GALAXY SYSTEM; JUPITER

K111055 · Personal Health Institute International · OLV · Dec 21, 2011 · Neurology

Device Facts

Record IDK111055
Device NameGALAXY SYSTEM; JUPITER
ApplicantPersonal Health Institute International
Product CodeOLV · Neurology
Decision DateDec 21, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesPediatric

Indications for Use

The Galaxy System (which includes the Jupiter and Galaxy software) is intended for use as a polysomnographic system to acquire record, display, print and store physiological parameters to assist clinicians/physicians. These parameters are presented graphically on a computer screen for diagnostic review, similar in application to the use of a traditional paper based polygraph recorder. The device will be used in hospitals, institutions, sleep centers, clinics or other test environments where adults or infants require the documentation of sleep or other physiological disorders. The Galaxy system does not provide alarms and is not intended for use as an automated apnea monitor. Caution: Federal law restricts this device to sale by or on the order of a Physician.

Device Story

Galaxy System is a polysomnography device for acquiring, recording, displaying, and storing physiological parameters (EEG, EOG, EMG, ECG, respiratory effort, nasal flow, body position, snoring, oximetry). System comprises Jupiter headbox/amplifiers (models 1166/1142), ISO101 isolator, and PC-based Galaxy software. Input signals are digitized and transmitted via Ethernet to a PC. Software provides manual review, annotation, and report generation tools for clinicians. Used in hospitals, sleep centers, and clinics by physicians/technicians. Device does not provide automated scoring, alarms, or apnea monitoring. Output is graphical display of physiological traces for diagnostic review, aiding clinical decision-making for sleep disorders.

Clinical Evidence

Bench testing only. Verification and validation included amplifier functionality, impedance testing, noise, grounding, and oximeter interface. Software verified per ANSI/AAMI SW68:2001. EMC and safety testing performed per IEC 60601 series.

Technological Characteristics

Polysomnography system; 32 or 64 AC channels; 8 DC channels (model 1142). Connectivity: Ethernet/TCP/IP to PC. Power: AC-DC adapter with ISO101 patient isolation. Digital resolution: 16-bit. Sampling: 32,768 Hz (AD), 1024 Hz (output/storage). Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-26, ANSI/AAMI SW68:2001.

Indications for Use

Indicated for adults and infants requiring documentation of sleep or other physiological disorders in clinical settings (hospitals, sleep centers, clinics). Not for use as an automated apnea monitor or life-support equipment.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K111055 ## 510(k) Summary ## Submitter DEC 2 1 2011 Personal Health Institute (PHI) international Rijswijkstraat 141e 1062 ES Amsterdam The Netherlands Phone: 011 31 646104625 011 31 207071538 Fax: #### Registration Number: Will apply ## Contact person Anand Kumar #### Preparation Date December 21, 2011 Device Trade Name: Galaxy System Classification Name: Standard polysomnograph with electroencephalograph Regulation Number: 882.1400 Product Code: OLV Class II Device Class: Classification Panel: Neurology #### Predicate Devices Neurolink IP Model PK1117 by Natus Medical Product code: GWQ 510(k) number: K100683 Alice 5 by Respironics Product code :GWQ 510(k) number: K04059 {1}------------------------------------------------ # Device Description The Galaxy system (which includes the Jupiter and Galaxy software) is a Polysomnography System that is intended to acquire, record, display and print physiological information to clinicians/physicians. These parameters are presented graphically on a computer screen for diagnostic review, similar in application to the use of a traditional paper based polygraph recorder. The device will be used in hospitals, institutions, sleep centers or clinics, or other test environments where adults or infant patients require the documentation of various sleep or other physiological disorders. The system can record, monitor, store and transfer of up to 42 channels of biophysical parameters. Generally the device is capable of acquiring and displaying the following parameters: - EEG . - EOG (eye movement) ● - Chin EMG . - . Leg EMG (leg movement) - ECG (single channel) ● - . Chest respiratory effort - . Abdomen respiratory effort - . Nasal Flow Thermistor - . Nasal Flow Pressure - . Body position - . Snoring - Oximeter . - . Patient "Event" button The components of the Galaxy system include: - Headbox/Amplifiers (i.e., "BrainBox EEG amplifier, "Touchproof connector box", "Jupiter 1. amplifier") – Galaxy can support two amplifier/headbox models, 1166 and 1142. Both models are electrically exactly the same, except for the differences between the number of channels. Channel characteristics of each of the two available amplifiers are as follows: | | Model 1166 | Model 1142 | |----------------------------------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------| | AC Channels (e.g., EEG, EOG, EMG, nasal flow<br>pressure, body position, respiratory effort) | 64 | 32 | | DC Channels (e.g., body position) | 0 | 8 | | Oximetry and Event button connection | Yes | Yes | | Total number of connectable electrodes/ sensors<br>used for data collection | 66 | 42 | | Number of grounds | 2 (G1, G2 - for<br>"common<br>reference<br>and<br>ground" | 2 (G1, G2 - for<br>"common<br>reference<br>and<br>ground" | {2}------------------------------------------------ - 2. Ethernet Interface connects the power supply and IP connections with the amplifiers (through the "Isolator") for converting serial data of the amplifier to the Ethernet data of the PC. - Isolator (ISO101) Isolates the AC power and Ethernet signals for patient safety. 3. - Connecting Wires the amplifiers and isolator are connected with a "BrainBus" serial 4. interface; the Isolator and Ethernet Interface are connected with a "BrainNet Connector". - Desktop Computer Receives the EEG data through the Ethernet and TCP/IP connection from 5. the "Amplifier" and then stores and displays it to the user. It also serves as the user interface for the device (receiving and implementing commands from the user). - 6. Electrodes Third party electrodes and sensors such as surface electrodes compatible with 32 AC inputs by means of touch-proof connectors, "Nonin oximeter (XPOD3012)," and "respiratory effort sensors" can be used with the system, but these sensors are not provided with the device. Only sensors specifically tested and verified for use with the device should be used and are listed in the user's manual. The interconnections of the components of the Amplifier are shown below: Image /page/2/Figure/6 description: The image shows a setup of EEG equipment. There is an isolator labeled ISO-101, and a main amplifier labeled EEG 1142 or PSG 1142. There is also an input box labeled INBOX-1142 A. Cables connect the devices, including an Ethernet cable connected to a PC and a cable connecting to Isolator ISO-101. Figure 1: Jupiter Hardware Set-up All the connections and functions are exactly similar to the predicate device Natus Neurolink IP 1117. The Galaxy software does not support the control of the flash unit and the digital I/O. - 7. Galaxy Software The Galaxy software's main functionalities include the following: - a. Record and display signals Collect and display PSG parameters. Plot acquired data on screen (within a time window of 30 seconds) and store on hard-disk of computer. {3}------------------------------------------------ - b. Provide tools for Manual Review of Data Allow the user to review and manually analyze data, edit this analysis and delete the entered events if needed. Software also displays the user annotations along with the signal traces, as trend overview over the night and as a list, and allows the user to review the acquired data after the completion of the recording to examine and annotate afterwards (offline). - c. Generate Patient Reports The software calculates summaries of the manually scored data and print them as tabulated reports and shows simple computer calculations like average value, rates etc ... Galaxy also has the following additional functions: - changing the mode of the amplifier for calibration, impedance check or data acquisition . - sending commands to set sample rate of the digitization, to specify the recording montage . and to receive data from the Jupiter amplifier via Ethernet interface. Image /page/3/Figure/5 description: The image shows a polysomnography (PSG) report, which is a sleep study. The report is divided into several panes, including a label pane, a signal pane, a trend pane, an event pane, and a hypnogram. The signal pane displays various physiological signals, such as EKG, SaO2, and EEG. The hypnogram shows the different sleep stages throughout the night, and the event pane lists various events that occurred during the sleep study, such as desaturations and arousals. A screen shot below shows the features of Galaxy. Figure 2: Sample Galaxy Software output screen {4}------------------------------------------------ The device does not provide any automatic scoring algorithms. ### Intended Use The Galaxy System (which includes the Jupiter and Galaxy software) is intended for use as a polysomnographic system to acquire record, display, print and store physiological parameters to assist clinicians/physicians. These parameters are presented graphically on a computer screen for diagnostic review, similar in application to the use of a traditional paper based polygraph recorder. The device will be used in hospitals, institutions, sleep centers, clinics or other test environments where adults or infants require the documentation of sleep or other physiological disorders. The Galaxy system does not provide alarms and is not intended for use as an automated apnea monitor. Caution: Federal law restricts this device to sale by or on the order of a Physician. ## Technological Characteristics The comparison table is provided as a summary of the technological characteristics relative to the predicate devices. The items like Intended use, principle of operation, are compared. The summary of this comparison table demonstrates that the Galaxy System has no significant differences from the predicate devices that would adversely affect product safety and effectiveness. #### Testing The Galaxy System has been tested and verified in various phases, internal testing, verification and validation as well as external testing and validation. The device passed verification and validation testing that includes tests for amplifier voltages and functioning, flash memory, pulse width specifications, impedance testing, channel outputs with and without input signals, noise, electrode grounding, and oximeter interface functionality. The design was verified throughout the design process. Risk analysis was done, appropriate measures were implemented and their effectiveness verified. The external test house DARE was used to confirm compliance to EMC requirements. Safety Tests have been performed to verify compliance with IEC 60601-1-1 and IEC 60601-2-26 to ensure that there are no potential hazards on patients, other persons, or the surroundings. Electromagnetic Compatibility tests according to IEC 60601-1-2 have been performed to ensure no intolerable magnetic disturbances are introduced into its electromagnetic environment. Immunity tests to IEC 60601-1-2 have been performed to ensure that the EEG equipment has the ability to operate satisfactorily in its electromagnetic environment. The Galaxy system was tested for displaying and printing of signals and scoring. These tests were performed at the work-bench by developers, in the factory by developers and in the field by sleep-technicians and researchers. The standard ANSI/AAMI SW68: 2001 Medical Device Software - Software Life Cycle Processes was used as advisory standard for the development and testing of all software functions. {5}------------------------------------------------ # Substantial Equivalence Comparison Table | Characteristics | Galaxy | Alice 5 | Neurolink IP Model<br>PK 1117 | |----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | K100683 | | 510 (K) ID | K111055 | K040595 | K100683 | | Device<br>Classification | II | II | II | | Product Code | OLV | GWQ | GWQ | | Classification<br>Panel | Neurology | Neurology | Neurology | | Intended use | The Galaxy System(Software<br>and Jupiter Amplifier) is<br>intended for use as a<br>polysomnographic system to<br>acquire record, display; print<br>and store physiological<br>parameters to assist<br>clinicians/physicians. These<br>parameters are presented<br>graphically on a computer<br>screen for diagnostic review,<br>similar in application to the<br>use of a traditional paper<br>based polygraph recorder.<br>The device will be used in<br>hospitals, institutions, sleep<br>centers or clinics or other test<br>environments where adults<br>or infants require the<br>documentation of sleep or<br>other physiological disorders.<br>The Galaxy system does not<br>provide alarms and is not<br>intended for use as an<br>automated apnea monitor.<br>Caution: This device is to be<br>used under the supervision of<br>a physician. | The Alice 5 System is a<br>Polysomnographic System<br>that is intended to record,<br>display, and print<br>physiological parameters to<br>clinicians/physicians. These<br>parameters are presented<br>graphically on a computer<br>screen for diagnostic review,<br>similar in application to the<br>use of a traditional paper<br>based polygraph recorder. The<br>device will be used in<br>hospitals, institutions, sleep<br>centers or clinics, or other test<br>environments where adult or<br>infant patients require the<br>documentation of various<br>sleep or other physiological<br>disorders.<br>The device does not provide<br>alarms and is not intended for<br>use as an automated apnea<br>monitor.<br>This device is to be used<br>under the supervision of a<br>physician. | Neurolink IP model PK1117<br>is intended to be used as an<br>electroencephalograph to<br>acquire, digitize and<br>transmit<br>electroencephalographic<br>and other physiological<br>signals (such as pulse and<br>oximetry) for EEG in<br>research and clinical<br>environments.<br>This device is to be used<br>under the supervision of a<br>physician. | | Note | Electrophysiological characteristics of the Jupiter Amplifier are technically the same to both<br>predicates.<br>The software functions of Galaxy are functionally equivalent to Alice5. | | | | Warning | Do not use in conjunction<br>with a defibrillator and<br>stimulators.<br>Do not use in conjunction<br>with medical imaging devices | Not known | Do not use in conjunction<br>with a defibrillator and<br>stimulators.<br>Do not use in conjunction<br>with medical imaging<br>devices. | | Contra<br>indications | This device does not provide<br>alarms and is not intended<br>for use as an automated<br>apnea monitor. The software<br>is not intended for use as a<br>life-support equipment like<br>vital signs monitoring | This device does not provide<br>alarms and is not intended<br>for use as an automated<br>apnea monitor. | The device is not intended<br>for use as a life support<br>equipment such as vital<br>signs monitoring in<br>intensive care units | | Characteristics | Galaxy | Alice 5 | Neurolink IP Model<br>PK 1117 | | 510 (K) ID | K111055 | K040595 | K100683 | | Prescription use | Yes | Yes | Yes | | Contact of device<br>with patient body | None | None | None | | Environment | The device can be used in<br>hospitals, institutions, sleep<br>centers or other similar<br>environments where patients<br>require the documentation of<br>various sleep or other<br>physiological disorders. | The device will be used in<br>hospitals, institutions, sleep<br>centers or clinics, or other<br>test environments where<br>adults or infant patients<br>require the documentation<br>of various sleep or other<br>physiological disorders. | Neurolink can be used in<br>hospital environment and<br>clinics. | | Environmental<br>Conditions | Normal: +5 to +40°C, max<br>80% rH non-condensing,<br>700-1060hPa | Not known | Normal: +5 to +40°C, max<br>80% rH non-condensing,<br>700-1060hPa | | Data input types | ECG, neurological, respiratory | ECG, neurological,<br>respiratory | Neurological, othe<br>physiological | | No. of AC<br>Channels | 32 or 64 (neurological or<br>physiological) | 26 neurological, 10<br>physiological | 64 neurological | | AD sample rate | 32,768 | 2000 | 32,768 | | Output Sample<br>Rate | 1024 | 2000 | 1024 | | Storage rate | 1024 | 200 | 1024 | | Digital resolution | 16bits | 16bits | 16bits | | Oximeter channel | Yes | Yes | Yes | | No. of DC<br>Channels | 8 (for patient safe sensors<br>like body position) for 1142<br>model. | 12 | None | | Connection to<br>patient | By means of sensors like EEG,<br>ECG, EMG electrodes to AC<br>inputs In addition, some of<br>the AC inputs are used to<br>connect to external patient<br>safe sensors like respiration | By means of sensors like<br>EEG, ECG, EMG electrodes to<br>AC inputs.<br>In addition, some of the AC<br>inputs are used to connect to<br>external patient safe sensors<br>like respiration. | By means of sensors like<br>EEG, electrodes to AC inputs. | | Connections for<br>42 channel<br>amplifier | 32 AC connections for<br>EEG/EOG/EMG/ECG. Active<br>sensors can be connected to<br>AC channels<br>8 DC channels for other<br>physiological sensors<br>Oximeter<br>Event | 26AC channels for<br>EEG/EOG/EMG/ECG. Active<br>sensors can be connected to<br>AC channels<br>10 DC channels for other<br>physiological signals<br>Oximeter<br>Event | Not applicable<br>-no-<br>Oximeter<br>Event | | Connections for<br>64 AC channel | 64 AC connections for<br>EEG/EOG/EMG/ECG. Active<br>sensors can be connected to<br>AC channels<br>NO DC channels<br>Oximeter | Not Applicable<br>Not Applicable<br>Oximeter | 64 AC connections for EEG<br>NO DC channels<br>Oximeter | | List of | | | | | Characteristics | Galaxy | Alice 5 | Neurolink IP Model | | | | | PK 1117 | | 510 (K) ID | K111055 | K040595 | K100683 | | Main amplifier<br>unit | JUPITER: Separate Brainbox<br>EEG-1166 for 64channel AC,<br>oximetery and event<br>This can be used in<br>neurological disorders during<br>sleep. | None | Brainbox EEG1166 for 64<br>channel AC, oximetery and<br>event | | Main amplifier<br>unit | Brainbox EEG1142 for 32<br>Channel AC and 8 Channel<br>DC, oximetery and event | Base Station | Not supported | | Patient<br>connection box | 1142 or 1166 input box | Patient headbox…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...