AURAGEN CORTICAL SURFACE ELECTRODES

K082250 · Integra Neurosciences · GYC · Oct 10, 2008 · Neurology

Device Facts

Record IDK082250
Device NameAURAGEN CORTICAL SURFACE ELECTRODES
ApplicantIntegra Neurosciences
Product CodeGYC · Neurology
Decision DateOct 10, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1310
Device ClassClass 2

Indications for Use

The cortical surface electrodes are intended for the intraoperative recording of EEG signals at the cortical surface of the brain.

Device Story

Auragen Cortical Surface Electrodes consist of platinum/iridium contact arrays (4 to 64 contacts) designed for intraoperative monitoring. Electrodes are placed on exposed cortical surface or inserted into subdural space to capture electrical activity. Device connects to Integra recording cables and hospital EEG equipment. Used by neurosurgeons/clinicians during surgery to identify epileptogenic foci. Output consists of raw EEG signals displayed on hospital monitoring equipment, enabling real-time clinical decision-making regarding surgical resection or mapping. Benefits include precise localization of seizure-generating brain tissue.

Clinical Evidence

Bench testing only. Testing included MRI safety assessments (RF heating, displacement force, torque, and artifact evaluation) and dimensional, physical, and electrical performance testing. Device is sterile and non-pyrogenic.

Technological Characteristics

Platinum/iridium contacts; 4 to 64 contact configurations; MR Conditional per ASTM F 2503-05; sterile; non-pyrogenic; passive electrode array; connects to external EEG recording equipment.

Indications for Use

Indicated for intraoperative recording of EEG signals at the cortical surface of the brain to define the location of epileptogenic foci in patients undergoing neurosurgical procedures.

Regulatory Classification

Identification

A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KOT2STO # OCT 1 0 2008 # Auragen™ Cortical Surface Electrodes 510(k) SUMMARY ## Submitter's name and address: Integra NeuroSciences Implants SA 2905 Route des Dolines 06921 Sophia Antipolis Cedex, France ## Contact person and telephone number: Anne Bigeard, Regulatory Affairs Manager Telephone: +33 (0)4 93 95 5667 Facsimile: +33 (0)4 93 65 4030 ## Date summary was prepared: September 22, 2008 ## Name of the device: | Proprietary Name: | Cortical Surface Epilepsy Electrode | |----------------------|-------------------------------------| | Common Name: | Cortical Electrodes | | Classification Name: | Cortical Electrodes, GYC | ## Substantial Equivalence: The modified cortical surface electrodes are substantially equivalent to the cortical surface electrodes cleared by the FDA under 510(k) (K926424). The design, intended use, materials of composition, functional/performance specifications, manufacturing process are equivalent to the previously cleared cortical surface electrodes. #### Intended use: The cortical surface electrodes are intended for the intraoperative recording of EEG signals at the cortical surface of the brain. ### Device Description: The cortical surface electrodes are designed for intraoperative monitoring of cortical electrical activity in order to define the location of epileptogenic foci. They are placed on the exposed cortical surface or inserted into the subdural space to contact the cortical surface. Cortical electrodes vary in size according to the number of contacts. The type of cortical electrode used is dependent on the procedure and the size of area being tested. The Auragen Cortical Electrodes are available with platinum/iridium contacts, with 4 to 64 contacts. The electrodes are connected to Integra recording cables themselves connected to the hospital EEG recording equipment. # Safety and Effectiveness: An array of cortical electrodes were subjected to MRI testing, which included radio frequency induced heating, magnetically induced displacement force and torque. MR imaging artifacts were assessed and the artifacts that appeared were shown as localized signal voids that were considered small in size in relation to the size and shape of the product. Results of the testing supports a "MR {1}------------------------------------------------ Conditional" labeling claim as defined in ASTM F 2503-05, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. The cortical electrodes are provided sterile and are non-pyrogenic. The cortical electrodes have been tested for dimensional, physical and electrical performance and meet the defined requirements of the products specification. # Conclusion : The modified cortical electrodes defined in this submission are substantially equivalent to the unmodified cortical electrodes cleared by the FDA under 510(k) (K926424). The modifications do not affect the intended use, the fundamental scientific technology of the device, and do not raise new issues of safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is rendered in black and white, giving it a formal and official appearance. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Integra Neurosciences % Integra Lifesciences Corporation Mr. Jon Caparotta Director, Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536 # OCT 1 0 2008 Re: K082250 Trade/Device Name: Auragen Cortical Surface Electrodes Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: Class II Product Code: GYC Dated: September 22, 2008 Received: September 23, 2008 Dear Mr. Caparotta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Jon Caparotta forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Survcillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours. Mark N Wilkinson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): Device Name: Auragen™ Strip and Grid Electrodes Indications For Use: The Auragen™ Cortical Surface Electrodes are intended for the intraoperative recording of EEG signals at the cortical surface of the brain. AND/OR Over-The-Counter Use Prescription Use x (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ই (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** 1<0 82250
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...