iovera System

K220656 · Pacira Biosciences, Inc. · GXH · May 20, 2022 · Neurology

Device Facts

Record IDK220656
Device Nameiovera System
ApplicantPacira Biosciences, Inc.
Product CodeGXH · Neurology
Decision DateMay 20, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The iovera System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera System is not indicated for treatment of central nervous system tissue. When stimulation compatible components are used , the iovera System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator.

Device Story

Portable cryogenic surgical device; uses nitrous oxide (N2O) to cool Smart Tip via Joule-Thomson effect/latent heat of vaporization; creates ice ball to destroy tissue or block peripheral nerve pain; used in clinical settings by healthcare providers; battery-powered handpiece with LCD feedback; connects to single-patient-use stainless steel Smart Tips; optional nerve stimulation support via 3rd party stimulator; provides targeted cryoneurolysis; benefits include non-systemic pain relief for knee osteoarthritis.

Clinical Evidence

Bench testing only. Verification and validation testing included functional performance, user interface (IEC 62366-1), software (IEC 62304), electrical/EMC safety (IEC 60601-1/1-2), mechanical/thermal safety, sterility (ISO 11135), packaging (ASTM D4332), and biocompatibility (ISO 10993-1). All tests met predetermined acceptance criteria.

Technological Characteristics

Cryogenic surgical device; nitrous oxide (N2O) propellant; stainless steel Smart Tips; Joule-Thomson cooling principle; battery-powered handpiece (3.6V Li-ion); LCD user interface; sterilization via ISO 11135; biocompatible materials per ISO 10993-1; electrical safety per IEC 60601-1.

Indications for Use

Indicated for patients requiring tissue destruction during surgery, peripheral nerve lesioning for pain blocking, and relief of pain/symptoms associated with knee osteoarthritis for up to 90 days. Not indicated for central nervous system tissue.

Regulatory Classification

Identification

A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 20, 2022 Pacira Biosciences, Inc. Danny Rivera Sr. Director, Quality and Regulatory Affairs 10450 Science Center Drive San Diego, California 92121 Re: K220656 Trade/Device Name: iovera System Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH, ETN Dated: April 20, 2022 Received: April 21, 2022 ### Dear Danny Rivera: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220656 Device Name iovera® System #### Indications for Use (Describe) The iovera System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera" System is not indicated for treatment of central nervous system tissue. When stimulation compatible components are used , the iovera System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator. Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-------------------------------------------------------------------------------------------------------| | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Device Trade Name: | iovera º System | |----------------------|-------------------------------------------------------------------------------------------------------------------------------| | Device Common Name | Cryogenic Surgical Device | | Device Class: | II | | Classification Name: | Cryosurgical Unit and Accessories Device<br>Surgical Nerve Stimulator | | Regulation No.: | 21 CFR Part 882.4250 and 874.1820 | | Product Code: | GXH, ETN | | Predicate Device: | ioveraº System K173763 and K211334 | | Owner/Submitter: | Pacira Biosciences, Inc.<br>10450 Science Center Drive<br>San Diego, California 92121 | | Regulatory Contact: | Danny Rivera<br>Senior Director, Quality and Regulatory Affairs<br>Tel: 8586252414 Ext 3175<br>Email: Danny.Rivera@Pacira.com | | Date: | May 20, 2022 | {4}------------------------------------------------ ### DEVICE DESCRIPTION The iovera? System is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N2O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization. The iovera" System may be used in conjunction with standard of-the-shelf nerve stimulator device in applications where precise nerve location is desired. ### Device Design The device comprises four main components: - A reusable Handpiece - A Charging Dock - An assortment of single-patient-use Smart Tips ● - A Cartridge containing nitrous oxide ● The iovera® System Handpiece is battery powered and provides feedback to the user during device preparation and use. The Handpiece connects to both the Cartridge and to the Smart Tip. The user activates a treatment cycle through a control on the Handpiece, which starts and stops the treatment. The Handpiece contains an LCD display for providing feedback to the user when the device is ready to use. The Charging Dock stores the Handpiece between uses and provides power for charging the battery. An assortment of Smart Tips is available for the ioveraº system. All Smart Tip needles are made of stainless steel and have a closed-end design that fully contains the cryogen so that it does not enter the target tissue. The Smart Tip is the only patient contacting component of the iovera® System. The user removes the Smart Tip from its sterile packaging and attaches it to the Handpiece immediately before use. Certain Smart Tip designs provide a connection to an external nerve stimulator to facilitate nerve location prior to treatment. The iovera System uses a commercially available nitrous oxide cylinder. The Cartridge is filled with pure N20. #### Device Functionality/Scientific Concepts: The device functionality is based on the user introducing the Smart Tip into the selected treatment area or the target nervous tissue. The user then initiates the flow of cryogen by pressing the on/off button. Liquid cryogen flows from the Handpiece into the closed-end Smart Tip. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization; as the liquid cryogen expands into a gas, the temperature drops around the external surface of the Smart Tip, causing the surrounding tissue to freeze. The treatment is completed after a pre-programmed amount of time, at which time the user can safely remove the Smart Tip from the treatment site. {5}------------------------------------------------ #### INTENDED USE/INDICATIONS FOR USE The iovera? System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera® System is not indicated for treatment of central nervous system tissue. When stimulation compatible components are used, the iovera? System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator. ### SUMMARY/COMPARISON OF IOVERA® SYSTEM DEVICE CHARACTERISTICS The iovera System is substantially equivalent in intended use, technology, design, and materials to the above listed legally marketed predicate devices. | Parameter | Proposed<br>iovera o System | Predicate 1<br>iovera o System<br>K211334 | Predicate 2<br>iovera o System<br>K173763 | |--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Destroy tissue through<br>freezing | Same | Same | | Indications for Use | The ioverao System is used<br>to destroy tissue during<br>surgical procedures by<br>applying freezing cold. It<br>can also be used to produce<br>lesions in peripheral nervous<br>tissue by the application of<br>cold to the selected site for<br>the blocking of pain. It is<br>also indicated for the relief<br>of pain and symptoms<br>associated with osteoarthritis<br>of the knee for up to 90<br>days. The ioverao system is<br>not indicated for treatment<br>of central nervous system<br>tissue.<br>When stimulation | The ioverao System is used to<br>destroy tissue during surgical<br>procedures by applying<br>freezing cold. It can also be<br>used to produce lesions in<br>peripheral nervous tissue by<br>the application of cold to the<br>selected site for the blocking<br>of pain. It is also indicated for<br>the relief of pain and<br>symptoms associated with<br>osteoarthritis of the knee for<br>up to 90 days. The ioverao<br>system is not indicated for<br>treatment of central nervous<br>system tissue. | The ioverao System is used to<br>destroy tissue during surgical<br>procedures by applying<br>freezing cold. It can also be<br>used to produce lesions in<br>peripheral nervous tissue by<br>the application of cold to the<br>selected site for the blocking<br>of pain. It is also indicated for<br>the relief of pain and<br>symptoms associated with<br>osteoarthritis of the knee for<br>up to 90 days. The ioverao<br>system is not indicated for<br>treatment of central nervous<br>system tissue.<br>The ioverao System's<br>"1x90" Smart Tip | | | compatible components are<br>used, the ioverao System can<br>also facilitate target nerve<br>location by conducting<br>electrical nerve stimulation<br>from a 3rd party nerve<br>stimulator. | | configuration (indicating<br>one needle which is 90<br>mm long) can also<br>facilitate target nerve<br>location by conducting<br>electrical nerve<br>stimulation from a<br>separate nerve stimulator. | | Anatomical Sites | Peripheral nerves | Same | Same | | Parameter | Proposed<br>iovera o System | Predicate 1<br>iovera o System<br>K211334 | Predicate 2<br>iovera o System<br>K173763 | | Technology | Cryogenic surgical device<br>with needle which<br>penetrates treatment area | Same | Same | | Cryogen cartridge size | 21ml (14.3 gram fill)<br>nitrous oxide cartridge | Same | 10ml (7.5 gram fill) nitrous<br>oxide cartridge | | Cryogen cartridge loading | Cartridge is loaded via a<br>hinged door on the side of<br>the device. Closing the<br>door provides the force<br>necessary to pierce the<br>cartridge. | Same | Cartridge is loaded from the<br>back of the device. Screwing<br>on the cap provides the force<br>necessary to pierce the<br>cartridge. | | Cryogen cartridge<br>piercing point/filter | In-line piercing point and<br>filter assembly resides in a<br>replaceable assembly<br>within the device. | Same | In-line piercing point and<br>filter are part of cryogen<br>cartridge assembly. | | Human Factors | Hand-held device containing<br>cryogen. Detachable<br>cryoprobes (Smart Tips). | Same | Same | | Information display | Color graphic LCD on rear<br>of device. Additional<br>information regarding<br>failure mode is also | Same | Array of different colored<br>LEDs located around the<br>device. | | Charging Dock display | LED light bar indicating<br>Charging Dock is receiving<br>power from AC-DC<br>adapter. Light bar changes<br>color when Charging Dock<br>is supply charging current<br>to Handpiece. | Same | LED indicating Charging<br>Dock is receiving power<br>from AC-DC adapter. | | Digital Interfaces | USB and SD card. The SD<br>card is not part of the<br>product offering to the<br>customer. This tool is used<br>for troubleshooting during<br>in-house failure<br>investigation. The SD card<br>has no impact on | Same | USB | | Method of Smart Tip<br>attachment | Push-on/pull-off design | Same | Screw-on tip attachment | | Skin-warmer heating | Semiconductor device used<br>as heating element - this<br>element is thermally<br>connected to heating block<br>and monitored with<br>redundant thermistors. | Same | Resistors used as heating<br>element - this element is<br>thermally connected to<br>heating block and monitored<br>with redundant thermistors. | | Parameter | Proposed<br>iovera® System | Predicate 1<br>iovera® System<br>K211334 | Predicate 2<br>iovera® System<br>K173763 | | Cryogen valve | Valve comprises needle<br>valve in compliant sealing<br>element.<br>Valve is controlled by<br>stepper-motor-based linear<br>actuator.<br>Switch detects valve open<br>position. | Same | Valve comprises face-to-<br>face seal against compliant<br>sealing element.<br>Valve is controlled by<br>stepper-motor-based linear<br>actuator. | | Power Source | Battery powered<br>Battery type: Single cell<br>Lithium Ion, 3100mAh<br>Battery voltage: 3.6Volts<br>Main powered Charging<br>Dock | Same | Same | | Operating Principle | Joule-Thomson Effect/<br>Latent Heat of<br>Vaporization | Same | Same | | Needle Size | 20G, 27G | Same | 20G | | Catalog Numbers | STT2309, STT2190, and<br>STT2190STIM | STT2309 and STT2190 | STT0811 | | Needle Working Length | 8.5mm, 90mm | Same | 90mm length | | Needles/Tip | 1 x 20G (90mm)<br>3 x 27G (8.5mm) | Same | 1 x 20G (90mm) | | Nerve Stimulation Support | Yes¹ | N/A | Yes | | Pre-heat time (min) | 12 seconds (8.5mm)<br>1 second (90mm) | Same | 1 second (90mm) | | Cooling Time (max) | 33 seconds (8.5mm)<br>60 seconds (90mm) | Same | 60 seconds (90mm) | | Post -heat Time | 20 seconds (8.5mm)<br>45 seconds (90mm) | Same | 45 seconds (90mm) | | Skin Warmer Set Point | 30°C | Same | Same | | Clinical Effect | Cellular death through<br>cryoneurolysis and<br>Wallerian degeneration<br><br>Second degree nerve<br>injury (axonotmesis) | Same | Same | | Treatment Temperature | -55° C to -75°C (-67°F to -<br>103°F) | Same | Same | | Parameter | Proposed<br>iovera o System | Predicate 1<br>iovera o System<br>K211334 | Predicate 2<br>iovera o System<br>K173763 | | Treatment Time | Preprogrammed into the<br>smart tip. | Same | Same | | Treatment Options | Multiple Smart Tip<br>options are available for<br>producing different<br>cryozone (ice ball) sizes. | Same | Same | | Single-patient Use | Smart Tips are single-patient<br>use. Handpiece is reusable. | Same | Same | | Sterility Assurance Level | Smart Tip individually<br>packaged and sterile with<br>a SAL of 10-6. | Same | Same | | Shelf Life | 37-month | Same | Same | | Biocompatible | Biocompatible. All patient<br>contacting materials, tested<br>to ISO 10993-1 | Same | Same | ### Substantial Equivalence Comparison Table {6}------------------------------------------------ {7}------------------------------------------------ <sup>4</sup> Nerve-stim tips have a thin, electrically insulating coating (Parylene C) over most of the needle and include an electrical connection between the needle and an external, off-the-shelf PENS device. All other aspects of the design are identical to the non-nerve-stim version of the tip. {8}------------------------------------------------ ## SUMMARY OF PERFORMANCE TESTING Design Verification testing was performed for the ioveras System to demonstrate that the device meets product specifications for its intended clinical use. Using a risk-based approach for the modifications, design verification testing was performed according to recognized standards, and is consistent with the predicate device and test methods described in previous 510(k) submissions for the iovera system. The verification methods used and applied are appropriate for the changes. The verification and validation testing assures the modified subject device meet design input requirements, product specifications and relevant standards. - Functional and Product Performance Testing of the Handpiece and Cartridge - Smart Tip Functional Testing - User Interface Testing: IEC 62366-1:2015 ● - Software Testing: Conforms to IEC 62304:2006/A ● - Electrical and EMC Safety Testing: Conforms to IEC-60601-1: A1 2012 and 60601-1-2:2014 ● - Mechanical and Thermal Safety Testing: Conforms to IEC 60601-1: A1:2012 ● - Sterility Testing: Conforms to ASTM F1980-16 and ISO 11135:2014/A ● - Packaging Testing: Conforms to ASTM D4332-14. ● - Biocompatibility Testing: Conforms to ISO 10993-1 ● Verification and validation testing performed on the subject device supports the substantial equivalence of the modified ioveraº System to the predicate device. The subject ioveraº System device met the design verification and validation inputs, passing all predetermined acceptance criteria. No new issues of safety or effectiveness were identified during design verification and validation testing. {9}------------------------------------------------ ## CONCLUSION The subject ioveraº System, with the same intended use, indications for use, design, materials, technological features, and principles of operation as the cleared ioveraº Systems (K211334 and K173763), is substantially equivalent to these predicate devices. The modification to the Indications for Use proposed in this 510(k) does not affect the safety and effectiveness of the device as the intended therapeutic use for the creation of lesions with the application of cold in peripheral nerves to block pain has not changed. The modification described, and the information presented, in this special premarket notification demonstrates that ioveraº System is safe and effective, does not introduce or raise new questions of safety and effectiveness, and is substantially equivalent in intended use, technology, design, and materials to the legally marketed predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...