← Product Code [GWK](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWK) · K052489

# ACTIHEART (K052489)

_Respironics, Inc. · GWK · Sep 27, 2005 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWK/K052489

## Device Facts

- **Applicant:** Respironics, Inc.
- **Product Code:** [GWK](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWK.md)
- **Decision Date:** Sep 27, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1845
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Actiheart is an ambulatory Heart Rate and Activity Recorder. Actiheart may be used to quantifiably measure Heart Rate, Activity, and estimates of Caloric Expenditure. This device is not intended for use as an ECG monitor.

## Device Story

Actiheart is a compact, wearable, battery-operated physiological data recorder used in home or hospital environments. It captures ECG signals via standard chest-worn electrodes and motion data via an internal accelerometer. A differential amplifier senses and amplifies the ECG signal, while proprietary algorithms process the digitized signals to detect heart rate (BPM) and activity counts. Energy expenditure is derived from these inputs using proprietary software algorithms. Data is logged on-board and later transferred to a PC via a reader for display and export. The device features a rechargeable lithium battery and a green LED indicator for heart beats. It assists healthcare providers by providing quantifiable physiological data for clinical assessment. It is not an ECG monitor.

## Clinical Evidence

Bench testing only. Performance validated against ANSI/AAMI EC13:2002 standards. Testing included tall T-wave rejection (sensitivity >98%, specificity 100%), and heart rate detection accuracy across multiple waveform libraries (MIT-BIH, European ST-T, AAMI Normal Sinus Rhythm) with sensitivity >99% and specificity >97-99%. Device performance was compared side-by-side with the predicate device using library recordings of known heart rates and simulated recordings. Results demonstrated comparable performance in heart rate measurement, range, and accuracy.

## Technological Characteristics

Polycarbonate/ABS plastic case; stainless steel attachment clip. Sensing: differential amplifier for ECG, accelerometer for motion. Energy: 3.0V rechargeable lithium battery. Dimensions: 195mm length, 38mm diameter, 12.5mm tail clip. Connectivity: wired data transfer to PC via reader. Software: proprietary algorithms for ECG feature detection and caloric expenditure derivation. Sterilization: none required.

## Regulatory Identification

A physiological signal conditioner is a device such as an integrator or differentiator used to modify physiological signals for recording and processing.

## Predicate Devices

- Mini Logger Series 2000 ([K991045](/device/K991045.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a sequence of alphanumeric characters, specifically "K052489". The characters are written in a clear, sans-serif font, and they appear to be part of a label or identification code. The characters are black against a white background.

June 15, 2005

Image /page/0/Picture/2 description: The image shows the text "Page 1 of 8 Actiheart". The number 1 is circled. The text appears to be handwritten.

SEP 27 2005

## 510(K) SUMM/

Date of Submission

Dollars

Official Contact / Address of Manufacturing facility

Zita A. Yurko Manager, Regulatory Affairs Respironics Inc. 1001 Murry Ridge Lane Murrysville, PA 15668

Phone: 724-387-4120 Fax: 724-387-4216 Zita.yurko@respironics.com

Heart Rate Monitor

| Proprietary Name |  | Actiheart |
|------------------|--|-----------|
|------------------|--|-----------|

Common/Usual Name

Classification

Predicate Devices

Physiological Signal Recorder Device Classification Name

Neurology

21 CFR 882.1845 Classification Reference

Class II

Appropriate Classification Panel

GWK Product Code

Mini Logger Series 2000 (Mini Mitter Co., Inc.)

New device Reason for submission

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the text "Page 2 of Actineart". The number 2 is circled. The text is handwritten and appears to be part of a document.

#### Substantial Equivalence

This premarket notification submission demonstrates that the Actiheart system is substantially equivalent to the Mini Logger Series 2000 (K991045).

The design of the Actiheart was verified through the use of design verification and validation testing. The Hazards Control Measures Traceability Matrix provided in the Risk Analysis assured that all hazards identified by the risk analysis were successfully mitigated.

This submission is seeking to obtain clearance to market a new product that is substantially equivalent to the Mini Logger Series 2000. The new product (Actiheart) performs substantially equivalent measurements of Heart Rate and Activity as the Mini Logger Series 2000.

#### Indications for Use

The Actiheart is an ambulatory Heart Rate and Activity Recorder. Actiheart may be used to quantifiably measure Heart Rate, Activity, and estimates of Caloric Expenditure. This device is not intended for use as an ECG monitor.

#### Device Description

Actiheart is a compact, ambulatory, physiological Heart Rate and Activity data recorder. Actiheart may be attached to the chest surface through the use of standard ECG electrodes. Recorded data may be transferred later to a PC for display and conversion for export to other programs.

#### Basic scientific concepts

The device acquires and logs digital data whose values represent the rate of Heart Beat (BPM), based on a proprietary analysis of ECG complexes. (Actiheart is not an ECG monitor.) The scientific concepts and technologies that are used to sense the signals are summarized in Table 10A.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the text "Page 3 of 8 Actiheart". The number 3 is circled. The text appears to be handwritten. 

# TABLE 10A. BASIC TECHNOLOGIES USED FOR PHYSIOLOGICAL SIGNAL RECORDING IN

### ACTIHEART RECORDERS

| Physiological<br>Parameter       | Where Obtained | Technology                                                                                                                                                                    | Value Obtained                                                        |
|----------------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Heart Rate                       | Actiheart      | Differential amplifier<br>senses and<br>amplifies ECG<br>signal. Proprietary<br>code transforms<br>digitized signal and<br>detects location of<br>features in ECG<br>complex. | Digital value<br>corresponding to<br>heart Beats Per<br>Minute (BPM). |
| Activity                         | Actiheart      | Signal from<br>accelerometer<br>amplifier is digitized<br>and analyzed using<br>proprietary<br>algorithms to detect<br>motion.                                                | Digital value<br>corresponding to<br>Activity Counts                  |
| Caloric Expenditure<br>(derived) | Software       | Conversion of<br>Activity Counts and<br>HR to<br>corresponding value<br>of Calories<br>Expended<br>(proprietary<br>algorithm)                                                 | Value<br>corresponding to<br>Calories Expended                        |

### Physical characteristics of Actiheart

Pertinent physical characteristics of the Actiheart are shown in Table 10B.

## TABLE 10B. PHYSICAL CHARACTERISTICS OF ACTIHEART RECORDER

| Parameter                | Value                                                      | Condition/Note                  |
|--------------------------|------------------------------------------------------------|---------------------------------|
| Size                     | 195mm length, overall<br>38mm diameter<br>12.5mm tail clip | Outer dimensions                |
| Weight                   | 12 grams                                                   | With no ECG electrodes attached |
| Case material            | Polycarbonate/ABS                                          | Flammability rating UL94 V-0    |
| Attachment clip material | Stainless steel                                            |                                 |
| Attachment type          | Adult ECG snap                                             |                                 |
| Battery type             | 3.0 volt lithium rechargeable                              | Not user replaceable            |
| Indicators               | Green LED                                                  | Coincident with heart beat      |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the words "page" and "of" with the numbers 4 and 8 in between. The number 4 is inside of a circle. Below the words "of" and the number 8, the word "Actiheart" is written.

#### Functional Description

Actiheart is intended for the measurement, storage, and display of physiological data (Heart Rate (HR) and Activity (ACT)). In addition, Actiheart derives an approximate value of Energy Expenditure from the values of HR and ACT. Actiheart can be attached to the subject's chest using standard ECG electrodes. Through the use of various types of standard ECG electrodes, the Recorder may be worn for varying lengths of time, dependent on the amount of data logging desired and the recommendations of the ECG electrode manufacturer. Parameter data are logged on board the Recorder and later transferred through a cable to a personal computer via the Reader. The PC Software can be used to display the physiological data and store the data for future comparison. The Multicharger is used to re-charge the Recorder's onboard rechargeable lithium battery.

#### Technological characteristics, comparison to predicate device

Actiheart and the Mini-Logger Series 2000 Heart Rate Probe (K991045) are each diagnostic test systems based upon the concept of an ambulatory, unattended physiological recorder that logs physiological data to the logging device. Each of these devices is a solid-state Recorder with differential amplifiers, data collection and analysis algorithms, and with the ability to store data until it is transfered into the PC. Actiheart and the Mini-Logger Series 2000 Heart Rate Probe are of similar size and weight. Both devices measure and record Heart Rate and Activity.

The physiological data logging device presented in this 510(k) submission (Actiheart) is substantially equivalent to the Heart Rate Probe currently marketed as part of the Mini-Logger® Series 2000 Physiological Data Logging Device (K991045).

The Actiheart and the Mini-Logger® Series 2000 Heart Rate Probe are of similar size and weight. Both devices have the ability to measure Heart Rate. Both devices are surface-mounting, chest-worn, and ambulatory.

The performance characteristics are illustrative of comparable devices. Similarities and differences between Actiheart and the predicate device are presented below in Table 10C. Where differences exist, they reflect reduced indications for use and improvements in convenience of use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the text "Page 5 of 8" in a handwritten style. Below the page number is the word "Actiheart" in a smaller, sans-serif font. The number 5 is circled.

| COMPARISON<br>PARAMETER                                            | Actiheart                                                                                                                                                           | Mini-Logger® Series 2000 with<br>Heart Rate Probe<br>(K991045)                                                                                                               |
|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description                                                 | Compact, wearable, battery-<br>operated physiological data<br>recorder                                                                                              | Compact, wearable, battery-<br>operated physiological data<br>recorder                                                                                                       |
| Where used                                                         | Actiheart may be used where<br>quantifiable measurement of<br>human Heart Rate and/or Activity is<br>needed. It may be used in the<br>home or hospital environments | The Mini-Logger® may be used in<br>any instance where quantifiable<br>analysis of physiological data is<br>desirable. It may be used in the<br>home or hospital environments |
| Visual appearance and<br>physical description of<br>Recorder/probe | 195mm length<br>38mm diameter<br>12.5mm tail clip                                                                                                                   | 145 mm length<br>31 mm wide                                                                                                                                                  |
| HR channels                                                        | 1                                                                                                                                                                   | 1                                                                                                                                                                            |
| Materials<br>(Recorder)                                            | Polycarbonate/ABS plastic case                                                                                                                                      | ABS plastic case                                                                                                                                                             |
| Mode of operation<br>(Recorder)                                    | Recorder input                                                                                                                                                      | Wireless Recorder input and<br>Direct-wired Recorder input                                                                                                                   |
| Design features<br>(System)                                        | Maximum recording time 14 days                                                                                                                                      | Maximum recording time 88 days                                                                                                                                               |
| Memory<br>(system)                                                 | 1 Megabit                                                                                                                                                           | 128 Kilobyte or 1 Megabyte                                                                                                                                                   |
| Performance                                                        | Comparable (see included)                                                                                                                                           | Comparable (see included)                                                                                                                                                    |
| Heart Rate range                                                   | 35 to 255 BPM                                                                                                                                                       | Up to 250 BPM                                                                                                                                                                |
| Heart Rate resolution                                              | 1 BPM                                                                                                                                                               | 1 BPM                                                                                                                                                                        |
| Heart Rate accuracy*                                               | ±10% of value, not to exceed 5<br>BPM<br>(Meets EC-13-4.2.7)                                                                                                        | Unpublished                                                                                                                                                                  |
| R-wave detection                                                   | dv/dt with variable threshold,<br>sampling at 128 Hz                                                                                                                | Performed by the heart rate probe.<br>Method is unpublished.                                                                                                                 |
| Sterility<br>(Recorder)                                            | None required                                                                                                                                                       | None required                                                                                                                                                                |
| Biocompatibility<br>(Recorder)                                     | Through the use of ECG electrodes                                                                                                                                   | Through the use of ECG<br>electrodes                                                                                                                                         |
| Human factors<br>(Recorder)                                        | Attaches to ECG electrodes                                                                                                                                          | Waist pack carrying case, direct-<br>wire attachment to ECG<br>electrodes, or telemetric link                                                                                |
| Electrical safety<br>(Recorder)                                    | Battery operated                                                                                                                                                    | Battery operated                                                                                                                                                             |
| Power used<br>(Recorder)                                           | 3.0 volt rechargeable lithium battery<br>(1 each)                                                                                                                   | 3.6 volt disposal lithium battery<br>(2 each)                                                                                                                                |

### TABLE 10C. SUBSTANTIAL EQUIVALENCE COMPARISON

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the text "Page 6 of 8" in a handwritten style. The number 6 is circled. Below the page number is the word "Actiheart".

#### Performance testing

An extensive collection of tests has been conducted and successfully completed, including safety, performance and comparative tests. Declarations of conformance to the FDA Recognized list of consensus standards, as well as FDA reviewers guidance and Applicable voluntary standards have been provided in support of the safety and effectiveness of the Actiheart System. This list of performance testing included/will include:

- IEC 60601-1: 1988 + A1: 1991 + A2: 1995, Medical Electrical Equipment Part 1: General . Requirements for Safety - Tab 11A
- IEC 60601-1-2: 2001 Medical Electrical Equipment Part 1-2: Collateral Standard: Electromagnetic . Compatibility - Requirements and tests - Tab 11B
- ISO-10993-1: Biological Evaluation of Medical Devices Evaluation and Testing Tab 9 .
- FDA Reviewer's Guidance (#G95-1, 5/1/95) Biological Evaluation of Medical Devices; Use of ISOt 10993 – Tab 9
- Actiheart to Mini Logger Series 2000 bench test comparison report; assesses the Heart Rate . detection performance characteristics of both devices side-by-side against a set of library recordings of known Heart Rate. Additional testing with the use of simulated recordings per ASTM EC13:2002.
- Actiheart Software Functional Requirements Test Procedure/Report; assesses the features of the . Actiheart to ensure compliance with the Software requirements.
- Actiheart Shock & Vibration Test .
- Actiheart Temperature & Humidity Test .

Additional testing has been performed in to ensure safety & effectiveness of the Actiheart System. The following set of standards and guidance documents have been used in the Actiheart System. These include:

Required Standards – A Declaration of Conformity for each of these standards is provided in the submittal:

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the text "Page 7 of 8". The number 7 is circled. Below the numbers is the word "Actiheart".

- IEC 60601-1: 1988 + A1: 1991 + A2: 1995, Medical Electrical Equipment Part 1: General . Requirements for Safety
- IEC 60601-1-2: 2001 Medical Electrical Equipment Part 1-2: Collateral Standard: Electromagnetic . Compatibility - Requirements and tests
- ISO-10993-1: Biological Evaluation of Medical Devices Evaluation and Testing {Biocompatibility has . been established by the device clearances associated with the patient applied parts – no separate DOC is provided to demonstrate compliance with this standard}

Voluntary Standards - Guidance, when applicable, has been adopted from the following standards.

- ANSI/ASTM EC13:2002 Cardiac monitors, heart rate meters, and alarms [applicable clauses] .
Reviewers Guidance - Guidance, when applicable, has been adopted from each of the FDA Reviewers Guidance documents

- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, t May 2005
- FDA Reviewer's Guidance (#G95-1, 5/1/95) Biological Evaluation of Medical Devices; Use of ISOt 10993
- FDA Reviewer's Guidance for Premarket Notification Submissions, Appendix A, November 1993 .
- FDA Reviewer's Guidance General Principles of Software Validation; Final Guidance for Industry and ● FDA Staff. January 2002

#### Assessment of non-clinical performance data

The performance of the Actifieart Recorder has been tested in accordance with applicable clauses of ANSI/AAMI EC13:2002, American National Standard for Heart Rate (HR) Meters. The categories of tests applied to the submitted Recorder are listed as follows:

- . Tall T-wave rejection capability
- Measurement of HR sensitivity and specificity for several libraries of ECG waveforms .
- Measurement of HR in the presence of line frequency and drift noise .
- Measurement of range and accuracy of ECG detection .
- Measurement of HR detection over a range of QRS amplitudes .
- Accuracy and response to irregular rhythms .

A detailed disclosure of the test results has been provided in Tab 5 of this Application. For all the categories, we have determined that Actiheart performance meets that of the currently-marketed predicate device, Mini-Logger® Series 2000 Heart Rate Probe. Neither the submitted device (Actiheart) nor the predicate device (Mini-Logger® Series 2000 Heart Rate Probe) is capable of detecting all types of irregular rhythms as specified in ANSI/AAMI EC13, Clause 4.1.2.1(e). However, per ANSI/AAMI EC13, a note has been placed in the Actiheart Instruction Manual indicating such. A summary of our results appears in Table 10D.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the text "page 8 of 8" with the word "Actiheart" below it. The number 8 is written inside of a circle. The text appears to be handwritten.

| Figure of Merit                                 | Requirement                                                                                               | Actiheart<br>Performance                                                                        |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Tall T-wave Rejection<br>Capability             | Reject Tall T-Waves up to 1.2<br>mV with ±10% accuracy or ±5<br>BPM (greater of two)                      | Sensitivity is better than 98 %<br>Specificity is 100%                                          |
| MIT-BIH Normal Sinus Rhythm<br>waveform library | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two)                                          | Sensitivity is better than 99%<br>Specificity is better than 98%                                |
| European ST-T waveform<br>library               | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two)                                          | Sensitivity is better than 99%<br>Specificity is better than 97%                                |
| AAMI Normal Sinus Rhythm<br>waveform library    | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two)                                          | Sensitivity is better than 99%<br>Specificity is better than 99%                                |
| Line frequency tolerance and<br>drift tolerance | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two) when drift or line noise<br>present      | No difference between HR<br>measurement for waveforms<br>with and without superimposed<br>noise |
| Range and accuracy of HR<br>meter               | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two) over specified range of 32<br>to 255 BPM | Within limits for all values of HR                                                              |
| Amplitude of QRS waveform                       | Detect HR with accuracy of<br>±10% or ±5 BPM (greater of<br>two) for amplitudes  0.5 mV to<br>5 mV        | Within limits for all values of<br>amplitude between 0.45 mV and<br>9 mV.                       |

## TABLE 10D. ACTIHEART HEART RATE MEASUREMENT CHARACTERISTICS

### Conclusion

It is the conclusion of Respironics that the Actiheart system is substantially equivalent to Mini Logger Series 2000 (K991045) and presents no new concerns about safety and effectiveness.

(End of Tab.)

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized representation of three human profiles facing to the right, stacked on top of each other.

SEP 2 7 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics, Inc. c/o Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, Massachusetts 01719

Re: K052489 Trade/Device Name: Actiheart Regulation Number: 21 CFR 882.1845 Regulation Name: Physiological signal conditioner Regulatory Class: II Product Code: GWK Dated: September 8, 2005 Received: September 12, 2005

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorobate) is registy
 commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 1977, the accordance with the provisions of the Federal Food, Drug, devices mat nave occh rechasined in asseroval of a premarket approval application (PMA). allo Cosment Act (71ct) that do not requently of the general controls provisions of the Act. The I ou may, merelore, manel act include requirements for annual registration, listing of general controls provisions of mactice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassined (600 as controls. Existing major regulations affecting your device can may be subject to such additional controlio "Entraligning to the 898. In addition, FDA may be round interents concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Tease oe advised that I Dri Visuan to syour device complies with other requirements of the Act that I DA has made a determinations administered by other Federal agencies. You must of any I cachi statutes and regalancents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It I art 607), adomig (21 CFR Part 820); and if applicable, the electronic forth in the quality Systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

Page 2- Mr. Neil E. Devine, Jr.

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin marketing of substantial equivalence of your device to a legally premarket notification: "The FDA miding of basically for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r the results of the may on 1011 at the subscription on to the receiled If you desire specific advice for your ac not our sales. Also, please note the regulation entitled, conlact the Office of Comphance at (216) 276 - 310 - 410 Part 807.97). You may obtain Misorallumg by reference to premainer to premainterial in Act from the Division of Small other general International and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark A. Millken

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

## Indications for Use

510(k) Number (if known): _KOS 2489

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The Actiheart is an ambulatory Heart Rate and Activity Recorder. Actifieart may be used to quantifiably measure Heart Rate, Activity, and estimates of Caloric Expenditure. This device is not intended for use as an ECG monitor.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark N. Milliken

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** K052484

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWK/K052489](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWK/K052489)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
