Geister K-Rex rongeurs

K242759 · Geister Medizin Technik GmbH · HAE · Jun 6, 2025 · Neurology

Device Facts

Record IDK242759
Device NameGeister K-Rex rongeurs
ApplicantGeister Medizin Technik GmbH
Product CodeHAE · Neurology
Decision DateJun 6, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4840
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GEISTER® K-Rex™ rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.

Device Story

Geister K-Rex rongeurs are reusable, manually operated surgical instruments designed for cutting or biting bone during skull or spinal procedures. The device consists of a stainless steel body with a TiAlN coating, featuring a detachable design with an integrated ejector mechanism. It is provided non-sterile and requires sterilization by the user. The instrument is operated by surgeons in an operating room environment. The device functions through manual mechanical force applied by the clinician to the handles, which actuates the jaw to cut bone. The output is the physical removal of bone tissue, which facilitates surgical access or decompression. The device benefits patients by providing a reliable, biocompatible tool for precise bone resection during neurosurgical or spinal interventions.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), reprocessing validation (ANSI/AAMI ST98:2022, ISO 17665, ISO 17664), and comparative cutting performance testing (10,000 cycles) against the predicate device.

Technological Characteristics

Manual surgical rongeur; stainless steel (ISO 7153-1, ASTM F899); TiAlN surface coating; detachable design with ejector; bite sizes 1–6 mm; shaft lengths 180–280 mm; jaw opening 12 mm; jaw angles 40° and 90° up; non-sterile; reusable; steam sterilization.

Indications for Use

Indicated for cutting or biting bone during surgery involving the skull or spinal column.

Regulatory Classification

Identification

A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 6, 2025 Geister Medizin Technik GmbH Florian Herb QM/RA Manager Foehrenstrasse 2 Tuttlingen, Baden 78532 Germany Re: K242759 Trade/Device Name: Geister K-Rex rongeurs Regulation Number: 21 CFR 882.4840 Regulation Name: Manual rongeur Regulatory Class: Class II Product Code: HAE Dated: May 6, 2025 Received: May 6, 2025 Dear Florian Herb: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K242759 - Florian Herb Page 2 Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K242759 - Florian Herb Page 3 Sincerely, Adam D. Pierce -S Digitally signed by Adam D. Pierce Date: 2025.06.06 14:56:19 -04'00' Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K242759 Device Name Geister® K-Rex™ rongeurs Indications for Use (Describe) The GEISTER® K-Rex™ rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} Geister Medizintechnik GmbH Rongeur 510(k) Premarket Notification K242759 K242759 510(k) Summary DATE OF APPLICATION: June 5, 2025 APPLICANT: Geister Medizintechnik GmbH Föhrenstrasse 2 78532 Tuttlingen Germany Tel.: +49 7461 96624 0 Fax: +49 7461 96624 22 E-Mail: info@geister.com www.geister.com CONTACT PERSON: Florian Herb Quality Manager Tel.: +49 7461 96624102 E-Mail: Florian.Herb@geister.com Page 1 of 4 {5} Geister Medizintechnik GmbH Rongeur 510(k) Premarket Notification K242759 # 1. Device Name Trade Names: Geister® K-Rex™ rongeurs Regulation description: Manual rongeur # 2. Classification Product Code / Subsequent Code ## 2.1. Rongeur | Device | Review Panel | Product Code | Device Class | Regulation Number | | --- | --- | --- | --- | --- | | Rongeur, Manual | Neurology | HAE | 2 | 882.4840 | # 3. Predicate Device Geister Rongeur are equivalent to the following predicate devices: | Geister Product | Predicate Device | 510(k) Number | 510(k) Holder | | --- | --- | --- | --- | | Geister® K-Rex™ rongeurs (new device) | CERAMO® KERRISION (primary predicate) | K153243 | Fehling instruments | | | SQ.line KERRISON | K223596 | Aesculap Inc. | # 4. Description of the Device Geister Geister® K-Rex™ rongeurs are reusable stainless steel instruments that are coated with TiAlN that are sterilizable and packaged non-sterile. ## 4.1. Device specifications | Bite Size: | 1 – 6 mm | | --- | --- | | Length: | 180 – 280 mm | | Jaw opening | 12 mm | | cutting angles | 40° and 90° up | | Footplates | regular footplates and thin footplates | | Materials | The products are made of stainless steel according to ISO 7153-1 and ASTM F899 | | Coatings | TiAlN | | Further characteristics | Detachable, with Ejector | # 5. Indications for Use The GEISTER® K-Rex™ rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Page 2 of 4 {6} Geister Medizintechnik GmbH Rongeur 510(k) Premarket Notification K242759 6. Substantial Equivalence Discussion | # | NEW DEVICE | PRIMARY PREDICATE DEVICE | PREDICATE DEVICE | | --- | --- | --- | --- | | 510(k)Submitter/Holder | Geister Medizintechnik GmbH | Fehling instruments | Aesculap, Inc. | | Trade Name | K-Rex™ Rongeur | CERAMO® KERRISION | SQ.line KERRISONS | | Illustration | | | | | Device | Rongeur, Manual | Rongeur, Manual | Rongeur, Manual | | Regulation Number | 882.4840 | 882.4840 | 882.4840 | | Device Class | II | II | II | | Classification Product Code | HAE | HAE | HAE | | 510(k) Number | new | K153243 | K223596 | | Indications for use | The GEISTER® K-rex™ rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. | Fehling rongeurs (bone punches) are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. | The SQ.line KERRISONS (bone punches) are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. | | Sterility | Non-sterile condition | Non-sterile condition | Non-sterile condition | | Prescription Use | Yes | Yes | Yes | | Material | Stainless steel 420, 303, and 301 (ASTM F-899); Surface coatings: black coating (TiAlN) | 420 and 304 Stainless Steels (ASTM F-899); Surface coatings: CERAMO® (TiAlN); | Stainless Steels (ASTM F-899); Surface coatings: DLC black coating | | Shaft lengths | 180 – 280 mm | 110 – 400 mm | 180 – 280 mm | | Jaw sizes | 1 – 6mm | 0.8-8 mm | 1 – 6mm | | Jaw angles | 40° and 90° up cutting angles | Straight and 40 – 90° (up & down) | 90 – 130° (up & down) | | Jaw opening | 12 mm | 9 – 19 mm | 10 – 15 mm | | Ejector | Yes | Yes | Yes | | detachable | Yes | Yes | Yes | | Profile footplates | Regular and thin | Regular and thin | Regular and thin | Geister K-Rex™ Rongeurs have the same intended use, similar performance characteristics, are manufactured from similar materials and are similar in design to the predicate devices. Page 3 of 4 {7} Geister Medizintechnik GmbH Rongeur 510(k) Premarket Notification K242759 ## 7. Non-Clinical Performance Data Biocompatibility: Biocompatibility end points were evaluated in accordance with ISO 10993. Biocompatibility testing for cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, interactions with blood, hemolysis and material mediated pyrogenicity were performed. Test results indicate that the GEISTER® K-rex™ rongeurs are biocompatible. Reprocessing, Cleaning and Sterilization: Reprocessing, Cleaning and Sterilization validation were performed according to ANSI/AAMI ST98:2022, ISO 17665 and ISO 17664. Cutting performance test: A cutting performance test was performed on the GEISTER® K-rex™ rongeur and the predicate device. A cutting test of 10,000 cycles was performed. Compared to the predicate device, the GEISTER® K-rex™ rongeur demonstrates a similar cutting performance. ## 8. Clinical Performance Data There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. ## 9. Substantial Equivalence Summary / Conclusion By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device. It has been shown in these 510(k) submissions that the difference between the Geister K-Rex™ Rongeurs and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the Geister K-Rex™ Rongeurs are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use. Geister K-Rex™ Rongeurs are determined to be substantially equivalent to the referenced predicate devices. Page 4 of 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...