K993202 · General Devices · GZO · Dec 17, 1999 · Neurology
Device Facts
Record ID
K993202
Device Name
EIM-107 PREP-CHECK
Applicant
General Devices
Product Code
GZO · Neurology
Decision Date
Dec 17, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1540
Device Class
Class 2
Indications for Use
The EIM-107 is indicated for use whenever it is desired to measure or assess the impedance and/or offset voltage of surface biopotential electrodes applied to the skin of a patient.
Device Story
EIM-107 Prep-Check Electrode Impedance Meter; measures impedance and offset voltage of surface biopotential electrodes. Used by clinicians to verify electrode-skin interface quality. Input: electrical signals from surface electrodes. Output: impedance/voltage measurements. Assists in ensuring signal integrity for biopotential monitoring; helps optimize electrode placement and skin preparation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Electrode impedance and offset voltage measurement device. Class II, Product Code IKJ. Operates as a diagnostic tool for biopotential electrode assessment.
Indications for Use
Indicated for measuring or assessing impedance and/or offset voltage of surface biopotential electrodes applied to patient skin.
Regulatory Classification
Identification
A galvanic skin response measurement device is a device used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9.
Related Devices
K980410 — SLE ELECTRODES AND ACCESSORIES · Specialized Laboratory Equipment, Ltd. · Oct 22, 1998
K022197 — SURFACE ELECTRODE FOR ELECTROENCEPHALOGRAPHY · The Electrode Store, Inc. · Aug 13, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
DEC 1 7 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Michael Smith President General Devices 1000 River Street Ridgefield. New Jersey 07657
Re: K993202
Trade Name: EIM-107 Prep-Check Electrode Impedance Meter Regulatory Class: II Product Code: IKJ Dated: September 22, 1999 Received: September 24, 1999
Dear Mr. Smith:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{1}------------------------------------------------
Page 2 - Mr. Michael Smith
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Kleine Zimme
& - James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
PMN 510(k) Number: K 993202
Device Name: EIM-107 Prep-Check Electrode Impedance Meter
## Indications For Use:
The EIM-107 is indicated for use whenever it is desired to measure or assess the impedance and/or offset voltage of surface biopotential electrodes applied to the skin of a patient.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
und "June fy J 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / 2 / r Division Sign-Off) ision of General Restorative Devices 16993003 510(k) Number
Prescription Use ర్లు (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.