Mediracer NCS

K190536 · Mediracer OY · JXE · Sep 25, 2019 · Neurology

Device Facts

Record IDK190536
Device NameMediracer NCS
ApplicantMediracer OY
Product CodeJXE · Neurology
Decision DateSep 25, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1550
Device ClassClass 2

Indications for Use

The Mediracer NCS is intended to measure sensory and motor nerve conduction from peripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician.

Device Story

Hand-held nerve conduction study (NCS) device; stimulates peripheral nerves via skin electrodes; records nerve responses; transmits data via Bluetooth to computer running Mediracer Analysis Center (MAC) software. Used in point-of-care settings by physicians or technicians under physician order. Device applies dual-phase rectangular wave electrical stimulus; records resulting nerve action potentials. MAC software stores/handles measurement data; allows cursor placement adjustment for latency analysis. Physician reviews output to determine if findings are abnormal and grade severity of nerve entrapment. Non-invasive; aids in clinical evaluation of focal neuropathies; provides objective latency data independent of skin temperature.

Clinical Evidence

Clinical validation conducted in two studies (Finland, n=194 patients/95 controls; UK, n=63 patients). Compared Mediracer NCS against traditional NCS. Results showed positive percent agreement ~90% and negative percent agreement 99-100%. High concordance in latency difference parameters observed. No adverse effects reported. Bench testing on 7 subjects confirmed results matching conventional EMG devices within deviation limits.

Technological Characteristics

Hand-held device; NiMH battery (6V, 730 mAh); Bluetooth connectivity. Stimulus: dual-phase rectangular wave (max 163V, 2Hz). Recording: 1 channel, 10-bit A/D resolution, 14.6 kHz sampling, 1.0 µV RMS noise, 95-100 dB CMRR. Electrodes: stimulation ring and recording electrode with conductive hydrogel (ISO 10993 compliant). Software: MAC analysis software. Standards: IEC 60601-1, IEC 60601-2-40, EN 60601-1-2.

Indications for Use

Indicated for patients suspected of having focal neuropathies, including Carpal Tunnel Syndrome (median nerve entrapment) and ulnar nerve entrapment at the elbow. Used for measuring sensory and motor nerve conduction from peripheral nerves.

Regulatory Classification

Identification

A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient's peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. Mediracer Ov Lassi Laitinen Product Specialist R&D Rantakatu 4 Oulu, 90100 FI Re: K190536 Trade/Device Name: Mediracer NCS Regulation Number: 21 CFR 882.1550 Regulation Name: Nerve Conduction Velocity Measurement Device Regulatory Class: Class II Product Code: JXE Dated: August 26, 2019 Received: August 30, 2019 Dear Lassi Laitinen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190536 Device Name Mediracer NCS Indications for Use (Describe) The Mediracer NCS is intended to measure sensory and motor neripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------------------------| | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for a company called "mediracer". The logo consists of a pink abstract shape on the left, followed by the word "mediracer" in black and gray letters. Below the word "mediracer" is the phrase "POINT OF CARE DIAGNOSTIC TESTING" in smaller, gray letters. The logo is simple and modern, and the colors are eye-catching. 510(K) SUMMARY | Submission type: | Traditional 510(k) | |-----------------------|-----------------------| | Basis for submission: | New device introduced | | Submitter information | | | Sponsor: | Mediracer Oy (Ltd.) | | Address: | Rantakatu 4 | | | 90100 Oulu | | | Finland | | Contact person: | Georg Meissner, CEO | | Phone: | +358 40 517 0182 | | Submission date: | 19/09/2019 | Device information | Trade name: | Mediracer NCS | |----------------------|-----------------------------------| | Common name: | Nerve Conduction Testing System | | Classification name: | Nerve Conduction Study Measuring | | Device Class: | Device Class II (21 CFR 882.1550) | | Product Code: | JXE | | Predicate device: | | | NeuroMetrix ADVANCE | K070109 | ### 1 Device Description The Mediracer NCS is intended to measure sensory and motor nerve conduction from peripheral nerves, when there is reason to suspect nerve damage. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. Current application allows examination of Carpal Tunnel Syndrome related median nerve entrapment and ulnar nerve entrapment at the elbow. Test results can be used to determine if the findings are abnormal and in CTS, grade the severity of the nerve entrapment. The Mediracer NCS is a hand-held NCS Device to stimulate and record the nerve responses from patient. The NCS Device is connected to the disposable surface electrodes with cables provided with the NCS Device. The NCS Device communicates via Bluetooth with a computer which is using The Mediracer Analysis Center (MAC) software. The computer {4}------------------------------------------------ uses the MAC software to store and handle the patient measurement data. With the NCS Device comes a charger and a docking station for charging the NCS Device. Also, solution includes a test module for checking the functionality of the whole test system. Mediracer NCS is used by physician or by technician on the order of physician for measuring sensory or motor nerve conductions from distal nerves. The measured data is transferred during the measurement to a computer via Bluetooth connection. The data measured is reviewed and interpreted by a specially trained physician or specialist in neurophysiology. ## 2 Intended Use The Mediracer NCS is intended to measure sensory and motor nerve conduction from peripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician. ### 3 Comparison of technological characteristics with the predicate device Mediracer NCS which is the subject of this Premarket Notification is substantially equivalent in sensory nerve conduction studies to the predicate device. NeuroMetrix ADVANCE. All the other functions of the predicate device are ignored in the comparison. Neuromuscular signal measurement is the technological principle for both the subject and predicate devices. It is based on stimulating the nerve with electric stimulus on skin above the nerve and the activated nerve response is measured on skin from the same nerve at some distance of the stimulus. Difference between the devices are mainly in the design and the technological specifications due to the design. None of the different features of Mediracer NCS compared to predicate device have altered the basic performance, operation or intended use. At high level, the subject and predicate devices are based on the following same technological elements: - stimulating and recording electrodes - stimulation and recording cables - hand-held, point-of-care, device that performs the measurement - . Bluetooth for data transfer The following technological differences exist between the subject and predicate devices: - Electrodes differ in design and placement. ● - The cables differ in design, the predicate uses the Proximal adapter in which the stimulation cable and recording cable are connected whereas the subject device has two separate cables which are connected straight to the device. - . Hand-held device differs from the display of data, the predicate uses touch screen for display of data whereas the subject device uses small LCD screen. - . The subject device must be used in conjunction with the Mediracer Analysis Center (MAC) whereas predicate device can be used without any external software. {5}------------------------------------------------ Table 1. Device Comparison | | Advance NCS | Mediracer NCS | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | Device for measuring<br>neuromuscular signals that<br>are useful as an aid in<br>diagnosing and evaluating<br>patients suspected of<br>having focal or systemic<br>neuropathies. | Device for stimulate and<br>measure neuromuscular<br>signals that are useful in<br>diagnosing and evaluating<br>systemic and entrapment<br>neuropathies. | | Power supply | High Capacity Lithium Ion<br>Battery | Battery: Nickel Metal<br>Hybrid NiHM battery 6V,<br>730 mAh | | Common Mode Rejection<br>Ratio | 90 dB min | 95...100 dB at 1 kHz | | Noise | <2µV RMS | 1.0 µV RMS (from 2 Hz to<br>10 kHz) | | Sampling Frequency | 20 kHz | 14.6 kHz | | A/D Resolution | 24 bits | 10 bits | | Stimulus Type | Monophasic or biphasic<br>square wave | Dual phase, rectangular<br>wave | | Stimulator Max Voltage | 420 V | 163 V | | Stimulation Frequency | 2 Hz | 2 Hz | | Recording Channels | 1 | 1 | | Connectivity | Bluetooth | Bluetooth | | Latencies | Motor and sensory<br>latencies | Motor and sensory<br>latencies | | Dimensions | 18.034 x 8.128 cm | 18.99 x 6.971 cm | | Nerves Tested | Median and ulnar nerves | Median and ulnar nerves | | Test Signal Control<br>During Test | Automatic or manual<br>adjustment depending on<br>the used electrodes | Manual adjustment | | Assessment time | 10 minutes | 15 minutes (both hands) | | Access to test data<br>adjustment | Cursor placing adjustment | Cursor placing adjustment | | Electrodes | Nerve specific wrist and<br>digit electrodes or standard<br>electrodes (also elbow and<br>foot electrodes available) | Stimulation Ring Electrode<br>and Recording Electrode | | Patient Data Handling<br>and Storage | On the Device. All patient<br>and test data included. Can<br>be uploaded to a Data<br>Server. | On the software included<br>on the system. All patient<br>and test data included. | | Operator Skill<br>Requirements | Trained technician<br>supervised by a physician<br>or a physician | Trained technician<br>supervised by a physician<br>or a physician | | Interpretation of the Test<br>Results | Physician or specialist<br>locally | Physician or specialist<br>locally or remotely | | Temperature | Temperature integrated sensor | Temperature measurement<br>with a separate device | | Regulations | CE -Marked; 60601-1<br>compliance | CE -Marked; 60601-1<br>compliance | {6}------------------------------------------------ ### 4 Performance data The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing The Mediracer NCS component relevant for biocompatibility evaluation is the Electrode conductive hydrogel, which is in direct contact of patient skin while conducting the NCS test. Conductive hydrogel was assessed according to International Organization of Standardization 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. Performance Testing includes the biocompatibility test, bench test and clinical tests. Detailed descriptions of the biocompatibility test, bench test and clinical tests are included in the submission. The testing included the following tests: - Cytotoxicity ● - Irritation - Sensitization The hydrogel passed all above ISO 10993 testing. ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Mediracer NCS device, consisting of the NCS device (ENG-CC-02), patient stimulator (PS-01), stimulation cable (SC-01) and recording cable (CC-02). The system complies with the IEC 60601-1, IEC 60601-2-40 standards for safety and the EN 60601-1-21 standard for EMC. ## Software Verification and Validation Testing Software verification and validation testing for the Mediracer NCS and MAC, were conducted, and documentation was provided as recommended by FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software of Mediracer NCS device and the MAC software was considered "Moderate" level of concern, since there is a risk of minor adverse events associated with Mediracer NCS and the MAC software is an accessory to a Medical Device (Mediracer NCS). <sup>4</sup> The European version (60601-1-2:2015) of the standard and the IEC-version (60601-1-2:2014) are equivalent to each other. In the European version the differences to its IEC counterpart are the exception of references of the 61000-4-x series and the addition of an Essential Requirements annex. The technical content of the IEC version is identical to the EN version. {7}------------------------------------------------ ## Performance testing Performance testing as a bench-testing, was conducted to the subject device to show equivalence to a previously approved device and to verify Mediracer NCS device accuracy. The performed test was same for both devices and the testing was performed by a trained operator. The testing was performed to seven people, which six was within the measurement criteria set before testing. The six results taken into account with proper placement show significantly similar nerve response action potential values. The results show that the Mediracer NCS device results compared to conventional EMG device were matching inside the given deviation. Based on this comparison test it can be said that Mediracer NCS device is equivalent to a conventional EMG device and shows accurate readings. ## Clinical Studies The clinical validation was made against traditional nerve conduction test devices in two separate studies; one in Finland and the other in the United Kingdom. All the subjects examined in these tests had AANEM 2002 recommended set of traditional nerve conduction studies (NCS) and measurement with the Mediracer NCS done. In the Finnish multicenter study, there were 194 recruited patients and 95 healthy control volunteers tested. The participants went through a clinical examination, Mediracer NCS measurement and the traditional NCS. In the United Kingdom study, the devices were compared only in the patient population, total of 63 patients participated in the study. The participants went through clinical assessment, Mediracer NCS measurement and the traditional NCS. In the Finnish multicenter study, the number of patients studied were 194, aged 17-85. mean 49 years. Of these patients 147 (75.77 %) were female. Exclusion criteria for the study were prior surgery for CTS (the measured hand), neurological disorder that may produce numbness or paraesthesias in hand and those with diagnosed diabetes mellitus. For the study control values of 95 healthy volunteers aged from 16-80, mean 41 years, were examined. Of these volunteers 52 (54.73 %) were female. In the United Kingdom the patients were recruited from the normal referral stream to Leicester Carpal Tunnel Service, total of 104 referrals were selected and invited to a clinical study. No other exclusion criteria. Of the invited patients 65 fulfilled the clinical CTS diagnosis after symptom inquiry and clinical assessment. On the day of the study two participants cancelled their involvement, thus the total participation number were 63 patients. Of these patients 47 (74.60 %) were female. ## Safety and effectiveness In both studies positive per cent agreement between the traditional and the Mediracer NCS devices was around 90 % and negative per cent agreement 99-100 %. The new device study missed abnormalities mainly in cases with very mild conduction slowing and due to technical (one broken cable) or biological causes (patient was unable to relax). A high concordance between the used latency difference parameters between the traditional and the new device study were obtained. Moreover, these parameters were not dependent on skin temperature as absolute conduction velocities clearly are. The method is non-invasive and is safe to preform for the patients, no adverse effects or complications were reported during these studies. ## Summary {8}------------------------------------------------ Based on the two studies performed, the Mediracer NCS measurement device seems to reliably detect the median nerve lesion in CTS. In the studies were found high degree of agreement between the systems compared, the new and traditional methods of NCS testing. ### 5 Conclusions The non-clinical data support the safety of the device and the Mediracer NCS and MAC software verification and validation demonstrate that the Mediracer NCS device should perform as intended in the normal use conditions. The clinical data and bench-testing demonstrates that nerve conduction measurements obtained using the Mediracer NCS are comparable to those obtained using conventional nerve conduction measurement equipment.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...