The Xpert Xpress CoV-2 plus test, performed on the GeneXpert Dx and GeneXpert Infinity Systems, is a rapid real-time RT-PCR test intended for the qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions.
Device Story
Rapid, automated in vitro diagnostic test for SARS-CoV-2 RNA detection; utilizes real-time RT-PCR. Input: nasopharyngeal or anterior nasal swab specimens in VTM/UTM/eNAT. Operation: performed on GeneXpert Instrument Systems (Dx, Infinity-48s, Infinity-80); integrates sample prep, nucleic acid extraction, amplification, and detection in single-use disposable cartridges. Cartridge contains syringe drive, ultrasonic horn for cell lysis, and I-CORE thermocycler. Output: qualitative SARS-CoV-2 result. Used in laboratory settings by trained professionals. Results aid clinical diagnosis of COVID-19; positive results indicate SARS-CoV-2 RNA presence; negative results do not preclude infection. Benefits: rapid (~30 min) detection, automated workflow, minimized cross-contamination via self-contained cartridges.
Clinical Evidence
Multi-site, observational method comparison study (32 US sites). Evaluated 3,750 valid specimens (1,879 NPS, 1,871 NS) from symptomatic individuals. Compared against FDA-cleared molecular respiratory panel. Overall PPA 98.1% (95% CI: 96.7%–98.9%) and NPA 98.3% (95% CI: 97.7%–98.7%). Non-determinate rate 1.6% (NPS) and 1.4% (NS).
Technological Characteristics
Real-time RT-qPCR assay. Targets: SARS-CoV-2 E, N2, and RdRP genes. Includes Sample Processing Control (SPC) and Probe Check Control (PCC). Form factor: single-use disposable cartridge. Instrument: GeneXpert Systems (automated fluidics, ultrasonic lysis, I-CORE thermocycler). Connectivity: networked/standalone instrument systems. Software: preloaded on system computer.
Indications for Use
Indicated for qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens from symptomatic individuals with signs/symptoms of respiratory tract infection. For use as an aid in COVID-19 diagnosis in conjunction with clinical/epidemiologic/laboratory findings.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
K242109 — Xpert® Xpress CoV-2 plus (XPRS-COV2-10) · Cepheid® · Jan 15, 2025
K240867 — cobas® SARS-CoV-2 Qualitative for use on the cobas® 5800/6800/8800 Systems · Roche Molecular Systems, Inc. · Feb 11, 2025
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K230440
B Applicant
Cepheid
C Proprietary and Established Names
Xpert Xpress CoV-2 plus
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QQX | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology |
| OOI | Class II | 21 CFR 862.2570 - Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for the Xpert Xpress CoV-2 plus test performed on the Cepheid GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s and GeneXpert Infinity-80 systems).
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Measurand:
Nucleocapsid (N2), envelope (E), and RNA-dependent RNA polymerase (RdRP) genes of SARS-CoV-2.
C Type of Test:
The Xpert Xpress CoV-2 plus test is a real-time reverse transcription polymerase chain reaction (RT-PCR) based in vitro diagnostic test for the qualitative detection of viral RNA from SARS-CoV-2 in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens.
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Xpert Xpress CoV-2 plus test, performed on the GeneXpert Dx and GeneXpert Infinity Systems, is a rapid real-time RT-PCR test intended for the qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection.
The Xpert Xpress CoV-2 plus test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens.
Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use Only
D Special Instrument Requirements:
The Xpert Xpress CoV-2 plus test is performed on the Cepheid GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s and GeneXpert Infinity-80s systems).
IV Device/System Characteristics:
A Device Description:
The Xpert Xpress CoV-2 plus test is a rapid, automated in vitro diagnostic test for the qualitative detection of viral RNA from SARS-CoV-2 in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens obtained from individuals with signs and symptoms of respiratory tract infection.
The Xpert Xpress CoV-2 plus test is performed on the Cepheid GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s and GeneXpert Infinity-80 systems), which consist of an
K230440 - Page 2 of 24
{2}
instrument, computer and preloaded software for running tests and viewing the results. The GeneXpert Instrument Systems automate and integrate sample preparation, nucleic acid extraction and amplification, and detection of the target sequences in simple or complex samples using real-time reverse transcription (RT)-polymerase chain reaction (PCR) and PCR assays. Depending on the instrument, the GeneXpert Instrument Systems can have from 1 and up to 80 randomly accessible modules, each capable of performing separate sample preparation and real-time RT-PCR and PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE thermocycler for performing real-time RT-PCR and PCR as well as detection. The systems require the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between cartridges during the testing process is minimized.
The Xpert Xpress CoV-2 plus test includes reagents for the detection of viral RNA from SARS-CoV-2 in NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2 plus test are designed to amplify and detect sequences in the genes that encode the following SARS-CoV-2 proteins: nucleocapsid (N2), envelope (E), and RNA-dependent RNA polymerase (RdRP). A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.
The Xpert Xpress CoV-2 plus test is designed for use with NPS or NS specimen collected with nylon flocked swabs and placed into a viral transport medium (VTM), Universal Transport Medium (UTM) or eNAT.
## B Principle of Operation:
The Xpert Xpress CoV-2 plus test utilizes automated reverse transcription cDNA synthesis from RNA and real-time polymerase chain reaction (PCR) for gene specific sequence amplification. Detection of the amplified DNA is by fluorogenic target-specific probe hybridization followed by 5'-nuclease cleavage of the probe to release the fluorophore (TaqMan).
The technology utilizes lyophilized reagents in the form of freeze-dried beads, which provide all the components (enzymes, primers, probes, salts, buffers and divalent metal ion) to support PCR when reconstituted.
## C Instrument Description Information:
1. Instrument Name: Cepheid GeneXpert Instrument Systems [GeneXpert Dx (versions 4.7b or higher), GeneXpert Infinity-48s and GeneXpert Infinity-80s systems (versions 6.4b or higher)].
2. Specimen Identification: Anterior nasal (NS) or nasopharyngeal (NPS) swab specimen identification is automated.
K230440 - Page 3 of 24
{3}
K230440 - Page 4 of 24
3. Specimen Sampling and Handling:
The NPS or NS specimen is collected and placed into a transport tube containing 3 mL of Viral Transport Medium (VTM)/ Universal Transport Medium (UTM) or 2 mL of eNAT. The specimen is briefly mixed by rapidly inverting the collection tube 5 times. Using the supplied transfer pipette, the specimen is transferred to the sample chamber of the Xpert Xpress CoV-2 plus cartridge. The GeneXpert cartridge is loaded onto the GeneXpert instrument, which performs hands-off, automated specimen processing, and real-time RT-PCR for detection of viral RNA.
4. Calibration:
GeneXpert instruments are calibrated at the factory. Routine calibration of the GeneXpert Instrument systems may be performed by Cepheid Field Service Engineers during annual maintenance.
5. Quality Control:
Internal Controls: Each cartridge includes a Sample Processing Control (SPC) and Probe Check Control (PCC).
- Sample Processing Control (SPC) – The SPC is an encapsulated RNA pseudovirus that ensures adequate lysis of target virus, monitors the presence of PCR inhibitors, and verifies the use of proper PCR conditions. The SPC passes if it meets the validated acceptance criteria. The SPC result should be PASS in a sample that is negative for all three SARS-CoV-2 target analytes and can be NEGATIVE or POSITIVE in a sample containing detectable levels of one or more of the SARS-CoV-2 target analytes.
- Probe Check Control (PCC) – Before the start of the PCR reaction, the GeneXpert System measures the fluorescence signal from the probes to monitor bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC passes if it meets the validated acceptance criteria.
External Controls:
External controls should be used in accordance with local, state, and federal accrediting organizations as applicable.
V Substantial Equivalence Information:
A Predicate Device Name(s):
BioFire COVID-19 Test 2
B Predicate 510(k) Number(s):
K211079
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K230440 | K211079 |
| --- | --- | --- |
| Device Trade Name | Xpert Xpress CoV-2 plus | BioFire COVID-19 Test 2 |
| Product Code | Same | QQX |
{4}
K230440 - Page 5 of 24
| | | Respiratory Specimen Nucleic Acid SARS-CoV-2 Test |
| --- | --- | --- |
| Regulation Number/Name | Same | 21 CFR 866.3981
Devices to detect and identify nucleic acid targets in respiratory samples from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-analyte test |
| Device Class | Same | II (Special Controls) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Xpert Xpress CoV-2 *plus* test, performed on the GeneXpert Dx and GeneXpert Infinity Systems, is a rapid real-time RT-PCR test intended for the qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection.
The Xpert Xpress CoV-2 *plus* test is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens.
Negative results do not preclude SARS-CoV-2 infection. The results of this test should not be used as the sole basis for diagnosis and patient management decisions. | The BioFire COVID-19 Test 2 is a qualitative nested multiplexed RT-PCR in vitro diagnostic test intended for use with the BioFire FilmArray 2.0 and BioFire FilmArray Torch Systems. The BioFire COVID-19 Test 2 detects nucleic acids from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in nasopharyngeal swabs (NPS) from symptomatic individuals suspected of COVID-19 by their healthcare provider.
Results are for the identification of SARSCoV-2 RNA. The SARS-CoV-2 RNA is generally detectable in NPS specimens during the acute phase of infection. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive results do not rule out co-infection with other pathogens.
Results are meant to be used in conjunction with other clinical, epidemiologic, and laboratory data, in accordance with the guidelines provided by the relevant public |
{5}
| | | health authorities. The BioFire COVID-19 Test 2 is intended for use by trained medical and laboratory professionals in a laboratory setting or under the supervision of a trained laboratory professional. |
| --- | --- | --- |
| Analyte Target | Same | SARS-CoV-2 |
| Control | Same | Internal Controls |
| Test Format | Same | Single Use |
| Result Interpretation | Same | Automated |
| Test Result | Same | Qualitative |
| Automation | Same | Automated Nucleic Acid Extraction, Detection and Results Interpretation |
| Intended Environment for Use | Same | Professional use |
| Prescription Use Only | Same | Yes |
| General Device Characteristic Differences | | |
| Technology | Real-time reverse transcription polymerase chain reaction (RT-qPCR). | Nested multiplex RT-PCR followed by high resolution melting analysis to confirm identity of amplified product. |
| SARS-CoV-2 Target Type | E, N2, RdRP | Multiple Targets |
| Specimen Type | • Nasopharyngeal swab (NPS)
• Anterior nasal swab (NS) | Nasopharyngeal swab (NPS) |
| Transport Media | • Universal Transport Medium (UTM) / Viral Transport Medium (VTM)
• eNAT | • Transport Medium
• Saline |
| Instrumentation | Cepheid GeneXpert Instrument Systems | FilmArray 2.0 or FilmArray Torch systems |
| Internal Control | • Sample Processing Control (SPC)
• Probe Check Control (PCC) | • Sample Processing Control
• PCR and Melt Analysis Control |
| Time to Result | ~ 30 minutes | ~ 45 minutes |
VI Standards/Guidance Documents Referenced:
- Class II Special Controls as per 21 CFR 866.3981
- Guidance for Industry and FDA Staff: General Principles of Software Validation, issued January 11, 2002.
K230440 - Page 6 of 24
{6}
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005.
- Guidance for Industry and FDA Staff: Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests, issued on March 13, 2007.
- Clinical and Laboratory Standards Institute (CLSI) documents: EP05-A3 (Evaluation of Precision of Quantitative Measurement Procedures, 3rd Edition).
- Clinical and Laboratory Standards Institute (CLSI) documents: EP07-A3 (Interference Testing in Clinical Chemistry, 3rd Edition).
- Clinical and Laboratory Standards Institute (CLSI) documents: EP12-A2 (User Protocol For Evaluation Of Qualitative Test Performance; Approved Guideline, $2^{\text{nd}}$ Edition).
- Clinical and Laboratory Standards Institute (CLSI) documents: EP17-A2 (Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, 2nd Edition).
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
# 1. Reproducibility and Within-Laboratory Precision Study:
# a) Reproducibility Study
The reproducibility of the Xpert Xpress CoV-2 plus test was established at three sites (two external and one internal) using a three-member panel, including one negative sample, one low positive ( $\sim 1.5\mathrm{x}$ LoD) sample and one moderate positive ( $\sim 3\mathrm{x}$ LoD) sample. The negative sample consisted of simulated NS/NPS matrix without target microorganism or target RNA. The positive samples were contrived samples in a simulated NS/NPS matrix prepared by spiking the NATtrol inactivated SARS-CoV-2 virus strain USA-WA1/2020 (ZeptoMetrix). Testing was conducted over six days with the GeneXpert Dx and Infinity systems, using three lots of Xpert Xpress CoV-2 plus cartridges at three participating sites each with two operators to yield a total of 144 observations per panel member (3 Sites x 2 Operators x 3 Lots x 2 Days/Lot x 2 Runs x 2 Replicates = 144 observations/panel member). The external controls, consisting of one positive and one negative control, were tested prior to testing panel samples at the site on each day of the study.
The percent agreement of the qualitative results for SARS-CoV-2 detection for each panel member analyzed by each of the six operators and each site is shown in Table 1. In addition, the overall percent agreement for each sample (% total agreement) and the two-sided Wilson Score confidence intervals (CI) are presented in the last column.
Table 1: Summary of the Reproducibility Study Results - % Agreement
| Panel Member | Site 1 | | | Site 2 | | | Site 3 | | | % Total Agreement and 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Op1 | Op2 | Site | Op1 | Op2 | Site | Op1 | Op2 | Site | |
| Negative | 100% (24/24) | 95.8% (23/24) | 97.9% (47/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (23/23) * | 100% (47/47) | 99.3% (142/143) [96.1% - 99.9%] |
| SARS-CoV-2 Low Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) [97.4% - 100%] |
K230440 - Page 7 of 24
{7}
The evaluation of reproducibility and within-laboratory precision of the underlying Ct values for E, N2 and RdRP obtained in the Xpert Xpress CoV-2 plus test was performed. The mean, standard deviation (SD), and coefficient of variation (CV) between-sites, between-operators, between-lots, between-days, between-runs and within-run for each panel member are presented in Table 2.
Table 2: Summary of Reproducibility Results – Nested ANOVA by Coefficient of Variation
| Sample | Analyte | N | Mean Ct | Between Sites | | Between Operators | | Between Lots | | Between Days | | Between Runs | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| Negative | SPC | 143 | 31.1 | 0.29 | 0.9 | 0.16 | 0.5 | 0.42 | 1.3 | 0 | 0 | 0 | 0 | 1.47 | 4.7 | 1.56 | 5.0 |
| SARS-CoV-2 Low Pos | E | 144 | 34.0 | 0.06 | 0.2 | 0.19 | 0.6 | 0 | 0 | 0.14 | 0.4 | 0 | 0 | 0.55 | 1.6 | 0.60 | 1.8 |
| | N2 | 144 | 37.5 | 0.12 | 0.3 | 0 | 0 | 0.2 | 0.5 | 0.11 | 0.3 | 0 | 0 | 0.68 | 1.8 | 0.73 | 1.9 |
| | RdRP | 144 | 36.3 | 0.07 | 0.2 | 0 | 0 | 0.25 | 0.7 | 0.06 | 0.2 | 0 | 0 | 0.61 | 1.7 | 0.66 | 1.8 |
| SARS-CoV-2 Mod Pos | E | 144 | 32.5 | 0.31 | 0.9 | 0.03 | 0.1 | 0 | 0 | 0.47 | 1.4 | 0 | 0 | 1.86 | 5.7 | 1.95 | 6.0 |
| | N2 | 144 | 36.2 | 0.16 | 0.4 | 0 | 0 | 0.29 | 0.8 | 0.17 | 0.5 | 0 | 0 | 0.65 | 1.8 | 0.75 | 2.1 |
| | RdRP | 144 | 34.8 | 0.33 | 1.0 | 0 | 0 | 0 | 0 | 0.62 | 1.8 | 0 | 0 | 2.18 | 6.2 | 2.29 | 6.6 |
b) Within-Laboratory Precision Study
The within-laboratory precision of the Xpert Xpress CoV-2 plus test was evaluated at a single site using a three-member panel, including one negative sample, one low positive (~1.5x LoD) sample and one moderate positive (~3x LoD) sample. The negative sample consisted of simulated NS/NPS swab matrix without target microorganism or target RNA. The positive samples were contrived samples in a simulated NS/NPS matrix prepared by spiking the NATtrol inactivated SARS-CoV-2 virus strain USA-WA1/2020 (ZeptoMetrix).
Testing was conducted over 20 days, using one lot of Xpert Xpress CoV-2 plus cartridges at a single site and with one operator to yield a total of eighty observations per panel member (1 Site x 1 Operator x 1 Lot x 20 Days x 2 Runs x 2 Replicates = 80 observations/panel member). The agreement of the qualitative results for SARS-CoV-2 detection for each panel member, including the overall percent agreement for each sample (% total agreement) and the two-sided Wilson Score confidence intervals (CI), are presented in Table 3. The mean, standard deviation (SD), and coefficient of variation (CV) between-days, between-runs, and within-run for each panel member are presented in Table 4.
Table 3: Summary of the With-in Laboratory Precision Results – % Agreement
| Panel Member | Agreement | % Total Agreement and 95% CI by Panel Member |
| --- | --- | --- |
| Negative | 80/80 | 100% (95.4%-100.0%) |
| SARS-CoV-2 Low Positive | 80/80 | 100% (95.4%-100.0%) |
| SARS-CoV-2 Mod Positive | 80/80 | 100% (95.4%-100.0%) |
K230440 - Page 8 of 24
{8}
Table 4: Within-Laboratory Precision - Nested ANOVA by Coefficient of Variation
| Sample | Analyte of Ct Values | N | Mean Ct | Between Days | | Between Runs | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV | SD | CV | SD | CV | SD | CV |
| Negative | SPC | 80 | 30.34 | 0.26 | 0.8 | 0.00 | 0.0 | 1.03 | 3.4 | 1.06 | 3.5 |
| SARS-CoV-2 ~1.5 x LoD | E | 80 | 34.12 | 0.00 | 0.0 | 0.21 | 0.6 | 0.5 | 1.5 | 0.55 | 1.6 |
| | N2 | 80 | 37.66 | 0.00 | 0.0 | 0.11 | 0.3 | 0.57 | 1.5 | 0.58 | 1.5 |
| | RdRP | 80 | 36.77 | 0.00 | 0.0 | 0.00 | 0.0 | 0.66 | 1.8 | 0.66 | 1.8 |
| SARS-CoV-2 ~3 x LoD | E | 80 | 33.37 | 0.00 | 0.0 | 0.00 | 0.0 | 1.03 | 3.1 | 1.03 | 3.1 |
| | N2 | 80 | 36.89 | 0.17 | 0.4 | 0.00 | 0.0 | 1.02 | 2.8 | 1.03 | 2.8 |
| | RdRP | 80 | 35.99 | 0.27 | 0.7 | 0.37 | 1.0 | 0.95 | 2.6 | 1.05 | 2.9 |
*CV = SD/Mean Ct * 100
# 2. Linearity:
Not applicable; this is a qualitative assay.
# 3. Analytical Reactivity:
# a) In silico Inclusivity Analysis
i) The inclusivity of Xpert Xpress CoV-2 plus primers was evaluated on June $30^{\text{th}}$ , 2022 using in silico analysis of the assay amplicons in relation to 11,650,640 SARS-CoV-2 sequences available in the GISAID gene database for three targets, E, N2 and RdRP. The 11,650,640 SARS-CoV-2 sequences were separated into the lineages of interest based on the Pango Lineage assigned to each genome by GISAID, and those with ambiguous nucleotides were removed. Thus, the following inclusivity analyses focuses on the combined, non-ambiguous sequences from the variants of interest and variants of concern as of June $30^{\text{th}}$ , 2022. These constituted 10,469,612 sequences for the E target, 10,587,381 sequences for the N2 target and 10,333,656 sequences for the RdRP target. Table 5 summarizes the effective predicted inclusivity for E, N2 and RdRP amplicons for the variants of interests and concern.
Table 5: Predicted Inclusivity for E, N2 and RdRP Amplicons for SARS-CoV-2 Variants of Interests and Concern
| SARS-CoV-2 Target Amplicon | Exact Match | 1 Mismatch a | 2 or More Mismatches | Predicted Inclusivity |
| --- | --- | --- | --- | --- |
| E | 10,420,248 of 10,469,612 total (99.5%) | 48,562 (0.5%) | 802 (0.01%) | 100% |
| N2 | 10,386,068 of 10,587,381 total (98.1%) | 196,336 (1.9%) | 4,977 (0.05%) | 99.95% |
| RdRP | 10,247,146 of 10,333,656 total (99.2%) | 85,373 (0.8%) | 1,137 (0.01%) | 100% |
a. Single-nucleotide mismatches are predicted to not impact the performance of the test.
K230440 - Page 9 of 24
{9}
The in silico inclusivity of the Xpert Xpress CoV-2 plus probe oligonucleotides for E, N2 and RdRP were also assessed against the top 20 most frequent matches in the GISAID EpiCoV sequence database as of June 15th, 2022, which constituted 10,310,839 for the E target, 10,428,014 for the N2 target and 10,178,602 for the RdRP target. For each of the probe oligonucleotides used in the Xpert Xpress CoV-2 plus test, Table 6 summarizes the number sequences as well as the corresponding percentage of sequences from this dataset with exact match, 1 mismatch/insertion, and 2 or more mismatches/insertions in the alignment.
Table 6: Predicted Inclusivity for E, N2 and RdRP Probes for SARS-CoV-2 Variants of Interests and Concern
| SARS-CoV-2 Target Probe | Exact Match | 1 Mismatch/Insertiona | 2 or More Mismatches/Insertions | Predicted Inclusivity |
| --- | --- | --- | --- | --- |
| E | 10,300,688 of 10,310,839 total (99.9%) | 9,853 (0.1%) | 22 (0.0002%) | 100% |
| N2 | 10,351,581 of 10,428,014 total (99.3%) | 72,957 (0.7%) | 0 (0%) | 100% |
| RdRP | 0 | 10,140,254 of 10,178602 total (99.6%) | 37,492 (0.4%) | 99.6% |
a. Single-nucleotide mismatches/insertions are predicted to not impact the performance of the test.
ii) Additional evaluation of inclusivity of Xpert Xpress CoV-2 plus primers was conducted on July 31, 2023 using in silico analysis of the assay amplicons in relation to SARS-CoV-2 sequences available in the GISAID gene database for three targets, E, N2 and RdRP. U.S.-specific sequences were then partitioned into the four time periods of 14, 30, 90 and 180 days (Tables 7-10) prior to the cutoff date of July 31, 2023 and were stratified by number and frequency of sequences with exact or mismatched sequences. The in silico inclusivity statistics for each time period are presented in Tables 7-10. In each of these tables, the results for the two N2 probes were combined into a single row which represents the best match of either probe.
Table 7: Inclusivity Statistic - Sequences Collected Within 14 Days Prior to July 31, 2023
| SARS-CoV-2 Target | | In Silico Inclusivity Statistics | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| E | Forward | 97.7% (173/177) | 2.3% (4/177) | 0.0% (0/177) | 0.0% (0/177) | 0.0% (0/177) |
| | Probe | 100.0% (177/177) | 0.0% (0/177) | 0.0% (0/177) | 0.0% (0/177) | 0.0% (0/177) |
| | Reverse | 100.0% (177/177) | 0.0% (0/177) | 0.0% (0/177) | 0.0% (0/177) | 0.0% (0/177) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| N2 | Forward | 97.7% (172/176) | 2.3% (4/176) | 0.0% (0/176) | 0.0% (0/176) | 0.0% (0/176) |
| | Probe | 100% (176/176) | 0.0% (0/176) | 0.0% (0/176) | 0.0% (0/176) | 0.0% (0/176) |
| | Reverse | 99.4% (175/176) | 0.6% (1/176) | 0.0% (0/176) | 0.0% (0/176) | 0.0% (0/176) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| RdRP | Forward | 99.4% (172/173) | 0.6% (1/173) | 0.0% (0/173) | 0.0% (0/173) | 0.0% (0/173) |
| | Reverse | 100.0% (173/173) | 0.0% (0/173) | 0.0% (0/173) | 0.0% (0/173) | 0.0% (0/173) |
| | Probe | 1 Insertion* | 1 Insertion + 1 Mismatch | 1 Insertion + 2 Mismatches | 1 Insertion + 3 Mismatches | 1 Insertion + > 3 Mismatches |
| | | 99.4% (172/173) | 0.6% (1/173) | 0.0% (0/173) | 0.0% (0/173) | 0.0% (0/173) |
* RdRp probe design lacks dC at position 14 from the 5'-end of the oligonucleotide.
As shown in Table 7, there are two instances where a specific oligonucleotide was observed to have a mismatch frequency greater than $1\%$ in the 14-day period. The total number of sequences collected in the 14-day period is small, which can be a source of bias and may not be representative of the test performance. Consistent with this account, there were no instances where a specific primer/probe sequence was observed with greater than $1\%$ frequency of
K230440 - Page 10 of 24
{10}
mismatch in any of the longer time stratifications (30 days, 90 days and 180 days), which would warrant a risk assessment for adverse impact to the test performance. Based on the built-in redundancy of the Xpert Xpress CoV-2 plus test's SARS-CoV-2 amplification system (i.e., 3 independent targets, only 1 of 3 must be detected to assign a positive result), it is not anticipated that any of the evaluated sequences would be missed by the Xpert Xpress CoV-2 plus test.
Table 8: Inclusivity Statistic - Sequences Collected Within 30 Days Prior to July 31, 2023
| SARS-CoV-2 Target | | In Silico Inclusivity Statistics | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| E | Forward | 99.2% (2505/2526) | 0.8% (21/2526) | 0.0% (0/2526) | 0.0% (0/2526) | 0.0% (0/2526) |
| | Probe | 99.8% (2522/2526) | 0.2% (4/2526) | 0.0% (0/2526) | 0.0% (0/2526) | 0.0% (0/2526) |
| | Reverse | 100.0% (2525/2526) | 0.0% (1/2526) | 0.0% (0/2526) | 0.0% (0/2526) | 0.0% (0/2526) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| N2 | Forward | 99.2% (2509/2528) | 0.8% (19/2528) | 0.0% (0/2528) | 0.0% (0/2528) | 0.0% (0/2528) |
| | Probe | 99.8% (2522/2528) | 0.2% (6/2528) | 0.0% (0/2528) | 0.0% (0/2528) | 0.0% (0/2528) |
| | Reverse | 99.3% (2510/2528) | 0.7% (18/2528) | 0.0% (0/2528) | 0.0% (0/2528) | 0.0% (0/2528) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| RdRp | Forward | 99.6% (2488/2499) | 0.4% (11/2499) | 0.0% (0/2499) | 0.0% (0/2499) | 0.0% (0/2499) |
| | Reverse | 99.9% (2497/2499) | 0.1% (2/2499) | 0.0% (0/2499) | 0.0% (0/2499) | 0.0% (0/2499) |
| | Probe | 1 Insertion* | 1 Insertion + 1 Mismatch | 1 Insertion + 2 Mismatches | 1 Insertion + 3 Mismatches | 1 Insertion + > 3 Mismatches |
| | | 99.5% (2486/2499) | 0.5% (13/2499) | 0.0% (0/2499) | 0.0% (0/2499) | 0.0% (0/2499) |
* RdRp probe design lacks $\mathrm{dC}$ at position 14 from the $5^{\prime}$ -end of the oligonucleotide.
Table 9: Inclusivity Statistic - Sequences Collected Within 90 Days Prior to July 31, 2023
| SARS-CoV-2 Target | | In Silico Inclusivity Statistics | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| E | Forward | 99.3% (25156/25321) | 0.7% (165/25321) | 0.0% (0/25321) | 0.0% (0/25321) | 0.0% (0/25321) |
| | Probe | 99.9% (25292/25321) | 0.1% (29/25321) | 0.0% (0/25321) | 0.0% (0/25321) | 0.0% (0/25321) |
| | Reverse | 99.9% (25304/25321) | 0.1% (16/25321) | 0.0% (1/25321) | 0.0% (0/25321) | 0.0% (0/25321) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| N2 | Forward | 99.2% (25272/25483) | 0.8% (211/25483) | 0.0% (0/25483) | 0.0% (0/25483) | 0.0% (0/25483) |
| | Probe | 99.6% (25389/25483) | 0.4% (93/25483) | 0.0% (1/25483) | 0.0% (0/25483) | 0.0% (0/25483) |
| | Reverse | 99.5% (25353/25483) | 0.5% (130/25483) | 0.0% (0/25483) | 0.0% (0/25483) | 0.0% (0/25483) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| RdRp | Forward | 99.7% (25216/25302) | 0.3% (86/25302) | 0.0% (0/25302) | 0.0% (0/25302) | 0.0% (0/25302) |
| | Reverse | 99.8% (25242/25302) | 0.2% (60/25302) | 0.0% (0/25302) | 0.0% (0/25302) | 0.0% (0/25302) |
| | Probe | 1 Insertion* | 1 Insertion + 1 Mismatch | 1 Insertion + 2 Mismatches | 1 Insertion + 3 Mismatches | 1 Insertion + > 3 Mismatches |
| | | 99.6% (25192/25302) | 0.4% (110/25302) | 0.0% (0/25302) | 0.0% (0/25302) | 0.0% (0/25302) |
* RdRp probe design lacks $\mathrm{dC}$ at position 14 from the $5^{\prime}$ -end of the oligonucleotide.
K230440 - Page 11 of 24
{11}
Table 10: Inclusivity Statistic - Sequences Collected Within 180 Days Prior to July 31, 2023
| SARS-CoV-2 Target | | In Silico Inclusivity Statistics | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| E | Forward | 99.5% (147492/148212) | 0.5% (717/148212) | 0.0% (3/148212) | 0.0% (0/148212) | 0.0% (0/148212) |
| | Probe | 99.8% (147963/148212) | 0.2% (249/148212) | 0.0% (0/148212) | 0.0% (0/148212) | 0.0% (0/148212) |
| | Reverse | 99.9% (148106/148212) | 0.1% (105/148212) | 0.0% (1/148212) | 0.0% (0/148212) | 0.0% (0/148212) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| N2 | Forward | 99.1% (149625/150916) | 0.9% (1287/150916) | 0.0% (4/150916) | 0.0% (0/150916) | 0.0% (0/150916) |
| | Probe | 99.8% (150597/150916) | 0.2% (315/150916) | 0.0% (4/150916) | 0.0% (0/150916) | 0.0% (0/150916) |
| | Reverse | 99.6% (150373/150916) | 0.4% (543/150916) | 0.0% (0/150916) | 0.0% (0/150916) | 0.0% (0/150916) |
| | | Exact Match | 1 Mismatch | 2 Mismatches | 3 Mismatches | 3+ Mismatches |
| RdRp | Forward | 99.8% (146465/146703) | 0.2% (238/146703) | 0.0% (0/146703) | 0.0% (0/146703) | 0.0% (0/146703) |
| | Reverse | 99.9% (146489/146703) | 0.1% (214/146703) | 0.0% (0/146703) | 0.0% (0/146703) | 0.0% (0/146703) |
| | Probe | 1 Insertion* | 1 Insertion + 1 Mismatch | 1 Insertion + 2 Mismatches | 1 Insertion + 3 Mismatches | 1 Insertion + > 3 Mismatches |
| | | 99.7% (146204/146703) | 0.3% (499/146703) | 0.0% (0/146703) | 0.0% (0/146703) | 0.0% (0/146703) |
* RdRp probe design lacks dC at position 14 from the $5^{\prime}$ -end of the oligonucleotide.
# b) Inclusivity Wet-Testing
In addition to the in silico analysis of the SARS-CoV-2 primers and probes, the inclusivity of the Xpert Xpress CoV-2 plus test was evaluated by wet testing against multiple strains of SARS-CoV-2 at $\sim 3\mathrm{x}$ LoD. A total of 61 strains comprised of 23 intact SARS-CoV-2 viral particles (18 inactivated viral cultures and 5 BSL-3 live viral cultures), and 38 SARS-CoV-2 in vitro RNA transcripts representing variant strains were tested in this study with the Xpert Xpress CoV-2 plus test. Three replicates were tested for each strain. All SARS-CoV-2 strains tested positive in all three replicates. Results are shown in Table 11.
Table 11: Analytical Reactivity (Inclusivity) of the Xpert Xpress CoV-2 plus Test
| Count No. | SARS-CoV-2 Strain | Tested Concentration | SARS-CoV-2 Test Results | Number of Positive Results Obtained out of the Total Number of Valid Replicates | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | E | N2 | RdRP |
| 1 | 2019-nCoV/Italy-INMI1a | 5 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 2 | England/204820464/2020a | 0.5 TCID50/mL | POSe | 3/3 | 3/3 | 3/3 |
| 3 | Hong Kong/VM20001061/2020a | 0.25 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 4 | South Africa/KRISP-K005325/2020a | 0.25 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 5 | USA/CA_CDC_5574/2020a | 0.25 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 6 | USA/MD-HP01542/2021a | 1.3e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 7 | USA/GA-EHC-2811C/2021a | 1.3e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 8 | NY-Wadsworth-103677-01/2020a | 0.15 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 9 | NY-Wadsworth-21006055-01/2021a | 0.04 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 10 | NY-Wadsworth-21025952-01/2021a (Isolate 1) | 0.625 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 11 | NY-Wadsworth-21018781-01/2021a (Isolate 2) | 0.625 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 12 | NY-Wadsworth-21033899-01/2021a | 1.25 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 13 | NY-Wadsworth-33126-01/2020a | 0.15 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
K230440 - Page 12 of 24
{12}
K230440 - Page 13 of 24
| Count No. | SARS-CoV-2 Strain | Tested Concentration | SARS-CoV-2 Test Results | Number of Positive Results Obtained out of the Total Number of Valid Replicates | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | E | N2 | RdRP |
| 14 | USA/CA-Stanford-15_S02/2021^{a} | 0.04 TCID_{50}/mL | POS | 3/3 | 3/3 | 3/3 |
| 15 | USA/PHC658/2021^{a} | 25 TCID_{50}/mL | POS | 3/3 | 3/3 | 3/3 |
| 16 | Japan/TY7-503/2021^{a} | 0.75 TCID_{50}/mL | POS | 3/3 | 3/3 | 3/3 |
| 17 | hCoV-19/USA/MD-HP30386/2022^{a} | 27.5 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 18 | hCoV-19/USA/COR-22-063113/2022^{a} | 19 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 19 | Australia/VIC01/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 20 | Belgium/ULG/10004/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 21 | England/205041766/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 22 | England/MILK-9E05B3/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 23 | England/SHEF-C05B2/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 24 | France/HF2393/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 25 | Hong Kong/HKU-211129-001/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 26 | Iceland/5/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 27 | India/CT-ILSGS00361/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 28 | India/MH-NCCS-P1162000182735/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 29 | India/MH-SEQ-221_S66_R1_001/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 30 | Japan/Hu_DP_Kng_19-020/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 31 | Japan/IC-0564/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 32 | Portugal/PT9543/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 33 | South Africa/KRISP-EC-K005299/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 34 | Taiwan/NTU02/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 35 | USA/CA9/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 36 | USA/CA-PC101P/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 37 | USA/CA-CZB-12943/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 38 | USA/MN2-MDH2/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 39 | USA/NY-MSHSPSP-PV24650/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 40 | USA/TX1/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 41 | USA/WA2/2020^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 42 | USA/WA-CDC-UW21061750277/2021^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 43 | Wuhan-Hu-1^{b} | 1.2e3 copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 44 | Germany/BavPat1/2020^{c} | 1.2e3 genome copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 45 | Singapore/2/2020^{c} | 300 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 46 | USA/AZ1/2020^{c} | 300 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 47 | USA/CA1/2020^{c} | 1.2e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 48 | USA/CA2/2020^{c} | 1.2e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
{13}
| Count No. | SARS-CoV-2 Strain | Tested Concentration | SARS-CoV-2 Test Results | Number of Positive Results Obtained out of the Total Number of Valid Replicates | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | E | N2 | RdRP |
| 49 | USA/CA3/2020c | 400 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 50 | USA/CA4/2020c | 1.2e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 51 | USA/IL1/2020c | 150 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 52 | USA/New York-PV08001/2020c | 1.2e3 genome copies/mL | POS | 3/3 | 3/3 | 3/3 |
| 53 | USA/New York-PV08410/2020c | 300 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 54 | USA/New York-PV08449/2020c | 300 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 55 | USA/New York-PV09158/2020c | 300 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 56 | USA/WI1/2020c | 1.2e3 genome equivalents/mL | POS | 3/3 | 3/3 | 3/3 |
| 57 | hCoV-19/USA/MD-HP38861/2022d | 0.29 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 58 | hCoV-19/USA/MD-HP38960/2022d | 0.043 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 59 | hCoV-19/USA/CO-CDPHE-2102544747/2021d | 0.028 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 60 | hCoV-19/USA/MD-HP38288/2022d | 0.24 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
| 61 | hCoV-19/USA/MD-HP40900/2022d | 0.72 TCID50/mL | POS | 3/3 | 3/3 | 3/3 |
a. Inactivated viral culture fluid (intact viral particles)
b. Synthetic, in vitro RNA transcript, TWIST controls
c. Genomic RNA
d. BSL-3 live viral culture fluid (intact viral particles)
e. One of 3 replicates reported ERROR. The run was successfully repeated to obtain 3 valid replicates.
## 4. Analytical Specificity/Interference:
### a) In Silico Analysis
The analytical specificity/cross-reactivity of the Xpert Xpress CoV-2 plus plan included evaluation of the SARS-CoV-2 test primer and probes with potentially cross-reactive microorganisms by in silico analysis. The analysis was conducted by mapping the primers and probes of the Xpert Xpress CoV-2 plus individually to the microorganism sequences downloaded from the NCBI database. E primers and probes are not specific for SARS-COV-2 and will detect Human and Bat SARS-coronavirus. No potential unintended cross reactivity with other organisms listed in Table 12 is expected based on the in silico analysis.
Table 12: Microorganisms Analyzed in the in silico Analysis for the SARS-CoV-2 Target
| Microorganisms from the Same Genetic Family | High Priority Organisms |
| --- | --- |
| Human coronavirus 229E | Adenovirus (e.g., C1 Ad. 71) |
| Human coronavirus OC43 | Cytomegalovirus |
| Human coronavirus HKU1 | Enterovirus (e.g., EV68) |
| Human coronavirus NL63 | Epstein-Barr virus |
| SARS-coronavirus | Human Metapneumovirus (hMPV) |
| MERS-coronavirus | Influenza A & B |
| Bat coronavirus | Measles |
| | Mumps |
| | Parainfluenza virus 1-4 |
K230440 - Page 14 of 24
{14}
| Microorganisms from the Same Genetic Family | High Priority Organisms |
| --- | --- |
| | Parechovirus |
| | Respiratory syncytial virus |
| | Rhinovirus |
| | Bacillus anthracis (Anthrax) |
| | Bordetella pertussis |
| | Bordetella parapertussis |
| | Chlamydia pneumoniae |
| | Chlamydia psittaci |
| | Corynebacterium diphtheriae |
| | Coxiella burnetii (Q-Fever) |
| | Escherichia coli |
| | Fusobacterium necrophorum |
| | Haemophilus influenzae |
| | Lactobacillus sp. |
| | Legionella non-pneumophila |
| | Legionella pneumophila |
| | Leptospira |
| | Moraxella catarrhalis |
| | Mycobacterium tuberculosis |
| | Mycoplasma genitalium |
| | Mycoplasma pneumoniae |
| | Neisseria elongata |
| | Neisseria meningitidis |
| | Pneumocystis jirovecii (PJP) |
| | Pseudomonas aeruginosa |
| | Staphylococcus aureus |
| | Staphylococcus epidermidis |
| | Staphylococcus salivarius |
| | Streptococcus pneumoniae |
| | Streptococcus pyogenes |
| | Aspergillus sp |
| | Candida albicans |
b) Cross-reactivity Wet-Testing
In addition to *in silico* analysis of the SARS-CoV-2 primers and probes for cross-reactivity, the analytical specificity of the Xpert Xpress CoV-2 *plus* test was evaluated by bench-testing a panel of 62 microorganisms comprising (4) human coronaviruses, (1) MERS-coronavirus, (1) SARS-coronavirus, (20) other respiratory viruses, (30) respiratory bacteria, (2) yeast strains, (3) fungal strains, and (1) human nasal wash fluid representing a diverse microbial flora in the human respiratory tract.
The intact viruses were tested at concentrations of $\geq 10^{5}$ TCID$_{50}$/mL, $\geq 10^{5}$ CEID$_{50}$/mL, and $\geq 10^{5}$ copies/mL. Bacteria and yeast were tested at $\geq 10^{6}$ CFU/mL. The bacteria *Chlamydia pneumoniae* and *Mycoplasma pneumoniae* were tested at concentrations of $\geq 10^{6}$ IFU/mL and $\geq 10^{6}$ CCU/mL, respectively. The live strains of *Coxiella burnetii*, *Lactobacillus reuteri*, and *Neisseria meningitides* were not available, therefore genomic DNA at $\geq 10^{6}$ genome equivalent copies/mL was used. The *in vitro* transcribed (IVT) RNA sample was tested at $\geq 10^{6}$ genome equivalent/mL for human Coronavirus HKU1, and the genomic RNA sample was tested at $\geq 10^{6}$ genome equivalent/mL for SARS-coronavirus Urbani.
K230440 - Page 15 of 24
{15}
The microorganisms being evaluated for cross-reactivity were tested in groups or individually and spiked into negative simulated NPS/NS background matrix for testing. If a sample with grouped microorganisms produced a SARS-CoV-2 positive result, then each member of the group was tested separately. If the individual non-target organism yielded a positive result, retesting was performed at lower concentrations until a concentration that no longer produces a false positive result was identified.
The results of the cross-reactivity study demonstrated that the primer/probe sets included in the Xpert Xpress CoV-2 plus do not cross-react with the nucleic acids from non-intended respiratory microorganisms and phylogenetically related human coronavirus species. The one exception was the SARS-coronavirus Urbani which yielded SARS-CoV-2 POSITIVE test result. The cross-reactivity for the E gene target was expected with the SARS-coronavirus Urbani strain which has the highly conserved E gene target shared among coronaviruses in the B lineage Betacoronavirus. Results are shown in Table 13.
Table 13: Cross-reactivity testing of the Xpert Xpress CoV-2 plus Test
| Respiratory Microorganisms | Test Group | Tested Concentration | Final SARS-CoV-2 Result Call-Out | Number of Positive Results Obtained from Total Number of Replicates Tested | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | E | N2 | RdRP |
| Human coronavirus, 229E | G1 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Human coronavirus, OC43 | | 1.1e5 TCID50/mL | | | | |
| MERS-coronavirus | | 1.1e5 TCID50/mL | | | | |
| Human coronavirus, NL63 | G2 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Human coronavirus, HKU1a | G3 | 1.1e6 genome cp/mL | NEG | 0/3 | 0/3 | 0/3 |
| SARS-coronavirus, Urbani a | G4 | 1.1e6 genome cp/mL | POS | 3/3 | 0/3 | 0/3 |
| Influenza A H1N1 (pdm2009), Michigan/272/2017 | G5 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Influenza B (Victoria Lineage), Hawaii/01/2018 (NA D197N) | | 1.1e5 TCID50/mL | | | | |
| RSV-A, Strain: 4/2015 Isolate #1 | | 1.1e5 TCID50/mL | | | | |
| Adenovirus Type 1 | G6 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Adenovirus Type 7A | | 1.1e5 TCID50/mL | | | | |
| Cytomegalovirus | | 1.1e5 TCID50/mL | | | | |
| Echovirus | G7 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Enterovirus, D68 strain US/KY/14-18953 | | 1.1e5 TCID50/mL | | | | |
| Epstein Barr Virus (Human Herpes Virus 4 [Hhv-4]) | | 1.1e5 cp/mL | | | | |
| Herpes Simplex Virus (HSV) type 1 | | 1.1e5 TCID50/mL | | | | |
| Human metapneumovirus (hMPV-5, type B1) | | 1.1e5 TCID50/mL | | | | |
| Measles | | 1.1e5 TCID50/mL | | | | |
| Mumps virus | | 1.1e5 TCID50/mL | | | | |
| Human parainfluenza Type 1 | G8 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 2 | | 1.1e5 TCID50/mL | | | | |
| Human parainfluenza Type 3 | | 1.1e5 TCID50/mL | | | | |
| Human parainfluenza Type 4 | | 1.1e5 TCID50/mL | | | | |
| Rhinovirus, Type 1Ab | | 4.5e4 TCID50/mL | | | | |
| Acinetobacter baumannii | G9 | 1.1e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Burkholderia cepacia | | 1.1e6 CFU/mL | | | | |
| Candida albicans | | 1.1e6 CFU/mL | | | | |
| Candida parapsilosis | | 1.1e6 CFU/mL | | | | |
| Bordetella pertussis | | 1.1e6 CFU/mL | | | | |
| Chlamydia pneumoniae | | 1.1e6 IFU/mL | | | | |
| Citrobacter freundii | | 1.1e6 CFU/mL | | | | |
K230440 - Page 16 of 24
{16}
| Respiratory Microorganisms | Test Group | Tested Concentration | Final SARS-CoV-2 Result Call-Out | Number of Positive Results Obtained from Total Number of Replicates Tested | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | E | N2 | RdRP |
| Corynebacterium xerosis | G10 | 1.1e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Escherichia coli | | 1.1e6 CFU/mL | | | | |
| Enterococcus faecalis | | 1.1e6 CFU/mL | | | | |
| Hemophilus influenzae | | 1.1e6 CFU/mL | | | | |
| Legionella sppd | | 1.1e6 CFU/mL | | | | |
| Moraxella catarrhalis | | 1.1e6 CFU/mL | | | | |
| Mycobacterium tuberculosis (avirulent) | G11 | 1.1e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Mycoplasma pneumoniae | | 1.1e6 CCU/mL | | | | |
| Neisseria mucosa | | 1.1e6 CFU/mL | | | | |
| Propionibacterium acnes (= Cutibacterium acnes) Z144 | | 1.1e6 CFU/mL | | | | |
| Pseudomonas aeruginosa, Z139 | | 1.1e6 CFU/mL | | | | |
| Staphylococcus aureus | | 1.1e6 CFU/mL | | | | |
| Staphylococcus epidermidis | G12 | 1.1e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Staphylococcus haemolyticus | | 1.1e6 CFU/mL | | | | |
| Streptococcus agalactiae | | 1.1e6 CFU/mL | | | | |
| Streptococcus pneumoniae | | 1.1e6 CFU/mL | | | | |
| Streptococcus pyogenes | | 1.1e6 CFU/mL | | | | |
| Streptococcus salivarius | | 1.1e6 CFU/mL | | | | |
| Streptococcus sanguinis | | 1.1e6 CFU/mL | | | | |
| Pneumocystis jirovecii (PJP) | | 1.1e6 CFU/mL | | | | |
| Lactobacillus reuteri, F275c | G13 | 1.1e6 genome cp/mL | NEG | 0/3 | 0/3 | 0/3 |
| Neisseria meningitidesc | | 1.1e6 genome cp/mL | | | | |
| Pooled human nasal wash | G14 | N/A | NEG | 0/3 | 0/3 | 0/3 |
| Influenza C (Taylor/1233/1947) | G15 | 1.1e5 CEID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Rhinovirus, Type 1Ab | G16 | 1.1e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Aspergillus flavus | G17 | 1.35e7 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Aspergillus fumigatus | G18 | 3.67e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Coxiella burnetii c | G19 | 2.5e6 genome cp/mL | NEG | 0/3 | 0/3 | 0/3 |
| Leptospira broomii | G20 | 2.9e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Parechovirus type 1 | G21 | 3.39e5 TCID50/mL | NEG | 0/3 | 0/3 | 0/3 |
| Fusobacterium necrophorum | G22 | 6.52e6 CFU/mL | NEG | 0/3 | 0/3 | 0/3 |
| Mycoplasma genitalium c | G23 | 3.6e6 genome cp/mL | NEG | 0/3 | 0/3 | 0/3 |
a. Microorganism in the form of genomic RNA were tested in Tris-EDTA+ $\left[\mathrm{NH}_{4}\right]_{2} \mathrm{SO}_{4}$ buffer using an assay definition file (ADF) without sample preparation.
b. Rhinovirus, Type 1A was initially tested at $4.5\mathrm{e}4\mathrm{TCID}_{50} / \mathrm{mL}$ in test group G8 using virus lot 325725. It was re-tested individually (G16) at a higher concentration of $1.1\mathrm{e}5\mathrm{TCID}_{50} / \mathrm{mL}$ using virus lot 326527.
c. Microorganisms in the form of genomic DNA were tested in simulated NPS/NS background matrix using the ADF with full sample preparation.
d. Legionella pneumophila was tested in this study.
# c) Microbial Interference
A microbial interference study was performed to assess the inhibitory effects of commensal microorganisms potentially encountered in upper respiratory tract specimens on the performance of the Xpert Xpress CoV-2 plus test. A panel of 18 commensal microorganisms, consisting of 15 viral strains and 3 bacterial strains was tested. Contrived samples consisted of SARS-CoV-2 virus spiked at $1.15\mathrm{x} - 3\mathrm{x}$ LoD into simulated NPS/NS matrix in the presence of fifteen (15) commensal virus strains and three (3) commensal bacterial strains (Haemophilus influenzae, Staphylococcus aureus, Staphylococcus epidermidis) spiked at their respective concentrations listed in Table 14. SARS-CoV-2 virus and each potential microbial interference strain combination was tested in replicates of eight. All eight positive replicate samples were correctly identified as SARS-CoV-2 POSITIVE using the Xpert
K230440 - Page 17 of 24
{17}
Xpress CoV-2 plus test. No interference by the listed above commensal viral or bacterial strains was reported at the concentrations tested.
Table 14: Microbial Interference Study Results
| Commensal Strain | Concentration Tested | Number of Correct Test Results/ Number Tested |
| --- | --- | --- |
| Adenovirus Type 1Ca | 1x10^{5} TCID50/mL | 8/8 |
| Adenovirus Type 2Cb | 1x10^{5} TCID50/mL | 8/8^{c} |
| Adenovirus Type 3Bb | 1x10^{5} TCID50/mL | 8/8 |
| Human Coronavirus-OC43a | 1x10^{5} copies/mL | 8/8 |
| Human Coronavirus-229Eb | 1x10^{5} TCID50/mL | 8/8 |
| Human Coronavirus-NL63b | 1x10^{5} TCID50/mL | 8/8 |
| Human Coronavirus-HKU1b | 1x10^{5} copies/mL | 8/8 |
| Metapneumovirus 5, Type B1a | 1x10^{5} TCID50/mL | 8/8 |
| Parainfluenza Type 1a | 1x10^{5} TCID50/mL | 8/8 |
| Parainfluenza Type 2a | 1x10^{5} TCID50/mL | 8/8 |
| Parainfluenza Type 3a | 1x10^{5} TCID50/mL | 8/8 |
| Rhinovirus Type 1Aa | 1x10^{5} PFU/mL | 8/8 |
| Influenza A H1N1b | 1x10^{5} CEID50/mL | 8/8 |
| Influenza Bb | 1x10^{5} TCID50/mL | 8/8^{d} |
| Respiratory Syncytial Virus Ab | 1x10^{5} TCID50/mL | 8/8 |
| Haemophilus influenzaea | 1x10^{7} CFU/mL | 8/8 |
| Staphylococcus aureus | 1x10^{7} CFU/mL | 8/8 |
| Staphylococcus epidermidis | 1x10^{7} CFU/mL | 8/8 |
a. These commensal strains were tested with SARS-CoV-2 at a concentration of 3x LoD.
b. These commensal strains were tested with SARS-CoV-2 at a concentration of 1.15x LoD.
c. Two of 8 replicates reported as ERROR. The runs were successfully repeated to obtain 8 valid replicates.
d. One of 8 replicates reported as ERROR. The run was successfully repeated to obtain 8 valid replicates.
d) Endogenous/Exogenous Interference Substances
An interfering substances study was performed to evaluate if substances naturally present or artificially introduced into the nasal cavity or nasopharynx interfere with the Xpert Xpress CoV-2 plus test performance. Positive and negative samples were prepared in simulated NPS/NS matrix. Each interfering substance was tested at the concentration listed below (Table 15) in the absence of SARS-CoV-2 and when SARS-CoV-2 is present at 3x LoD. All specimens were tested in eight replicates on the Xpert Xpress CoV-2 plus test according to the test procedure. For substances that resulted in an INVALID test result, the concentration of the substance was reduced by half and re-tested.
Expected results were obtained for all interfering substances tested, except for Mucin (Type I-S) (2.5 mg/mL) and Fluticasone Propionate Nasal Spray (5 µg/mL). Table 15 below presents the concentration at which no interference was seen for each substance tested.
K230440 - Page 18 of 24
{18}
Table 15: Interfering Substances Study Results
| Substance | Concentration Tested | Number SARS-CoV-2 Negative / Number of Valid Replicates Tested | Number SARS-CoV-2 Positive / Number of Valid Replicates Tested |
| --- | --- | --- | --- |
| SARS-CoV-2 Negative (No Substance) Control | 100% (v/v) | 8/8 | - |
| SARS-CoV-2 Positive (No substance) Control | 100% (v/v) | - | 8/8 |
| Afrin | 15% (v/v) | 8/8 | 8/8 |
| Albuterol Sulfate | 0.83 mg/mL | 8/8 | 8/8 |
| BD Universal Transport Medium | 100 (v/v) | 8/8 | 8/8 |
| Blood | 2% (v/v) | 8/8 | 8/8 |
| Copan Swab M | 100 (v/v) | 8/8 | 8/8 |
| FluMist | 6.7% (v/v) | 8/8 | 8/8 |
| Fluticasone Propionate Nasal Spray | 5 μg/mL | 7/8a | 7/8a |
| | 2.5 μg/mL | 8/8 | 8/8 |
| Ibuprofen | 21.9 mg/dL | 8/8 | 8/8 |
| Leukocytes | 1 x 10^{6} cells/mL | 8/8 | 8/8 |
| Menthol | 1.7 mg/mL | 8/8 | 8/8 |
| Mucin (Type II) | 0.1% (w/v) | 8/8 | 8/8 |
| Mucin (Type I-S) | 2.5 mg/mL | 7/8a | 8/8 |
| | 1.25 mg/mL | 8/8 | - |
| Mupirocin | 10 mg/mL | 8/8 | 8/8 |
| Human Peripheral Blood Mononuclear Cells (PBMC) | 1 x 10^{3} cells/μL | 8/8 | 8/8 |
| PHNY | 15% (v/v) | 8/8 | 8/8 |
| Remel M4RT | 100% (v/v) | 8/8 | 8/8 |
| Remel M5 | 100% (v/v) | 8/8 | 8/8 |
| Saline | 15% (v/v) | 8/8 | 8/8 |
| Snuff | 1% (w/v) | 8/8 | 8/8 |
| Tamiflu | 7.5 mg/mL | 8/8 | 8/8 |
| Tobramycin | 4 μg/mL | 8/8 | 8/8 |
| Zicam | 15% (w/v) | 8/8 | 8/8 |
| Zinc | 0.1 μg/mL | 8/8 | 8/8 |
a. One of 8 replicates reported INVALID test result, indicating interference from the substance. The substance was subsequently tested with 8 replicates at half the initial concentration and no interference was observed.
5. **Assay Reportable Range:**
Not applicable; this is a qualitative assay.
6. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
a) **Assay Controls**
Please refer to Instrument Description Information (Section C.5) above for assay controls.
b) **Specimen Stability Studies**
**Room Temperature (15 to 30°C) and Refrigerated (2 to 8°C) Stability Studies**
A study was conducted to evaluate the stability of NPS specimens stored in UTM/VTM, followed by testing with the Xpert Xpress CoV-2 plus test. Clinical NPS specimens collected in UTM/VTM and eNAT, and confirmed SARS-CoV-2 negative were pooled for this study. The positive sample consisted of the NATtrol inactivated SARS-CoV-2 virus strain USA
K230440 - Page 19 of 24
{19}
WA1/2020 (ZeptoMetrix Inc., USA) spiked at 2.75× LoD (1108 copies/mL) into the pooled negative clinical NPS in UTM/VTM and into the pooled clinical NPS in eNAT. The same negative clinical matrix pool was used to provide the negative sample. Aliquots of each positive and negative sample were stored at 15°C and 30°C to cover the extremes of the recommended room temperature storage temperature range, and at 2°C and 8°C to cover the extremes of the recommended refrigerated storage temperature range. All four storage temperatures were monitored continuously by automated temperature monitoring and access control system. Eight (8) replicates of the positive sample and 4 replicates of the negative sample were tested at T=0, 4 hours, 8 hours, 24 hours, 48 hours and 50 hours after storage at 15°C and 30°C. Eight (8) replicates of the positive sample and 4 replicates of the negative sample were tested at T=0, 1 day, 2 days, 4 days, 6 days, 7 days, and 8 days after storage at 2°C and 8°C.
External positive and negative controls, run daily during the study, reported the expected Xpert Xpress CoV-2 plus test results. Aliquots of the negative and positive specimens stored at 15°C, 30°C, 2°C and 8°C, and tested at their respective timepoints gave expected results.
The study data supports the following specimen stability claims:
- NPS and NS specimens can be stored in UTM/VTM or in eNAT matrix at refrigerated (2°C - 8°C) temperatures for 7 days and at room temperature (15°C - 30°C) for up to 48 hours until testing is performed on the GeneXpert Instrument Systems.
Freeze/Thaw Testing for Samples Frozen at -20°C or -80°C in VTM or eNAT
A study was performed to determine equivalency of frozen SARS-CoV-2 samples compared to fresh samples when tested with the Xpert Xpress CoV-2 plus test. Positive samples used for the fresh and frozen specimen equivalency study were prepared by spiking the NATtrol SARS-CoV-2 (USA-WA1/2020) strain into pooled negative clinical NPS-UTM/VTM and NPS-eNAT at concentrations of < 1× LoD, 1-3× LoD and 5× LoD. A negative sample consisting of negative clinical NPS-UTM/VTM and NPS-eNAT matrices were also included in the study. The positive and negative samples were tested fresh, and each sample was subsequently allowed to undergo two freeze-thaw cycles consisting of freezing at -80°C for at least 24 hours and thawing at room temperature prior to testing. Thawed samples were kept on ice prior to testing.
The study data supports the following specimen stability claims:
Frozen clinical NPS-UTM/VTM, clinical NS-UTM/VTM and clinical NPS-eNAT specimens positive for SARS-CoV- 2 can undergo one cycle of freeze-thaw before testing with the Xpert Xpress CoV-2 plus test.
7. Analytical Sensitivity/Limit of Detection (LoD):
The LoD of the Xpert Xpress CoV-2 plus test was established using NATtrol inactivated SARS-CoV-2 virus strain USA-WA1/2020 (ZeptoMetrix Inc., USA) diluted into pooled negative clinical NPS matrix and negative clinical NS matrix. Clinical NPS specimens were collected in VTM/UTM and in eNAT, while clinical NS specimens were collected in VTM/UTM only. The LoD is defined as the lowest concentration for SARS-CoV-2 strain at which 95% (19/20) of replicates yield a positive result.
K230440 - Page 20 of 24
{20}
The virus strain was initially estimated in a range finding study at >10 concentrations in replicates of 20 per concentration of virus. Range finding was performed with two cartridge lots. Probit analysis was performed on the range finding results to estimate the LoDs, and the estimated LoDs were subsequently confirmed using two lots of Xpert Xpress CoV-2 plus cartridges. For the confirmatory study, testing was also performed in replicates of 20 per cartridge lot. The highest (least sensitive) LoD value for the two lots evaluated in the confirmatory study was reported as the final LoD for the Xpert Xpress CoV-2 plus test. The claimed LoD for SARS-CoV-2 is 403 copies/mL in Clinical NPS-UTM/VTM (Table 16). An equivalent LoD (403 copies/mL) was also established for SARS-CoV-2 in Clinical NPS-eNAT. The claimed LoD is 462 copies/mL in Clinical NS-UTM/VTM (Table 17).
Table 16: Results for Verification of SARS-CoV-2 LoD in Clinical NPS-UTM/VTM
| Virus Strain | Reagent Lot | LoD (copies/mL) | Positive Results out of # Valid Replicates | % Positive | Mean E Ct | Mean N2 Ct | Mean RdRP Ct |
| --- | --- | --- | --- | --- | --- | --- | --- |
| NATtrol SARS-CoV-2 | Reagent Lot 1 | 403 | 20/20 | 100% | 34.3 | 38.3 | 36.6 |
| | Reagent Lot 2 | 403 | 20/20 | 100% | 34.1 | 37.7 | 36.5 |
Table 17: Results for Verification of SARS-CoV-2 LoD in Clinical NS-UTM/VTM
| Virus Strain | Reagent Lot | LoD (copies/mL) | Positive Results out of # of Valid Replicates | % Positive | Mean E Ct | Mean N2 Ct | Mean RdRP Ct |
| --- | --- | --- | --- | --- | --- | --- | --- |
| NATtrol SARS-CoV-2 | Reagent Lot 1 | 462 | 20/20 | 100% | 34.4 | 38.9 | 36.8 |
| | Reagent Lot 2 | 462 | 20/20 | 100% | 34.1 | 37.6 | 36.1 |
The analytical sensitivity of the Xpert Xpress CoV-2 plus test was also estimated by testing WHO First International Standard for SARS-CoV-2 RNA in pooled negative clinical NS-UTM/VTM matrix. The study was performed in a random and blinded fashion. The LoD for the First WHO International Standard for SARS-CoV-2 RNA tested in NS-UTM/VTM matrix with the Xpert Xpress CoV-2 plus test was 1000 IU/mL (Table 18).
Table 18: LoD Concentration, Positivity Rate and Mean Ct Values for SARS-CoV-2 for the First WHO International Standard for SARS-CoV-2 RNA in Clinical NS-UTM/VTM Matrix
| Virus | LoD Concentration | SARS-CoV-2 Positive Results / Test Replicates | E Target | | N2 Target | | RdRP Target | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Positive Results / Test Replicates | Mean Ct | Positive Results / Test Replicates | Mean Ct | Positive Results / Test Replicates | Mean Ct |
| 1st WHO International Standard for SARS-CoV-2 RNA Virus | 1000 IU/mL | 20/20 | 20/20 | 34.3 | 19/20 | 39.5 | 20/20 | 36.5 |
8. Assay Cut-Off:
K230440 - Page 21 of 24
{21}
The Xpert Xpress CoV-2 plus test detects defined cycle threshold (Ct) ranges for the SARS-CoV-2 targets and SPC. The Ct cut-offs are hard coded values and are included as automatic calculations in the assay definition file (ADF) of the Xpert Xpress CoV-2 plus test.
## 9. Carry-Over Contamination Study:
A study was conducted to assess whether the single-use, self-contained Xpert Xpress CoV-2 plus cartridge prevents specimen and amplicon carryover by testing a negative sample immediately after testing of a very high positive sample in the same GeneXpert module. The study utilized two GeneXpert instrument modules; both modules were cleaned prior to starting the study. The negative sample used in this study consisted of simulated NPS/NS matrix and the positive sample consisted of high SARS-CoV-2 virus concentration (inactivated SARS-CoV-2 USA-WA1/2020 at 5×10⁴ copies/mL) spiked into negative NPS/NS matrix. The negative sample was tested in a GeneXpert module at the start of the study. Following the initial testing of the negative sample, the high positive sample was processed in the same GeneXpert module immediately followed by another negative sample. This was repeated 20 times in the same module, resulting in 20 positives and 21 negatives for the module. The study was repeated using a second GeneXpert module for a total of 40 positive and 42 negative samples. All 40 positive samples were correctly reported as SARS-CoV-2 POSITIVE and all 42 negative samples were correctly reported as SARS-CoV-2 NEGATIVE with the Xpert Xpress CoV-2 plus test. No specimen or amplicon carry-over contamination was observed in this study.
## B. Comparison Studies:
### 1. Method Comparison with Predicate Device:
Not applicable. Please refer to the Clinical Studies Section of this document.
### 2. Matrix Comparison:
The following studies were performed to establish equivalent performance of Xpert Xpress CoV-2 plus between (a) clinical nasopharyngeal swab (NPS), clinical anterior nasal swab (NS) and simulated NPS/NS matrices in Universal Transport Medium (UTM)/Viral Transport Medium (VTM), and (b) clinical NPS in UTM/VTM and eNAT.
Aliquots of clinical NPS-UTM/VTM, clinical NS-UTM/VTM, and simulated NPS/NS matrices were screened with the Xpert Xpress CoV-2 plus test and negative clinical samples were pooled. NATtrol inactivated SARS-CoV-2 virus strain USA-WA1/2020 (ZeptoMetrix Inc., USA) was spiked into the negative matrices at 0.05x, 1.5x, 5x, and 10x LoD. Aliquots of the negative and SARS-CoV-2 positive matrices were tested in replicates 10 (except for 30 replicates in case of 1.5x LoD). Expected results were obtained for all replicates tested.
A similar study as described above, except the simulated NPS/NS matrix, was used to establish equivalent performance between clinical NPS-UTM/VTM and clinical NPS-eNAT. Expected results were obtained for all replicates tested.
## Clinical Studies:
The clinical performance of the Xpert Xpress CoV-2 plus test was evaluated in a multi-site, observational and method comparison study in 2022 that included 32 geographically diverse sites in the United States (US). Of the 32 sites, 5 sites participated in specimen collection only, 26 sites performed Xpert testing and specimen collection, and 1 site performed Xpert testing as well as comparator and discrepant testing.
K230440 - Page 22 of 24
{22}
To be enrolled in the study, patients had to be presenting at the study centers showing signs and symptoms of respiratory infection. One nasal (NS) or nasopharyngeal (NPS) swab collected in a randomized and blinded fashion from each patient was tested using the Xpert Xpress CoV-2 plus test side by side with an FDA-cleared molecular respiratory panel that includes SARS-CoV-2. The demographics of study participants are presented below (Table 19). Of the 4047 specimens collected from individuals showing signs and symptoms of respiratory infection, a total of 3075 specimens (869 NPS and 1871 NS) passed all study criteria and were used for performance calculations. The Positive Percent Agreement (PPA) and Negative Percent Agreement of Xpert Xpress CoV-2 plus test with the FDA cleared molecular comparator test are presented in Table 20 below. Based on the specimens that yielded non-determinate results (INVALID, ERROR, and NO RESULT) with Xpert Xpress CoV-2 plus, the non-determinate rate was $1.6\%$ (32/2029) with NPS and $1.4\%$ (28/2018) with NS specimens.
Table 19: Demographic Information of Participants Symptomatic of Respiratory Infection
| | NPS (N=1879) | NS (N=1871) | Overall (N=3750) |
| --- | --- | --- | --- |
| Gender | | | |
| Female | 1107 (58.9%) | 1171 (62.6%) | 2278 (60.7%) |
| Male | 772 (41.1%) | 700 (37.4%) | 1472 (39.3%) |
| Age Group (Years) | | | |
| ≤5 | 8 (0.4%) | 64 (3.4%) | 72 (1.9%) |
| 6-21 | 389 (20.7%) | 380 (20.3%) | 769 (20.5%) |
| 22-59 | 1228 (65.4%) | 1167 (62.4%) | 2395 (63.9%) |
| ≥60 | 254 (13.5%) | 260 (13.9%) | 514 (13.7%) |
| Race | | | |
| American Indian or Alaska Native | 4 (0.2%) | 4 (0.2%) | 8 (0.2%) |
| Asian | 44 (2.3%) | 44 (2.4%) | 88 (2.3%) |
| Black or African American | 525 (27.9%) | 524 (28.0%) | 1049 (28.0%) |
| White | 1236 (65.8%) | 1222 (65.3%) | 2458 (65.5%) |
| Native Hawaiian or Other Pacific Islander | 4 (0.2%) | 0 (0%) | 4 (0.1%) |
| Black or African American, White | 7 (0.4%) | 8 (0.4%) | 15 (0.4%) |
| Other Mixed (each N≤5) | 7 (0.4%) | 2 (0.1%) | 9 (0.2%) |
| Participant declined to answer, or unknown | 52 (2.8%) | 67 (3.6%) | 119 (3.2%) |
| Ethnicity | | | |
| Hispanic | 170 (9.0%) | 165 (8.8%) | 335 (8.9%) |
| Non-Hispanic | 1681 (89.5%) | 1668 (89.2%) | 3349 (89.3%) |
| Participant declined to answer, or unknown | 28 (1.5%) | 38 (2.0%) | 66 (1.8%) |
K230440 - Page 23 of 24
{23}
| Specimen Testing | | | |
| --- | --- | --- | --- |
| Fresh | 1852 (98.6%) | 1845 (98.6%) | 3697 (98.6%) |
| Frozen | 27 (1.4%) | 26 (1.4%) | 53 (1.4%) |
| Testing Environment | | | |
| Laboratory/NPT | 996 (53.0%) | 971 (51.9%) | 1967 (52.5%) |
| CW | 883 (47.0%) | 900 (48.1%) | 1783 (47.5%) |
| Vaccine Status | | | |
| Vaccinated | 1336 (71.1%) | 1344 (71.8%) | 2680 (71.5%) |
| Not Vaccinated | 523 (27.8%) | 508 (27.2%) | 1031 (27.5%) |
| Unknown | 20 (1.1%) | 19 (1.0%) | 39 (1.0%) |
Table 20: Performance of Xpert Xpress CoV-2 plus in Symptomatic Individuals (Nasal and Nasopharyngeal Swabs Combined)
| | Comparator Result | | |
| --- | --- | --- | --- |
| Xpert Xpress CoV-2 plus | Positive | Negative | Total |
| Positive | 574 | 55 | 629 |
| Negative | 11 | 3110 | 3121 |
| Total | 585 | 3165 | 3750 |
| PPA (95% CI) | 98.1% (96.7% - 98.9%) | | |
| NPA (95% CI) | 98.3% (97.7% - 98.7%) | | |
C Clinical Cut-Off:
Not applicable.
D Expected Values/Reference Range:
Expected values as determined by Xpert Xpress CoV-2 plus are presented for nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimen types (Table 21).
Table 21: Positivity Rate by Specimen Type
| Specimen Type | |
| --- | --- |
| NPS | NS |
| 17% (320/1879) | 16.5% (309/1871) |
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K230440 - Page 24 of 24
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.