ENTERPRISE ELECTROPHYSIOLOGY SYSTEM

K110967 · Neuralynx, Inc. · OMC · Jun 30, 2011 · Neurology

Device Facts

Record IDK110967
Device NameENTERPRISE ELECTROPHYSIOLOGY SYSTEM
ApplicantNeuralynx, Inc.
Product CodeOMC · Neurology
Decision DateJun 30, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2

Indications for Use

The intended use of the Neuralynx Enterprise Electrophysiology System is for temporary (<30 days) recording and monitoring of brain electrical activity.

Device Story

Enterprise Electrophysiology System is an EEG device for recording and monitoring brain electrical activity. System components include a clinical headstage, interface cabinet, and monitoring station PC. It supports up to 160 channels of recording. The device does not include electrodes; it connects to third-party electrode arrays (AdTech or PMT). The headstage connects to the interface cabinet, which acquires unprocessed data; data is transmitted to the monitoring station PC via fiber-optic Ethernet. Software on the PC controls signal display and processing. Raw data channels are stored without manipulation. Used in clinical, operating room, and epilepsy monitoring unit environments by healthcare professionals. The system facilitates routine recording, monitoring, and retrieval/replay of brain activity recordings to assist in clinical assessment.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including verification of design specifications and adherence to international safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26) and software verification.

Technological Characteristics

EEG system with 160-channel capacity. Features 24-bit A/D sampling at 32,556 samples/sec/channel. Analog bandwidth DC to 8KHz. Input range +/- 132mV. Common mode rejection 100dB. Connectivity via fiber-optic Ethernet between interface cabinet and PC. Electrical classification: Class 1, Type BF protection. Software supports acquisition, display, and storage of raw data. No electrodes included; compatible with third-party arrays.

Indications for Use

Indicated for temporary (<30 days) recording and monitoring of brain electrical activity in clinical, operating room, and epilepsy monitoring unit environments.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K110967 JUN 30 2011 # 510(k) Summary This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date April 4, 2011 [21 CFR 807.92(a)(1)]. A. Applicant Name and Address [21 CFR 807.92(a)(1)] Neuralynx Inc. 105 Commercial Drive Bozeman, MT 59715-2217 Tel: (406) 585-4542 Fax: (866) 585-1743 B. Contact Information Neuralynx Inc. 105 Commercial Drive Bozeman, MT 59715-2217 Tel: (406) 585-4542 Fax: (866) 585-1743 Contact person: Michael Johnson M.D. Email: mikejohnson@neuralynx.com Page 1 of 5 {1} C. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] Trade Name: Enterprise Electrophysiology System Device Common Name: Electroencephalograph Classification Name: Electroencephalograph (per 21 CFR 882.1400) Product Code: OMC Panel: Neurology Device Classification: Class II D. Predicate Devices [21 CFR 807.92(a)(3)] The Enterprise Electrophysiology System uses similar technology and has similar physical output characteristics as the following predicate device: k060523 NeuroPort System manufactured by Cyberkinetics Neurotechnology Systems. | Parameters | Neuralynx Enterprise Electrophysiology System | Cyberkinetics NeuroPort System | | --- | --- | --- | | 510k number | K110967 | K060523 | | Indications for Use | The intended use of the Neuralynx Enterprise Electrophysiology System is for temporary (<30 days) recording and monitoring of brain electrical activity. | The intended use of the Cybernetics Neurotechnology Systems, Inc. NeuroPort System is for temporary (<30 days) recording and monitoring of brain electrical activity. | | Location of Use | Clinical, operating room, and epilepsy monitoring unit environments. | Clinical, operating room, and epilepsy monitoring unit environments. | {2} | Physical Characteristics | Neuralynx Enterprise | Cyberkinetics NeuroPort System | | --- | --- | --- | | Amplifier | Interface Cabinet | Front End Amplifier / Digitizer | | Signal processor | Interface Cabinet | Neural Signal Processor | | Display | Monitoring Station PC | Display PC | | Electrodes | Neuralynx Enterprise | Cyberkinetics NeuroPort System | | Electrodes supplied by Manufacturer | No Uses AdTech (k053363) and PMT (k082474) electrodes | Yes Uses NeuroPort Electrode Array(k042384, k070272) | | Anatomical Sites of Use | No direct patient contact | Manufactures electrodes which contact patient | | System connects to electrodes | Yes | Yes | | Function | Neuralynx Enterprise | Cyberkinetics NeuroPort System | | Number of Channels | Up to 160 channels | Up to 128 channels | | Sampling Resolution | 24 bit A/D | 16 bit A/D | | Analog Bandwidth | DC to 8KHz | unknown | | Sampling Frequency | 32,556 samples per second per channel | 30,000 samples per second per channel | | Analog Input Range | +/- 132 millivolts 1 microvolt per bit at bit #18 | 8.1 mv to -8.1 mv | | Input Noise Levels | 8 microvolt peak to peak 1.3 microvolt RMS | 14 microvolt peak to peak 3 microvolts RMS | | | | | | Common mode rejection | 100db | 90 db | | CMR Input Range | +/- 10v | +/- 3v | | Number of references | 1 per channel | 1 per bank of 32 channels | | | | | | Interface Cabinet to PC data communications connection | Fiber Optic Cable | Fiber Optic Cable | | | | | | Electrical Classification | Class 1, continuous operation | Class 1, continuous operation | | Degree of protection against electrical shock | Type BF | Type BF | Page 3 of 5 {3} Page 4 of 5 ## E. Device Description [21 CFR 807.92(a)(4)] The Enterprise Electrophysiology System is an electroencephalograph (EEG) device intended for recording and monitoring electrical brain activity. Functions include routine recording of brain activity, monitoring of brain activity, and retrieval and replay of recordings. The Enterprise System supports 160 channels of recording. The Enterprise System does not include electrodes. The system includes the Clinical Headstage, Interface Cabinet, and a Monitoring Station PC. The Enterprise System software supports acquisition and display. ## F. Device Specifications [21 CFR 807.92(a)(6)] The Neuralynx Enterprise System does not include the electrode arrays. These are supplied by AdTech (K053363) or PMT (K082474). The Enterprise System includes the headstage connection to the electrode arrays. The headstage connects to the Interface Cabinet which controls the acquisition of unprocessed data. The Interface Cabinet connects to the Monitoring Station PC by way of a fiberoptic Ethernet connection. The software in the Monitoring Station PC controls the display and processing of signal channels. Raw data channels are stored without manipulation. {4} G. Indications for Use [21 CFR 807.92(a)(5)] The intended use of the Neuralynx Enterprise Electrophysiology System is for temporary (&lt;30 days) recording and monitoring of brain electrical activity. H. Performance Data [21 CFR 807.92(b)(2)] Testing was conducted to verify that the Enterprise System met design specifications and was substantially equivalent to the predicate devices. Testing consisted of third party verification of adherence to IEC 60601-1, IEC 60601-1-2, IEC 606012-2-26. Software verification was also performed. No Clinical testing is required. I. Conclusion [21 CFR 807.92(b)(3)] Technologically, the Enterprise System was found to be substantially equivalent to the currently cleared k060523/k042626 NeuroPort manufactured by Cyberkinetics Neurotechnology Systems Inc. The indications for use are identical to the previously cleared device. The risks and benefits for the Enterprise System are argued to be comparable to the predicate devices. We believe that there are no new questions of safety or efficacy raised by the introduction of the Enterprise System. The nonclinical performance testing demonstrates that the Enterprise System is as safe, as effective, and performs at least as safely and effectively as the legally marketed device. {5} DEPARTMENT OF HEALTH &amp; HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Neuralynx, Inc. Michael Johnson M.D. Clinical and Regulatory Manager 105 Commercial Drive Bozeman, MT 59715 JUN 30 2011 Re: K110967 Trade/Device Name: Enterprise Electroencephalographic System Regulation Number: 21 CFR 882.1400 Regulation Name: Reduced-Montage Standard Electroencephalograph Regulatory Class: Class II Product Code: OMC Dated: April 4, 2011 Received: April 6, 2011 Dear Dr. Johnson We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {6} Page 2 – Dr. Michael Johnson comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health’s (CDRH’s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ![img-0.jpeg](img-0.jpeg) Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health {7} # Indications for Use 510(k) Number (if known): K110967 Device Name: *Enterprise Electrophysiology System* Indications for Use: The intended use of the Neuralynx *Enterprise Electrophysiology System* is for temporary (&lt;30 days) recording and monitoring of brain electrical activity. Prescription Use ☑ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ☐ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K110967
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...