bones and teeth in a wide variety of surgical procedures, including general orthopedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngological, reconstructive and spine surgery.

Device Story

Synthes Air Pen Drive (APD) System is a pneumatic-powered surgical drill system. Components include foot pedal, drive unit, hand switch, and irrigation control unit. Drive unit is pen-shaped, connecting to foot pedal via sterilizable air hose. System accepts various quick-connect attachments and cutting tools (drill bits, burrs, saw blades). Rotation speed controlled via foot pedal or hand switch (0-60,000 rpm at 8 bar; 0-80,000 rpm at 12 bar). Used in OR settings by surgeons for bone/metal cutting, drilling, and shaping. Output provides mechanical power for surgical tools, facilitating bone/tissue modification. Benefits include precise surgical site preparation and hardware placement.

Clinical Evidence

Bench testing only. Testing included duty cycle, wear and leak testing of couplings, valves, hoses, and seals, endurance testing of switches and valve control, noise testing, and temperature testing. Engineering performance evaluation compared output torque and power to the predicate device.

Technological Characteristics

Pneumatic-powered surgical drill system. Pen-shaped drive unit with quick-connect coupling for attachments. Operates at 0-60,000 rpm (8 bar) to 80,000 rpm (12 bar). Components include foot pedal, hand switch, and irrigation control unit. Sterilizable air hose connection. No electronic software or algorithm components.

Indications for Use

Indicated for patients requiring bone or tooth surgical procedures including screw insertion, pin/wire placement, cutting, drilling, reaming, decorticating, shaping, and smoothing. Applicable across general orthopedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngological, reconstructive, and spine surgery.

Regulatory Classification

Identification

A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ PAGE 1 of ① K 893361 Image /page/0/Picture/8 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. There is a horizontal line underneath the word "SYNTHES". APR 1 4 2010 ## 3.0 510(k) Summary Page _ of __ of ___________________________________________________________________________________________________________________________________________________________________ 1 | Sponsor:<br>Contact:<br>Date of Preparation: | Synthes (USA)<br>1301 Goshen Parkway<br>West Chester, PA 19380 Phone: (610) 719-6538<br>Jill R. Yelton, Regulatory Compliance Manager<br>October 26, 2009 | |----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Synthes Air Pen Drive (APD) System | | Classification: | 21 CFR 872.4120: Drill, bone, powered<br>21 CFR 874.4250: Drill, surgical, ENT (Electric or Pneumatic<br>21 CFR 882.4310: Powered simple cranial drills, burrs, trephines, and<br>accessories<br>21 CFR 882.4370: Pneumatic cranial drill motor | | Predicate Device: | Synthes Electric Pen Drive (EPD) System (K043310)<br>Midas Rex Legend System (K020069) | | Device Description: | The Synthes Air Pen Drive (APD) System is a pneumatic powered<br>system that consists of a Foot Pedal, Drive Unit, Hand Switch and<br>Irrigation Control Unit that may be used with various commercially<br>available attachments and cutting tools. The Drive Unit is pen-shaped<br>and is connected to the Foot Pedal via a sterilizeable air hose. The<br>rotation speed of the Drive Unit may be controlled from 0 to 60,000<br>rpm at 8 bar and from 0 to 80,000 at 12 bar via the Foot Pedal or the<br>removable Hand Switch. Multiple attachments are available that have<br>a quick-connect coupling to attach to the Drive Unit. The attachments<br>accept various cutting tools including drill bits, burrs and saw blades. | | Intended Use: | The Synthes Air Pen Drive (APD) is indicated for screw insertion, pin<br>and wire placement, cutting of bone and metal, drilling, reaming,<br>decorticating, shaping and smoothing of bones and teeth in a wide<br>variety of surgical procedures, including general orthopedic trauma,<br>foot, hand, maxillofacial, neurosurgical, oral, otolaryngological,<br>reconstructive and spine surgery. | | Substantial<br>Equivalence: | Information presented supports substantial equivalence of the<br>proposed device to the predicate device. The proposed device has the<br>same indications for use, incorporates the same fundamental product<br>technology and is composed of the same materials. Non-clinical<br>testing was performed in accordance with current standards for<br>functionality and reliability of the proposed device. Testing includes;<br>duty cycle testing, wear and leak testing of couplings, valves hoses<br>and seals, endurance testing of switches and valve control, noise<br>testing and temperature testing. Additionally, engineering<br>performance evaluation and analysis was performed comparing output<br>torque and power of the proposed device to the predicate device. | . . {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird in flight, with three wing-like shapes suggesting movement. The bird is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged around the circumference of the circle, with the bird facing towards the text. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 APR 1 4 2010 Synthes (USA) % Ms. Jill R. Yelton Regulatory Compliance Manager 1301 Goshen Parkway West Chester, Pennsylvania 19380 Re: K093361 Trade/Device Name: Synthes Air Pen Drive (APD) System Regulation Number: 21 CFR 872.4120 Regulation Name: Regulatory Class: Class II Product Code: HBB Dated: April 07, 2010 Received: April 08, 2010 Dear Ms. Yelton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {2}------------------------------------------------ Page 2 - Ms. Jill R. Yelton CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a design inside. A registered trademark symbol is located to the right of the word. A horizontal line is located underneath the word. Page 1 of 1 **2.0 Indications for Use** | 510(k) Number (if known): | k093361 | |---------------------------|------------------------------------| | Device Name: | Synthes Air Pen Drive (APD) System | | Indications for Use: | | The Synthes Air Pen Drive (APD) is indicated for screw insertion, pin and wire placement, cutting of bone and metal, drilling, reaming, decorticating, shaping and smoothing of bones and teeth in a wide variety of surgical procedures, including general orthopedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngological, reconstructive and spine surgery X Prescription Use (Per 21 CFR 801.109) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Nui R.P. Oglumfsmkn (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K093361 000004
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...