SOLOPASS 2.0 System

K251317 · Intravent Medical Partners · HAW · Sep 5, 2025 · Neurology

Device Facts

Record IDK251317
Device NameSOLOPASS 2.0 System
ApplicantIntravent Medical Partners
Product CodeHAW · Neurology
Decision DateSep 5, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The SOLOPASS 2.0 System is a tool that obtains ultrasound images and positional data to provide intra-procedural, image guided localization and navigation, to aid in the frontal placement of an intra-ventricular catheter.

Device Story

Neuronavigational system for intra-ventricular catheter (EVD) placement. Inputs: 2D ultrasound images from cranial burr-hole probe; positional data from skull-mounted Patient Interface Device (PID) encoders. Operation: System integrates imaging and positional data to construct 3D anatomical model without preoperative imaging; user selects catheter trajectory; PID locks to guide catheter. Used in OR or patient bedside by surgeons/qualified HCPs. Output: 3D visualization on workstation monitor; guidance for catheter insertion. Benefits: Improved targeting accuracy (+/-2mm) and imaging accuracy (+/-1.5mm) compared to predicate; facilitates precise catheter placement.

Clinical Evidence

Bench testing only. No clinical data provided. Performance validated via ISO 10993 biocompatibility testing, IEC 60601 electrical/mechanical safety, and verification of imaging/targeting accuracy. Design validation study performed under simulated use conditions confirmed user needs were met.

Technological Characteristics

System includes skull-mounted PID (gamma-sterilized, single-use), reusable burr-hole phased array ultrasound probe (VHP-sterilized), and workstation. Materials conform to ISO 10993-1. Connectivity: Standalone workstation. Imaging: B-mode, 5 MHz. Targeting accuracy: +/-2mm. Imaging accuracy: +/-1.5mm. Software development conforms to IEC 62304. Mechanical testing per ASTM F543.

Indications for Use

Indicated for patients requiring frontal placement of an intra-ventricular catheter, utilizing intra-procedural image-guided localization and navigation.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION September 5, 2025 inTRAvent Medical Partners % Connie Qiu Senior Medical Technology Regulatory Consultant ProPharma MedTech 107 West Hargett St. Raleigh, North Carolina 27601 Re: K251317 Trade/Device Name: SOLOPASS 2.0 System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW, IYN Dated: April 29, 2025 Received: August 11, 2025 Dear Connie Qiu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251317 - Connie Qiu Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251317 - Connie Qiu Page 3 Sincerely, Yen-chih Lin -S Digitally signed by Yen-chih Lin -S Date: 2025.09.05 14:29:54 -04'00' For Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K251317 Device Name SOLOPASS® 2.0 System Indications for Use (Describe) The SOLOPASS 2.0 System is a tool that obtains ultrasound images and positional data to provide intra-procedural, image guided localization and navigation, to aid in the frontal placement of an intra-ventricular catheter. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K251317, Page 1 of 6 510(k) Summary - K251317 Sponsor: inTRAvent Medical Partners, LP 1730 Walton Road, Suite 110 Blue Bell, PA 19422 Contact: Connie Qiu ProPharma MedTech 107 West Hargett St. Raleigh, NC 27601 Date Prepared: August 27, 2025 Trade Name: SOLOPASS® 2.0 System Common Name: Neurological stereotaxic instrument Classification: II Product Code: (Primary) HAW, 21 CFR 882.4560, Neurological Stereotaxic Instrument IYN, 21 CFR 892.1550, System, Imaging, Pulsed Doppler, Ultrasonic Predicate Device: SOLOPASS® System (K203251) Description of Device: The SOLOPASS® 2.0 System is the next generation product of the predicate device, SOLOPASS® System. The subject device maintains the same indications for use and operating principles as the predicate device. The SOLOPASS® 2.0 System is a neuronavigational system that collects intraoperative ultrasound imaging referenced to a skull mounted fixation device, allowing the user to plan the desired placement for external ventricular drain (EVD). The system utilizes two-dimensional imaging data with simultaneously captured location data to build a three-dimensional model of the anatomy without the use of preoperative imaging. Once the user has chosen a catheter placement location, the fixation device is locked in place to guide a catheter towards the intended anatomic location. The SOLOPASS® 2.0 System consists of three main sub-systems: {5} K251317, Page 2 of 6 1. The Patient Interface Device (PID): A skull-mounted fixation device that translates mechanical motion into digital position and secures the Ultrasound Probe and Catheter Guide. 2. The Ultrasound Probe "The Probe": A custom cranial "burr-hole" style probe used to collect intraoperative ultrasound image data from the patient. 3. The Workstation: A custom, portable unit that includes a dedicated operating system, imaging software application, and 27" monitor for displaying the User Interface. The Workstation is the primary interface of the other subsystems and is controlled by the included foot pedal. ## Indications for Use: The SOLOPASS® 2.0 System is a tool that obtains ultrasound images and positional data to provide intra-procedural, image guided localization and navigation, to aid in the frontal placement of an intraventricular catheter. ## Comparison to Predicate Device: Comparison of intended use and technological characteristics between the SOLOPASS® 2.0 System to the predicate device, SOLOPASS® System K203251 is presented below. Table 1 Comparison of Subject and Predicate Devices | | Subject Device SOLOPASS® 2.0 System | Predicate Device SOLOPASS® System | Substantial Equivalence Comparison | | --- | --- | --- | --- | | Product Codes | HAW, IYN | HAW, IYN | Same | | Indications for Use | The SOLOPASS® 2.0 system is a tool that obtains ultrasound images and positional data to provide intra-procedural, image guided localization and navigation, to aid in the frontal placement of an intraventricular catheter. | The SOLOPASS® system is a tool that obtains ultrasound images and positional data to provide intra-procedural, image guided localization and navigation, to aid in the frontal placement of an intraventricular catheter. | Same | | Planned Use Environment | Operating Room (OR), Patient Bedside outside of OR | Operating Room (OR), Patient Bedside outside of OR | Same | | Planned User | Surgeon or qualified HCP | Surgeon or qualified HCP | Same | | Anatomic region | Cranial | Cranial | Same | {6} K251317, Page 3 of 6 | Use For Neurosurgical Catheter/Instrument Placement | Yes | Yes | Same | | --- | --- | --- | --- | | Main System Components | Ultrasound Imaging System; Skull mounted mechanical module; Software Module for trajectory planning; Workstation/ Display; Cart. | Ultrasound Imaging System; Skull mounted mechanical module; Software Module for trajectory planning; Workstation/ Display; Cart. | Same | | Image Guidance | Ultrasound Intra-op 2D imaging data with simultaneously captured location data is used to build 3D model of anatomy | Ultrasound Intra-op 2D imaging data with simultaneously captured location data is used to build 3D model of anatomy | Same | | Trajectory Guide Function | Yes | Yes | Same | | Patient Fixation | Skull-mounted frame (Patient Interface Device, PID) serves as reference for instruments and catheters | Skull-mounted frame (Patient Interface Device, PID) serves as reference for instruments and catheters | Same | | Manually Operated | Yes | Yes | Same | | Instrument/Catheter Compatibility | Catheters with 3.4mm or 2.8mm outer diameter | Catheters with 3.4mm or 2.8mm outer diameter | Same | | Localization Method | Encoders on Patient Interface Device to track motion of instruments/catheters | Encoders on Patient Interface Device to track motion of instruments/catheters | Same | | Transducer Type | Phased Array Probe | Phased Array Probe | Same | | Transducer Frequency | 5 Mhz 1 Transducer | 5 Mhz 1 Transducer | Same | | Transducer Style | “Burr-Hole” Style (Craniotomy) | “Burr-Hole” Style (Craniotomy) | Same | | Acoustic Output Display & FDA Limits | Track 3 | Track 3 | Same | | Imaging Mode | B Mode | B Mode | Same | {7} K251317, Page 4 of 6 | General Safety and Effectiveness Information | 1. Total Image Depth 0-9 cm 2. Optimal Image Range 2.5 – 8 cm | 1. Total Image Depth 0-10 cm 2. Optimal Image Range 2.5 – 8 cm | Similar. Total image depth was reduced by 1cm to enlarge useful area shown on screen. | | --- | --- | --- | --- | | Accuracy | Targeting Accuracy: +/-2mm Imaging Accuracy: +/-1.5mm | Targeting Accuracy: +/-3mm Imaging Accuracy: +/-2mm | Targeting Accuracy improved from +/-3mm to +/-2mm. Imaging Accuracy improved from +/-2mm to +/-1.5mm | | User Interfaces | Graphical Touch Screen, Foot Switch | Graphical Touch Screen, Foot Switch | Same | | Electrical Safety | Conformity to IEC 60601-1 IEC 60601-2 | Conformity to IEC 60601-1 IEC 60601-2 | Same | | Biocompatibility Patient Contacting Components | Conformity To ISO 10993-1 Limited Contact (<24 Hours) • Patient Interface Device • Ultrasonic Probe Used With Sheath And Acoustic Gel | Conformity To ISO 10993-1 Limited Contact (<24 Hours) • Patient Interface Device • Ultrasonic Probe Used With Sheath And Acoustic Gel | Same | | Sterilization | PID: Sterile, single-use, Gamma Ultrasound Probe: Reusable, sterilized by vapor hydrogen peroxide (VHP) by the end user | PID: Sterile, single-use, Gamma Ultrasound Probe: Reusable, sterilized by vapor hydrogen peroxide (VHP) by the end user | Same | ## Nonclinical Testing Summary: The following performance data are provided in support of the substantial equivalence determination between the proposed device, SOLOPASS® 2.0 System to the first-generation predicate device, SOLOPASS® System K203251. Table 2 Summary of Non-Clinical Performance Data | TEST | TITLE/TEST METHOD SUMMARY | RESULTS | | --- | --- | --- | | Biocompatibility | | | | ISO 10993-5 | Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity | Non-cytotoxic | {8} K251317, Page 5 of 6 | ISO 10993-10 | Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization | Non-sensitizing Non-irritating | | --- | --- | --- | | ISO 10993-11 | Biological evaluation of medical devices — Part 11: Tests for systemic toxicity | Non-pyrogenic Negative for acute systemic toxicity | | ANSI/AAMI ST72, USP <85>, USP <161> | Bacterial endotoxins test | Pass, all samples demonstrated less than 2.15 Eu/device required for devices with cerebrospinal fluid contact | | **Thermal, Electrical, Mechanical Safety** | | | | IEC 60601-1/ ANSI AAM ES 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Pass | | IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral | Pass | | IEC 60601-2-37 | Particular Requirements for the safety of ultrasonic medical diagnostic and monitoring equipment. | Pass | | AIM 7351731 | Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers | Pass | | IEC/EN 60529 | Degrees of protection provided by enclosures (PIX2, IPX7) | Pass | | **Cleaning, Disinfection, Sterilization** | | | | Ultrasound Probe Cleaning Validation | Validation of cleaning and disinfection method for reusable ultrasound probes. | Pass | | Ultrasound Probe VHP Sterilization Validation | Validation of VHP sterilization method for reusable ultrasound probes. | Pass, SAL 10^{-6} | | AAMI/ANSI/ISO 11137-1, 11137-2 | Validation of gamma sterilization method for single-use PID. | Pass, SAL 10^{-6} | | Ship and Shelf Life Functional Test | Verify functional performance of device components following applicable testing per ISTA 3A, ISO 11607-1, and shelf-life aging. | Pass PID Shelf life: 12 months | | **Verification Bench Testing** | | | | 2D Imaging Qualification | Verification of ultrasound requirements including imaging depth, image accuracy, active element check and other specifications. | Pass | | System Targeting Accuracy | Measure targeting accuracy. Acceptance criteria defined based on predicate device. | +/- 2.0mm target at 6cm | | System Imaging Accuracy | Measure imaging accuracy. | +/- 1.5mm target at 6cm | {9} K251317, Page 6 of 6 | Cranial Mounting Mechanical Testing | Verify performance of SOLOPASS® anchor assembly. Test methods based on ASTM F543. | Met acceptance criteria for: mean pullout strength of anchor. | | --- | --- | --- | | Hardware Verification | Verify performance of system electrical design requirements in addition to electrical safety and EMC. | Pass | | Software Verification and Validation | | | | Software Verification and Validation | Demonstrate that all software requirements were appropriately implemented in the software. Software development process demonstrates conformity to IEC 62304. | Pass | | Design Validation | | | | Design validation study | Validation study in simulated use conditions to demonstrate that SOLOPASS® 2.0 System final design met user needs. | Pass, user needs were successfully validated | ## Conclusions: In summary, the SOLOPASS® 2.0 System and predicate device, SOLOPASS® System (K203251), are substantially equivalent with respect to intended use and performance testing. Non-clinical testing results support that the subject and predicate devices are substantially equivalent in function for use as the predicate. The differences between the two devices do not raise new questions of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...