K120415 · Myoscience, Inc. · GXH · Jun 22, 2012 · Neurology
Device Facts
Record ID
K120415
Device Name
CRYO-TOUCH III
Applicant
Myoscience, Inc.
Product Code
GXH · Neurology
Decision Date
Jun 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Myoscience Cryo-Touch III is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch III is not indicated for treatment of central nervous system tissue.
Device Story
Portable cryogenic surgical device; utilizes liquid nitrous oxide (N2O) to cool a needle probe 'Tip' via Joule-Thomson effect and/or Latent Heat of Vaporization. Used to destroy tissue or create peripheral nerve lesions for pain management. Comprised of re-usable control unit, re-usable handpiece, and single-patient-use tip. May be used with standard off-the-shelf nerve stimulators for precise nerve localization. Operated by clinicians in surgical settings. Output is localized extreme cold applied to target tissue. Benefits include precise, controlled tissue destruction or nerve block.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, and ASTM F882-84.
Technological Characteristics
Portable cryogenic surgical unit. Components: re-usable control unit, re-usable handpiece, single-use tip. Cooling mechanism: Joule-Thomson effect/Latent Heat of Vaporization using N2O. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, ASTM F882-84. Sterilization: ISO 11137-1 compliant.
Indications for Use
Indicated for surgical tissue destruction and peripheral nerve lesioning for pain blocking. Contraindicated for central nervous system tissue.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
K102021 — CRYO-TOUCH II · Myoscience, Inc. · Sep 17, 2010
K123516 — CRYO-TOUCH IV · Myoscience, Inc. · Jan 10, 2013
K133453 — IOVERA · Myoscience, Inc. · Mar 24, 2014
K142866 — Myoscience iovera system · Myoscience, Inc. · Jan 21, 2015
K173763 — iovera system · Myoscience, Inc. · Feb 28, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
## myoscience
# Cryo-Touch III
### Premarket Notification
### Section 5: 510(k) Summary
#### Device Information:
| Category | Comments |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor / Submitter: | Myoscience, Inc<br>525 Chesapeake Drive<br>Redwood City, CA 94063<br>(650) 474-2600<br>(650) 474-2700 |
| Correspondent Contact<br>Information: | Tracey Henry<br>Vice President RAQA, Clinical Compliance<br>525 Chesapeake Drive<br>Redwood City, CA 94063<br>(650) 474-2600<br>(650) 474-2900 |
| Device Common Name: | Cryogenic Surgical device |
| Device Classification & Code: | Class II, GXH |
| Device Classification Name: | Cryosurgical unit and accessories (21 CFR 882.4250) |
| Device Proprietary Name: | Myoscience Cryo-Touch III |
#### a. Predicate Device Information:
The Cryo-Touch III is substantially equivalent to the following currently legally marketed devices:
| 510(k) Number | Product | Sponsor |
|---------------|---------------|-----------------|
| K083493 | Cryo-Touch | Myoscience, Inc |
| K102021 | Cryo-Touch II | Myoscience, Inc |
#### b. Date Summary Prepared
February 8, 2012
#### c. Description of Device
The Myoscience Cryo-Touch III is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a needle probe, "Tip" cooled by the cryogenic fluid (liguid nitrous oxide, N₂O) to a selected area. The Tip is cooled by the Joule-Thomson effect and/or Latent Heat of Vaporization. The Cryo-Touch III may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.
#### Device Design
The device is comprised of four main components:
- 1. a re-usable Control Unit,
- 2. a re-usable Handpiece,
- a single patient use Tip, and 3.
Page
{1}------------------------------------------------
# myoscience
| Test Performed | Result |
|-----------------------------------------|---------------------------------------------|
| Temperature<br>reproducibility | PASS, Substantially equivalent to predicate |
| Mechanical<br>Integrity for system | PASS, Substantially equivalent to predicate |
| Disable valve<br>reliability | PASS, Substantially equivalent to predicate |
| Thermal fuse<br>testing | PASS, Substantially equivalent to predicate |
| Nitrous exposure | PASS, Substantially equivalent to predicate |
| Cryozone size | PASS, Substantially equivalent to predicate |
| Needle Integrity | PASS, Substantially equivalent to predicate |
| Device Durability | PASS, Substantially equivalent to predicate |
| Sterilization and<br>Shelf Life Testing | PASS, Substantially equivalent to predicate |
| Electrical Safety<br>Testing | PASS, Substantially equivalent to predicate |
| Software Testing | PASS, Substantially equivalent to predicate |
| Safety Testing | PASS, Substantially equivalent to predicate |
| Biocompatibility<br>Testing | PASS, Substantially equivalent to predicate |
This performance testing demonstrated that the device is in compliance with pertinent standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, and ASTM F882-84), the product labeling, and is substantially equivalent to the predicates.
Clinical Testing Submitted: None
#### g. Conclusion
Myoscience concludes that the Cryo-Touch III described in this submission is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol, with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MyoScience, Incorporated % Ms. Tracey Henry Sr. Director, Regulatory Affairs and Quality Assurance 525 Chesapeake Drive Redwood City, CA 94063
JUN 2 2 2012
Re: K120415
Trade/Device Name: Myoscience Cryo-touch III Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: II Product Code: GXH Dated: June 13, 2012 Received: June 15, 2012
Dear Ms. Henry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for answer to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability additeration. Thease now u, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 toure be actived a determination that your device complies with other requirements of the Act that I DA has made a actoring administered by other Federal agencies. You must
{3}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you attp://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 1000 the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Cryo-Touch III Premarket Notification
### Section 4: Indications for Use Statement
510(k) Number:
Device Name: Myoscience Cryo-Touch III
#### Indications for Use:
The Myoscience Cryo-Touch III is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch III is not indicated for treatment of central nervous system tissue.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Daryl Kausin
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices
K120415 510(k) Number.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.