← Product Code [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ) · K203684

# Neurosign V4 Intraoperative Nerve Monitor (K203684)

_Magstim Company, Ltd. · PDQ · Mar 17, 2021 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K203684

## Device Facts

- **Applicant:** Magstim Company, Ltd.
- **Product Code:** [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ.md)
- **Decision Date:** Mar 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1820
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The Neurosign® V4 Intraoperative Nerve Monitor is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal nerve roots.

## Device Story

Neurosign V4 is a multi-channel intraoperative nerve monitor (IONM) used in ENT and neurosurgery. It collects EMG signals from muscles via needle or surface electrodes connected to a 4-channel or 8-channel pre-amplifier. The device detects muscle contractions resulting from mechanical manipulation or electrical stimulation of nerves. It amplifies these signals, providing audible feedback and visual waveform displays on a 15-inch color touchscreen. A stimulator pod allows surgeons to map and locate nerves during procedures. A mute sensor can silence audio during electrocautery interference. Operated by clinical staff in surgical environments, the device assists in nerve identification and preservation, potentially reducing the risk of iatrogenic nerve injury. The system includes hardware, software, and accessories.

## Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), alarm systems (IEC 60601-1-8), and biocompatibility (ISO 10993-1, -5, -10). Usability/human factors testing (AAMI/ANSI HE75, IEC 62366-1) confirmed the safety and effectiveness of the 8-channel configuration. No clinical data was required or provided.

## Technological Characteristics

Integrated system with 15-inch color touchscreen, rotary controls, and 4 or 8-channel pre-amplifier. Stimulator supports constant current (10µA–10mA) or constant voltage (10mV–10V) modes. Connectivity via electrode inputs. Compliant with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-8. Biocompatible materials per ISO 10993.

## Regulatory Identification

A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.

## Predicate Devices

- Neurosign® V4 Intraoperative Nerve Monitor ([K181559](/device/K181559.md))
- Neurosign® Nerve Monitor, Model Neurosign® 400 and Neurosign® 800 ([K053141](/device/K053141.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 1, 2021

Magstim Company Ltd. Tom Campbell Chief Quality & Regulatory Affairs Officer Spring Gardens Whitland, Carmarthenshire SA34 0HR United Kingdom

Re: K203684

Trade/Device Name: Neurosign V4 Intraoperative Nerve Monitor Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: PDQ

Dear Tom Campbell:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 17, 2021. Specifically, FDA is updating this SE Letter due to typographical errors in the Indications for Use Statement as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jay Gupta, OHT5: Office of Neurological and Physical Medicine Devices, Jay Gupta, 301-796-2795, jay.gupta@fda.hhs.gov.

Sincerely,

# Jay R. Gupta -S

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.

March 17, 2021

Magstim Company Ltd. Tom Campbell Chief Quality & Regulatory Affairs Officer Spring Gardens Whitland, Carmarthenshire SA34 0HR United Kingdom

Re: K203684

Trade/Device Name: Neurosign V4 Intraoperative Nerve Monitor Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ Dated: December 17, 2020 Received: December 17, 2020

Dear Tom Campbell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{2}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Jay R. Gupta -S

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203684

Device Name Neurosign® V4 Intraoperative Nerve Monitor

Indications for Use (Describe)

The Neurosign® V4 Intraoperative Nerve Monitor is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal nerve roots.

Indications for Neurosign® V4 EMG Monitoring Procedures include: Intracranial, Extracranial, Extratemporal, Neck Dissections, Thoracic Surgeries, and Upper and Lower Extremities.

Indications for spinal procedures which may use Neurosign® V4 EMG monitoring include: Degenerative Treatments, Pedicle Screw Procedures, Fusion Cages, Rhizotomy, Orthopedic Surgery and Open and Percutaneous Lumbar Procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# K203684 Traditional 510(k) SUMMARY Magstim's Neurosign® V4 Intraoperative Nerve Monitor

Prepared according to the requirements outlined in 21 CFR 807.92

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Magstim Company Limited Spring Gardens, Whitland, Carmarthenshire SA34 OHR, United Kingdom

Phone: +44 (0) 1994 240798 Facsimile: +44 (0) 1994 240061

Contact Person: Tom Campbell, Chief Quality & Regulatory Affairs Officer requlatory@magstim.com

Date Prepared: March 17, 2021

#### Trade Name of Device

Neurosign® V4 Intraoperative Nerve Monitor

#### Common or Usual Name

Neurosurgical Nerve Locator

#### Classification

Neurosurgical Nerve Locator

21 CFR 874.1820, Class II, product code PDQ

#### Predicate Devices

K181559 Neurosign® V4 Intraoperative Nerve Monitor. The Magstim Company Limited (Primary Predicate Device), 21 CFR 874.1820, PDQ, ETN

K053141 Neurosign® Nerve Monitor, Model Neurosign® 400 and Neurosign® 800, The Magstim Company Limited (Secondary Predicate), 21 CFR 874.1820, ETN

#### Device Description

The Neurosign® V4 Intraoperative Nerve Monitor (IONM) is a multi-channel nerve monitor designed for use in Ear, Nose and Throat (ENT) surgery and neurosurgery. The subject device differs from its primary predicate due to the introduction of improved Pre-amplifier and Stimulator pod cables, as well the addition of an 8-channel Pre-amplifier variant.

{5}------------------------------------------------

Using needle or surface electrodes 1 connected to the Pre-amplifier, the nerve monitor detects electromyographic (EMG) signals within muscles caused by the contraction of the muscle due to mechanical manipulation or electrical stimulation of the nerve controlling it. The Neurosign® V4 then amplifies the EMG data into an audible signal and displays the data as a waveform on the active monitoring screen.

The Neurosign® V4 Intraoperative Nerve Monitor is used for patient treatment by prescription only under the supervision of a licensed physician.

The Neurosign® V4 is an integrated system consisting of a combination of hardware, software, and accessories. Its technological characteristics are described in further detail below.

## Intended Use

The Neurosign® V4 Intraoperative Nerve Monitor is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal nerve roots.

## Technological Characteristics

The Neurosign® V4 is comprised of following components:

- 1. Neurosign® V4 Intraoperative Nerve Monitor
- 2. Neurosign® V4 Pre-amplifier:
	- a. Neurosign® V4 Pre-amplifier 4-Channel
	- b. Neurosign® V4 Pre-amplifier 8-Channel
- 3. Neurosign® V4 Stimulator Pod
- 4. Neurosign® V4 Mute Sensor

The Neurosign® V4 Intraoperative Nerve Monitor (IONM) includes a user interface comprised of an audio output, a color graphics display with a touch screen and dedicated rotary controls for frequently adjusted parameters.

EMG signals are collected from the patient using needle and surface electrodes [1] connected to the Neurosian® V4 Pre-amplifier. The pre-amplifier is available in a 4-channel or 8-channel variant, allowing the user to monitor up to eight separate neural pathways depending on the pre-amplifier used. The Neurosign® V4 Pre-amplifier collects, processes and transmits the EMG signals to the IONM for display and for an audio output.

The Neurosign® V4 Stimulator Pod allows the simultaneous connection of two stimulating probes for the mapping and locating of nerves within tissue, and to test the nerve activity at various stages during surgery.

Software documentation for a "major" level of concern has been provided.

<sup>1</sup> Single use electrodes and probes not subject to this submission

<sup>2</sup> Single use electrodes and probes not subject to this submission

{6}------------------------------------------------

# Non-Clinical Testing Table 1: Summary of Non-Clinical Testing

| Test                                   | Method                                                                                                                                                                                                                                                                                                                       | Results/ Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety<br>Mechanical Safety | ANSI/AAMI ES60601-1:2005/(R)2012 and<br>A1:2012-<br>Medical electrical equipment - Part 1:<br>General requirements for<br>basic safety and essential performance;<br>FDA Recognition Number: 19-4                                                                                                                            | A sample Neurosign® V4 (4-Channel<br>and 8-Channel Variants) has been<br>tested and found to be compliant to the<br>requirements of IEC 60601-1 by<br>independent test laboratory Element<br>Materials Technology, to demonstrate<br>safety and effectiveness of the system<br>following incorporation of new/different<br>characteristics as compared to the<br>predicate device.<br><br>Following the design changes to the Pre<br>amplifier and Stimulator Pod cables,<br>non-clinical bench testing was performed<br>by Magstim confirming that there is no<br>discernable difference in the isolation<br>parameters of the new cable design in<br>comparison to the old. |
| Electromagnetic<br>Compatibility       | IEC 60601-1-2: 2014 -<br>Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance -<br>Collateral standard: Electromagnetic<br>compatibility -Requirements and tests;<br>FDA recognition number: 19-8                                                                  | A sample Neurosign® V4 (4-Channel)<br>has been tested and found to be<br>compliant to the requirements of EN<br>60601-1-2 by independent test<br>laboratory Eurofins Hursley, to<br>demonstrate safety and effectiveness o<br>the system following incorporation of<br>new/different characteristics as<br>compared to the predicate device.<br><br>Following the design change to the Pre-<br>amplifier and Stimulator Pod cables,<br>non-clinical bench testing was performed<br>by Magstim concluding there is no<br>discernable difference in the EMC<br>performance of the new cable design in<br>comparison to the old.                                                 |
| Alarm Systems                          | IEC 60601-1-8 -<br>Medical electrical equipment - Part 1-8:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: General requirements, tests and<br>guidance for alarm systems in medical<br>electrical equipment and medical electrical<br>systems; FDA Recognition Number: 5-76 | A sample Neurosign® V4 has been<br>tested and found to be compliant to the<br>requirements of IEC 60601-1-8 by<br>independent test laboratory Element<br>Materials Technology, thus<br>demonstrating the Neurosign® V4 is<br>substantially equivalent to the legally<br>marketed predicate device.                                                                                                                                                                                                                                                                                                                                                                            |
| Biocompatibility                       | ISO 10993-1:2009 - Biological Evaluation<br>of Medical Devices - Part 1: Evaluation and<br>testing within a risk management process                                                                                                                                                                                          | The Neurosign®V4 Nerve Monitor is no<br>intended to come into contact with the<br>patient, as patient contact is achieved                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Human Factors Testing                  | ISO 10993-5:2009 - Biological evaluation<br>of medical devices -- Part 5: Tests for in<br>vitro cytotoxicity; FDA Recognition<br>Number: 2-245                                                                                                                                                                               | via use of single use electrodes and<br>probes.<br>However, the Stimulator Pod, Pre-<br>amplifier and connecting leads may<br>come into contact with the patient.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                        | ISO 10993-10:2010 - Biological evaluation<br>of medical devices - Part 10: Tests for<br>irritation and skin sensitization; FDA<br>Recognition Number: 2-174                                                                                                                                                                  | Samples of these materials have been<br>tested and found to be compliant to the<br>requirements of ISO 10993-1, ISO<br>10993-5 and ISO 10993-10 by an<br>independent test laboratory, thus<br>demonstrating the Neurosign® V4 is<br>substantially equivalent to the legally<br>marketed predicate device.                                                                                                                                                                                                                                                                                                                                                                     |
|                                        | AAMI/ANSI HE75 - Human Factors<br>Engineering - Design of Medical Devices;<br>FDA Recognition Number: 5-57                                                                                                                                                                                                                   | Usability testing was performed to<br>analyze the user experience of the 8-<br>Channel Neurosign® V4.<br>The Human Factors Engineering report<br>verifies the Neurosign V4, using the 8-<br>channel pre-amplifier, to be safe and<br>effective for the intended users, uses,<br>and use environments thus<br>demonstrating the Neurosign® V4 is<br>substantially equivalent to the legally<br>marketed predicate device.                                                                                                                                                                                                                                                      |
|                                        | IEC 62366-1 - Medical Devices - Part 1:<br>Application of Usability Engineering To<br>Medical Devices; FDA Recognition<br>Number: 5-114                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{7}------------------------------------------------

The software verification and validation testing further demonstrated that the software performs as intended and in accordance with specifications. In accordance with ISO14971, risks associated with the Neurosign® V4 have been identified, assessed and, where necessary, mitigated with risk control measures to an acceptable level.

# Substantial Equivalence Discussion

The Neurosign® V4 (subject of this submission) is substantially equivalent to the primary predicate device, the Neurosign® V4 Intraoperative Nerve Monitor (K181559) and the secondary predicate device, the Neurosign® Nerve Monitor Model Neurosign® 400 and Model Neurosign® 800® (K053141).

The intended use and indications for use are identical between the Neurosign® V4 (subject of this submission) and the primary predicate device, the Neurosign® V4 Intraoperative Nerve Monitor (K181559).

All devices have the same four principal components:

- (1) Intraoperative Nerve Monitor,
- (2) Pre-amplifier,
- (3) Stimulator and
- (4) Mute Sensor.

The notable difference between the subject device Neurosign® V4 and that which was cleared earlier in K181559 includes the redesign of the Pre-amplifier and Stimulator pod

{8}------------------------------------------------

cables to improve reliability and malleability during use, as well the addition of an 8-channel Pre-amplifier variant, enabling the user to monitor up to eight separate neural pathways at risk during surgery. The Neurosign® V4 8-Channel Pre-amplifier is substantially equivalent to the Neurosign® V4 4-Channel Pre-amplifier previously cleared under K181559, with the addition of 4 additional electrode channels.

Furthermore, the addition of an 8-channel Pre-amplifier is supported by the clearance of the secondary predicate device (K053141), as the secondary predicate includes the option to monitor up to eight separate neural pathways at risk during surgery.

The Principles of Operation are identical to the primary predicate device, the Neurosign® V4 Intraoperative Nerve Monitor (K181559). Both the subject device and the predicate device systems are all systems that are intended for use in collecting, processing and representing EMG signals from nerves at of choice during surgery.

Both the subject and predicate devices use a stimulator for the mapping and locating of nerves within tissue and testing the nerve activity at various stages during surgery. All offer the option to use a mute sensor to silence the audio output during episodes of RF interference from electrocautery.

A summary of the similarities and minor differences between the Neurosign® V4, the previous Neurosign® V4 (primary predicate device) and the Neurosign® Nerve Monitor Model Neurosign® 400 and Model Neurosign® 800 (Secondary Predicate Device) are described in Table 2.

### Conclusions

In summary, the subject and predicate devices are substantially equivalent based on their identical intended use and similar technological characteristics, which have been tested according to the applicable non-clinical performance testing.

Non-clinical test data demonstrates that the Neurosign® V4 is as safe and effective as the predicate devices.

Thus, the Neurosign® V4 Intraoperative Nerve Monitor is substantially equivalent to the primary predicate device, the Neurosign® V4 Intraoperative Nerve Monitor (K181559), and the secondary predicate device, the Neurosign® Nerve Monitor, Model Neurosign® 400.

{9}------------------------------------------------

# Table 2: Substantial Equivalence Summary

| Criteria of<br>Comparison                    | K203684<br>Neurosign® V4 Intraoperative<br>Nerve Monitor<br>(Subject of this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K181559<br>Neurosign® V4<br>Intraoperative Nerve<br>Monitor<br>(Primary Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K053141<br>Neurosign® Nerve Monitor,<br>Model Neurosign® 400 and<br>Model Neurosign® 800®<br>(Secondary Predicate<br>Device)              |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                 | Magstim Company Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Magstim Company Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Magstim Company Limited                                                                                                                   |
| Device Name                                  | Neurosign® V4 Intraoperative<br>Nerve Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Neurosign® V4<br>Intraoperative Nerve<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Neurosign® Nerve Monitor,<br>Model Neurosign® 400                                                                                         |
| Product code(s)<br>& regulation              | PDQ<br>21 CFR 874.1820                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | PDQ, ETN<br>21 CFR 874.1820                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ETN<br>21 CFR 874.1820                                                                                                                    |
| Intended Use/<br>Indications for<br>Use      | The Neurosign® V4<br>Intraoperative Nerve Monitor<br>is intended for locating and<br>identifying cranial and<br>peripheral motor and mixed<br>motor-sensory nerves during<br>surgery, including spinal<br>nerve roots.<br>Indications for Neurosign®<br>V4 EMG Monitoring<br>Procedures include:<br>Intracranial, Extracranial,<br>Intratemporal, Extratemporal,<br>Neck Dissections, Thoracic<br>Surgeries, and Upper and<br>Lower Extremities.<br>Indications for spinal<br>procedures which may use<br>Neurosign® V4 EMG<br>monitoring include:<br>Degenerative Treatments,<br>Pedicle Screw Procedures,<br>Fusion Cages, Rhizotomy,<br>Orthopedic Surgery and<br>Open and Percutaneous<br>Lumbar Procedures. | The Neurosign® V4<br>Intraoperative Nerve<br>Monitor is intended for<br>locating and identifying<br>cranial and peripheral motor<br>and mixed motor-sensory<br>nerves during surgery,<br>including spinal nerve roots.<br>Indications for Neurosign®<br>V4 EMG Monitoring<br>Procedures include:<br>Intracranial, Extracranial,<br>Intratemporal,<br>Extratemporal, Neck<br>Dissections, Thoracic<br>Surgeries, and Upper and<br>Lower Extremities.<br>Indications for spinal<br>procedures which may use<br>Neurosign® V4 EMG<br>monitoring include:<br>Degenerative Treatments<br>Pedicle Screw Procedures,<br>Fusion Cages, Rhizotomy,<br>Orthopedic Surgery and<br>Open and Percutaneous<br>Lumbar Procedures. | To locate and identify<br>cranial motor nerves during<br>ENT and intra-cranial<br>procedures.                                             |
| Intraoperative Nerve Monitor                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                           |
| Type of monitor                              | EMG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | EMG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EMG                                                                                                                                       |
| System<br>hardware                           | Digital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Digital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Digital                                                                                                                                   |
| Display                                      | 15" color touchscreen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 15" color touchscreen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 125mm x 95mm<br>electroluminescent display                                                                                                |
| Criteria of<br>Comparison                    | K203684<br>Neurosign® V4 Intraoperative<br>Nerve Monitor<br>(Subject of this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K181559<br>Neurosign® V4<br>Intraoperative Nerve<br>Monitor<br>(Primary Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K053141<br>Neurosign® Nerve Monitor,<br>Model Neurosign® 400 and<br>Model Neurosign® 800®<br>(Secondary Predicate<br>Device)              |
| Stimulator<br>adjust                         | By control knob                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | By control knob                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | By control knob                                                                                                                           |
| Volume adjust                                | By control knob; individual<br>channels by touchscreen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | By control knob; individual<br>channels by touchscreen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | By control knob                                                                                                                           |
| Needle<br>placement<br>impedance<br>checking | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No                                                                                                                                        |
|                                              | Amplifier                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                           |
| Number of<br>channels                        | 4 (4-Channel Pre-amplifier)<br>8 (8-Channel Pre-amplifier)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 4 (Neurosign® 400)<br>8 (Neurosign® 800)                                                                                                  |
| Color coded<br>electrode<br>sockets          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                       |
| Bandwidth                                    | 10Hz - 1kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 10Hz - 1kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 10Hz - 5kHz                                                                                                                               |
| Common mode<br>rejection ratio               | > 90dB @ 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | > 90dB @ 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | > 95dB @ 50/60Hz                                                                                                                          |
|                                              | Stimulator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                           |
| Stimulator<br>channels                       | Channel 1: 2 probes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Channel 1: 2 probes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Channel 1: 2 probes                                                                                                                       |
| Operational<br>mode                          | Constant Current or<br>Constant Voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Constant Current or<br>Constant Voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Constant Current or<br>Constant Voltage                                                                                                   |
| Current ranges                               | 10μΑ – 10mA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 10μA - 10mA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 50µA - 10mA                                                                                                                               |
| Voltage ranges                               | 10mV - 10V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 10mV - 10V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 50mV - 10V                                                                                                                                |
| Accuracy                                     | Current Stimulating Mode<br>± 10% (1mA – 10mA), into<br>1kΩ load<br>± 25% (10μA - 990 μA), into<br>1kΩ load<br><br>Voltage Stimulating Mode<br>± 10% (1V - 10V), into 1kΩ<br>load<br>± 25% (10mV - 990 mV), into<br>1kΩ load                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Current Stimulating Mode<br>± 10% (1mA – 10mA), into<br>1kΩ load<br>± 25% (10μA - 990 μA), into<br>1kΩ load<br><br>Voltage Stimulating Mode<br>± 10% (1V - 10V), into 1kΩ<br>load<br>± 25% (10mV - 990 mV),<br>into 1kΩ load                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Current Stimulating Mode<br>± 10% (50µA - 10mA), into<br>1kΩ load<br><br>Voltage Stimulating Mode<br>± 10% (50mV – 10V), into<br>1kΩ load |
|                                              | Mute Sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                           |
| Available                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                       |
|                                              | Operational Environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                           |
| Criteria of<br>Comparison                    | K203684<br>Neurosign® V4 Intraoperative<br>Nerve Monitor<br>(Subject of this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K181559<br>Neurosign® V4<br>Intraoperative Nerve<br>Monitor<br>(Primary Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K053141<br>Neurosign® Nerve Monitor,<br>Model Neurosign® 400 and<br>Model Neurosign® 800®<br>(Secondary Predicate<br>Device)              |
| Operating<br>temperature                     | 5°C to 40°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 5°C to 40°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 5°C to 30°C                                                                                                                               |
| Relative<br>humidity                         | 10% to 80% (non-<br>condensing)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 10% to 80% (non-<br>condensing)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 30% to 70% (non-<br>condensing)                                                                                                           |
| Atmospheric<br>pressure                      | 70kPa to 106kPa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 70kPa to 106kPa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 50kPa to 106kPa                                                                                                                           |

{10}------------------------------------------------

{11}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K203684](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K203684)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
