Anchor Bolt (as an accessory to Depth Electrodes)

K181544 · Ad-Tech Medical Instrument Corporation · GZL · Aug 9, 2018 · Neurology

Device Facts

Record IDK181544
Device NameAnchor Bolt (as an accessory to Depth Electrodes)
ApplicantAd-Tech Medical Instrument Corporation
Product CodeGZL · Neurology
Decision DateAug 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1330
Device ClassClass 2

Indications for Use

The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.

Device Story

Anchor Bolts are mechanical accessories for Depth Electrodes; used in intraoperative and neurological monitoring settings. Device is threaded into a pre-drilled hole in the skull; provides a secure access point for electrode insertion; stabilizes electrode position; includes an inner lumen silicone gasket to minimize CSF leakage. Operated by physicians/surgeons; requires a placement/removal wrench. Output is physical stabilization of the electrode; benefits patient by reducing infection risk and CSF leakage during intracranial monitoring. Device is disposable; single-patient use.

Clinical Evidence

Bench testing only. Performance verified via torque insertion tests, axial tension pull tests (anchor bolt from skull and electrode from anchor bolt), and CSF leakage resistance testing at 15 mmHg hydrostatic pressure. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, genotoxicity, hemocompatibility, neurotoxicity, and implantation testing per ISO standards.

Technological Characteristics

Materials: Titanium, Silicone (inner lumen gasket), Parylene. Dimensions: 13-26 mm length; compatible with 0.86-1.3 mm diameter electrodes. Mechanical fixation via threaded screw into pre-drilled skull hole. Sterilization: Ethylene Oxide (EtO) per ANSI/AAMI/ISO 11135-1:2007. Single-patient use, disposable.

Indications for Use

Indicated for use as an accessory to Depth Electrodes to provide an access point through the skull and stabilize the electrode while minimizing CSF leakage and infection risk. Contraindicated in patients with skull softening, low skull bone density, or scalp infection.

Regulatory Classification

Identification

A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, which features a stylized depiction of human profiles. To the right of this symbol, there is the FDA logo, which includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. The text is arranged in two lines, with 'U.S. FOOD & DRUG' on the top line and 'ADMINISTRATION' on the bottom line. August 9, 2018 Ad-Tech Medical Instrument Corporation Gary Syring Principal Consultant Quality and Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589 Re: K181544 Trade/Device Name: Anchor bolt (as an accessory to Depth Electrodes) Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: June 8, 2018 Received: June 12, 2018 Dear Mr. Syring: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jay R. Gupta -S For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181544 Device Name Anchor Bolts as Accessories to Depth Electrodes Indications for Use (Describe) The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes. Type of Use (Select one or both, as applicable) | <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☑ </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |---------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☐ </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 5.0 510(k) Summary This summary is provided to support the 510(k) pre-market notification for the Anchor Bolts, as accessories to Depth Electrodes. | Company Name: | Ad-Tech Medical Instrument Corporation<br>400 West Oakview Parkway<br>Oak Creek, WI 53154<br>Phone: (262) 634-1555 | |------------------------|--------------------------------------------------------------------------------------------------------------------| | Company Contact: | Lisa Theama, Chief Operating Officer | | Date Summary Prepared: | June 8, 2018 | | Trade Name: | Anchor Bolts | | Common Name: | Anchor Bolt (Accessory to Depth Electrodes) | | Classification Name: | Depth Electrode<br>21 CFR 882.1330<br>Product Code: GZL<br>Class II | | Predicate Device: | K970418 Anchor Bolt, Accessory Depthalon® Depth Electrode<br>PMT Corporation | #### Product Description 5.1 The device under review is a family of Anchor Bolts. Anchor Bolts are optional accessories to Depth Electrodes. These Anchor Bolts provide an optional access point through the skull and stabilization support for Depth Electrodes. #### 5.2 Intended Use of the Device The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes. #### 5.3 Summary of Technological Characteristics The following table provides a side-by-side comparison of the Anchor Bolts to the predicate device applied to support this pre-market notification. {4}------------------------------------------------ | Table 5.3-1: Substantial Equivalence Technical Characteristics | | | | |----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | Anchor Bolts<br>(Under Review) | Anchor Bolt, Accessory<br>Depthalon® Depth Electrode<br>(Predicate K970418) | Comment | | Indications<br>for Use | The Ad-Tech Anchor Bolts are<br>optional accessories for use with<br>Depth Electrodes. The Anchor Bolts<br>may be applied when it is desired to<br>minimize concerns about potential<br>cerebrospinal fluid (CSF) leakage and<br>infection of the subdural space while<br>stabilizing the electrode.<br>Anchor Bolts are secured in the skull<br>to provide an access point for Depth<br>Electrodes. | The PMT depth electrode anchor<br>bolt is an accessory used in cases<br>where it is necessary to minimize<br>concerns about potential<br>cerebrospinal fluid leakage and<br>infection of the subdural space will<br>stabilizing the electrode during<br>intraoperative electro clinical<br>characterization. The PMT depth<br>electrode anchor bolts rest on the<br>skull with the insertion of the<br>anchor bolt tapered screw threads<br>into a predrilled burr hole through<br>a skin incision. | Same.<br>Both Anchor Bolts<br>are applied to a<br>pre-drilled hole in<br>the skull. The<br>Anchor Bolts<br>support Depth<br>Electrode<br>placement. | | Clinical<br>Application | Threaded into a pre-drilled hole in the<br>skull. | Threaded into a pre-drilled hole in the<br>skull. | Same | | Duration of<br>use | < 30 days | < 30 days | Same | | Contra-<br>indications | Anchor Bolts should not be used on<br>any patient whom the physician/<br>surgeon considers at risk for infection<br>or on whom the use cannot be<br>performed safely.<br><br>The Anchor Bolt should not be used<br>with patients that have softening of<br>the skull or low skull bone density. | Depthalon® Depth Electrodes<br>should not be used in the presence<br>of any infection of the scalp. Other<br>infections may be considered a<br>contraindication if determined by<br>the attending physician.<br><br>The Anchor bolt accessory should<br>not be used if the patient has a<br>softening of the skull or low skull<br>bone density. | Equivalent<br>Anchor Bolts are<br>selected and<br>applied by a<br>physician<br>knowledgeable in<br>their use. | | Anchor Bolt<br>Single patient<br>use,<br>Disposable | Yes | Yes | Same | | Provided<br>Sterile | Yes<br>(Anchor Bolts provided sterile,<br>optional for Placement/Removal<br>Wrench ) | Yes<br>(Anchor bolts can be provided<br>sterile or non-sterile) | Equivalent | | User<br>Sterilizable | Yes<br>(Placement/Removal Wrench only) | Yes<br>(Anchor Bolts supplied non-sterile) | As a convenience<br>to the user, the<br>Placement /<br>Removal Wrench<br>can be re-<br>sterilized. | | Environment<br>of Use | Intraoperative and Neurological<br>monitoring locations | Intraoperative and Neurological<br>monitoring locations | Same | | Duration of<br>Use | < 30 days | < 30 days | Same | | Patient<br>contact<br>material | Titanium<br>Silicone (inner lumen gasket)<br>Parylene | Titanium Alloy<br>Silicone (inner lumen gasket) | Equivalent | | Length | 13 mm to 26 mm | 25 mm to 40 mm | Equivalent | | Table 5.3-1: Substantial Equivalence Technical Characteristics | | | | | Feature | Anchor Bolts<br>(Under Review) | Anchor Bolt, Accessory<br>Depthalon® Depth Electrode<br>(Predicate K970418) | Comment | | Compatible<br>Depth<br>Electrode<br>Body<br>Diameter | 0.86 mm to 1.3 mm | 1.27 mm | Equivalent | | Depth<br>Electrode<br>Retention<br>Force | > 100 grams | > 100 grams | Same | | Placement /<br>Removal<br>Wrench | Yes | Unknown | Given the unique<br>design of Anchor<br>Bolts it is very<br>likely that the<br>predicate device<br>requires a<br>placement and<br>removal tool. Hand<br>tightening an<br>Anchor Bolt into a<br>pre-drilled skull<br>hole is not<br>probable. | | | | | The length of the<br>Anchor Bolt is<br>based upon need<br>and user<br>preferences. | {5}------------------------------------------------ ### 5.4 Non-clinical Performance Tests to Demonstrate Substantial Equivalency To establish the technical equivalency of the Anchor Bolts, evaluations were conducted to confirm compliance with performance requirements, including: | Test | Test Method Summary | Result | |-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------| | Torque required to insert<br>to pre-drilled skull hole. | With a torque wrench, verify the torque required to insert<br>the Anchor Bolt into pre-drilled hole in 40 pcf polyurethane<br>foam (simulated skull bone) with the Placement/Removal<br>Wrench is ≤ 8 in-lbs. (N=10 samples). | Pass<br>All results ≤<br>1 in-lbs. | | Axial tension to pull<br>Anchor Bolt out of pre-<br>drilled skull hole. | With a tensile tester, measure the axial tension to pull the<br>Anchor Bolt out of a pre-drilled hole in 40 pcf polyurethane<br>foam is ≥ 40 lbf. (N=10 samples) | Pass<br>All results<br>> 48 lbf. | | Axial tension to pull<br>Depth Electrode from the<br>Anchor Bolt. | With a tensile tester, measure the axial tension to pull a<br>Depth Electrode from the Anchor Bolt, ≥ 100 grams. (N=10<br>samples). | Pass<br>All results<br>> 140<br>grams. | | Anchor Bolt gasket resist<br>CSF leakage. | Insert Depth Electrode into the Anchor Bolt secured in 40<br>pcf polyurethane foam. Apply a metal tube with a diameter<br>larger than the Depth Electrode and Anchor Bolt body,<br>around both. Seal the metal tube to the polyurethane foam.<br>Fill the metal tube with water to a hydrostatic pressure of | Pass<br>No leaks<br>observed | | Test | Test Method Summary | Result | | | 15 mmHg. Verify no leakage at the Depth Electrode,<br>Anchor Bolt after 24 hours. (N=22 samples) | after 24<br>hours. | | Sterility | Anchor Bolts and Placement/Removal Wrenches are<br>terminally ethylene oxide sterilized devices by a validated<br>EtO sterilization cycle, per ANSI/AAMI/ISO 11135-1:2007<br>half-cycle method. The sterility assurance level (SAL) is ≤<br>SAL of 10-6. | Pass | | Sterile barrier integrity | The sterile barrier package is a double Tyvek pouch. The<br>methods applied to evaluate the Tyvek sterile barrier<br>package integrity included post terminal sterilization:<br>simulated distribution, seal peel and bubble emission<br>testing. | Pass | | Anchor Bolt silicone<br>gasket shelf life | The ability of real time aged Anchor Bolts with Silicone<br>Gaskets to meet Depth Electrode retention requirements<br>and minimize cerebrospinal fluid (CSF) leakage were<br>evaluated to the methods above for Axial tension to pull<br>Depth Electrode from the Anchor Bolt and Anchor Bolt<br>gasket resist CSF leakage. | Pass<br>The same<br>acceptance<br>criteria were<br>met. | | Biocompatibility | The following test methods were applied to final finished<br>form Anchor Bolts (patient contacting materials), with<br>passing results:<br>Cytotoxicity: L929 MEM Elution Test – ISO Sensitization: Kligman Maximization Test - ISO Intracutaneous Reactivity: Intracutaneous Injection Test - ISO Acute Systemic Toxicity: 28-Day Systemic Toxicity in Rabbits via Subcutaneous Implantation Material-Mediated Pyrogenicity: Rabbit Pyrogen Test (Material Mediated) - ISO Genotoxicity: Salmonella Typhimurium and Escherichia Coli Reverse Mutation Assay - ISO and Mouse Lymphoma Mutagenesis Assay - ISO Hemocompatibility: Rabbit Blood Hemolysis Test (Indirect Contact) – ASTM Neurotoxicity by: SK-N-MC MEM Elution Test. Subacute / Subchronic Toxicity: Systemic Injection Test - ISO Implantation: Neurological (Brain) Implantation Test - ISO. | Pass | {6}------------------------------------------------ ### 5.5 Conclusion The Anchor Bolts meet performance requirements equivalent to the predicate device. The intended use and technology of the Anchor Bolts are the same as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...