Exacta External Drainage and Monitoring System, Exacta Pole with Laser Level, Exacta System Replacement Drainage Bag, Replacement Laser Level

K201118 · Medtronic, Inc. · GWM · May 26, 2020 · Neurology

Device Facts

Record IDK201118
Device NameExacta External Drainage and Monitoring System, Exacta Pole with Laser Level, Exacta System Replacement Drainage Bag, Replacement Laser Level
ApplicantMedtronic, Inc.
Product CodeGWM · Neurology
Decision DateMay 26, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1620
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to: 1. Reduce ICP, e.g., pre-, intra- or postoperative. 2. Monitor CSF chemistry, cytology, and physiology. 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. Monitoring of ICP is indicated in selected patients with: 1. Severe head injury. 2. Subarachnoid hemorrhage graded III, IV or V preoperatively. 3. Reyes syndrome or similar encephalopathies. 4. Hydrocephalus. 5. Intracranial hemorrhage. 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions.

Device Story

Exacta EDMS is a closed system for temporary CSF drainage and ICP monitoring from lateral ventricles or lumbar subarachnoid space. System components include a single-use drainage assembly (patient line, stopcock, graduated cylinder, drainage bag) and a reusable blue pole clamp. Gravity and intracerebral pressure drive CSF drainage; rate depends on system height relative to patient anatomy. An optional battery-operated laser level accessory assists clinicians in aligning the system to the patient's Foramen of Monro or lumbar catheter exit site. Used in clinical settings (e.g., ICU, OR) by healthcare professionals. Laser level ensures accurate leveling, which is critical for correct pressure monitoring and drainage rate control. Benefits include therapeutic ICP reduction and diagnostic monitoring of CSF physiology.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including mechanical strength, beam uniformity, laser accuracy, laser safety, and auto-shutoff verification.

Technological Characteristics

System includes single-use drainage assembly and reusable pole clamp with laser level accessory. Laser level: battery-operated, Class 3R laser (<5.0 mW, 650nm), two level bubbles. Sterilization: Ethylene Oxide. Compliance: IEC 60601-1, IEC 60601-1-2, IEC/EN 60825-1.

Indications for Use

Indicated for patients requiring CSF drainage or ICP monitoring due to severe head injury, subarachnoid hemorrhage (grade III-V), Reyes syndrome, hydrocephalus, intracranial hemorrhage, or space-occupying lesions; also for temporary drainage in infected CSF shunts.

Regulatory Classification

Identification

An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. May 26, 2020 Medtronic, Inc. Jen Correa Program Manager, Regulatory Affairs 5290 California Avenue Irvine, California 92617 Re: K201118 Device Name: Exacta External Drainage and Monitoring System (EDMS) Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM, HCA Dated: April 24, 2020 Received: April 27, 2020 Dear Jen Correa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Xiaolin Zheng -S Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use ### 510(k) Number (if known) K201118 ### Device Name Exacta External Drainage and Monitoring System (EDMS) ### Indications for Use (Describe) Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to: - 1. Reduce ICP, e.g., pre-, intra- or postoperative. - 2. Monitor CSF chemistry, cytology, and physiology. - 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. Monitoring of ICP is indicated in selected patients with: - 1. Severe head injury. - 2. Subarachnoid hemorrhage graded III, IV or V preoperatively. - 3. Reves syndrome or similar encephalopathies. - 4. Hydrocephalus. - 5. Intracranial hemorrhage. - 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary ## Medtronic Exacta External Drainage and Monitoring System (EDMS) ## 510(k) Summary May 19, 2020 - I. Medtronic Neurosurgery Company: 5290 California Ave Irvine, CA 92617 USA Contact: Jen Correa Program Manager, Regulatory Affairs Jennifer.L.Correa@medtronic.com Telephone Number: 949-297-5494 #### II. Establishment Registration Number: 3015531529 - III. Proprietary Trade Name: Exacta External Drainage and Monitoring System (EDMS) - IV. Regulator Classification: II #### V. Primary Classification Name: Intracranial pressure monitoring device Product code: GWM Regulation: 21 CFR 882.1620 ### Secondary Classification Name: Ventricular Catheter Product code: HCA Regulation: 21 CFR 882.4100 - VI. Product Description: The Exacta External Drainage and Monitoring System (EDMS) is provided as a complete closed system for the drainage and monitoring of cerebrospinal fluid (CSF) flow from the lateral ventricles or the lumbar subarachnoid space. The system is offered in various kit configurations for various clinical applications. The Exacta EDMS product family is comprised of a single use drainage system, a reusable blue pole clamp and a laser level accessory. The single use drainage assembly is comprised of a patient line, main system stopcock, graduated cylinder and drainage bag. The single use drainage assembly is mounted on the reusable blue pole clamp. The reusable blue pole clamp secures the svstem to an I.V. pole and includes the system pressure scale and holds an optional laser level accessory. The optional laser level accessory assists the user in leveling the single use drainage system to the patient's Foramen of Monro or lumbar catheter exit site. {4}------------------------------------------------ #### VII. Indications for Use: Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to: - 1. Reduce ICP, e.g., pre-, intra- or postoperative. - 2. Monitor CSF chemistry, cytology, and physiology. - 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. Monitoring of ICP is indicated in selected patients with: - 1. Severe head injury. - 2. Subarachnoid hemorrhage graded III, IV or V preoperatively. - 3. Reyes syndrome or similar encephalopathies. - 4. Hydrocephalus. - 5. Intracranial hemorrhage. - 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions. - VIII. Summary of the Technological Characteristics: The subject laser level device shares the same features and use as the predicate laser level. The rest of the Exacta EDMS device has the same technological characteristics as the predicate device. | | Subject Device | Predicate Device | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | | | K983799 | | Intended<br>use/Indications<br>for Use | Draining and monitoring of CSF flow from<br>the lateral ventricles or lumbar<br>subarachnoid space is indicated in selected<br>patients to:<br>1. Reduce ICP, e.g., pre-, intra- or<br>postoperative.<br>2. Monitor CSF chemistry, cytology, and<br>physiology.<br>3. Provide temporary CSF drainage in<br>patients with infected cerebrospinal<br>fluid shunts.<br>Monitoring of ICP is indicated in selected<br>patients with:<br>1. Severe head injury.<br>2. Subarachnoid hemorrhage graded III,<br>IV or V preoperatively.<br>3. Reyes syndrome or similar<br>encephalopathies.<br>4. Hydrocephalus. | Same | | | Subject Device | Predicate Device | | | | K983799 | | | 5. Intracranial hemorrhage. | | | | 6. Miscellaneous problems when<br>drainage is to be used as a therapeutic<br>maneuver.<br>Monitoring can also be used to evaluate the<br>status pre- and postoperatively for space-<br>occupying lesions. | | | Operating<br>Principle | External drainage is temporary drainage of<br>cerebrospinal fluid (CSF) from the lateral<br>ventricles of the brain, or the lumbar space<br>of the spine, into an external collection<br>bag. The Exacta EDMS drains CSF by<br>using a combination of gravity and<br>intracerebral pressure. The drainage rate<br>depends on the height at which the system<br>is placed relative to the patient's anatomy. | Same | | Materials | There are no materials with direct patient<br>contact | Same | | Laser Level<br>Accessory<br>Features and<br>Use | Battery-operated reusable non-sterile<br>accessory that is used to align the external<br>reference point with the zero reference<br>point of the EDMS device. | Same | | Laser Level<br>Accessory<br>Design | Level Bubbles: Two, located at top and<br>bottom of device | Level Bubbles: Same | | | Maximum Output: <5.0 mW<br>Wavelength: 650nm<br>Laser Classification: 3R | Maximum Output: <1 mW<br>Wavelength: 670nm<br>Laser Classification: 2 | | Anatomical<br>Sites | Drainage and monitoring of CSF flow from<br>the lateral ventricles or lumbar<br>subarachnoid space. | Same | | Biocompatibility | No direct patient contact<br>In contact with CSF only | Same | | Sterilization<br>Method | Ethylene Oxide | Same | | Pyrogenicity | Non pyrogenic | Same | | Shelf life | 2 yr. | Same | {5}------------------------------------------------ ### IX. Identification of Legally Marketing Devices: Medtronic PS Medical Niveau Drainage System, K983799. {6}------------------------------------------------ - X. Discussion of the Performance Testing: In accordance with the risk assessment of the change it was determined that dimensional verification, and design verification testing of the laser level was necessary. As the changes to the laser level design potentially impact EMC and electrical safety, the design verification testing was completed to demonstrate compliance to the following electrical standards: - . IEC 60601-1:2005 + AMD1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. - . IEC 60601-1-2:2014 / EN 60601-1-2:2015. Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests In addition, laser testing was completed in compliance with IEC/EN 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements. The successful results of the testing demonstrated that the changes do not raise questions of safety and effectiveness, supporting the substantial equivalence to the predicate device. | Test | Test Method Summary / Purpose | Results | |------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | Mechanical<br>Strength | Verify laser does not present a safety hazard<br>after push test, impact test and dropping. | The laser level device<br>met the acceptance<br>criteria for mechanical<br>strength. | | Beam Uniformity | Measure and record the beam uniformity to<br>determine offset | The laser level device<br>met the acceptance<br>criteria for beam<br>uniformity | | Laser Accuracy | Calculate and record the laser accuracy at a set<br>distance | The laser level device<br>met the acceptance<br>criteria for laser<br>accuracy | | Laser Safety | Verify laser power output complies with<br>exposure limits for Class 3R Laser | The laser level device<br>met the acceptance<br>criteria for laser safety | | Auto Shut Off | Measure and record the time laser is "on" after<br>activation | The laser level device<br>met the acceptance<br>criteria for auto shut off | ## Performance Data – Bench {7}------------------------------------------------ ### Performance Data - Animal The risk assessment of the proposed modifications to the laser level did not require animal testing. Determination of substantial equivalence for the design change is based upon the design verification bench testing. ## Performance Data - Clinical The risk assessment of the proposed modifications to the laser level did not require clinical testing. Determination of substantial equivalence for the design change is based upon the design verification bench testing. - XI. Conclusions: The information provided in this submission demonstrates that the subject device, the Exacta EDMS, has the same intended use/indications for use as the predicate device and the differences in technological characteristics introduced by the proposed changes to the laser level device do not raise any questions of safety and effectiveness. Based on the information provided in this submission the subject Exacta EDMS device is considered substantially equivalent to the previously cleared predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...