← Product Code [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ) · K131304

# SENTIO MMG (K131304)

_Innovative Surgical Solutions, LLC ("Iss") · PDQ · May 29, 2014 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K131304

## Device Facts

- **Applicant:** Innovative Surgical Solutions, LLC ("Iss")
- **Product Code:** [PDQ](/submissions/EN/subpart-b%E2%80%94diagnostic-devices/PDQ.md)
- **Decision Date:** May 29, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1820
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.

## Device Story

Sentio MMG is a multichannel intraoperative monitor used during surgeries where motor nerves are at risk. System inputs include mechanomyographic (MMG) signals from muscles innervated by the target nerve, triggered by either operator-applied electrical stimulus or direct mechanical stimulus during surgery. The system consists of a touchscreen PC control unit, patient module, sensor connector module, and disposable probes/sensors. The device processes these signals to assist surgeons in identifying and mapping nerves, alerting them when a nerve is activated to minimize trauma. Used in clinical/surgical settings by surgeons; output is displayed on the touchscreen to guide real-time surgical decision-making and nerve preservation.

## Clinical Evidence

No clinical data. Substantial equivalence was established through non-clinical comparative performance evaluation assessing statistical agreement between the subject and predicate devices.

## Technological Characteristics

Multichannel intraoperative monitor; 10-channel capacity. Uses constant electrical current (1-15 mA) for nerve stimulation; monophasic square pulse (100μs, 2Hz). Employs skin-surface mounted electromechanical MMG sensors. System includes a touchscreen PC control unit and patient module. Connectivity via sensor connector module. Sterilization method not specified.

## Regulatory Identification

A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.

## Predicate Devices

- NeuralMAS ([K100992](/device/K100992.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "sentio" in a sans-serif font. To the left of the word is a four-pointed star-like symbol. The word is in lowercase letters and appears to be a logo or brand name. The overall image has a simple and clean design.

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

- · 1. Submitter's Identification:
Aashish Shah Vice President of Quality & Operations Sentio, LLC 50461 W Pontiac Trail Wixom, MI 48393 Tel: (248) 595-0420 Fax: (248) 956-8123

Date Summary Prepared: April 29th, 2014

#### 2. Name of the Device:

Trade or Proprietary Name: Classification Name: Device Class: Classification: Product Code:

Sentio MMG® Neurosurgical Nerve Locator Class II 21 CFR §874.1820 PDQ

#### Common or Usual Name: 3.

Avoidance Intraoperative Mechanomyographic (MMG) Mapping and System, Nerve Monitor/Stimulator, Nerve Locator/Stimulator

#### Predicate Device Information: .4.

The subject device Sentio MMG® is substantially equivalent to the following predicate device:

K100992 NeuralMAS™, Innovative Surgical Solutions, LLC d/b/a Sentio, LLC, Wixom, Michigan

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "sentio" in a sans-serif font, with a four-pointed star-like symbol to the left of the word. The word is in lowercase letters, and the symbol is slightly larger than the letters. The image is in black and white.

#### Device Description: 5.

The Sentio MMG® system is a multichannel intraoperative monitor for use during surgeries in which a motor nerve is at risk. The Sentio MMG® system records mechanomyographic (MMG) signals from muscles innervated by the affected nerve, which may originate from operator applied electrical stimulus or from direct mechanical stimulus occurring during the course of surgery. The monitor will assist early nerve identification by providing the surgeon with a tool to help locate and identify the particular nerve at risk to minimize trauma by alerting the surgeon when a particular nerve has been activated.

The Sentio MMG® system consists of a reusable Control Unit comprised of a touch-screen PC and Patient Module, a reusable Sensor Connector Module and an assortment of disposable conductive probes, stimulators, sensors, electrodes and electrode leads.

#### Intended Use: ર.

This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.

#### 7. Comparison to Predicate Devices:

The subject device has indications for use which are the same as the predicate device, K100992, is composed of the same materials as the predicate commercially marketed device, has the same or equivalent design features as the predicate device, is substantially equivalent in its method of . use, incorporates the same fundamental scientific technology and has functional characteristics which are the same or equivalent to those of the predicate device. Due to the equivalency of indications for use, materials of composition, design features, method of use, scientific technology and functional characteristics, the device raises no new safety or effectiveness issues.

A comparison of the technological characteristics to the predicate device is provided below:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "sentio" in lowercase letters next to a four-pointed star-like symbol. The word "sentio" is written in a simple, sans-serif font. The star symbol is positioned to the left of the word, slightly above the baseline of the text. 

r

# Table 1: Overview & Indications

| Subject Device:                                                                                                                                                                                                             | Predicate Device:                                                                                                                                                                                                        |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sentio MMG® #K131304                                                                                                                                                                                                        | ISS NeuralMAS #K100992                                                                                                                                                                                                   |
| System Overview<br>Uses low level electrical current to stimulate<br>nerves sufficient to produce a detectable<br>mechanical signal in contracting muscle using<br>skin-surface mounted mechanomyographic<br>(MMG) sensors. | System Overview<br>Uses low level electrical current to stimulate nerves sufficient to produce a<br>detectable mechanical signal in contracting<br>muscle using skin-surface mounted<br>mechanomyographic (MMG) sensors. |

# Table 2: Physical Attributes

| Attribute                               | Sentio MMG®<br>#K131304  | ISS NeuralMAS<br>#K100992 |
|-----------------------------------------|--------------------------|---------------------------|
| Control Module: PC/Display              |                          |                           |
| Type:                                   | Touchscreen              | Touchscreen               |
| Number of Channels:                     | 10                       | 10                        |
| Channel<br>Enable/Disable<br>Controls:  | Touchscreen controlled   | Touchscreen controlled    |
| Event Threshold<br>Control and Display: | Touchscreen controlled   | Touchscreen controlled    |
| Stimulator                              |                          |                           |
| Type Constant:                          | Electrical Current (mA)  | Electrical Current (mA)   |
| Range: Control:                         | 1-15 mA                  | 1-15 mA                   |
| Waveform:                               | Digitally controlled     | Digitally controlled      |
| Pulse Duration:                         | Monophasic, square pulse | Monophasic, square pulse  |
|                                         | 100μς                    | 100μς                     |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "sentio" in a bold, sans-serif font. To the left of the word is a four-pointed star-like symbol. The word is in lowercase letters and appears to be a logo or brand name. 

| Rate:             | 2Hz                         | 2Hz                         |
|-------------------|-----------------------------|-----------------------------|
| Stimulus Probe:   | Monopolar                   | Monopolar                   |
| Activation:       | Touch screen or hand switch | Touch screen or hand switch |
| Monitoring Sensor |                             |                             |
| Function:         | Mechanomyographic (MMG)     | Mechanomyographic (MMG)     |
| Type: Attachment  | Electromechanical           | Electromechanical           |
| Site: Attachment  | Skin surface                | Skin surface                |
| Mode:             | Adhesive                    | Adhesive                    |

## 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

A comparative performance evaluation was conducted to assess statistical agreement between the subject device and the predicate. Results for positive, and overall percent agreement have established that 1) the principle functions of the Sentio MMG® system are effective, such that spinal nerve status and location may be ascertained and monitored, and 2) the data processed by the subject device following surgical use correlate well with those acquired by the predicate device.

#### 9. Discussion of Clinical Tests Performed:

Not Applicable

#### 10. Conclusions:

The conclusions drawn from our non-clinical testing (to include our market data evaluation) of the subject device demonstrates that the subject device is as safe, as effective and performs better than the legally marketed predicate device. Any noted differences between the devices are minor and do not raise new issues of safety and effectiveness. The subject device is substantially equivalent to the legally marketed predicate device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 29, 2014

Innovative Surgical Solutions, LLC Mr. Aashish Shah Vice President Quality & Operations 21520 Bridge Street Southfield, MI 48033

Re: K131304

Trade/Device Name: Sentio MMG Regulation Number: 21 CFR 874.1820 Regulation Name: Neurosurgical Nerve Locator Regulatory Class: Class II Product Code: PDQ Dated: April 29, 2014 Received: April 30, 2014

Dear Mr. Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{5}------------------------------------------------

Page 2 - Mr. Aashish Shah

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aquel - S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510{k} Number (if known) K131304

## Device Name

Sentio MMG

## Indications for Use (Describe)

This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMO) signals and electrical stimulus of nerves. This device is indicated for locating and identifying spinal nerve roots and peripheral motor nerves originating from spinal levels C3-T1 and L2-S2.

Type of Use (Select one or both, as applicable)

∑ Prescription Use (Part 21 CFR 801 Subpart D)

■ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# ം Date: 2014.05.29
22:30:24 -04'00'

EF PSC Publishing Scriptocs (30)) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K131304](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94diagnostic-devices/PDQ/K131304)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
