NIHON KOHDEN EEG-1200A SERIES NEUROFAX, MODEL EEG-1200A

K080546 · Nihon Kohden America, Inc. · OLT · Aug 29, 2008 · Neurology

Device Facts

Record IDK080546
Device NameNIHON KOHDEN EEG-1200A SERIES NEUROFAX, MODEL EEG-1200A
ApplicantNihon Kohden America, Inc.
Product CodeOLT · Neurology
Decision DateAug 29, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1400
Device ClassClass 2

Indications for Use

The EEG-1200A Series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in Sleep Disorder, Epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic or nursing home or outside of a medical facility under direct supervision of a medical professional.

Device Story

EEG-1200A Series Neurofax records, measures, and displays cerebral and extracerebral activity; inputs include EEG signals, ECG, EMG, respiration, ocular motility, SpO2, and CO2. Device transforms raw physiological signals into waveforms via amplifier and montage settings; data can be stored on electronic media, printed, or analyzed using integrated software. Operated by medical personnel in medical facilities, clinics, or nursing homes. Output displayed on screen for clinician review; aids in diagnosis of epilepsy and sleep disorders. Benefits include improved diagnostic capability through multi-parameter monitoring and digital data management.

Clinical Evidence

No clinical data. Bench testing only. Device underwent electromagnetic, environmental, safety, and performance testing. Software validation confirmed acquisition, processing, display, and recording functions met specifications.

Technological Characteristics

Electroencephalograph; 32-channel capacity; frequency response 0.08 to 300 Hz; noise level < 1.5uV p-p (0.53 to 60 Hz); DC standard high-pass filter. Power: 120V +/- 10% 50/60Hz. Non-sterile. Uses standard electrodes/sensors. Software-based signal processing and analysis.

Indications for Use

Indicated for patients requiring EEG or sleep studies to aid in the diagnosis of sleep disorders, epilepsies, and related neurological conditions. Intended for use by medical personnel in clinical, laboratory, or home settings under professional supervision.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # NIHON KOHDEN AMERICA, INC. ### SECTION 2- 510(K) SUMMARY ## Name and Address of Applicant Nihon Kohden America, Inc. 90 Icon Street Foothill Ranch, CA 92610 AUG 2 9 2008 Contact: Jack Coggan Director, Regulatory Affairs (949) 580-1555 ex. 3325 Fax: (949) 580-1550 ### Trade/Device Name: EEG-1200A Series Neurofax ### Common or Usual Name: Electroencephalograph (EEG) ### Classification Name: The device has been classified as Class 2 by the Neurology Device Classification Panel under 21 CFR Part 882.1400 Electroencephalograph per GWQ, OLLT C C V ### Legally Marketed Predicate Device: Nihon Kohden EEG-1100A Series and Accessories per 510(k) K992742 commercial distribution certification dated October 14, 1999. #### Intended Use: The EEG-1200A Series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in Sleep Disorder, Epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic or nursing home or outside of a medical facility under direct supervision of a medical professional. A summary of the technological characteristics of the device compared to the predicate device: The new EEG-1200A Series Neurofax technological characteristics differences compared to the predicate device are: | Comparison | New Device | Predicate Device | |----------------------------|----------------------------------|----------------------------------| | Number of channels | 32 | 40 | | Noise Level | < 1.5uV p-p(0.53 to 60 Hz) | < 3 uV p-p (0.53 to 60 Hz) | | Frequency Response | 0.08 to 300 Hz | 0.016 to 300 Hz | | High-pass Filter (Low-cut) | DC Standard | DC Option | | Power | 120 V +/- 10% 50/60 Hz<br>750 VA | 117 V +/- 10% 50/60 Hz<br>550 VA | All other characteristics are of the device are the same. {1}------------------------------------------------ ## 510(k) Summary: - The device is not sterile. ● - . The EEG-1200A Electroencephalograph is the instrument to provide the information for intervention to utilize derivation, recording, analysis, or each combination of the brain's action potential. The instrument can measure the vital signal (including ECG waveform, EMG waveform, respiration waveform, ocular motility, SpO2 and CO2) in relation to EEG examination and display the waveform on the screen changing the montage and amplifier conditions. In addition, measurement data is available to file into the electric media. Changing the montage and amplifier conditions, then record to the printer, previews the measured data filed into the electric media. EEG data analysis is also executed to use the analysis software. - The device does not directly contact patients. Accessories that contact patients, ● such as electrodes, are the same accessories as used with the predicate or are comprised of the same component material with the same design and manufacturing processes as the predicate accessories. The device may also use commercially available electrode and sensor products. Therefore, good laboratory practice studies were not required per 21 CFR Part 58. - . The EEG-1200A Series was subjected to electromagnetic, environmental, safety and performance testing procedures. These tests verified the operation of the device. Software validation tested the operation of the software functions for acquiring, processing, displaying and recording of all functions of the device. The results confirmed that the device performed within specifications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with a bird-like figure intertwined. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is simple and recognizable, representing the department's role in public health and welfare. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Nihon Kohden America % Mr. Jack Coggan Director of Regulatory Affairs/Quality Assurance 90 Icon Street Foothill Ranch, California 92610 APR - 9 2012 Re: K080546 Trade/Device Name: EEG-1200A Series Neurofax Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, OLV, GWQ Dated (Date on orig SE ltr): July 29, 2008 Received: (Date on orig SE ltr): July 30, 2008 Dear Mr. Coggan: This letter corrects our substantially equivalent letter of August 29, 2008 We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Kesia Alexander Image /page/3/Picture/8 description: The image contains a handwritten word "for" in cursive script. The letter 'f' is elongated and loops downwards, while the 'o' and 'r' are connected with a smooth curve. The writing appears to be done with a pen or marker, and the background is plain white. Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ NIHON KOHDEN AMERICA, INC. G. Indications for Use Statement: 510(k) Number (if known): KOSO546 Device Name: EEG-1200A Series Neurofax Indication of Use: The EEG-1200A Series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the oliar inn in Sleep Disorder, Epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic or nursing home or outside of a medical facility under direct supervision of a medical professional. Mark M. Miller (Division Sign-(Division Siga-o-al, Restorative, Division of Neurological Devices K080596 510(k) Number Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over The Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Page 8 of 22
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...